Page 1: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

For internal use only / Copyright © Siemens AG 2006. All rights reserved.For internal use only / Copyright © Siemens AG 2006. All rights reserved.

Possible Pitfalls of InSitu Measurements

Page 2: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 2 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Contents

Contents

Impact of hearing instrument setting / Connexx model 7

Impact of test setup and protocol 11

Impact of individual anatomy 19

Modifications and fixes 24

Verification of the modified Connexx model 26

Page 3: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 3 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Background

Deviations between NAL-NL1 target and REM gain during InSituverification of First Fit were reported.

Bad shape of curve between 1.5 and 3kHz“2kHz dip”Peaks at higher frequencies

Observed especially in UK and AustraliaObserved for various hearing instruments Observed for various test systems

Individual characteristics

Page 4: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 4 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Background

A project was started to

Investigate the reproducibility of the reported observations If applicable: find reasons for the deviationsIf applicable: perform fix and validate the fix

The actual presentation gives an overview about the investigations and actions performed within this project.

Page 5: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 5 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Agenda of actual presentation

Outcomes of investigations (reproducible impact factors)Connexx modelTest setupIndividual anatomy

Actions based on these findingsValidationFurther Recommendations

Questions and discussion at the end of each section

Page 6: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 6 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Investigated issues

Impact of hearing instrument setting / Connexx modelInternal clocking / Connexx ModelInput compression (AGCi)Output limiting (AGCo)

Impact of test setup and protocol StimulusSound source positionProbe tube PositionRoom acousticsMeasurement system

Impact of individual anatomyInter-individual variation of OEG and RECDHI tube lengthVolume of ear canal Diffractions of sound signal caused by head and shoulder

Page 7: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 7 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of hearing instrument setting / Connexxmodel

Wireless applications and standard on the ASIC make a 5% over-clocking of the chip necessary. However, all correction curves of the electro-acoustic elements stored in Connexx 5.3 are still based on the original clocking.The effect on the gain curve differs depending on the shape of the curve and is hardly predictable. In general, stronger effects can be expected for moderately sloping curves. Amount: about 2dB less gain for f<5kHz.

(Acuris S, First Fit Cluster3, pure tone stimulus)

LE65

-10

0

10

20

30

40

50

60

100 1000 10000Frequency (Hz)

gain

/ dB

UPLCxx

Deviation at LE65

-10

0

10

100 1000 10000

Frequency (Hz)

UPL

-Cxx

/ dB

Page 8: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 8 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of hearing instrument setting / Connexxmodel

An impact of sub-optimal AGCisettings (level meter) can be observed across the whole frequency range. Amount: up to 3dB @ 2kHz

(Acuris S, First Fit Cluster3, LTASS stimulus)

LE65

-10

0

10

20

30

40

50

60

100 1000 10000Frequency (Hz)

gain

/ dB

UPLCxx

Deviation at LE65

-20

-10

0

10

20

100 1000 10000

Frequency (Hz)

UPL

-Cxx

/ dB

Page 9: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 9 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of hearing instrument setting / Connexxmodel

The impact of AGCo (intended behavior) may cause large deviations from the Connexxsimulation of LE90 curve, if modulated noise is used as stimulus.Direct impact cannot be excluded for modulated stimuli and higher amplification levels.In fact, a reduction of about 6dB was observed for this study for 15dB linear gain with AGCo limiting while using the Aurins stimulus. When output liming was set to PC, correct gain curves were observed .(Acuris S, FOG curve for IEC-2 noise)

LE90

-10

0

10

20

30

40

50

60

100 1000 10000Frequency (Hz)

gain

/ dB

UPLCxx

Deviation at LE90

-10

0

10

100 1000 10000

Frequency (Hz)

UPL

-Cxx

/ dB

Page 10: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 10 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of hearing instrument setting

Impact of Connexx model and chip clocking

5% over-clocking was not accounted for in the Connexx model=> mis-shaping of curve especially between 1 and 2kHz, about 2dB

less gain for f<5kHz

AGCi parameters / level meters were sub-optimal=> up to 3dB less gain than expected

AGCo may influence measurement in some cases=> up to 6dB less gain was observed

Page 11: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 11 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

If two different stimuli are used for measurement, gain may be falsely estimatedby Unity1 using the 16bit software.Example: If Aurins is used for the measurement of insertion response and pink noise for the measurement of open ear response, the gain calculated by Unity1 will be over-estimated in the frequency range around 1kHz.

Comparison of Aurins and Pink Noise Spectra

0102030405060708090

100

100 1000 10000

Frequency / Hz

SPL

/ dB

Pink NoiseAurins

Page 12: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 12 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

If AURINS noise is used for the IG measurement, the measured curves must not be compared to the target curves displayed in the Unity, because the latter are NAL targets which are, by definition, based on an LTASS input. The correct targets are shown in Connexx and yield about 6dB less gain for Aurins between 1 and 2.5kHz (Cluster3, typical age related hearing loss).

