pharmacokinetic analysis of the mab adalimumab … notes/bioba-adalimumab-tech-note.… ·...
TRANSCRIPT
p 1
Pharmacokinetic Analysis of the mAb Adalimumab by ELISA
and Hybrid LBA/LC/MS: A Comparison Study
Featuring the SCIEX BioBA Solution
Xi Qu1, Shuyu Hou1, Meghan McCann1, Caroline Becker1, Xun Wang1, Zamas Lam1, Susan Zondlo1, John Kolman1, Lei Xiong2, Witold Woroniecki2, Hua-Fen Liu2, Ian Moore3
1QPS, Delaware, USA, 2SCIEX, Redwood Shores, USA, 3SCIEX, Concord, Canada Key Challenges in Biologics Quantitation
• Achieving wide linearity across the expected PK
sample range to minimize sample >ULOQ
reanalysis
• Hybrid LBA/LC-MS sample work-up requires
sourcing reagents from different vendors
• Lengthy and complicated procedures for
measuring active or free circulating drug are more
challenging in complex matrices
Key Features of the SCIEX BioBA Solution
• A complete solution for biologics bioanalysis by
immuno-affinity sample preparation
o BioBA kits provide all the reagents
necessary from high capacity streptavidin
beads to digestion enzyme
• Assays with mass spec detection offer wider LDR
than LBA
• High capacity streptavidin coated beads to
achieve high ULOQ
• Immuno enrichment allows enhanced sensitivity
for low abundance samples
• Ready to use vMethods reduce method
development time.
Introduction
The selection of a quantitative protein assay for biologics
bioanalysis in a pre-clinical study depends on what
platform will provide the right data for the drug under
development. Some questions to be considered when
choosing an assay platform include, is the total or free
drug concentration required? Are there in vivo structural
changes that might impact activity? Are these modified
proteins important to measure? Other considerations for
the chosen assay platform include reagents availability,
sensitivity and LDR requirements, sample throughput and
potential interferences.
The popular choice for protein quantitation has been
ELISA because of its high sensitivity, high throughput and
low per sample costs once the assay is developed and
validated. Limitations of the ELISA technique when
considering it as a platform include: lack of specificity in
primary or secondary antibody reagents, limited linear
dynamic range and interference due to ADA cross
reactivity. To avoid this cross reactivity, a more expensive
targeted antibody has to be used for pre-clinical studies.
The reasons for choosing a hybrid LBA and LC-MS assay
include: selectivity, broad LDR which reduces sample
dilution and the ability to multiplex a second analyte or
catabolite. Another reason to choose the hybrid
LBA/LC/MS approach for pre-clinical studies is the quick
method development turnaround time afforded by a
generic method that can be developed where the same
LBA capture reagent can be used across analytes and the
final selectivity of the assay is provided by the LC-MS
system.
The use of hybrid LBA LC/MS assays is growing and
SCIEX has developed the BioBA solution with ready to
use sample prep kits that include all the necessary
reagents and buffers which enables bioanalytical
scientists from all backgrounds to accomplish biologics
bioanalysis. Whether hybrid LBA LC/MS assays are used
BioBA reagents kit and SCIEX QTRAP® 6500 system.
p 2
instead of or as a complement to ELISAs in drug
development, the most important question is do they give
equivalent results?
In this tech note we have compared the PK parameters
generated from two different bioanalytical assays using
the two different platforms. Samples from a PK study of
the mAb drug adalimumab were split and tested with both
assays and the resulting sample concentrations and PK
parameters generated by the two assays were compared.
Methods
Dosing Study: Male Sprague-Dawley rats were given a
sub-cutaneous dose of adalimumab at 10 mg/kg and
samples were collected by tail-snip at: predose, 0.5, 2, 6,
24 h, 2, 3, 6, 8, 10, 14, 17, 21, 24 and 28 days, processed
into serum, split into 2 samples then frozen for ELISA and
LBA LC/MS analysis.
ELISA Sample Analysis: The samples were analyzed by
QPS using a previously validated ELISA assay in the
range of 250 to 10 000 ng/mL and samples were pre-
diluted 5 or 10 fold prior to analysis. A target specific
ELISA using TNFα coated plates was used for analysis.
Hybrid LBA LC/MS Sample Analysis: The samples were
analyzed by SCIEX Labs using the BioBA solution and
following the vMethod protocol for adalimumab using a
generic anti-human IgG capture antibody. The heavy
labeled antibody SILuMAB from was Sigma was added as
internal standard. The signature peptide APYTFGQGTK
was used for quantitation and the heavy labeled signature
peptide DTLMIS[R] from SILuMAB was used as the
internal standard.
Chromatography: Separation of the signature peptides of
the digested samples was performed on a Shimadzu
Prominence system using a Phenomenex 2.6 µm, Kinetex
C18 Column, (50 x 3.0 mm). A 7.0 minute gradient was
used and 20 µL of sample was injected onto the column.
Mass Spectrometry: The MRM analysis was performed
on a SCIEX QTRAP 6500® system equipped with an
IonDriveTM
Turbo V source.
Data Processing: MultiQuantTM
software was used for
peak integration, calibration and calculation of unknown
sample and QC concentrations.
Results
The serum concentration of adalimumab was measured
from two different dosing experiments. In dosing
experiment 1 four male Sprague-Dawley rats were used.
The adalimumab concentration from rats in experiment 1
as measured by a TNFα specific ELISA is shown in Figure
1A and the adalimumab serum concentration measured
using the generic hybrid LC-MS/MS assay where a
Figure 1A. Adalimumab serum concentration of Rats 5,6,7 and 8 measured by ELISA.
