performance testing of x-ray and gamma … testing of x-ray and gamma-ray detectors for imaging and...

122
PERFORMANCE TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department at the University of Surrey in partial fulfilment of the degree of Master in Physics. Department of Physics University of Surrey April 2008

Upload: hadat

Post on 17-Mar-2018

229 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

PERFORMANCE TESTING OF X-RAY AND GAMMA-RAY DETECTORS

FOR IMAGING AND SPECTROSCOPY

Richard Giordmaina

A dissertation submitted to the Physics Department at the University of Surrey in

partial fulfilment of the degree of Master in Physics.

Department of Physics

University of Surrey

April 2008

Page 2: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

ii

ABSTRACT

Spectroscopic detection of X and gamma (γ) radiation is of great importance in medical and security

applications. Research was undertaken at the University of Surrey and the Dstl Fort Halstead

laboratories to identify the imaging and spectroscopic capability with both existing and novel radiation

detection technology. There is a real need for spectroscopic information to be provided by radiation

detectors to offer the possibility for material discrimination, providing essential benefits that come with

this knowledge. Testing of the imaging and spectroscopic detection capabilities took place mainly

using a Cadmium Zinc Telluride (CZT) detector array, and novel Silicon Photomultiplier (SPM)

detector technology.

An initial review explored the various detection technologies which are currently available, populating

a graphical model to illustrate how these link together. Preparation and characterisation of CZT

radiation detectors was conducted to explore how raw detector materials can be tested and made ready

for use, and research using these detectors continued throughout the year, involving many experiments

to characterise and measure the performance of radiation detection equipment.

A CZT detector array was characterised by measuring the uniformity and energy resolution using

radioactive calibration sources to identify the spectroscopic capability of the detector. This was

repeated after a hardware upgrade to higher quality CZT, showing a noticeable improvement. It was

found that the CZT has energy resolutions of 7.91 0.03% at 59.5keV and 4.00 0.02% at 122keV

(an increase with energy as expected). The spectroscopic performance was then tested using two

calibration sources to determine the linearity of the counts recorded with time and inverse square law

with distance. The efficiency was found to be 72 1% at 122keV, proving that most photons will be

detected at 140keV, an energy commonly used in medical imaging.

The CZT detector was then used as a pinhole imager of backscattered X-ray photons from a variety of

objects, and the resulting images were compared to those obtained with an Intensified Charged

Coupled Device (ICCD) detector. These demonstrated that backscatter X-ray images can be produced

with both detectors. Analysis of the acquired images showed that the larger pixels of the CZT detector

make a more sensitive and better quality imaging camera, although with coarser images than the ICCD.

The majority of the research placement explored the characterisation and spectroscopic capability of

Silicon Photomultiplier (SPM) detectors, a relatively new technology, which are marketed to be a

possible replacement for Photomultiplier Tubes (PMTs) in many detection applications. A model

exploring the expected efficiency and energy resolution when different single pixel SPM detectors

(1mm and 3mm pixel sizes) are coupled to different scintillator crystals was produced to identify if

spectroscopy is theoretically possible. It was found that in certain scintillator, γ-source and SPM

combinations, spectroscopy would be possible as there would be enough photons remaining after losses

for an energy resolution of less than 15% to be obtained. The SPMs were then tested by measuring the

spectra produced when different scintillator crystals (CdWO4, CsI(Tl), BGO and LYSO) were coupled

to the three different SPMs and irradiated with radioactive γ-sources (241

Am, 57

Co, 22

Na and 137

Cs).

Many results found were directly comparable to the results expected using the model. From the

various measurements, energy resolutions were achieved, including 11.78 0.02% at 662keV. The

practical use for SPM detectors as small, fast counting spectroscopic radiation detectors has been

shown to be possible from the results obtained.

Page 3: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

iii

Acknowledgements

The author would like to take this opportunity to thank Dr Regan and Dr Sellin for organising the

placement which was a new, varied and exciting experience, which will no-doubt be of benefit in the

future. In addition, their assistance during the initial placement delays was greatly appreciated.

Dr Sellin has been very supportive as visiting tutor, by being available during the research year for

advice and help during my time away from the University.

The author would also like to show gratitude to his friends and colleagues at Dstl, especially Ian, Jane,

Dave and Paul for providing guidance and assistance, especially at the beginning of the new role.

Additionally thanks go to Paul for the use of portions of his existing programming code, which saved

're-inventing the wheel' in several cases, and for his help in checking through sections of new code.

Finally, the author would like to thank his friends and family, for their support, patience and

understanding during the research year and whilst writing this dissertation.

Author Declarations

Whilst preparing this dissertation, the author has not been registered for any other academic

qualifications, and this dissertation has only been submitted for the Master of Physics academic award.

Page 4: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

iv

List of Abbreviations

A list of all the abbreviations used throughout this dissertation is included here.

Abbreviation Description

APD Avalanche Photodiode

BGO Bismuth Germinate

CCD Charge Coupled Device

CdWO4 Cadmium Tungstate

Cs Cesium

CsI(Tl) Cesium Iodide (Thallium activated)

CT Computerised Tomography

CZT Cadmium Zinc Telluride

DAQ Data Acquisition System

DGF Digital Gamma Finder

Dstl Defence Science and Technology Laboratory

ehp(s) Electron-hole pair(s)

et al. and others

etc. et cetera

eV Electron volt

FF Fill Factor

FNA Fast Neutron Activation

FWHM Full Width at Half Maximum

GAPD Geiger mode Avalanche Photodiode

GMS Graphical Modelling System

IC Integrated Circuit

ICCD Intensified Charged Coupled Device

IV Current-Voltage

LED Light Emitting Diode

LYSO Lutetium Yttrium Silicon Dioxide

MCP Micro Channel Plate

M.Phys. Master in Physics

MCA Multi Channel Analyser

MOD Ministry of Defence

NaI Sodium Iodide

NIST National Institute of Standards and Technology

NQR Nuclear Quadrupole Resonance

PAB Probability to initiate Avalanche Breakdown

PDE Photon Detection Efficiency

PFNA Pulsed Fast Neutron Activation

PMT Photomultiplier Tube

ROI Region of Interest

SCA Single Channel Analyser

SPES Single Photoelectron Spectrum

SPM Silicon Photomultiplier

SNR Signal to Noise Ratio

TNA Thermal Neutron Activation

UK United Kingdom

XCOM Attenuation Database

XIA X-ray Instrumentation Associates

Page 5: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

v

List of Tables

A list of all the tables used in this dissertation is included here.

Table 2-1: Scintillator crystals and their key properties [12 (LYSO details from 13)]. ......................... 19

Table 3-1: Procured calibration source details. (*Used for CZT detector uniformity and energy

resolution experiments and were original laboratory sources.) ................................................... 47

Table 4-1: The results for the uniformity (after flat fielding) and energy resolution of the CZT detector

before and after the hardware upgrade. ...................................................................................... 56

Table 4-2: The uniformity of the ICCD detector for each source before and after applying the flat field

corrections. ................................................................................................................................ 56

Table 4-3: The effect of „flat fielding‟ on the CZT Gaussian peaks, for the upgraded detector. ............ 57

Table 4-4: A close match for the measured values to that expected from the model for energy resolution

measurements. ........................................................................................................................... 80

Table 4-5: The average measured energy resolutions (%). (Errors at the 95% confidence level). ....... 85

Table 5-1: The best obtainable spectroscopic energy resolutions (nearest %) for the two detectors at the

same energies. ........................................................................................................................... 96

Table 5-2: The best obtainable detection efficiencies (nearest %) for the two detectors at the same

energies. .................................................................................................................................... 96

Table 6-1: Resistivity values and results from the conducting foam experiment. ............................... 106

Page 6: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

vi

List of Figures

A list of all the figures in this dissertation is included here.

Figure 1-1: The summed attenuation in two regions for various materials over 2cm. ............................. 5

Figure 2-1: The electromagnetic spectrum (left) [4] and a diagram of an X-ray tube (right) [5]. ............ 6

Figure 2-2: A typical X-ray spectrum showing the characteristic peaks of the target material and the

Bremsstrahlung spectrum which accounts for most of the energy [4]. .......................................... 7

Figure 2-3: The energy level transitions in an atom. .............................................................................. 7

Figure 2-4: The main photon interaction mechanisms in matter varying with energy [6]. ...................... 9

Figure 2-5: Possible photon interactions in a detector (left) [6] and the Compton scattering of an

electron (right) [4]. ...................................................................................................................... 9

Figure 2-6: A typical X-ray transmission imaging arrangement. .......................................................... 12

Figure 2-7: An image of a brain tumour identified by 99

Tcm imaging [3]. ............................................ 13

Figure 2-8: Pinhole imaging providing sharper images for a smaller entrance aperture [8]. ................. 14

Figure 2-9: The operation of a semiconductor (CZT) radiation detector [6]. ........................................ 15

Figure 2-10: The drift velocity (Vd) as a function of applied electric field [6]. ..................................... 16

Figure 2-11: The stages in an ICCD detector [11]. .............................................................................. 18

Figure 2-12: The readout in a charged coupled device [6]. .................................................................. 18

Figure 2-13: The emission of scintillation crystals and the response of PMT devices [6]. .................... 19

Figure 2-14: The scintillation process from activated states [6]. .......................................................... 20

Figure 2-15: A photomultiplier tube with the amplification stages in the dynodes [3].......................... 21

Figure 2-16: A single pixel SPM (on a square base approximately 3x3cm) [20]. ................................. 22

Figure 2-17: A 1mm2 SPM pixel with many microcells on the top (left) [17] and the GAPDs in the

pixel are connected together to provide a photon proportional output [17] (right). ..................... 23

Figure 2-18: A graph containing the PDEs for 4V over bias for various SPM products [19]. ............... 23

Figure 2-19: The attenuation ratio I/I0 for the three detector materials explored for the project. ........... 25

Figure 2-20: The attenuation ratio over the energy range to be tested for the four scintillator crystals

used at 3mm thickness and the CZT detector at 5mm for direct comparison. ............................. 26

Figure 2-21: Operation of an MCA, an extension of many single channel analysers (SCAs) [6]. ......... 27

Figure 2-22: Typical pulse height spectra from γ-sources [6]. ............................................................. 28

Figure 2-23: An improved energy resolution is obtained with a thinner peak [6]. ................................ 29

Figure 2-24: The source emitting in 4π, where only a proportion of the activity is incident on the

detector at some distance (d) away............................................................................................. 30

Figure 3-1: The CZT detector array. ................................................................................................... 34

Figure 3-2: The experimental setup for the uniformity and energy resolution measurements. .............. 35

Figure 3-3: An image (based on detector area ~12.5cm2) from a γ-source illumination (left) and a

histogram from which the detector uniformity can be calculated (right). .................................... 35

Page 7: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

vii

Figure 3-4: The Gaussian fit to the 241

Am peak (at 0.5m for 10 minutes). ............................................ 37

Figure 3-5: The ICCD detector (with a 2mm pinhole plate attached). .................................................. 38

Figure 3-6: The housing applied to the CZT detector providing a pinhole aperture at 10cm from the

centre of the detector. ................................................................................................................ 39

Figure 3-7: Top view of the arrangement for X-ray backscatter pinhole imaging................................. 39

Figure 3-8: An image created using a pinhole receiving backscattered X-rays (left) and the effect

chamfering has on the pinhole (right-above before and right-below after chamfering). .............. 39

Figure 3-9: The leakage obtained before the additional shielding (detector area ~7.9x10-3

m2). ............ 40

Figure 3-10: The optimum shielding configuration (left) and the more uniformly distributed counts

with this shielding. .................................................................................................................... 41

Figure 3-11: The experimental arrangement for separation experiment (CZT pinhole covered for a

background measurement). ........................................................................................................ 42

Figure 3-12: The positions of the tubes of salt and sugar, moving closer together in 1cm steps. .......... 42

Figure 3-13: The line of data extracted for ICCD images (left) and CZT images (right). ..................... 43

Figure 3-14: The transistor astable circuits produced to pulse an LED [30] (left) and a circuit diagram

for the 555 IC used to pulse an LED [32] (right). ....................................................................... 45

Figure 3-15: The oscillator circuits produced to pulse an LED using transistors and capacitors (left) and

the pre-packaged oscillator in the IC NE555 timer (right). ......................................................... 45

Figure 3-16: Experimental arrangement for the SPM pulse linearity experiment. ................................ 47

Figure 3-17: The pulse exploration and SPM spectroscopy experimental arrangement. ....................... 49

Figure 3-18: The Xia Pixie-4 system used to acquire the spectrum from the SPM detectors. ............... 50

Figure 3-19: A summary of the planned SPM testing, showing each SPM coupled to each scintillator

crystal activated by each γ-source, and the background measurements. ...................................... 50

Figure 3-20: The single photoelectron spectrum possible with an SPM [21 modified]. ........................ 51

Figure 4-1: The activity calculator with extrapolated activity over two months. .................................. 52

Figure 4-2: A screenshot from the shielding requirements spreadsheet for pinhole imaging. ............... 52

Figure 4-3: An image (left) and histogram (right) from the 241

Am illumination (over 30 minutes) ....... 53

Figure 4-4: A histogram with a fitted Gaussian for the 30 minute integration of 241

Am. ...................... 53

Figure 4-5: Flat field corrections applied to the 10 minute data for 241

Am (left) and 57

Co (right). ........ 54

Figure 4-6: The energy resolution per pixel for the 57

Co 30 minute integration (left) and for the 241

Am

30 minute integration (right). ..................................................................................................... 54

Figure 4-7: Energy spectra using 57

Co for 30 minutes (left) and

241Am for 30 minutes (right). ............. 55

Figure 4-8: Illumination of the 30 minute 57

Co (left) and 241

Am for 30 minutes (right) showing crystal

artefacts. .................................................................................................................................... 55

Figure 4-9: Flat fielded image with 57

Co (left) and 241

Am (right) showing the artefact has now gone.