Comparison of Aurins and LTASS Target, Prisma2 Pro

-200

-100

0

100

200

300

400

100 1000 10000

Frequency

dB *

1/10

Aurins

LTASS

Page 13: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 13 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

The position of the reference microphone yields a different sound field compared to the field reference point. If the input level is regulated with respect to the reference microphone, less input than expected is presented to the hearing instrument

effective sound field deviation

-20

-10

0

10

20

100 1000 10000

Frequency (Hz)

Lo /

dBsp

l

diff Kemar 0°

diff Kemar 30°

diff Kemar 45°diff Kemar 60°

settings:MP1, Centra S

deviation causedby equalizationbelow Kemar pinna

Page 14: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 14 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

The sound source position at 45deg combined with the different sound field cause a deviation from the standard REUG as assumed by Connexx. This may cause a peak / dip pattern in the aided response between 2.5 and 6kHz.

(Acuris S, First Fit Cluster3, pure tone, KEMAR

measurement)

Deviation of REUG at 45deg from exprected value

-15

-10

-5

0

5

10

15

100 1000 10000f / Hz

dB

Page 15: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 15 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

The position of the reference microphone yields a different sound field compared to the field reference point. In addition, the MLE for 45° sound direction differs from the 0° MLE. This leads to a different input to the instrument and, as a consequence, to a different gain curve than expected.

effective HI input deviation (45°)

-20

-10

0

10

20

100 1000 10000

Frequency (Hz)

Lo /

dBsp

l

insitu HI input deviation

settings:MP1, Centra S

1) equalization below pinna !2) Kemar 45°

Page 16: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 16 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Summed effects of setup and Connexx model

If the impacts of AGCi, clocking, and test setup are regarded isolated from each other, they are all within the tolerance range. However, if they are combined, the total effect rises up to about 6dB.

-10

-8

-6

-4

-2

0

2

4

6

8

10

100.00 1000.00 10000.00

Frequency

dB

total difftest setupAGCiclocking

Page 17: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 17 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Example case

The stimulus spectrum was not equalized during the measurement. This caused a prominent dip at 2.5 kHz during the InSitu measurement which cannot be attributed to the instrument. Rather, this is a measurement artifact caused by diffractions at the head and the shoulders of the patient. Expected gain curve: about 20dB flat from 250Hz to 6kHz.

Artis P, subject RS

-20

0

20

40

60

80

100

100 1000 10000

Frequency / Hz

Leve

l / d

B

REIG REAR Aurins

stimulus at ref mic

aided response(InSitu)

insertion gain(InSitu)

Page 18: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 18 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of test setup and protocol

If two different stimuli (Pink Noise and Aurins) are used for measurement, this may result into an over-estimation of gain at 1kHz with old Unity1 software

The position of reference microphone below the ear may cause dip at 1.8kHz of about 3dB

The sound source position at 45° causes peaky shape for f>2kHz

NAL-NL1 is based on LTASS and does not support the AURINS stimulus, instead Connexx makes a calculation based on the LTASS target. Unity is unable to do this and always shows the LTASS targets.

For some stimuli (incl. AURINS) only the overall level is equalized during the measurement, but not the spectrum shape. For these stimuli, diffraction at head and shoulders may cause dip at 2.5kHz of 10dB. The amount and shaping of the dip are determined by the individual anatomy of the patient.

Page 19: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 19 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of individual anatomyLength of HI tube

For HI tubes longer than standard, a more peaky shape of the gain curve can be expected for f>1kHz.

Impact of HI tube length at given ear canal volume

-10

-8

-6

-4

-2

0

2

4

6

8

10

100 1000 10000Frequency (Hz)

dB

length 33mm

length 35mm

length 37mm

length 39mm

length 43mm

settings:

CIELO PFOG - 20dBLE70

audio input:Zwislocki (var. length)

Parameter:real ear tube length

Page 20: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 20 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of individual anatomyVolume of ear canal

Less overall gain is measured for identical HI settings in case of larger ear canal volumes (i.e. large ear canal diameter).

Impact of ear canal volume

-10

-8

-6

-4

-2

0

2

4

6

8

10

100 1000 10000Frequency (Hz)

dB

volume 0%

volume +20%

volume +40%

settings:

CIELO PFOG - 20dBLE70

audio input:Zw islocki (35mm tubing)

Parameter:ear canal volume

Page 21: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 21 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of individual anatomyIndividual OEG and RECD: one example

The target gain curve in this figure was 20dB flat from 250 Hz to 4kHz. The statistical OEG and RECD curves lead to deviations below 900Hz (leakage effects) and a dip between 1 and 2 kHzImporting individual OEGs or RECDs and using tight ear molds led to a satisfactory result in this case.