Figure 1B. Adalimumab serum concentration of Rats 5,6,7 and 8
measured by hybrid LC/MS/MS.
p 3
generic anti-human IgG antibody was used for the same
rats is shown in Figure 1B. The sample analysis was
carried out by QPS and SCIEX Labs separately as a
blinded study. It can be seen from inspecting both plots
that the two different techniques generated similar PK
plots for each rat. Furthermore the uniqueness of each
animal’s PK profile from the ELISA plot is also apparent
using the hybrid LC/MS/MS assay, e.g., the slower
elimination phase of adalimumab in rat 7. The average
adalimumab concentration across all four rats is plotted in
Figure 2 for both platforms and the resulting PK
parameters are in Table 1.
The data presented in figure 2 were generated in two
different labs, using two different assay platforms and by
two different analysts yet generate extremely similar PK
profiles. Closer inspection of the plot in Figure 2 reveals
the average adalimumab concentration (4 rats) as
measured by the ELISA assay is consistently higher at
each time point than the hybrid LC/MS/MS assay. There is
an approximately 20% bias in the ELISA data compared to
the hybrid LC/MS/MS data. This bias can also be seen in
the PK parameters generated from the two average PK
curves. The Cmax and AUC parameters are higher for the
ELISA measurement than the hybrid LC/MS/MS
measurement. The reason for the 20% bias was
investigated. Selectivity issues with the ELISA assay were
considered but eliminated because of the selective TNFα
ELISA assay used. Next the stock solutions used to
construct the calibration curves in the two labs were
compared. They were compared by LC/MS/MS and a
positive 15% bias was found in the ELISA stock solution
which contributed to the positive bias in the bioanalytical
results.
A second rat dosing study was then performed. The
adalimumab concentration from rats in dosing experiment
2 as measured by a TNFα specific ELISA is shown in
Figure 3A and the adalimumab serum concentration
measured using the generic hybrid LC-MS/MS assay for
the same three rats is shown in Figure 3B.
Figure 3A. Adalimumab serum concentration of Rats 17, 18 and 19 measured by ELISA.
Figure 2. The average adalimumab concentration across all four
rats measured by ELISA (closed circles) and hybrid LC/MS/MS (open triangles) from dosing experiment 1.
Table 1. The PK parameters of a 10 mg/kg dose of adalimumab
as determined by ELISA and hybrid LC/MS/MS from dosing experiment 1.
p 4
Figure 3B. Adalimumab serum concentration of Rats 17, 18 and 19 measured by ELISA.
The assay experiments (ELISA and hybrid LC/MS/MS)
from dosing study 2 were carried out in the same lab
(QPS) using the same stock solution. Again it can be seen
that the two different techniques generated similar PK
plots for each rat.
The average adalimumab concentration across all three
rats is plotted in Figure 4 for both platforms and the
resulting PK parameters are in Table 2.
It can be seen from Figure 4 that the two different assay
platforms generated almost identical PK profiles when
working from the same stock solution. Figure 5 shows the
excellent correlation between the two assay platforms for
each rat at each time point. With the smaller bias between
the two platforms in this dosing study the PK parameters
generated from the two average PK curves are the same
within experimental error.
Figure 4. The average adalimumab concentration across all
three rats measured by ELISA (closed circles) and hybrid LC/MS/MS (open triangles) from dosing experiment 2.
Table 2. The PK parameters of a 10 mg/kg dose of adalimumab
as determined by ELISA and hybrid LC/MS/MS from dosing experiment 2.
p 5
Figure 5. The correlation of each rat sample concentration
between each assay platform for dosing study 2. Excellent correlation was seen from dosing study 2 when the same stock solution was used to prepare calibrators for the two platforms
Conclusions
The objective of the animal dosing study was to determine
if the same samples tested using a generic hybrid
LC/MS/MS assay where a generic anti-human IgG
antibody was used would yield similar results to an ELISA
assay where a target specific immunocapture was used. It
can be seen from the PK plots in figures 2 and 4 that the
two platforms do give similar results. In animal dosing
study 2 both platforms gave the same PK parameters
(Cmax and AUC) which indicates that a generic hybrid
LC/MS/MS assay can be used as a selective method for
each antibody drug even before a target specific assay is
available. In animal dosing study 1 the measured
adalimumab concentrations were comparable within 20%.
After investigation the source of the bias was attributed to
a bias in the two different stock solutions used to prepare
calibrators in the two different labs.
It is important to know when choosing an assay platform
that it will deliver accurate results that the user can be
confident in. The reasons for choosing a hybrid LBA and
LC-MS assay over an ELISA assay include: selectivity,
wider LDR which reduces ULOQ sample dilutions, the
ability to multiplex a second analyte or catabolite and the
faster sample turnaround time afforded by a generic
method that can be applied to multiple studies. With the
BioBA solution SCIEX has provided a ready-to-use kit with
reagents that provide a generalized approach for the
immuno-capture and signature-peptide quantitation of
monoclonal antibody therapeutics that allow bioanalytical
scientists to easily benefit from the advantages of hybrid
LC/MS/MS assays. With the results of this tech note
bioanalytical scientists can be confident that these assays
will deliver accurate results from real animal or subject
samples.
AB Sciex is doing business as SCIEX.
© 2016 AB Sciex. For Research Use Only. Not for use in diagnostic procedures. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX™ is being used under license.
Biomek Method Launcher may not be compatible with Biomek Accounts and Permissions authentication. Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries. All trademarks are the property of their respective owners.
Document number: RUO-MKT-02-4865-B