Note the bright pixels (right) mask the true discontinuities image. ............................................. 55

Figure 4-10: The effect of flat fielding on the 30 minute flood illuminations for both sources 241

Am

(left) and 57

Co (right). ................................................................................................................ 57

Figure 4-11: A linear relationship successfully identified between increase in counts and integration

time, for two γ-sources. ............................................................................................................. 58

Page 8: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

viii

Figure 4-12: The energy spectrum after just one second, showing 241

Am (59.5keV) and 57

Co source

(122 and 136keV) peaks. ........................................................................................................... 59

Figure 4-13: The effect of time and distance on the maximum number of counts received for two of the

integrations times tested using the CZT detector, showing the inverse square law. ..................... 60

Figure 4-14: The measured energy spectrum (left) and one just published for 57

Co (right) [38]. .......... 60

Figure 4-15: The efficiency of the CZT detector with increasing source distance. ............................... 61

Figure 4-16: A direct comparison for the same object (sugar) with the 4mm and 2mm pinholes (before

flat field corrections) from the ICCD detector. ........................................................................... 62

Figure 4-17: Flat fielded images of cotton in a case, using CZT (left) and ICCD (right). ..................... 62

Figure 4-18: Talc and aluminium powder on a plastic base (flat fielded) using CZT (left), ICCD

(middle), and before flat field corrections applied to the CZT (right). ........................................ 63

Figure 4-19: Images of the same object (but reversed) of the sections taken for analysis of the mean and

standard deviation (CZT left and ICCD right). ........................................................................... 64

Figure 4-20: A graphical representation of the data collected for the ratio of object to background each

image from both detectors. ........................................................................................................ 64

Figure 4-21: The number of counts from the average of 10 pixels in the CZT and ICCD detector for all

of the objects (error as the standard deviation). .......................................................................... 65

Figure 4-22: 1cm apart before flat field corrections (left) and after (right) for the CZT detector. ......... 66

Figure 4-23: 2-4cm (left to right) after flat field corrections for the CZT detector. .............................. 66

Figure 4-24: 1cm apart before flat field corrections (left) and after (right) for the ICCD detector. ....... 66

Figure 4-25: 2-4cm (left to right) after flat field corrections with 57

Co 30 minutes for the ICCD

detector. .................................................................................................................................... 66

Figure 4-26: The separation seen using the CZT (top) and ICCD (bottom) detectors over 5cm, where

the background for the ICCD is always higher than CZT. .......................................................... 67

Figure 4-27: The separation displayed as the change in normalised intensity for both detectors (error

bars as smallest unit taken from the trough height). ................................................................... 68

Figure 4-28: Results for the modelled energy resolutions for each crystal and SPM over the energy rage

to be tested. Clockwise from top left: CsI(Tl), BGO, LYSO and CdWO4. ................................. 69

Figure 4-29: The 1mm SPM pulses with no sources present seen in oscilloscope mode on the DAQ

(left) and expanded on another oscilloscope showing distinguishable dark photons (right). ........ 70

Figure 4-30: The dark counts in the 3mm 20μm SPM (top) and 3mm 35μm SPM (bottom), showing

noise photons not as clearly defined as the 1mm SPM. .............................................................. 70

Figure 4-31: 3mm 35µm SPM at 32V bias using the 555 IC timer (left) and 3mm 35µm 32V (right) for

an LED pulse where the SPM response (pink) to the LED pulse (blue) ...................................... 71

Figure 4-32: The onset times of all three SPMs using pulsed LED circuits to show that the pulses are

indeed produced in around 12ns. ............................................................................................... 71

Figure 4-33: A pulse caused by the pulsing of a red square LED, with black tape around the four

exposed sides on the 1mm SPM................................................................................................. 72

Figure 4-34: The SPM response to an LED pulse not fully recorded. .................................................. 72

Figure 4-35: The counting of the SPM to be at least 2x105Hz. ............................................................ 73

Page 9: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

ix

Figure 4-36: The effect of resistance (and therefore LED power from the 555IC timer) on the size of

the SPM pulses produced. .......................................................................................................... 74

Figure 4-37: A scintillation pulse from the 3mm 20µm SPM (at 32V bias) using CsI(Tl) with the 22

Na

source. ....................................................................................................................................... 74

Figure 4-38: A comparison of the voltage pulses from scintillator crystals and sources using the 3mm

20μm SPM at 32V. .................................................................................................................... 75

Figure 4-39: The CdWO4 response to 22

Na on the 3mm 20μm SPM, the decay lasting much less than

the 20µs expected. ..................................................................................................................... 75

Figure 4-40: Comparing the response of the pulse preamplifier which cuts off the full pulse duration

(left) with the transimpedance amplifier (right). ......................................................................... 76

Figure 4-41: The effect of bias on the SNR for each SPM. .................................................................. 77

Figure 4-42: The energy spectrum using CsI(Tl) with the 22

Na source, providing an energy resolution

of 14% using the 3mm 35μm SPM. ........................................................................................... 78

Figure 4-43: The modelled energy resolutions (top) and measured average energy resolutions using

CsI(Tl) for each SPM (bottom). ................................................................................................. 79

Figure 4-44: An energy spectrum from the 3mm 20μm SPM using BGO and 137

Cs. ............................ 79

Figure 4-45: Two separate SPM linearity experiments; using the 3mm 20μm SPM with 137

Cs and

CsI(Tl), and the 3mm 35μm SPM with 22

Na and BGO. .............................................................. 80

Figure 4-46: A ln-ln plot of energy vs. energy resolution using BGO on the 3mm 35μm SPM. ........... 81

Figure 4-47: The average measured energy resolution results using BGO, LYSO and CsI(Tl)

scintillator crystals on the 3mm 20µm SPM. .............................................................................. 81

Figure 4-48: The average measured energy resolution results from several scintillator crystals coupled

to the 3mm 35µm SPM. ............................................................................................................. 82

Figure 4-49: The average measured energy resolution results for the CsI(Tl) scintillator crystal

compared to the modelled result for the 1mm SPM. ................................................................... 83

Figure 4-50: The measured energy resolutions with the 1mm SPM and the modelled results based on a

complete match in areas and at a 1/9 area match. ....................................................................... 83

Figure 4-51: A comparison of the measured energy resolutions for the three SPMs using all of the

sources on the CsI(Tl) crystal. ................................................................................................... 84

Figure 4-52: The extrapolated energy resolution (based on measured results) to 140keV for eight

scintillator and SPM combinations. ........................................................................................... 85

Figure 4-53: A 137

Cs Spectrum for two minutes on the 3mm 35μm SPM (left) and the effect seen when

241Am (further back at 5cm) from the crystal is added to the experiment (right). ........................ 86

Figure 4-54: The 122keV peak clearly to the right of the cursors showing the position around the

59.5keV peak. ........................................................................................................................... 87

Figure 4-55: Illumination with 57

Co (left) and both 241

Am and 57

Co (right) showing a broadening due to

the 59.5keV source from 33% to 52% at 150 counts. ................................................................. 87

Figure 4-56: Both high energy sources (511 and 662keV) when integrating for two minutes. .............. 88

Figure 4-57: The energy spectrum for 22

Na (511keV) without the 662keV source, the peak is missing

when integrating for two minutes............................................................................................... 88

Page 10: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

x

Figure 4-58: BGO with 137

Cs for two minutes on the 3mm 35μm SPM. .............................................. 89

Figure 4-59: The addition of 22

Na to the 137

Cs source for two minutes. ................................................ 89

Figure 4-60: Integrating 22

Na and 137

Cs for 30 minutes better defines the spectrum. ........................... 89

Figure 4-61: The beta spectrum from LYSO taken for 30 minutes with no additional radioactive

sources using the 3mm 20μm SPM. ........................................................................................... 90

Figure 4-62: The linear increase of counts with integration time for the 3mm 35µm using 57

Co and

CsI(Tl). ..................................................................................................................................... 90

Figure 4-63: A comparison of the system efficiencies for each SPM using CsI(Tl). ............................ 91

Figure 5-1: An array of 16 3mm pixel SPMs [34]. .............................................................................. 97

Figure 6-1: The system start-up screen. ............................................................................................... 99

Figure 6-2: The Pixie4 Run Control menu......................................................................................... 100

Figure 6-3: The positions of the cursors to find the energy resolution. ............................................... 101

Figure 6-4: The cursors around the peak to provide the energy resolution and value of the peak. ...... 101

Figure 6-5: The attenuation jumpers for each channel of the acquisition system. ............................... 102

Figure 6-6: GMS with no object selected. ......................................................................................... 103

Figure 6-7: Identifying who takes Mathematics by moving the mouse over Mathematics. ................. 103

Figure 6-8: The „Weld View‟ of the Students Tutorial. ..................................................................... 104

Figure 6-9: A schematic of the conducting foam used [35]. ............................................................... 105

Figure 6-10: The experimental arrangement to find the CZT sample resistivity. ................................ 106

Figure 6-11: The change of source energy from 662keV (top) to 59.5keV (bottom). ......................... 108

Page 11: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

xi

Contents

ABSTRACT ................................................................................................................................... ii

Acknowledgements ........................................................................................................................ iii

Author Declarations ....................................................................................................................... iii

List of Abbreviations....................................................................................................................... iv

List of Tables ................................................................................................................................... v

List of Figures ................................................................................................................................. vi

Contents .......................................................................................................................................... xi

Chapter 1 : Introduction and Background .............................................................................................. 1

1.1 Background ..................................................................................................................... 1

The Research Problem ............................................................................................................. 2

Detection Techniques Review .................................................................................................. 3

Material Modelling .................................................................................................................. 4

Chapter 2 : Theory ................................................................................................................................ 6

2.1 Production and Properties of X and Gamma-rays ............................................................. 6

X-rays ...................................................................................................................................... 6

Radioactive Sources ................................................................................................................. 8

Photon Interactions in Matter ................................................................................................... 8

Photon Attenuation ................................................................................................................ 11

2.2 Detection Technologies ................................................................................................. 12

X-ray Transmission Imaging .................................................................................................. 12

Medical Imaging .................................................................................................................... 12

Backscatter X-ray Imaging ..................................................................................................... 13

2.3 Radiation Detection Systems ......................................................................................... 14

Cadmium Zinc Telluride (CZT) Detectors .............................................................................. 14

Intensified Charge Coupled Device Detectors ........................................................................ 17

Scintillator Crystal Properties ................................................................................................. 18

The Silicon Photomultiplier (SPM) ........................................................................................ 21

2.4 Radiation Measurement and Spectroscopy ..................................................................... 26

What Happens to the Detected Radiation? .............................................................................. 26

Pulse Height Spectra .............................................................................................................. 27

Detector Calibration ............................................................................................................... 28

How is Spectroscopic Capability Determined? ....................................................................... 28

Error Analysis ........................................................................................................................ 32

Page 12: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

xii

Chapter 3 : Experiments and Work Conducted .................................................................................... 33

3.1 Calculations and Modelling ........................................................................................... 33

Radioactive Source Calculator ............................................................................................... 33

CZT Shielding Requirements ................................................................................................. 33

3.2 CZT Experiments .......................................................................................................... 34

The CZT Detector .................................................................................................................. 34

Uniformity and Energy Resolution ......................................................................................... 34

CZT Spectroscopy ................................................................................................................. 37

CZT Efficiency Measurements ............................................................................................... 37

The ICCD Detector ................................................................................................................ 38

CZT and ICCD X-ray Backscatter Pinhole Imaging ............................................................... 38

CZT and ICCD Angular Resolution ....................................................................................... 42

3.3 SPM Experiments .......................................................................................................... 43

Scintillator and SPM Energy Resolution Model ..................................................................... 43

Light Emitting Diode (LED) testing ....................................................................................... 45

SPM Pulse Linearity .............................................................................................................. 46

SPM Pulse Observations ........................................................................................................ 47

SPM Spectroscopy ................................................................................................................. 49

SPM Detector Efficiency ....................................................................................................... 51

The Single Photo Electron Spectrum ...................................................................................... 51

Chapter 4 : Results and Analysis ......................................................................................................... 52

4.1 Modelling Results.......................................................................................................... 52

Radioactive Source Calculator ............................................................................................... 52

CZT Shielding Results ........................................................................................................... 52

4.2 CZT Experimentation Results ........................................................................................ 53

Uniformity and Energy Resolution ......................................................................................... 53

CZT Spectroscopy ................................................................................................................. 58

CZT Efficiency ...................................................................................................................... 61

CZT and ICCD X-ray Backscatter Imaging ............................................................................ 61

Angular Resolution ................................................................................................................ 65

4.3 Scintillator and SPM Experiments ................................................................................. 68

Scintillator and SPM Energy Resolution Model ..................................................................... 68

Scintillator and SPM: Preliminary Test Results ...................................................................... 70

Page 13: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

xiii

LED Testing .......................................................................................................................... 70

SPM Pulse Linearity .............................................................................................................. 73

SPM Pulse Observations ........................................................................................................ 74

SPM Spectroscopy ................................................................................................................. 77

Observing Complex Spectra using SPMs ............................................................................... 86

SPM Detection Time .............................................................................................................. 90

SPM Efficiency Results ......................................................................................................... 91

Chapter 5 : Review, Conclusions and Further Work ............................................................................ 92

6.1 Conclusions ................................................................................................................... 92

6.2 Further Research ........................................................................................................... 96

References .......................................................................................................................................... 98

Chapter 6 : Appendices ....................................................................................................................... 99

APPENDIX I: OBTAINING A SPECTRUM ............................................................................ 99

APPENDIX II: GMS ............................................................................................................... 103

APPENDIX III: Initial Research: Introduction and CZT Resistivity ........................................ 105

APPENDIX IV: Experimental Equipment List ........................................................................ 107

Page 14: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

1

Chapter 1 : Introduction and Background

This dissertation describes the research conducted in 2007 into current and novel spectroscopic and

imaging radiation detectors, which could have widespread applications.

1.1 Background

The research placement at the Defence Science and Technology Laboratory (Dstl) started in April.

Dstl is a trading fund of the Ministry of Defence (MOD), comprising over 3500 employees in several

locations across the UK providing “essential, impartial, high quality, timely advice on science and

technology issues”. Working for the “UK Armed Forces, the MOD or other government departments,

Dstl does not engage in work that can be done outside Government and therefore does not compete for

business with Industry” [1]. Dstl is divided into 14 departments (including Electronics, Biomedical

Sciences and Physical Sciences) and teams [1]. The research placement was undertaken in the Physical

Detection team of the Energetics Department. The broad aims of the team are to investigate, develop

and refine the sensors used in non-invasive detection. Technologies using X-rays, neutrons and nuclear

quadrupole resonance are some of those of interest.

There is a vital need to detect materials non-invasively, and an obvious example of this is aviation

security which has the requirement for rapid threat detection, due to a high volume of items.

Frequently, imaging using transmitted X-rays is performed, which relies upon the attenuation of

photons due to different absorption in different materials to produce an image. When materials

overlap, differentiation is more difficult as a thick amount of one material could be similar to a thinner

amount of another [2]. A key question is „could detection systems provide any more information?‟

For example, if spectroscopic information was obtained, there could be the possibility to identify

materials based on their elemental composition, rather than just an image to be interpreted.

