possible calibration methods for the final lxe calorimeter a. papa 01/20/2004
TRANSCRIPT
Possible calibration methods for the final LXe calorimeter
A. Papa 01/20/2004
The motivations
a frequent and precise check of the calorimeter stabilityeven during the normal data acquisition
An energy resolution of
Causes for gain instabilities:
• Beam intensity variations• Variable background rates (photons and neutrons in the experimental hall)• Effects of the temperature T variation on the photocathode Q.E. and resistivity• Effects due to the capacitive coupling• Possible hysteresis phenomena as a function of T
4%E/E (FWHM) means:
Thermal neutron capturePossible source:
Pulsed neutron generator (commercially produced)(switchable on-off)
Am/Be (~10 KBq)
• Continuous n spectrum
Moderator: ~10-20 cm of the polyethylene
γ shield:
• 40% thermalized n• 10% n captured inmoderator~ 3 cm of the tungsten
•Neutron lines: 4.5 or 14 MeV (d-d or d-t reaction respectively) • Typical intensity is 106 n/s or 108 n/s• Typical pulse rate and pulse width 10 Hz and 1 μs • Possible separation of direct from delayed reactions • Price ~ 10000 $
Counts
0 10 MeV
Precise calibration rarely performed
• γ‘s from decay (E(γ) ~ 54.9 MeV): use of a liquid hydrogen target
• Optional calorimeter calibration over range of γ energies:γ‘s from a tagged electron beam (small magnet + MWPC’s)
θ (degrees)
E (MeV)
Calibrations to be frequently performed (A)
Thermal neutron in Xe• Absorption length ~ 3 cm• Capture close to calorimeter walls• Multi γ, Σ E(γ) = 9.3 MeV• Possible spill-out
Capture on Ni plate on calorimeter wall • Single γ emission highly probable 52.7%• E(γ) = 9.0 MeV (used in SK)52.7% 25.6% 4.65% 1.28%
9.0 9.0 8.534 8.122 7.698 MeV
0
Neutron in the Large Prototyperecent measurement
γ energy spectrum
ADC
ADC
Without moderator (paraffin)
With moderator
Peak at 9.3 MeV
The neutron source is Am/Be (2 KBq) + diffused thermal neutron background in the experimental hall ( (?) note TN022 )
It can be improved: test with source on the calorimeter back and thicker moderator
γ energy spectrum
12216 scmn
Calibration to be frequently performed (B)
Other possibility (less recommanded):
Isotope activation far from detector with neutron generator and energetic neutron sources
• E(γ) = 6.13 MeV• Decay constant τ = 7.2 s
Possible reaction: or
Nitrogen laser UV: emission line at ~ 300 nm; use of the optical fibre and a small diffusor
• Gain and relative QE measurements
is PMT independent?
• No neutron on calorimeter (apart from hall background)
)175(
)330(
nmQE
nmQE
Conclusion
• Possible calibration methods were examined• Extremely important for calorimeter stability checks • Improvements studies on geometry, moderators, sources, reactions, etc under way