Acuris

-20

-10

0

10

20

30

40

100 1000 10000

Frequency

dB

stat. curvesind. OEGind. RECDind. OEG & RECD

Page 22: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 22 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of individual anatomyAverage individual RECD

Measurements of individual RECDs with 50 subjects at an individual research site yielded an average RECD very much comparable to the standard RECD that is actually used in Connexx (HA2).

-20

-10

0

10

20

30

40

100 1000 10000

Frequency

dB

Median RECD in OL, Unity2UK RECDHA2 RECD

Page 23: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 23 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Impact of individual anatomy

Individual factors

Longer HI tube may cause dip at 2kHz

Larger ear canal volumes may cause less overall gain

Diffraction of stimulus at patient’s shoulder may cause dip at 2.5kHz

Individual REUG and RECD may have unpredictable influence which may be compensated easily by importing individual curves. However, the standard RECD used by Connexx fits very good to the average individual RECD.

Page 24: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 24 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Modifications and fixesConnexx 5.4

Compensation of clock trimming: included in Connexx 5.4

Optimisation of AGCi level meter: included in Connexx 5.4

Activate PC instead of AGCo during test setting for InSitu and coupler measurements: included in Connexx 5.4

Adjustment to modified sound field in case of InSitu fitting (reference mic and sound source positions): Connexx 5.4 allows this to be optimised via country specific SiFit.

Page 25: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 25 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Modifications and fixesUnity

Future versions of Unity software will be able to import the Connexx Aurinstarget curve. Until then

Compare to the Connexx targets to be met, not the Unity targets (on-top modus does not work!) or

Use Icra1 stimulus for InSitu measurements

Page 26: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 26 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Verification of the modified Connexx model Coupler measurements (UPL)

typical age related hearing lossmodulated speech-like stimulus (IEC2)Prisma2 Pro

Page 27: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 27 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Verification of the modified Connexx model Coupler measurements (UPL)

typical age related hearing lossmodulated speech-like stimulus (IEC2)Prisma2 Pro

Page 28: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 28 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Verification of the modified Connexx model First Fit Accuracy of Prisma2 Pro

Median REM insertion gain of 6 subjects

Median REM Gain Curves

-20

-10

0

10

20

30

40

100 1000 10000

Frequency

Inse

rtion

Gai

n (d

B)

LTASS Target

Prisma2 Pro

Page 29: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 29 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Verification of the modified Connexx model First Fit Accuracy of Prisma2 Pro P

Median REM insertion gain of 6 subjects

Median REM Gain Curves

-20

-10

0

10

20

30

40

100 1000 10000

Frequency

Inse

rtion

Gai

n (d

B)

LTASS Target

Prisma2 Pro P

Page 30: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 30 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Verification of the modified Connexx model First Fit Accuracy of Prisma2 Pro P

Median REM insertion gain of 6 subjects

Median REM Gain Curves

-20

-10

0

10

20

30

40

100 1000 10000

Frequency

Inse

rtion

Gai

n (d

B)

Prisma2 Pro DSP

LTASS Target

Page 31: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 31 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Summary

Deviations between target gain and REM gain as reported in the past are the result of various effects. Both instrument settings and measurement artefacts play a role.

Several impacts of HI setting / Connexx model were corrected for in Connexx 5.4.

Individual corrections that are needed due to test setup can be implemented in country specific SiFit (e.g. NHSFit).

The impact of very individual anatomy cannot be excluded. In some cases, it may be useful to import individual curves to Connexx in order to enhance the quality of the fitting. However, this does not always work with Unity1using the 16bit software => switch to 32bit software recommended

Page 32: Possible Pitfalls of InSitu Measurements - Sivantos · InSitu Measurements. Page 2 Siemens AudiologiDr. Heike Heuermann sche Technik, Erlangen Contents ... The actual presentation

Page 32 Siemens Audiologische Technik, ErlangenDr. Heike Heuermann

Recommendations

Hearing instrument settingsSwitch to test settings or switch off all adaptive parameters manually before REM measurement

Test setupUse the same stimulus for the measurement of unaided and aided responsesUse LTASS or ICRA instead of AURINSIf other stimuli are used: do not use the NAL-NL1 target displayed in Unity, but compare with target displayed in the fitting software (import of Connexx targets into Unity will be offered soon)Unity32 or Unity1 with 32bit software is recommended

Individual factorsFor patients with longer HI tubes, 2kHz dips should be expected For patients with large ear canal, less overall gain should be expectedFor patients with very unusual REURs, import OEG and/or RECD into Connexx(applicable approach for Unity2)