The need for non-invasive detection in also vital for diagnostic medicine using passive γ-radiation

detectors to identify tumours in Positron Emission Technology (PET) and Single Photon Emission

Computerised Tomography (SPECT), where it is not always required to operate on a person to

diagnose certain tumours. By inducing a γ-emitter which is taken up by a tumour, it is possible to build

Page 15: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

2

up a picture of the inner body from the outside. The energy range of the diagnostic medicine includes

140keV and 511keV (higher for some radiological treatments).

The possibility for improved detectors could have important benefits including rapid detection (to

reduce exposure time and dose required) and imaging in medical and threat detection applications.

The placement has explored three strands of detection:

active transmission X-ray modelling to determine how well metals and organics can be

separated;

active backscatter X-ray imaging to identify imaging when there is only access to one side of

an object;

passive γ-ray spectroscopy and detector characterisation.

The Research Problem

The research has been driven by the requirement to improve the current sensors for radiation detection

measurements, by exploring the characteristics of new detector materials and their ability to provide

spectroscopy. There is the potential to improve the information provided by detection systems.

Detailed detector characterisation was conducted by:

identifying the detector performance of a CZT detector array (before and after a hardware

upgrade), in terms of the uniformity (and comparing this to existing imaging technology using

an ICCD detector) and spectroscopic energy resolution to determine the level of peak

separation possible;

exploring the detection efficiency of the CZT detector using single and multiple γ-sources (to

include linearity of count rate over integration time and confirming the inverse square law

with increasing distance from the detector);

demonstrating backscatter X-ray imaging through a pinhole using the CZT detector with a

specifically designed graded material housing, and comparing these images to ones taken with

existing technology (an ICCD detector);

Page 16: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

3

characterisation of three novel single pixel Silicon Photomultipliers (SPMs) to identify

detection efficiency, proportionality of the light output and the noise of the detector to

establish the performance possible;

spectroscopic testing of the SPMs with several different scintillator crystals in combination

with a large energy range of γ-sources;

comparing the different detection systems tested in terms of detection efficiency, energy

resolution, portability and linearity of the detectors.

All of this research focuses on identifying the characteristics and capabilities of two main detector

systems detectors (CZT, and scintillators with SPMs) as spectroscopic imagers to quantify their current

characteristics and to determine the suitability of these in passive γ-ray detectors, active transmission or

active backscatter detectors. Should these new detector systems have superior qualities, they could

replace existing technology in many applications.

Detection Techniques Review

The need for the research was identified by conducting a review of current detection techniques used,

providing an understanding of the systems that currently exist, and their method of operation. This

included metal detectors, X-ray detection (such as transmission and backscatter), neutron-based

detectors (such as Thermal, Fast and Pulsed Fast Neutron Analysis (TNA, FNA, PFNA)), vapour

detection, millimetre wave radiation and Nuclear Quadrupole Resonance (NQR). Information was

found using journals, texts and the Internet (product manufacturer web pages) for current and relevant

information. Where a piece of equipment using the technique existed, it was explored for advantages

and disadvantages.

Using a graphical modelling system (GMS 4.2 produced by the Office of Naval Research) a model was

produced summarising this information based on equipment identified, showing graphically the

different detection techniques. The software allows a technique to be selected, to display equipment

available using the technique, or select an item to be detected (e.g. metals) to display methods to detect

this. The report and model were designed to give an understanding of the techniques available and

how the research over the placement would aid the area of detection by adding spectroscopic sensors.

Page 17: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

4

Material Modelling

To create a complete picture of the three detection methods, X-ray transmission imaging was explored

to identify the level of material discrimination possible. The attenuation data of nine common

materials (aluminium, copper, iron, salt, iron, Perspex, cotton, water and sucrose) was investigated

using XCOM (an Internet based attenuation calculator). The ratio of X-ray intensities through to that

absorbed was calculated for a range of energies for a range of material thickness. This explored how

the X-ray attenuation in these materials varies with X-ray energy and material thickness over a 1-

140keV (every keV) range over a material thickness of 2mm up to 2cm thickness (every 2mm).

A pre-existing modelled X-ray spectrum was normalised for intensity at each energy, and was

multiplied into the material attenuation data in a spreadsheet to produce attenuation curves (or „banana

curves‟) for each material. Attenuation curves were made using data at specific attenuation energies;

80 and 140keV, and 90 and 120keV. Then by dividing the 140keV energy range into sections (1-

85keV and 86-140keV) the sum of the attenuations for each thickness in each section was plotted. The

level of material discrimination possible is determined by how close the data points lie together.

Multi-energy attenuation was explored by further dividing the 140keV energy range into three equal

sections and sums and averages were taken of the attenuation in each material in each energy section,

for thicknesses of 0 to 2cm. To visualise three-dimensional plots, the values of summed attenuation

were read into a program, which showed organic materials, such as sucrose and water are very different

to inorganic materials such as iron and aluminium when explored at these energies.

By further dividing the energy 140keV range into 4, 5 and 6 energy ranges, multi-energy material

discrimination could be explored. However attenuation curves cannot be created for more than three

dimensions, so at the time of writing, various mathematical methods were being explored to analyse the

data collected, including Principle Component Analysis. The results of the mathematical analysis of

this should determine the ability for differentiation based on more than two energies used.

Page 18: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

5

Figure 1-1: The summed attenuation in two regions for various materials over 2cm.

Figure 1-1 shows the attenuation curves for common materials and that there is clear separation

between metals and organic materials. The results of this study showed that material discrimination is

possible based on the attenuation of X-rays through a material. With this introductory work completed,

the next section of the project was to experiment with X-ray backscatter imaging and spectroscopy

using newly acquired experimental equipment.

Page 19: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

6

Chapter 2 : Theory

This chapter contains relevant theory used over this research project.

2.1 Production and Properties of X and Gamma-rays

X-rays

X-rays were first discovered by Röntgen in 1895 [3], and applications for them has continued to

increase since their ability to penetrate matter was identified. They are used in medical X-rays and

Computerised Tomography (CT) scans, and industrial applications such as fault finding in structures.

X-rays are part of the high energy section of the continuous electromagnetic spectrum comprising radio

waves (wavelength (λ) ~ 108m) through to X-rays (λ~10

-10m) and γ-rays (λ~10

-15m).

Figure 2-1: The electromagnetic spectrum (left) [4] and a diagram of an X-ray tube (right) [5].

To produce X-rays, an electron beam can be produced by thermionic emission (heating off electrons)

from a cathode. By applying a high potential difference between the anode and cathode, an electric

field is produced (Equation 13) and the electrons are rapidly accelerated towards a target metal

(commonly tungsten). In the metal, electrons are excited into higher energy levels which promptly

decay, producing X-ray photons with energy directly proportional to the difference between energy

levels. This process takes place under vacuum to remove the air in the path of the electrons which

would otherwise cause a breakdown of the air inside the generator due to the high voltages applied (a

160kV potential produces 160keV photons).

Page 20: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

7

X-rays are the result of atomic de-excitation, and are distinguished from γ-rays which are the result of

nuclear de-excitation. There are two types of X-ray. Bremsstrahlung radiation is a range of energies

caused by electrons scattering and changing velocity due to a nucleus, producing Bremsstrahlung

photons. These are seen as a range of energies in a typical X-ray spectrum (Figure 2-2).

Figure 2-2: A typical X-ray spectrum showing the characteristic peaks of the target material and

the Bremsstrahlung spectrum which accounts for most of the energy [4].

Characteristic X-rays are material specific (in terms of energy) X-rays, generated from transitions in

energy levels of the atom. By irradiating a material with high energy photons, enough energy can be

applied to remove an electron in the target material from its orbit, creating a vacancy. Depending on

which shell the electron falls from determines the energy of the resulting X-ray, and these can be

detected and analysed. The levels have historically been assigned letters starting with K for n=1.

Figure 2-3: The energy level transitions in an atom.

1. An electron decaying from the M to the L shell produces a Lα X-ray.

2. An electron decaying from the M to the K shell produces a Kβ X-ray.

3. An electron decaying from the L to the K shell produces a Kα X-ray.

A branch of spectroscopy explores the peaks corresponding to these transitions which appear in the

energy spectrum, which are specific to each element and can therefore be used to identify that element.

Page 21: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

8

Radioactive Sources

Radiation comes in many forms, from charged alpha and beta particles to radiation comprised of

uncharged photons such as X-rays and γ-rays. In fact, the use of alpha particles helped physicists to

formulate a model of the atom in the famous „gold leaf experiment‟. X-rays can be artificially

produced and switched on and off as required, making them a useful probing tool. Radioactivity from

natural sources is a random process which has no such capability and is therefore stored in lead when

not required. Radiation detection can be active or passive. Active imaging requires additional

radiation to be supplied externally such as in transmission X-rays, whilst passive detection makes used

of existing radiation being emitted, such as in thermal imaging, which makes use of radiation already

emitted by a body. The decay constant (λ) in Equation 1 gives the probability of a decay per second,

which is inversely proportional to the half life (t1/2).

Equation 1

2/1

2ln

t

The activity (A) in Equation 2 is the number of decays per second in Becquerel (Bq) (where 1Bq is 1

decay per second). Activity decreases exponentially with time.

Equation 2 teAA 0

The half life (Equation 3) is a measure of the time it takes for half of a sample to decay, and this varies

widely for different sources. For the sources to be used in the experiments, it is required to determine

the change in source activity over the time of the project. The half life can be derived using the

exponentially decaying number N0 or activity A0 as they are simply related by the decay constant,

reducing to N0 /2 or A0/2.

Equation 3 2ln

2

1t

Photon Interactions in Matter

A photon is defined as a quantum of electromagnetic radiation (Equation 4) with an energy E (where h

is the Planck constant and υ is the frequency of the radiation [5]). There are three main photon

Page 22: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

9

interactions in matter which are, in ascending order of energy, the Photoelectric effect, Compton

scattering and Pair production.

Equation 4 hc

hfE

The following interaction mechanisms have different probabilities of occurring (determined by the

cross section (σ)) which relate to the atomic number (Z) of the detector material seen in Figure 2-4.

The photoelectric effect and Compton scattering are most relevant to the energies explored.

Figure 2-4: The main photon interaction mechanisms in matter varying with energy [6].

Figure 2-5: Possible photon interactions in a detector (left) [6] and the Compton scattering of an

electron (right) [4].

Page 23: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

10

Photoelectric Effect (or Photoelectric Absorption) – An incident photon surrenders its energy to eject

an inner electron from an atom, and a photoelectron with kinetic energy T is produced from the energy

of the photon (with energy hυ) once the energy required to eject the electron (the material‟s work

function Ф) is gained (Equation 5). The vacancy from the ejected electron is filled from the de-

excitation of a higher shell electron, with the energy released being an X-ray photon with energy equal

to the difference between the two levels.

Equation 5 Th

The likelihood of an interaction via the photoelectric effect (P) occurring is approximated in Equation

6, where Z is the atomic number (to the power of between 4 and 5) of the material and Eγ is the energy

of the incident photon. An interaction via this method is more likely for higher Z materials.

Equation 6 5.3

5.4~

PE

Z

Compton Scattering – An incident photon collides with a stationary electron in a material and the

photons scatters (now carrying a reduced energy 'E in Equation 7) with the remaining energy given to

the electron, causing it to carry a momentum (Figure 2-5 right). The photons which interact in the

detector by this method scatter at the different angles (θ) depositing only some of the energy. The

probability of scatter is a function of the incident energy, depositing more energy with a larger angle of

scatter. When only a portion of the photon energy stays inside the detector, energies less than the peak

are recorded. The range of energies appears as the Compton continuum in Figure 2-22.

Equation 7

2

0

'

)cos1(1cm

E

EE

Pair Production – An electron-positron pair is produced equally sharing the γ-ray energy in the

presence of the nuclear electric field. The positron loses energy in scattering events and then produces

two 511keV (almost) back-to-back photons following annihilation with a free electron. An initial

photon energy of at least 1.022MeV is therefore required for pair production to occur and is unlikely to

Page 24: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

11

be seen in these measurements. Equation 8 shows how the energy of the incident photon is distributed

in pair production, where e

T and e

T are the kinetic energies of the electron and positron respectively,

and 2m0c2 is the rest mass energy of the electron and positron.

Equation 8 2

02 cmTTEee

When one annihilation photons leaves the detector, an escape peak appears 511keV less than the full

energy peak, or 1.022MeV less than the full energy peak if both annihilation photons escape (Figure

2-22).

Photon Attenuation

Attenuation is the reduction in number of photons in a beam due to passing though a material and all

materials (including air) attenuate photons. In Equation 9, Io is the initial number of photons, I is the

number remaining after attenuation through thickness (x), of the material. Colour coding based on the

attenuation detected is possible in some detector systems, separating metals into blue and organics into

orange for example, to help with identification of materials. The linear attenuation coefficient, µ, is a

measure of the attenuation in a medium, and it is constructed from the sum of the attenuations of the

three main photon interactions (Equation 10), where µ varies with photon energy. I / Io usefully

provides the ratio of photons through, regardless of the initial number of photons.

Equation 9 xeII 0

Equation 10 PPCSPE

Page 25: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

12

2.2 Detection Technologies

X-ray Transmission Imaging

The technique of using X-rays to produce an image (a radiograph or just an X-ray) of an object by

transmitting X-rays through it is well understood. In this „standard‟ transmission imaging, such as a

bone X-ray, an X-ray generator is placed on one side of an object and a detector (commonly silver

halide film [7]) is placed on the opposite side of the object. The X-ray photons pass through the object

and due to attenuation of the photons in the beam, fewer pass through the object to reach the detector.

Film darkens where a photon strikes it, so bones for example, which are more dense (higher µ) than the

surrounding tissue will absorb more photons, so a film based detector on the other side will show bone

as whiter as fewer photons have struck the detector. The attenuation of the X-ray photons is the main

factor to determine the image obtained. The detection of these photons used in transmission imaging

has advanced dramatically since photographic film, to include charge coupled devices (CCDs) and

research is constantly conducted into finding new detector materials.

Figure 2-6: A typical X-ray transmission imaging arrangement.

Medical Imaging

Common medical imaging techniques include Positron Emission Tomography (PET) and Single

Photon Emission Computed Tomography (SPECT). In PET, a natural radionuclide is induced into the

body (such as carbon oxygen or nitrogen depending on the region required to be imaged) which decay

by emitting a positron. The positron travels a short distance (millimetres) losing energy within the

body. When enough energy has been lost in scattering events, the positron annihilates with a free

electron creating a pair of 511keV γ-ray photons approximately back-to-back, which are detected from

Page 26: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

13

outside the body. Detecting the concentration of these photons allows an image to be constructed to

identify abnormalities.

In SPECT imaging, the collection of 140keV γ-ray photons emitted from inside the body is another

method to non-invasively image the internal body structure. Molybdenum decays by β- emission into

technetium (metastable or long-lived state), which then decays into technetium and a single 140keV γ-

ray photon (Equation 11). Choosing a pharmaceutical which is taken up by a specific organ allows

imaging of that region when the emitter is added to the solution. These photons can be detected after

collimation by scintillator crystals and a bank of photomultiplier tubes [3].

Equation 11 99

Mo 99

Tcm

99Tc + γ

Figure 2-7: An image of a brain tumour identified by 99

Tcm

imaging [3].

New sensors for the detection of these photons could make lower power, light and portable radiation

detectors a possibility, whilst retaining or improving the spectroscopic and imaging capability.

Backscatter X-ray Imaging

Compton scattered photons, cause a wide angle of scattered radiation, (determined from Equation 7)

can also be used to provide X-ray images. Backscatter (or single-sided) X-ray can be used when there

is only easy access to one side of an object, if for example the contents of a case are unknown and it

therefore may not be safe to move the item to view it in a traditional transmission X-ray system. This

can provide more information due to the increased photon flux. In this arrangement, a detector is

placed on the same side as the X-ray source, which records X-rays that have returned from the object.

The amount of energy from a backscattered X-ray is significantly less than the transmitted energy due

Page 27: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

14

to the scattering. Higher Z materials will absorb more of the X-rays (Equation 6) whilst lower Z

materials will scatter more.

The pinhole effect is a well understood method for imaging light rays, commonly by exposing film.

The recorded image will be a reflection of the true one as seen in Figure 2-8. When beams of light are

passed through smaller apertures, sharper images are obtained but will take more time to form, as a

lower photon flux is incident on the film.

Figure 2-8: Pinhole imaging providing sharper images for a smaller entrance aperture [8].

When the diameter of the pinhole aperture is of the order of the wavelength of the radiation, diffraction

occurs, which is the spreading of waves from a source. Maximum diffraction occurs at the point when

the size of the aperture is of the same order as the wavelength of the radiation (light). The wavelengths

being used in this application are 7.8x10-12

m (Equation 4) and therefore it is determined that diffraction

will not need to be compensated for, when using a 2mm pinhole.

2.3 Radiation Detection Systems

Cadmium Zinc Telluride (CZT) Detectors

Radiation detectors are broadly divided into two categories; „charge-based‟ and „light-based‟. The

proportional counter, Geiger-Muller tube and semiconductor detectors are charge-based detectors

working on the principle of electron and holes charges; whilst light-based detectors such as a

photomultiplier tubes (PMTs) or photodiodes operate by detecting and amplifying optical photons

produced by light-emitting (scintillator) materials. CZT is a semiconductor based radiation detector

which has gained a lot of attention over the past few years due to its room temperature (without the

need for liquid nitrogen cooling to reduce the thermally created charges) detection of X- and γ-ray

Larger

aperture

leads to

blurred

image

Page 28: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

15

photons. CZT is made from materials with high atomic numbers (Z) of 48, 30 and 52 respectively,

which have the ability to attenuate photons better than lower Z materials, and therefore detect an

interaction. Other materials which are used as direct semiconductor radiation detectors include silicon

(Z=14, suited for low energy photons) and germanium (Z=32, requires cooling) and research into new

materials for detecting radiation is constantly ongoing. CZT is more formally written as Cd1-xZnxTe

where x is the blending fraction of zinc telluride in cadmium [6]. A higher zinc concentration increases

the band gap of the material, allowing fewer thermally generated electrons to be detected as noise.

CZT resistivity values between 2.5x1010

Ωcm (4% zinc concentration) and 1.5x1011

Ωcm (20% zinc

concentration) are expected for CZT depending on the blending fraction of zinc in the sample [9]. The

value of resistivity is an important feature in the detection of radiation photons in the material, as a

higher resistivity implies a lower leakage current (as resistivity is inversely proportional to current)

providing a better quality material.

Incident radiation such as those from radioactive sources producing X- or γ-rays will deposit energy

creating electron-hole pairs (ehp) proportionally to the amount of incident energy (Figure 2-9). The

ehp pair creation energy (W) for a typical Cd0.9Zn0.1Te detector 4.64eV [10].

Figure 2-9: The operation of a semiconductor (CZT) radiation detector [6].

The capacitance (C) of the detector system is generally a constant, and includes the capacitance of the

detector and all leads (which should therefore be kept short). The voltage (V) detected is then directly

proportional to the charges produced by the radiation (Equation 12).

Page 29: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

16

Equation 12 )(VC

QV

Inside a semiconductor detector, an applied voltage (V) between the two electrodes (at thickness d)

creates an electric field (Equation 13) often 105Vm

-1, causing the charges created by radiation to drift

with a velocity υ. The electrons migrate to the positive electrode (anode) and the holes to the negative

electrode (cathode). The drift velocity (for electrons υe and holes υh in Equation 14) is proportional to

the mobility, μ, of the electrons and holes (how easily they can move in the material), and the strength

of the electric field. The charges created in the detector move between 106 and 10

7ms

-1 to the

collection points at the electrodes of the detector material.

Equation 13 illustrates how a smaller distance between the anode and cathode will linearly increase the

electric field. CZT and SPM detectors can be made very thin providing a very large electric field

(105Vm

-1). However, thinner detectors will be less able to attenuate radiation photons causing a lower

detection efficiency. Figure 2-10 shows how the increase in the electric field causes the drift velocity

to increase in silicon (in semiconductor detectors the electric field is approximately 103Vcm

-1).

Equation 13 d

VE (Vm

-1)

Equation 14 Eee and Ehh (ms-1

)

Figure 2-10: The drift velocity (Vd) as a function of applied electric field [6].

Page 30: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

17

From Ohm‟s Law (Equation 15) for a given voltage (V) and current (I) the resistance (R) of a detector

material can be found. By measuring the current at different voltages, a plot of current vs. voltage

provides resistance from the reciprocal of the gradient.

Equation 15 I

VR (Ω)

Resistivity ( ) is “a measure of a material‟s ability to oppose the flow of an electric current” and

increases with resistance (R) and cross sectional area, A [5]. It is commonly expressed in units of Ωcm

for CZT devices. An IV graph can be used to determine the sample‟s resistivity, where m in Equation

16 is the gradient of an IV graph is the reciprocal of resistance.

Equation 16 mL

A

L

RA (Ωm)

Intensified Charge Coupled Device Detectors

[11] describes the operation of Intensified Charge Coupled Device (ICCD) detectors. Figure 2-11

shows incident photons striking a photocathode, producing photoelectrons. A micro channel plate

(MCP) receives these photoelectrons and through the electric field, many more secondary electrons are

produced. These strike a fluorescent screen (such as phosphorus), producing optical light flashes

which are usually transferred by fibre optic cable. A CCD receives these photons which are converted

to charges, which are moved and read out row-by-row in the readout section (Figure 2-12).

Page 31: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

18

Figure 2-11: The stages in an ICCD detector [11].

Figure 2-12: The readout in a charged coupled device [6].

This device is used for imaging only, as only counts and no energy information is recorded. Having a

pixellated input phosphor allows for imaging, as variations in intensity over the pixels can be displayed

graphically as an image.

Scintillator Crystal Properties

A scintillator is a light-based radiation detector which emits photons due to excitation from the incident

radiation, and qualities such as high linear light yield (the number of photons created per unit of energy

deposited (usually given in MeV)) and a short decay time are desired, such that fast and efficient

detection is possible. Scintillators, such as cesium and sodium iodide, are commonly coupled to a

Page 32: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

19

Photomultiplier tube (PMT) increasing a low level of light photons into a much higher one. The PMT

turns the light produced from the scintillator into electrons in the photocathode by the photoelectric

effect, and the efficiency of this process is determined by the quantum efficiency of the photocathode.

There is a rapid increase the number of electrons using stages of dynodes [3]. According to [6], the

charge creation energy is 3.6eV for silicon and around 15-20 times that for scintillators, which only

convert 5-10% of energy to light. Scintillators are therefore intrinsically less efficient than

semiconductor detectors at stopping and detecting radiation photons. Radiation damage to scintillators

and other detectors is possible, as the lattice can be altered from the regular periodic structure,

compromising the ability to effectively detect and respond to radiation events [6].

Many materials scintillate in response to radiation. Scintillator materials are sub-divided into organic

(liquids and plastics) and inorganic materials, the latter being used for the project. Using Equation 4,

the energy of the optical photons from four scintillator crystals was determined.

Scintillator Light Yield /

MeV

Decay Time (μs) Peak λ (nm) Photon

Energy (eV)

BGO 9,000 0.3 480 2.57

CdWO4 13,000 20 520 2.19

LYSO 32,000 0.04 420 2.94

CsI(Tl) 52,000 1 565 2.38

Table 2-1: Scintillator crystals and their key properties [12 (LYSO details from 13)].

Based on these detector properties, these four scintillators were procured to provide a range of fast and

slow decay times, and high and low light yields to fully test the range of the SPM. Hygroscopic

crystals were not chosen to avoid degradation in crystal performance over the period of experiments.

Figure 2-13: The emission of scintillation crystals and the response of PMT devices [6].

Page 33: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

20

Figure 2-13 shows the different responses of PMTs and scintillator crystals (Figure 2-18 shows the

SPM detector response). The best response of a detector would be where the peak detector response

matches the peak crystal emission, however, this is rarely the case for many crystal-detector

combinations due to the range of scintillator wavelengths. The emission spectrum of light emission

from scintillator crystals and the spectrum that detectors are sensitive to, allows detection of light from

scintillator crystals using detectors which are not perfectly matched (albeit with lower sensitivity).

Lutetium Yttrium Silicon Oxyorthosilicate (LYSO) is a scintillator crystal which was chosen for its

very fast decay time and high light yield. However, Lutetium is itself radioactive decaying by β-

emission, causing a level of background radiation in all measurements [14]. Using values from [14]

and scaling these values for the scintillator size used in experiments, the intrinsic activity of the

scintillator due to lutetium is about 13.5kBq, whilst the sources used at the energy of interest are much

more active at approximately 370kBq.

Activation of a scintillator using Thallium is common for sodium and cesium iodide which adds sites in

the scintillator that can produce optical photons in the forbidden band [6]. This increases the

probability that de-excitation of an optical state due to radiation will lead to optical photons (Figure

2-14). Having a fast, bright scintillator crystal which has a high attenuation of radiation photons over a

large energy range (detection efficiency) with a peak emission wavelength of the detector used, is the

ideal scintillator crystal. As this doesn't exit, a compromise of detection properties is usually made.

Figure 2-14: The scintillation process from activated states [6].

The light yield of a scintillator material arises as more charges are successfully created from the

incident radiation. By having a linear light output the amount of incident radiation can be determined

to deduce the energy of the incident radiation. The rise time varies in a scintillator crystal is related to

Page 34: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

21

the mobility of the charges in the material [15, 16] where more prompt decay times are due to a better

charge mobility in the material.

[16] describes the process of fluorescence in a scintillator crystals. The absorption of radiation causes

electron-hole pairs to be created in the crystal at between two and seven times the band gap energy.

Electrons scatter to reduce in energy until they can excite luminescent centres which produce optical

photons. Due to the incident radiation, the scintillator emits photons in the visible part of the

electromagnetic spectrum (~400 – 700nm). The light produced from a scintillator can be recorded by a

photomultiplier, a photodiode, an avalanche photodiode and a new detector technology, the SPM.

The Silicon Photomultiplier (SPM)

The high gain, low noise and availability in large sizes are some of the reasons making the PMT system

a popular detector for scintillation light. There are however, several disadvantages to the technology;

fragility of the glass, large power requirements and sensitivity to magnetic fields [17]. For portable

radiation detectors, the large sizes of the PMTs can be a disadvantage.

Figure 2-15: A photomultiplier tube with the amplification stages in the dynodes [3].

Anode

Dynodes

Scintillator

Photocathode

Photomultiplier

tube

Page 35: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

22

Photodiodes have been used as an alternative to PMTs to detect optical photons from scintillator

crystals, which enter the photodiode material and create electron-hole pairs. Photodiode gain is not as

large as the PMT, so a better alternative to the PMT is the avalanche photodiode (APD) which can

provide a much larger gain (several hundred is possible [6]) than a standard photodiode. A rapid

increase in the number of electrons is created due to further collisions and the multiplicative process

occurs due to the presence of a large electric field [6]. The gain is still a limiting factor and alternative

methods to readout from scintillator crystals are being explored.

Figure 2-16: A single pixel SPM (on a square base approximately 3x3cm) [20].

SPMs (Figure 2-16) combine the features of PMTs and APDs providing a high gain (~10

6) whilst

requiring a low operating voltage [18] making portability a possibility. Geiger mode Avalanche

Photodiodes are (GAPDs) are operated with a bias above the breakdown voltage such that only one

carrier is required for „breakdown‟. This breakdown is stopped or „quenched‟ by a large resistance in

series with the GAPDs. By connecting the output of thousands of GAPDs together in parallel, a photon

flux causes a current which is “directly proportional to the number of incident photons” [17]. Each

GAPD is a „microcell‟, any many are tightly packed such that there are thousands in a single pixel.

Page 36: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

23

Figure 2-17: A 1mm2 SPM pixel with many microcells on the top (left) [17] and the GAPDs in the

pixel are connected together to provide a photon proportional output [17] (right).

When coupled to a scintillator crystal, the SPM detects the optical photons produced by radiation

incident on the scintillator. SPMs have intrinsic characteristics causing the loss of photons, including

the active area (the area able to detect a photon interaction), photon detection efficiency (PDE) and fill

factor (the ratio of the active area to the total area) which reduces the possible energy resolution as

fewer photons are present. The PDE (Equation 17) is the product of quantum efficiency (QE), the

probability an avalanche breakdown occurs (PAB), and the fill factor (FF). According to [19], the PDE

is “the probability that an incident photon produces a Geiger pulse from one of the microcells”, and

encompasses the probability of the photon initialising the avalanche breakdown and the quantum

efficiency of the detector.

Figure 2-18: A graph containing the PDEs for 4V over bias for various SPM products [19].

Equation 17 FF*PAB*QEPDE [22]

Page 37: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

24

With the number of photons remaining after the losses caused by the scintillator and SPM system, the

energy resolution (R) can be determined using Equation 18 and Equation 19.

Equation 18 available photons No.

FWHM R

Equation 19 available photons No.

available photons No.*35.2 R

It is known [3, 6] that the energy resolution improves (by reducing) with increasing energy (E)

provided by Equation 20, where K is a constant of proportionality.

Equation 20 E

K R

By taking logarithms of both sides, Equation 21 is produced, which should produce a straight line and

finding the constant K allows the statistical broadening of the peak to be found [6].

Equation 21 )0.5ln(E-ln(K)R)ln(

A pre-amplification („preamp‟) board is connected to the SPM which increases the signal to hundreds

of mV, with the brightest scintillators predicted to produce (negative) SPM pulses of the order of volts

out. [22] states that the dead time of the detector to be 0.1µs, leading to a maximum count rate of

10MHz. Equation 22 shows how SPM gain is linear with the bias voltage (Vo) after breakdown

voltage (Vbr) (the electron charge (e) and the capacitance of the system (C) are constants). [22] also

shows how the dynamic range (the maximum number of simultaneous photons which can be detected

is limited to the number of microcells) increases with the number of microcells. As the microcell size

reduces, the gain decreases, so there is an important trade-off between the gain and microcell size.

Equation 22 e

brV

oVC

G)(

Page 38: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

25

Two of the many possible areas where the SPMs have been proposed for use includes; [17]

Medical: PET scanning is dependant on fast coincidence timing which the SPM

should be capable of;

Portable security applications: such as portable radiation detectors due to the small

size and low power requirements.

Silicon has a low Z (14) which has a lower stopping power for X- and γ-ray photons than CZT (Figure

2-19). However, in SPMs, radiation detection is indirect, as optical photons come from the scintillator

crystals, not directly from the radiation photons. Scintillator materials have a good stopping ability of

low energy photons (especially 0-100keV). The noise in an SPM takes the form of „dark photons‟,

which are thermally generated (there is the option of cooling the SPM to reduce these). [33] shows that

the dark rate is 9 times higher in the 3mm SPM than the 1mm SPM and therefore the individual dark

photon pulses cannot be as clearly distinguished. It also states that the 1mm SPM has a faster onset

and decay time than the 3mm SPM. [23] describes how optical cross talk causes distinct „levels‟ of

noise present. This is where a photon is recorded across two or three microcells, with the probability

decreasing with the increased number of microcells. The onset time and dark photons were

investigated in the project.

Figure 2-19: The attenuation ratio I/I0 for the three detector materials explored for the project.

Using attenuation data collected from XCOM, Figure 2-19 shows how CZT attenuates far more

photons over the energy range and should therefore have the better detection efficiency of all the

detectors explored. (An I/Io value of 1 indicates that all of the photons pass through the material

Page 39: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

26

unattenuated). Silicon as a direct radiation detector is seen to be very inefficient at energies greater

than 100keV.

When comparing the efficiency of the CZT to the scintillator materials (in the thicknesses used for the

detector experiments) it is seen that the detection efficiency of CZT is comparable to the scintillator

crystals.

Figure 2-20: The attenuation ratio over the energy range to be tested for the four scintillator

crystals used at 3mm thickness and the CZT detector at 5mm for direct comparison.

2.4 Radiation Measurement and Spectroscopy

What Happens to the Detected Radiation?

Imaging is possible when more than one pixel is present, so differences in the radiation detected over

the pixels can be seen. At the detector, the radiation interaction of the photon in the material is

analysed by two common methods;

Photon counting is used for low numbers of photons where a background level is set and all

counts above this are amplified and recorded.

Charge integration is used where there is a much higher flux of photons. The number of

counts incident on each pixel can be found with knowledge of the time each pixel was

integrating for, and the current for that integration time per pixel. The electron-hole pair

creation energy (W) for the detector material is known and therefore the number of counts per

Page 40: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

27

pixel can be found. With this information, an image can be built up from the amount of

counts incident on each pixel in the detector by assigning a greyscale which typically lightens

with a higher number of counts in a pixel. The charge deposited is found by integrating the

current pulses produced in a given time according to Equation 23.

Equation 23

t

IdtQ0

Pulse Height Spectra

The energy spectrum from the radiation is typically displayed using a Multi Channel Analyser (MCA).

Here the channels (which are related to the energy of the radiation according to Equation 24) are

displayed on the x axis, and the number of counts at this channel on the y axis. In this way, a

distribution of counts over the number of available channels is produced, resulting in a pulse height

spectrum or simply „spectrum‟. The location of the peak determines the its energy, which can be used

to identify the material from the energy value. For example, a peak at 59.5keV would show that 241

Am

is present.

Figure 2-21: Operation of an MCA, an extension of many single channel analysers (SCAs) [6].

A pulse height spectrum (Figure 2-22) shows features such as the Compton edge, the peak centroid (or

full energy peak where the counts corresponding to the main γ-energy are), and escape peaks.

Page 41: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

28

Figure 2-22: Typical pulse height spectra from γ-sources [6].

Detector Calibration

When spectroscopic data is collected, the x axis is usually a channel which is not directly equivalent to

an energy. The y axis displays the number of counts received in each channel. By irradiating the

detector with calibration sources (with peaks at known γ-energies) the channels corresponding to the

peaks can be identified with energies according to Equation 24.

Equation 24 Energy = Calibration constant * Channel Number

At least two peaks are required so an accurate calibration constant for the detector can be calculated.

More points will provide a more accurate peak value when an unknown source is used.

How is Spectroscopic Capability Determined?

Energy resolution is the ability to distinguish between two close energy peaks seen in the energy

spectrum as separate. This is a very important basis for comparison between spectroscopic detectors,

and is found using Equation 27. Spectroscopy is possible when the energy resolution, commonly

expressed as a percentage at a given energy, is low enough so the separation of close energy peaks can

be identified. The following three equations show how the energy resolution is calculated to identify

the possibility for spectroscopy from the counts in the energy peak, where σ is the standard deviation

(and according to Poisson statistics is also the error) and FWHM is the Full Width at Half Maximum.

Page 42: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

29

Equation 25 counts of No.

Equation 26 *35.2FWHM

Equation 27 Centroid

35.2

Centroid

FWHMResolutionEnergy

Figure 2-23: An improved energy resolution is obtained with a thinner peak [6].

Ideally, a single line would be produced corresponding to the photon's γ-ray energy, and this is

approached with some very high resolution detectors, however, according to [6], the peak broadens

(Figure 2-23) due to statistical fluctuations in how the pulse is recorded, as there are inherent

fluctuations in the number carriers produced each time. For detectors with the same efficiency, the

better energy resolution will be taller (the efficiency is not dependant on the energy resolution). As

more charge carriers are produced, the error becomes less significant, so at higher energies, the energy

resolution is improved (the peak is less broad) as, for an equivalent number of photons, more charge

carriers are produced from higher energy photons, reducing the peak broadening.

The radiation peak produced is assumed to follow a Gaussian distribution (Equation 28), where the

parameters for A, B, and C correspond to the peak height, peak centroid and standard deviation

respectively.

Equation 28

2)(

2

1exp

C

BxAy

Page 43: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

30

The area of Gaussian peaks (Equation 29) provides the number of counts in them (to be used with

Equation 28) to find the detector‟s efficiency.

Equation 29 2πAArea

Other properties of a radiation detector determine its effectiveness:

Intrinsic efficiency of a detector (εint) (Equation 30) is a measure of how many photon

interactions the detector has recorded from the number of photons which can interact with the

detector, based on the area the detector occupies.

Equation 30 detectoron incident pulses No.

recorded pulses No.int

Figure 2-24 shows how the radioactive sources radiate in a sphere, where a detector usually occupies a

small area of this. The further back the source is from the detector, the smaller the area of the sphere

incident on the detector, where fewer counts are expected to reach the detector.

Figure 2-24: The source emitting in 4π, where only a proportion of the activity is incident on the

detector at some distance (d) away.

The sources used are at different activities, and as for most sources, the γ-decays of interest for

spectroscopy (59.5, 122, 511 and 662 keV) do not account for the whole source activity (% determined

from the decay scheme). The activity (and therefore the number of counts received by a detector)

reduces as the source is moved further away from a detector according to the inverse square law as

Page 44: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

31

seen in Equation 31, where C is the number of counts, K is a constant of proportionality for the detector

and d is the source to detector distance.

Equation 31 2d

KC

Once a number of counts recorded at a given distance is known, it is possible to determine how many

counts are expected at another source to detector distance by determining K. By using this method for

several points, the change in counts with distance to the source can be identified.

Saturation is the point at which no new counts can be detected due to a maximum number of

counts a detector can record being reached. It is an important quantity to know so the sources

which are used are kept below this level.

Dead time is the time that the detector cannot detect radiation (or count) due to the previous

counts being recorded and processed. This is equal to the difference between live time and

real time where the live time is the time the detector has actually been counting for, whilst the

real time is the time elapsed since the starting the experiment. Reducing the number of counts

(placing the source further from the detector) will be reduce the dead time.

Uniformity (Equation 32) is a measure of how similar the recorded radiation is over all the

pixels in a detector for the same incident radiation. σ is the standard deviation and centroid is

the peak energy.

Equation 32 Centroid

σUniformity

Gain is the amplification of a signal to utilise the full dynamic range of the detector. By

increasing the gain, the noise is also amplified, and therefore the level of the gain needs to be

carefully checked to ensure that any benefit in signal is not at the expense of more noise. The

signal to noise ratio (SNR) was determined for the SPM detector using Equation 33 at two

different bias voltages.

Equation 33 Size Noise

Size Pulse SNR

Page 45: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

32

Error Analysis

The standard error in the mean (sm) (Equation 34) was commonly used to provide an uncertainty

estimate where multiple measurements were taken, where σn-1 is the standard deviation and n is the

number of readings or samples.

Equation 34 n

s nm

1

Various techniques of error analysis were performed over the course of the project in order to give

results with an estimate of the associated error. Applying Poisson statistics allows the error in a

number of counts to be recorded simply as the square root of the value as seen in Equation 35,

therefore at larger numbers of counts the statistical error in the count rate is lower.

Equation 35 counts No.Error

Background radiation is a random factor adding to the uncertainty. It is naturally occurring, however,

in a radiation laboratory, levels of background are likely to be higher and therefore needed to be

accounted for by talking measurements without the additional radiation used in the experiments.

[6] describes the propagation of errors for a function (u) dependant on three variables (x, y and z).

Equation 36 2

2

2

2

2

2

2

zyxuz

u

y

u

x

u

When appropriate, a confidence level was applied to a measured value, the most common being 95%.

The coverage factor (k) is used to give the confidence in a measurement and commonly the intervals

used are k=2 (95% confidence) and k=3 (99% confidence) [24].

There are an almost infinite number of possible sources of uncertainty when taking a measurement

including the temperature, humidity, the time of the reading, the ability of the measurer, ambient light,

presence of drafts, fields and the list goes on. Therefore any uncertainty applied to a value is only an

estimate of the error, and it is almost impossible to get a perfect measure of the uncertainty.

Page 46: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

33

Chapter 3 : Experiments and Work Conducted Many programs, models and practical experiments were carried out to meet to aims of quantifying

spectroscopic and imaging performance of different radiation detector systems.

3.1 Calculations and Modelling

Radioactive Source Calculator

A spreadsheet was created to calculate the current activity of the radioactive sources used for the

experiments based on a known activity at some point in the past. By entering the half life of the

source, and the activity at a known date, the current activity was calculated. Using Equation 1 for each

source, the activity at any time was determined. By extrapolating the date over time from the current

activity, the future activity was calculated by Equation 2.

This was a useful spreadsheet to create, which was used several times over the project to determine

which and when sources need replacing, and the effect of diminishing activity over the measurement

period was determined.

CZT Shielding Requirements

The thickness of shielding required for the CZT detector housing needed to be calculated so the

detector could be used as a pinhole imager with a pinhole (at 10cm) away which can be added and

removed as required. All of the X-ray photons other than those from the pinhole needed to be

attenuated. The thickness of lead to do this was calculated using Equation 9. With values of „total

attenuation with coherent scattering‟, from XCOM, which varies with different materials and energies.

A spreadsheet was produced calculating the material thicknesses required (based on the absorption of

different materials) to reduce the photon flux to a desired amount. 5mm of lead would sufficiently

attenuate the X-rays, however, to reduce the fluorescence lines from the lead, the thickness of tin was

found, and finally the thickness of copper was found to reduce the fluorescence lines from the tin. The

exponentially decreasing intensity (I) was calculated as the material thickness (x) increases, reducing

the initial number of photons (I0) due to the type of material defined by the absorption coefficient (µ).

Page 47: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

34

3.2 CZT Experiments

The CZT Detector

Figure 3-1: The CZT detector array.

The CZT detector array (Figure 3-1) has a pixel size of 1.5mm on a 1.6mm pitch which is 5mm thick.

An electric field is produced with a 500V bias, resulting in an electric field of 105Vm

-1. With electron

mobility (µe) in CZT of 1000 cm2/Vs [25] this leads to a υe of 10

6ms

-1. According to the CZT detector

manufacturer, the detector can record at 100kcps, which corresponds to 32 cps / pixel, however, the

system could equally handle all 100kcps in one pixel if all the others had zero counts [26].

Uniformity and Energy Resolution

The CZT detector was irradiated with

241Am and

57Co γ-sources to determine the uniformity and energy

resolution of the detector. The purpose of these experiments was to identify any improvement (or

deterioration) in the CZT detector performance in terms of energy resolution and uniformity, after an

upgrade, to determine whether to fully populate the detector with the upgraded material.

Page 48: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

35

Figure 3-2: The experimental setup for the uniformity and energy resolution measurements.

The sources were in turn placed 10cm from the centre of the detector, integrations of the flood

illumination were taken over 10 and 30 minutes for each source as seen in Figure 3-2. The detector

records energy spectrum information per pixel. The analysis required code to be written in IDL (a

programming language similar to FORTRAN 90) to obtain a greyscale image of the pixel intensity. By

producing histograms (such as Figure 3-3) from the data producing this image, an inbuilt Gaussian fit

function was applied to provide values for standard deviation (C) and centroid (B). Using Equation 32,

the uniformity was calculated for both sources, to provide the spread of this data. „Binning‟ is a useful

tool, especially in with low count levels, which allows the available number of channels to be

combined such that a count with a small accepted range of energies is added to only one channel.

Figure 3-3: An image (based on detector area ~12.5cm2) from a γ-source illumination (left) and a

histogram from which the detector uniformity can be calculated (right).

Page 49: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

36

Code was then written to correct for the artefacts and discontinuities at the CZT module boundaries by

„flat fielding‟ which uses the data taken over a large time integration (30 minutes which is much longer

than data would be collected for) to provide a data set high statistical quality data. By programming an

iterative loop to calculate the average number of counts received per pixel for the 30 minute data

(performed for each source in turn), a factor was found giving the difference from the average number

of counts per pixel. This factor was multiplied to each pixel for any other image captured with the

detector (this was initially applied to the 10 minute source flood illumination data to produce flat

fielded images). A „flat‟ (uniform number of counts over the image) image should then be produced.

Flat fields (using both calibration sources) were available to correct for other artefacts in images

collected by the detector.

Importantly, using the flood illumination data, the energy resolution was found per pixel (for each

source separately), using Equation 27. The data was read out into a spreadsheet, where statistical

analysis on the energy resolution could take place (principally the average energy resolution per pixel,

standard deviation and error) to determine the spread of the data over the pixels.

The detector was then returned to its manufacturer to have all 25 modules (5x5 array with 6400 pixels)

of CZT replaced with 12 modules (3x4 array with 3072 pixels) of higher quality CZT (the exact details

are unknown but it is assumed better leakage and uniformity across the material). Following the

detector hardware upgrade, the CZT detector was again illuminated using the previous method with the

same calibration sources in turn at 10cm from the centre of the face of the detector. Again, 10 and 30

minute integrations were taken to check the functionality of the flat field program. All of the previous

IDL programs were modified to account for the change in size of the array.

It was hoped that the CZT detector will be an improvement to the imaging quality of the ICCD

detector, and as spectroscopic information can be obtained from the detector (the CZT records the

charge and the energy per pixel) a much more compact and useful system will then be available.

Page 50: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

37

CZT Spectroscopy

The spectroscopic capability of the detector was tested in experiments with the pinhole cover and

housing removed (to use the full area) using the 241

Am and 57

Co sources both individually and

separately. The time taken and the type of γ-spectrum which can be obtained from the CZT detector

was determined.

Using the

241Am and

57Co sources at 10cm and 12cm from the centre of the detector respectively, the

simultaneous energy spectrum (and counts recorded by the detector at each energy) were recorded.

The sources were then aligned at 15cm from the detector and were incrementally moved further from

the detector (every 5cm from 15cm to 30cm) and the energy spectrum was recorded in increasing time

increments from 1 to 10 seconds at each distance. This experiment was designed to quantify the

maximum number of counts received by the detector over the energy range to identify if the detector

counts linearly over time and that the inverse square law is followed.

CZT Efficiency Measurements

241

Am and 57

Co sources were placed separately at three distances (0.5, 1 and 1.5m) from the CZT

detector, without the housing and pinhole, and 10 minute integrations for each source at each distance

were taken. To determine the intrinsic efficiency (Equation 30), the number of counts in the energy

peak is required, which for the CZT detector was determined by calculation (Equation 29) using the

height and standard deviation values from the fitted Gaussian peak. (The energy peak is assumed to be

a Gaussian, where the fit is a good match to the data as seen in Figure 3-4).

Figure 3-4: The Gaussian fit to the 241

Am peak (at 0.5m for 10 minutes).

Energy (keV)

Sum

med

counts

Page 51: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

38

From the activity incident on the detector (based on the distance to the detector and source), the portion

of the sphere occupied by the detector created by the source, and the activity at the energy peak of

interest (59.5 and 122keV for the CZT detector) the detection efficiency was determined.

The ICCD Detector

The Intensified Charge Coupled Device (ICCD) detector has a thin (2.5mm thick) pixellated CsI(Tl)

crystal array converting X- and γ-ray photons into optical photons.

Figure 3-5: The ICCD detector (with a 2mm pinhole plate attached).

Removing the cover and pinhole plate exposes the front of the detector, which allowed the detector

area to be flooded with the same two calibration sources used for the CZT detector, in the same

experimental arrangement. Software used with the detector allows use of the previously acquired data

for flat field corrections. For a fair comparison to the CZT for the detector uniformity, flat field

measurements were made using the ICCD detector.

CZT and ICCD X-ray Backscatter Pinhole Imaging

The use of the CZT detector as an imager using a newly designed graded shielding was tested and

compared to the images obtained using an ICCD detector in the same experimental configuration.

Page 52: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

39

Figure 3-6: The housing applied to the CZT detector providing a pinhole aperture at 10cm from

the centre of the detector.

Figure 3-7: Top view of the arrangement for X-ray backscatter pinhole imaging.

By facing an X-ray generator towards a target material and irradiating it with 160keVp (peak energy of

160keV) X-rays for 10 seconds, the CZT detector (with the 2mm pinhole) received backscattered

photons from the target and surroundings.

Figure 3-8: An image created using a pinhole receiving backscattered X-rays (left) and the effect

chamfering has on the pinhole (right-above before and right-below after chamfering).

CZT detector

(with pinhole)

Target

material

X-ray

generator

Page 53: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

40

Figure 3-8 (right) shows that the area covered by the pinhole (exaggerated) which was not initially the

entire active area of the detector (in Figure 3-8 (left) black indicates where there are fewer photon

interactions due to the collimation effect). It was therefore necessary to chamfer behind the pinhole by

calculating the angle required to allow the whole detector area to be used.

With the chamfering successful, a background measurement with the pinhole covered and

backscattered X-rays being incident on the detector was made. This found that X-rays were leaking in

to the detector, causing the image shown in Figure 3-9.

Figure 3-9: The leakage obtained before the additional shielding (detector area ~7.9x10-3

m2).

Before any backscatter images were taken, this leakage needed to be eliminated to avoid the additional

effect on the measurements. The best method for additionally shielding the detector was established by

applying lead shielding all around the detector (including the pinhole) and sequentially removing the

lead in sections to identify where the leak was coming from, by measuring the average number of

counts per second in the imaging section of the CZT software for short (10-60 second) integrations. It

was found that the optimum arrangement (using the least additional lead to produce a uniform output)

of lead shielding requires blocks between the X-ray generator and the side of the CZT detector, and

additional blocks at the back of the detector to reduce X-rays getting in this way, as seen in Figure

3-10.

Page 54: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

41

Figure 3-10: The optimum shielding configuration (left) and the more uniformly distributed

counts with this shielding.

After warming up the X-ray generator (required to recondition the vacuum after a period of inactivity),

various objects were illuminated with 160keVp X-rays from a continuous X-ray generator for 10

seconds. The detector recorded all the time X-rays were being produced. The objects were placed

55cm from X-ray generator and 30cm from the pinhole aperture of the detector. These distances and

times were determined after looking at several settings and integration times, so the output images were

centred, a good size in proportion to the background and not saturated. The same experimental

arrangement (integration times, pinhole, distances and energies) was setup for the ICCD detector and

images were taken using the 2mm pinhole, once removed from the CZT detector housing.

Images were taken of several objects with different atomic numbers (Z) and densities (including salt,

sugar, cotton, talc powder, aluminium powder and tungsten) which were on their own, and then in a

typical luggage case. To elevate the objects that were not in a suitcase (for differentiation between the

floor and the object and to reduce scatter from the floor), the objects were placed on a plastic block

5cm high when out of case measurements were made for a direct comparison between the images. The

ICCD images had a „best fit‟ function for the brightness and contrast applied using Image Pro Plus

software. The CZT data was processed with IDL code to display the images. The flat field corrections

acquired prior to these measurements for both the CZT and ICCD detectors, (based on the 30 minute

57Co illumination) were used to „clean-up‟ the images by removing the artefacts caused by the detector.

Page 55: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

42

CZT and ICCD Angular Resolution

This experiment was conducted to provide another basis of comparison between the ICCD and CZT

detectors. A tube of salt and a tube of sugar (arbitrary objects 7.5cm long (up to the lid) with a

diameter of 2.5cm), were placed at 40cm from the both detectors, and the X-ray generator was placed

at 60cm from the objects to be imaged. Additional shielding was again added around the CZT detector.

The 2mm pinhole was used for both detectors, and as there is only one pinhole, the experiments were

run first for the CZT and then for the ICCD using the same experimental setup.

Figure 3-11: The experimental arrangement for separation experiment (CZT pinhole covered for

a background measurement).

The salt tube was placed on one circle and was fixed there, and the sugar was placed in decreasing

distances in increments of 1cm closer to the salt tube. At distances of 1 and 2cm from the tube of salt,

the sugar was placed further back from the salt (Figure 3-12) else both tubes would have overlapped.

Figure 3-12: The positions of the tubes of salt and sugar, moving closer together in 1cm steps.

CZT Detector

ICCD

Detector X-ray

Generator

Objects

Fixed

salt position 3 4 5

1 2

Page 56: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

43

Backscatter X-ray images were taken for 10 seconds using 160keVp X-rays. The images from the CZT

detector were then processed using IDL code to flat field each image using the 30 minute 57

Co data.

The ICCD images were also flat fielded using a pre-existing function in the image processing software,

also based on the 30 minute 57

Co illumination for the ICCD detector.

Using IDL code for each flat fielded image from each detector, each image was modified so a line of

data (or a profile), of counts (CZT) and intensity (ICCD) was stored to a file, recording the change

through the objects and background. (This line was identified by setting the pixel values along a line to

an arbitrary number, which was returned to its original value when the data was read out).

Figure 3-13: The line of data extracted for ICCD images (left) and CZT images (right).

This data was normalised for each image, and plots of the intensity with pixel value were made

separately for each image, combined for individual detectors and combined for both detectors to

numerically quantify how the images vary for each detector records over each image.

3.3 SPM Experiments

Scintillator and SPM Energy Resolution Model

A spreadsheet model was created to estimate the number of photons reaching each SPM and then being

lost, to find the energy resolutions possible using Equation 18 and Equation 19 over the energy range.

Starting from an initial energy deposited to each scintillator through the four calibration sources, this

study was performed to indicate if spectroscopy is theoretically possible from the energy resolutions

Page 57: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

44

obtained. For example, an energy resolution of 20% at 662keV would allow another peak to be

resolved below 530 and above 795keV. According to [28] values for the scintillator efficiency are

between 3 and 15%. The justification for using scintillator efficiencies for each of the scintillators of

5% comes from the fact that not all of this light produced will be transferred to the SPM (coupling

losses between the scintillator and SPM with an imperfect match) which will reduce this efficiency.

Numbers corresponding to each scintillator material and SPM are chosen by the user, and the

spreadsheet automatically looks up the values for light yield, active areas, and photon detection

efficiency from product manufacturer data. The number of photons remaining after each deduction can

be seen for each scintillator and SPM combination over the energy range, resulting in the overall

expected system efficiency and the expected energy resolution.

Multiplying the PDE (obtained from Figure 2-18) at each wavelength by the scintillator by the active

area (which varies between SPMs) gives a number of photons, which was used to find the energy

resolution for that combination. According to the manufacturer [27], Figure 2-18 could be used to

determine the PDEs for all of the SPMs by multiplying by a constant (1.5) from the 20µm PDE values

to find the PDE for the 35µm SPM. These values were used in the model to calculate the energy

resolution for each combination of scintillator, crystal and SPM when used with each source. The PDE

technical note was released in August 2007 showing the recent development in the technology. Plots

based on the model provided the expected energy resolution as a function of energy for each of the

scintillator crystals with each SPM. These were compared to the measured spectra produced later on.

The spreadsheet can easily be upgraded with new values or materials.

Saturation of the SPM occurs when all of the microcells simultaneously detect photons, and there are

1144 microcells on the 1mm 20μm SPM, and 3640 and 8640 in the 3mm 35μm and 20μm SPM [37].

Therefore the SPM with the highest PDE is the 3mm 35μm SPM, which will however saturate before

the 3mm 20μm SPM.

Page 58: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

45

Light Emitting Diode (LED) testing

Before the scintillator crystals arrived, experiments utilising LEDs were designed and built to simulate

light flashes. This series of experiments was used to identify the speed of the SPM response from an

off to an on state, and the size of the SPM response pulses for a given LED light pulse incident on the

SPM. Astable multivibrator (pulsing) circuits were setup using various combinations of components

(Figure 3-14) providing a selection of duty cycles, frequencies and durations to cause an LED to flash.

Figure 3-14: The transistor astable circuits produced to pulse an LED [30] (left) and a circuit

diagram for the 555 IC used to pulse an LED [32] (right).

The 555 Integrated Circuit (IC) was used to pulse an LED for the SPM experiments. Usually a 555

operates with a duty cycle (the amount of time the unit is on) of over 50%. Using a diode in parallel

with the R2 resistance allowed the duty cycle to be less than 50%, and the „on‟ time is then related to

R1*C, and the „off‟ time by R2*C. These were used to find values of resistors and capacitors to keep

the LED pulse on and off for the desired times (between 0.5 and 20µs).

Figure 3-15: The oscillator circuits produced to pulse an LED using transistors and capacitors

(left) and the pre-packaged oscillator in the IC NE555 timer (right).

Page 59: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

46

A 555 Integrated Circuit (IC) was used for the majority of the measurements with the timer output

pulsing an LED requiring fewer components to set the frequency than Figure 3-15 (left). An

oscilloscope placed in parallel with the LED, was used to capture the pulses the LED was receiving to

confirm the frequency, duration and size. The timer is capable of microsecond pulsing [31] and the

LED was then pulsed with the maximum repetition frequency of the 555 (on for 0.5μs).

The tests were initially conducted on the 1mm SPM at a bias of 30V. The LED was placed on top of

the SPM in contact with the glass, and secured there. The equipment was all housed in an aluminium

box and covered with black cloth to prevent as much extraneous light as possible reaching the SPM.

The rise time of the SPMs was found by using LED pulses incident on the SPM to determine the time

taken for the SPM pulse voltage to go from off to on (onset time). The SPM was connected to a

channel of the oscilloscope and another channel of the oscilloscope was connected in parallel with the

LED to allow the LED pulse and SPM response to be recorded simultaneously. Various amplitudes

and frequencies of light pulses were sent to the LED and the SPM responses were captured using a

printer connected to the oscilloscope (Lecroy 334AM). This was repeated for various frequencies for

each SPM.

SPM Pulse Linearity

Tests were again conducted on the 1mm SPM at a bias of 30V, to determine how the SPM responds to

the varying power of an LED (changing the number of photons produced). It is expected that the

parallel arrangement of the SPM microcells will provide a proportional output to the amount of

incident light. A variable resistor (potentiometer) was placed in series with the output of the 555 timer

circuit and a red 3mm LED (which was in contact with the SPM) to adjust the power to the LED. An

ohmmeter was placed across the resistor to determine its resistance, whilst simultaneously the current

through the LED was determined by placing an ammeter in series with the LED.

Readings were taken approximately every 10Ω, from 0 to 70Ω, however, it was extremely difficult to

get 10Ω increments accurately, due to a very small adjustment on the potentiometer changing the

Page 60: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

47

resistance by several ohms. Repeat measurements were therefore not possible using the same values of

resistance, however, by taking many measurements over the range from the LED on to off (around 10

measurements) a graphical fit can be obtained. The data collected (the current and voltage for each

resistance) was used to calculate the power going through the LED, so comparisons between the

resistance and LED voltage, resistance and SPM voltage, and power and SPM voltage can be made.

Figure 3-16: Experimental arrangement for the SPM pulse linearity experiment.

SPM Pulse Observations

To activate the scintillator crystals 57

Co, 22

Na and 137

Cs sources at high activity with different γ-

energies were selected and procured, providing peak γ-energies from 59.5 to 662keV which were used

with an existing 241

Am source.

Source Peak γ-Energy

(keV)

Activity (kBq)

(1/10/07)

Half life % of Source

Intensity

241Am* 59.5 465 432.2 yr 35.75

57Co 122 362 270 days 85.64

22Na 511 343 2.6 yr 180.00

137Cs 662 410 30.174 yr 85.05

57Co (original)* 122 30 270 days 85.64

Table 3-1: Procured calibration source details. (*Used for CZT detector uniformity and energy

resolution experiments and were original laboratory sources.)

Page 61: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

48

Four scintillator crystals (CsI(Tl), BGO, LYSO and CdWO4) which produce light at different

wavelengths with different light yields and decay times were supplied as 3x3x3mm cubes to couple to

the SPMs. These cubes were tightly wrapped with reflective material (Tyvek paper) around 5 of the

sides which was held in place with „plumbers tape‟, to ensure as much light as possible leaves through

the exposed aperture to the SPM. The free side was then coupled to the SPM surface using a tiny

amount of silicone grease (just enough to cover the exposed side of the scintillator) allowing

conduction of the light from the scintillator to the SPM, and providing enough friction to hold the

crystal on top of the SPM. The scintillator and SPM combination was then placed in the aluminium

box. Power and output to an oscilloscope were connected to the SPM through a small hole in this box,

and each γ-source was placed in turn at approximately 2.5cm from the scintillator crystal. The box was

then closed and covered with a black cloth (in the absence of a dark room). This setup can be seen in

Figure 3-17. The source caused the SPM to record pulses from the scintillator crystal light, the voltage

pulse size and decay times from the scintillator were found for each combination and directly printed

from the oscilloscope once a pulse had triggered (trigger set above the noise). To prevent radiation

damage to the crystals, sources were placed only placed near the crystals when necessary. The

printouts were used to find the decay time and size of pulses so the data acquisition system could be

setup to „look‟ for decay times and voltages of the pulses from each experimental combination. These

values for rise time and pulse height were entered to a spreadsheet, where the differences between the

sizes of the pulses from the different SPMs and crystals were quantified.

Using CsI(Tl) with 22

Na in the experimental arrangement above, the effect of the increase in the SPM

bias voltage from 30 to 32V was determined by exploring the sizes of the scintillation pulses printed

from an oscilloscope. Noise pulses were found by removing the source and printing the SPM

responses. The sizes of the pulses were read from the oscilloscope printouts for both the noise (dark

photons) and for the scintillation photons, and a comparison plot of the SNR (Equation 33) for all three

SPMs at the two bias voltages was produced.

Page 62: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

49

Figure 3-17: The pulse exploration and SPM spectroscopy experimental arrangement.

SPM Spectroscopy

Using a similar experimental setup for the pulse measurements, the wrapped scintillator cubes were

coupled to the SPMs using silicone grease (RS 494-124). The SPM output was then connected to a

DAQ system (Figure 3-18), which records and displays the energy spectrum produced from the SPM.

Parameters on the DAQ were set based on the decay of the crystals found previously and all settings

needed to be very carefully adjusted by collecting spectra for about 10 minutes using a range of rise

times and decay times to provide the best energy resolution. Once correctly setup, the spectra were

collected for 30 minutes for each radioactive source, and these measurements were repeated again to

allow for the standard error in the mean (Equation 34) to be applied to the energy resolutions measured.

Following these measurements, the source was changed such that all four sources were used on the

same scintillator crystal and SPM. The system was then setup again using the second of the four

scintillator crystals, and the four radioactive sources used in turn to activate them. The measurements

using this crystal were then repeated to allow an average and error to be applied. Finally, the SPM was

tested without any sources present, to quantify the background.

Power

SPM Wrapped

crystal

Source

Output

Page 63: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

50

Figure 3-18: The Xia Pixie-4 system used to acquire the spectrum from the SPM detectors.

The setup was then disassembled and the experiments repeated for the next SPM where all of the

measurements were repeated, and then finally for the third SPM. A summary of all the measurements

planned is included in Figure 3-19.

Figure 3-19: A summary of the planned SPM testing, showing each SPM coupled to each

scintillator crystal activated by each γ-source, and the background measurements.

For consistency, the same set of scintillator crystals (cubes 3x3x3mm) was used for each of the SPM

experiments even though there was a mismatch in sizes when using the 1mm SPM. For best light

transfer, the area of the detector and scintillator should match. It was initially estimated that 1 in 9 of

the photons would be detected due to the difference in scintillator area to the SPM area.

For 1mm spectra, another MCA channel (without a 7.5x attenuation) was used with a lower SPM bias

voltage (30V compared to 32V for the 3mm SPMs), and a higher threshold was set to eliminate the

noise pulses. MCA events seen in the MHz region (due to noise) were reduced to ~10Hz with the

correct threshold set, indicating that only scintillation events are recorded.

Page 64: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

51

SPM Detector Efficiency

Previously (Figure 2-20) it was seen that the ability of the scintillator crystals to attenuate at 59.5keV is

much higher than at 662keV. It was required to determine the SPM detector efficiency over a range of

energies.

Using several energy spectra found from in previous section, the number of counts in the photopeak for

the SPM detector was easier to determine than for the CZT. Using the MCA software to select a region

of interest (ROI) around the γ-peak to return the number of counts, Equation 30 was used to find the

system efficiency (as for the CZT efficiency seen previously, to find the area of a sphere occupied).

The Single Photo Electron Spectrum

The SPMs are able to detect single photons and the Single Photo Electron Spectrum (SPES) is

produced when the charge from the short light pulses are integrated [21]. Using the pulsed LED

circuits placed on top of the SPMs, attempts to collect the SPES were made. It was however very

difficult to perform this experiment due to the requirements to produce such a small number of

photons. The fastest the circuits made could pulse was in microseconds, whilst for SPES nanoseconds

are preferred to ensure fewer photons are created. The distance from the LED to the SPM was

maximised, the power to the LED was reduced as much as possible (using a resistor in series with the

LED) and varying thicknesses of paper were placed over the SPM to reduce the light. However, when

exploring the spectrum, blurring of the peaks was found, indicating the increased number of photons

from the LED. This spectrum has been successfully produced by others [21] (Figure 3-20) for the SPM

devices, showing three clear photoelectron peaks (corresponding to 1, 2 or 3 microcells firing due to an

incident photon) after the pedestal (integration of the noise of the system [17]) peak.

Figure 3-20: The single photoelectron spectrum possible with an SPM [21 modified].

Page 65: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

52

Chapter 4 : Results and Analysis

Vast amounts of data and spectra were collected from the experiments using the various detectors

which produced graphs, images and numerical values for comparison. These were mainly analysed

using IDL software, MCA software and spreadsheets.

4.1 Modelling Results

Radioactive Source Calculator

It can be seen from Figure 4-1, that the biggest decrease in radioactive source activity occurs with 57

Co

due to its relatively short half life. Additionally, the activity on the date of the measurements could be

found allowing the intrinsic detector efficiency calculations to be performed.

Figure 4-1: The activity calculator with extrapolated activity over two months.

CZT Shielding Results

A spreadsheet using data from XCOM was used to determine the amount of attenuation provided by

three different metals to attenuate the X-rays and the fluorescence lines from the materials.

Figure 4-2: A screenshot from the shielding requirements spreadsheet for pinhole imaging.

It can be seen that to reduce the number of photons to an acceptable level (taken as approximately 2%)

from the fluorescence lines of lead (Pb), 2mm of tin (Sn) is required, and to reduce the fluorescence

lines of tin, 1mm of copper (Cu) is required. At these thicknesses, the greatest contribution to a

spectrum would be from the 85keV peak from lead, which is only 2.3% of its original value. The final

design included an aluminium casing so that lead is not handled directly.

Page 66: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

53

4.2 CZT Experimentation Results

Uniformity and Energy Resolution

The results of the first 30 minute integrations for the original combination of the 5x5 CZT array clearly

show the module edges (Figure 4-3 is 241

Am). A greyscale image for the intensity at each pixel

(brighter pixels shows higher photon counts) and a histogram were produced for both sources.

Figure 4-3: An image (left) and histogram (right) from the 241

Am illumination (over 30 minutes)

Figure 4-4: A histogram with a fitted Gaussian for the 30 minute integration of 241

Am.

The CZT detector uniformity was found using Equation 32 using values from the fitted Gaussian in

Figure 4-4, where C is the standard deviation and B is the peak centroid. It was found that the

uniformity of the detector after applying flat fielding corrections was 8.7% for 241

Am and 9.8% for

57Co. With this low spread in the data we can determine that after these corrections, any images

produced should have a similar (within error from the flat field corrections) greyscale intensity for the

same incident radiation as the number of counts from an object. Flat fielding the 10 minute

illumination using the 30 minute flat field data for each illumination results in Figure 4-5 for both

sources, where the majority of the defects (module edges) in Figure 4-3 are removed. It should be

noted that the bright pixels (Figure 4-5 right) make the image appear more uniform or „flat‟ than they

really are, numerically there is not as much difference between the two which is shown in Table 4-1.

Number of counts

Num

ber

of

pix

els

wit

h t

his

count

inte

nsi

ty

Page 67: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

54

Figure 4-5: Flat field corrections applied to the 10 minute data for 241

Am (left) and 57

Co (right).

Based on the data collected, the energy resolution was calculated to be 4.16 0.05% for 57

Co and

8.14 0.05% for 241

Am, which is very encouraging, showing that spectroscopy should be possible at

these energies, with another energy peak able to be resolved above 127keV in the presence of a

122keV 57

Co peak. This was tested to determine if the 136keV peak in 57

Co was seen in the energy

spectrum.

Once upgraded, the CZT detector had 3072 pixels on a 3x4 array (reduced from 6400 on the 5x5

array). The energy resolution per pixel was plotted following the CZT detector upgrade based on the

30 minute integration data for 241

Am and 57

Co are shown in Figure 4-6.

Figure 4-6: The energy resolution per pixel for the 57

Co 30 minute integration (left) and for the 241

Am 30 minute integration (right).

When plotted, the summed energy spectrum from each pixel shows clear peaks at the energy of the

sources used (Figure 4-7) demonstrating spectroscopic capability. The 136keV is also present at 10%

the activity of the 122keV line as expected.

Bright pixels

masking true

discontinuities

Ener

gy R

esolu

tion (

%)

Channel Number

Ener

gy R

esolu

tion (

%)

Channel Number

Page 68: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

55

Figure 4-7: Energy spectra using 57

Co for 30 minutes (left) and

241Am for 30 minutes (right).

Plotting the intensity per pixel for the upgraded CZT detector clearly shows an artefact in the top

middle module and the module boundaries (Figure 4-8) which appear in images for both sources before

flat fielding. By applying the flat field corrections in IDL to the 10 minute integration image, these

artefacts are removed, and the flat fields were then applied to the future images taken. Without

applying the flat field, the artefacts will continue to be present in any images taken.

Figure 4-8: Illumination of the 30 minute 57

Co (left) and 241

Am for 30 minutes (right) showing

crystal artefacts.

Figure 4-9: Flat fielded image with 57

Co (left) and 241

Am (right) showing the artefact has now

gone. Note the bright pixels (right) mask the true discontinuities image.

Substantial

artefact in

this module

Energy (keV) (keV)

Sum

med

Ener

gy S

pec

trum

Energy (keV)

Sum

med

Ener

gy S

pec

trum

Page 69: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

56

Creating a histogram from these images and fitting a Gaussian curve in IDL produces values for the

standard deviation and centroid used to find the upgraded detector uniformity. (The uniformity ratio

encompasses an error as the standard deviation is the error in the centroid.)

Source Before Upgrade After Upgrade

Uniformity (%) Energy Resolution (%) Uniformity (%) Energy Resolution (%) 241

Am 8.69 8.14 0.05 7.04 7.91 0.03 57

Co 9.84 4.16 0.05 2.94 4.00 0.02

Table 4-1: The results for the uniformity (after flat fielding) and energy resolution of the CZT

detector before and after the hardware upgrade.

The 10 and 30 minute flood illuminations were also completed for the ICCD detector using the same

experimental setup and sources as the CZT detector measurements, so flat field corrections would be

available to correct for systematic artefacts in the images produced by the ICCD detector. Table 4-2

shows the effect the flat field corrections have on the 10 minute collected data compared to that

collected for 30 minutes for the ICCD detector. For both sources, the flat field causes the spread in the

data to be reduced by over a factor of 2.

The results in Table 4-2 show that on average the CZT detector has an overall better uniformity after

the flat field corrections, than the ICCD detector, which is thought to be due to the much larger pixel

size (1.6mm compared to 0.5mm) of the CZT collecting more charges reduces the statistical error on

the uniformity.

Source Before flat field

corrections

After flat field

corrections

% Improvement

241Am

Mean 1073.069 1510.657

SD 155.9503 98.334

SD/Mean 14.53% 6.51% 223% 57

Co

Mean 482.7793 1673.786

SD 83.526 138.8751

SD/Mean 17.30% 8.30% 209%

Table 4-2: The uniformity of the ICCD detector for each source before and after applying the flat

field corrections.

The errors in the CZT detector energy resolution were found using the average energy resolution per

pixel and calculating the standard error (Equation 34, where n is 6400 for the original detector and

Page 70: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

57

3072 for the upgraded detector). It can be seen that the error in the energy resolution is reduced for the

upgraded detector indicating there is less spread over all pixels.

It can be seen that a 4% improvement in energy resolution has been gained from the upgrade with the

57Co source, even with a more pronounced artefact in one of the top modules. With this performance

increase, better γ-peak separation (as the energy resolution is improved) and more uniform images

(more similar pixel values for the same incident radiation) should be obtained from this detector. It

was found that both the energy resolution and the uniformity of the new detector is improved compared

to the original detector. Initial observations showed that the number of obviously hot pixels (those with

a number of counts substantially higher than the mean) was also reduced (however, this is also because

there are fewer pixels available). To quantify the effect the flat field had in addition to Figure 4-8,

normalised Gaussian curves were plotted based on the parameters (A, B and C) provided by IDL for

both before and after the flat field for the upgraded CZT detector using both γ-sources.

Using the four sets of Gaussian parameters (both sources before and both sources after the flat field

corrections), four Gaussian curves were plotted using Equation 28. The curves were normalised so a

comparison for each source of before and after flat field corrections were applied to determine the

change in statistical spread in the data (as seen in Figure 4-10).

Source FWHM before corrections FWHM after corrections Improvement (%)

241Am 1975 136 1452%

57Co 382 44 868%

Table 4-3: The effect of ‘flat fielding’ on the CZT Gaussian peaks, for the upgraded detector.

Figure 4-10: The effect of flat fielding on the 30 minute flood illuminations for both sources 241

Am (left) and 57

Co (right).

Page 71: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

58

CZT Spectroscopy

To test the spectroscopic capability of the detector, in terms of linearity in the number of counts

recorded over time and the energy resolutions possible, radioactive sources were placed near the

detector. The results seen in Figure 4-11 show that for two different sources with different activities

there is linearity with a high goodness of fit, in the detection of the counts (taken as the maximum

number in the peak) at 59.5 and 122keV with time. The CZT detector background (cover off with no

sources) is insignificant compared to these values providing 14 1 counts per second (error taken as

nearest count from the average background measurements of 10 and 60 seconds).

Figure 4-11: A linear relationship successfully identified between increase in counts and

integration time, for two γ-sources.

The expected number of counts for each source can be confidently extrapolated, based on the data

collected due to the uniformity of the fit. Dividing the number of counts by the integration time, the

count rate (in counts per seconds) was determined for each source over the 10 second integration

period which was repeated for both sources. These were found to be 868 6 for 57

Co and 1103 4 for

241Am. When incrementally decreasing the integration time, it was possible to obtain a spectrum with

over a thousand counts in only one second using this detector, with well defined peaks at the two γ-

energies used. Figure 4-12 shows an energy spectrum with three clearly distinguishable peaks (59.5,

Page 72: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

59

122 and 136keV). The detector uniformity is indicated by the hits / channel chart on the bottom

showing nearly all pixels receive a similar number of counts (all but one pixel has approximately 10

counts).

Figure 4-12: The energy spectrum after just one second, showing 241

Am (59.5keV) and 57

Co

source (122 and 136keV) peaks.

It was found that the CZT detector counts linearly for both sources over the experimented time range,

(as the count rate remains constant when graphed they are within error bars).

As the source to detector distance increases, the number of counts decreases following the inverse

square law (Figure 4-13) as expected from Equation 31. The results in this section prove that the

detector can make a high resolution and fast spectroscopic detector when integrating for short times,

for passive γ-ray detection, currently operating in the range of ~20 – 200keV.

Page 73: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

60

Figure 4-13: The effect of time and distance on the maximum number of counts received for two

of the integrations times tested using the CZT detector, showing the inverse square law.

Figure 4-14 shows the excellent correspondence to recent (in the final weeks of the placement)

published results using 57

Co [38] also using a CZT detector array.

Figure 4-14: The measured energy spectrum (left) and one just published for 57

Co (right) [38].

The energy resolutions have also recently been explored by others [38], where an excellent match to

the measured results in addition to the measured spectra. It was found in the „Uniformity and Energy

Resolution‟ section that the upgraded CZT detector provides an energy resolution of 7.91 0.03 % at

59.5keV and 4.00 0.02 % at 122keV. [38] presents an average energy resolution of 5.5keV FHWM

at 122keV, equivalent to 4.5% which is strikingly similar to the results measured.

16.0 27.5

Page 74: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

61

CZT Efficiency

Efficiency is a measure of how well the detector records the incident radiation. The CZT detection

efficiency was measured to be 72 1% for the 241

Am source and 58 2% for using the 57

Co source

(calculated as the standard error in the measurements where error bars were found using the fractional

errors based on the error in the distance from the detector, and the error in the number of counts). This

corresponds to a 24% decrease in efficiency at the 122keV energy compared to at 59.5keV, where from

the attenuation of the material, only a 6% decrease is expected due to the increase in energy. The CZT

detector efficiency is less than expected, based on the attenuation of the CZT material for these energy

photons and is possibly due to not all of the CZT thickness able to detect photons.

Figure 4-15: The efficiency of the CZT detector with increasing source distance.

CZT and ICCD X-ray Backscatter Imaging

To determine the performance of the imaging quality using the CZT detector and for comparison with

the ICCD detector, X-ray backscatter images through a pinhole were taken. The same experimental

setup was used for both detectors using the newly created graded shielding with a 2mm pinhole

aperture with the CZT detector (along with the additional lead shielding required).

It was proved that imaging using both the ICCD and the CZT detector is clearly possible and 20 images

were taken for objects both in and out of a case. The items in a case made no significant impact on the

images using either detector, due to its thin size and low density. Due to the number of CZT pixels

being substantially less than in the ICCD (3072 vs. approximately 42,000) and much larger pixels

Page 75: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

62

being used in the CZT (CZT pixel size of 1.5mm (pitch 1.6mm) vs. 450μm (pitch 0.5mm) for the

ICCD), the images are not as clear using the CZT detector compared to the ICCD. However, the

difference in contrast initially appeared better in the CZT images, where there is an object and where

there is no object, which was quantified later on. The ICCD images sharpen as the pinhole reduces in

size from 4mm to 2mm for the same time integration as expected (Figure 4-16).

Figure 4-16: A direct comparison for the same object (sugar) with the 4mm and 2mm pinholes

(before flat field corrections) from the ICCD detector.

To compare the images between the CZT and ICCD detectors in the rest of this section fairly, the

images are shown with the peak of dynamic range selected for both, such that the contrast is not biased

towards either detector. The images however, are quantified numerically later on, where the effect of

the contrast range selected makes no difference.

Figure 4-17: Flat fielded images of cotton in a case, using CZT (left) and ICCD (right).

Page 76: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

63

Figure 4-18: Talc and aluminium powder on a plastic base (flat fielded) using CZT (left), ICCD

(middle), and before flat field corrections applied to the CZT (right).

After flat field correcting, („flat fielding‟) the images produced are much clearer (compare Figure 4-18

left and right). The blackened pixels in the CZT images are assumed to correspond to the pixels

switched off due to erroneously high count rates. Future work on the image processing can eliminate

the effect these pixels have on the images by programming code to identify these pixels (those with an

extremely low count) and setting them equal to the average of the surrounding pixels.

To compare the images produced from the backscattered X-rays on the CZT and ICCD images, it was

necessary to quantify the data numerically. The number of counts (CZT) and intensity (ICCD) of the

pixels were measured for the same images. Regions of the flat fielded images were selected to

encompass a portion of the object and then the same size portion of the background, by identifying an

area „by eye‟ which appears in images from both detectors, and selecting pixels in these areas.

Page 77: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

64

Figure 4-19: Images of the same object (but reversed) of the sections taken for analysis of the

mean and standard deviation (CZT left and ICCD right).

For the flat fielded CZT detector images, 20 pixels were manually chosen in these regions of objects

and the background and the count values were taken and recorded in a spreadsheet. This was repeated

so that there were two target objects and two background objects. The same regions were selected in

the ICCD images, where software automatically calculates the mean and standard deviation for all of

the pixels in the selected region.

Figure 4-20: A graphical representation of the data collected for the ratio of object to

background each image from both detectors.

By investigating the same two images for each detector, it was found that for each image, the ratio of

object to background was higher for the CZT detector (Figure 4-20). This means that the objects of

interest stand out better for this detector than for the ICCD detector. A surprisingly high fractional

error was found for the CZT detector for these measurements which was attributed to the vast

Page 78: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

65

difference in number of pixels selected for the CZT detector and ICCD detector, causing the standard

deviation to have a much greater effect on the associated error.

Based on the images produced and the improved energy resolution, it is recommended that the CZT

detector be fully populated with the upgraded material, to provide a wider imaging area, and therefore a

larger detection probability. In the future, reducing the CZT pixel size, clearer images will be obtained,

making the CZT better competition for the smoothness provided by the ICCD detector.

Figure 4-21: The number of counts from the average of 10 pixels in the CZT and ICCD detector

for all of the objects (error as the standard deviation).

Figure 4-21 shows how the highest Z element (tungsten) produces the fewest backscattered photons (as

expected as more X-rays are absorbed in the photoelectric effect at higher Z). The effect of the case

the objects were in was found to be negligible for the experiments. The error (taken as the standard

deviation of the data for both detectors) was greater for each of the ICCD images than for the CZT

detector, which reflects the larger pixel size of the CZT providing better sensitivity.

Angular Resolution

The resulting backscatter images from this experiment were processed and displayed using IDL (CZT

detector) and Image Pro Plus software (ICCD detector) to compare the two detectors. The images from

each detector are displayed in Figure 4-22 to Figure 4-25 (the 1cm images for both detectors being used

to compare the flat fielded images to the non-flat fielded images).

Norm

alis

ed a

ver

age

counts

in t

he

obje

ct

Page 79: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

66

Figure 4-22: 1cm apart before flat field corrections (left) and after (right) for the CZT detector.

Figure 4-23: 2-4cm (left to right) after flat field corrections for the CZT detector.

Figure 4-24: 1cm apart before flat field corrections (left) and after (right) for the ICCD detector.

Figure 4-25: 2-4cm (left to right) after flat field corrections with 57

Co 30 minutes for the ICCD

detector.

Page 80: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

67

Based on the diameter of the two tubes, the first position where the two tubes are physically separate is

at 3cm (the two radii of the tubes are 2.5cm therefore the first tested position where the objects should

be resolved is the 3cm position). A change in intensity indicating separation was found at this distance

for both detectors. For the ICCD detector, the overlapping of the lids in the 2cm image are an

indication that more than one object is present whilst for the CZT detector, the overlapping at 2cm

causes a wider object to be produced than at 1cm. The results from the profile for each separation

image for both detectors are shown in Figure 4-26, which were normalised for a direct comparison.

Figure 4-26: The separation seen using the CZT (top) and ICCD (bottom) detectors over 5cm,

where the background for the ICCD is always higher than CZT.

For both detectors, there is a trough in the normalised counts (CZT) and normalised intensity (ICCD) at

the 2cm separation image, even though there is no physical gap between the objects; the difference is

due to the thickness of the materials being detected due to a change in the number of photons. This is a

success for both detectors. It should be noted that there is a much higher background using the ICCD

detector (about 20% of the maximum intensity compared to less than 10% for the CZT). The

Page 81: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

68

separation of the materials was measured by the size of the trough between the two peaks was plotted

for each detector at each distance, and the results shown in Figure 4-27, giving a comparison between

the two. The high ICCD background level makes the objects stand out less than for the CZT which

follows the theory as smaller pixels give better spatial resolution but reduced sensitivity.

Figure 4-27: The separation displayed as the change in normalised intensity for both detectors

(error bars as smallest unit taken from the trough height).

This experiment showed that both detectors measure a difference in intensity where the objects begin to

overlap, which increases with the separation of the two objects. Importantly, the larger pixel size of the

CZT provides better sensitivity seen as a much lower background where no object is present, than the

ICCD detector which shows a steady 20% background (Figure 4-26 bottom).

4.3 Scintillator and SPM Experiments

Scintillator and SPM Energy Resolution Model

The model showed that coupling scintillators to SPMs would be an inefficient process with some

combinations of scintillator, SPM and γ-sources, resulting in a total loss of over 95% of optical photons

produced from the radiation photons, producing energy resolutions from around 145% to 11%.

As the photon detection efficiency is a function of wavelength, the energy resolutions possible are

dependant on the scintillator crystal used. The highest PDE for the SPMs is at a wavelength of

approximately 470nm [19], corresponding best to BGO (480nm) and worst for CsI(Tl) (565nm).

Page 82: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

69

However, the much lower light yield of BGO compared with CsI(Tl) means that the best modelled

energy resolution still occurs using the CsI(Tl) crystal.

Figure 4-28: Results for the modelled energy resolutions for each crystal and SPM over the

energy rage to be tested. Clockwise from top left: CsI(Tl), BGO, LYSO and CdWO4.

These modelled results correspond well with knowledge of the materials and the system, as the

scintillators with the highest light yield provide better energy resolutions. The model also shows how

the energy resolution improves with increasing source energy (more light flashes are produced).

Spectroscopy is at least theoretically possible based on these calculations especially at higher energies,

as some energy resolutions are estimated to be less than 20%.

In each case the energy resolution was modelled to be best for the 3mm 35μm SPM, as from the theory

the PDE is highest with this SPM, followed by the 3mm 20μm SPM with the 1mm 20μm SPM

consistently providing the lower energy resolutions as it has the lower PDE and smallest active area.

The energy resolution was found to vary with the square root of the energy was successfully verified

for the model, and the modelled results were compared to those measured experimentally.

En

ergy R

eso

luti

on

(%

)

En

ergy R

eso

luti

on

(%

)

En

ergy R

eso

luti

on

(%

)

En

ergy R

eso

luti

on

(%

)

Page 83: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

70

Scintillator and SPM: Preliminary Test Results

Figure 4-29: The 1mm SPM pulses with no sources present seen in oscilloscope mode on the DAQ

(left) and expanded on another oscilloscope showing distinguishable dark photons (right).

Figure 4-29 (left) shows „levels‟ of pulses are present, which correspond to optical cross talk and

individual photons are clearly visible on the 1mm SPM Figure 4-29 (right).

Figure 4-30: The dark counts in the 3mm 20μm SPM (top) and 3mm 35μm SPM (bottom),

showing noise photons not as clearly defined as the 1mm SPM.

The rate of the dark photons is known to depend on the SPM size, with the 3mm SPMs having a dark

rate 9 times as high. From Figure 4-29 and Figure 4-30 the dark rate can be approximated as; 1mm

SPM at 3MHz and 3mm SPM at 6MHz, proving the 3mm SPMs have a higher dark noise rate.

LED Testing

Preliminary tests explored the pulses produced by the SPM when subjected to light from LEDs. It was

found that the circuit using transistors and capacitors instead of a 555 IC took much longer to switch on

Page 84: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

71

(microseconds compared to nanoseconds with visible capacitor charging seen on the graphs) and

therefore the SPM response was also much slower (average of 95ns) so these results were not included

in Figure 4-32. All three SPMs respond to the LED circuits with a large negative pulse in the order of

volts (the maximum possible is 2V from the board [33]).

Figure 4-31: 3mm 35µm SPM at 32V bias using the 555 IC timer (left) and 3mm 35µm 32V

(right) for an LED pulse where the SPM response (pink) to the LED pulse (blue)

The time for the SPMs to change state from off to maximum pulse voltage (Figure 4-32) is between 10

and 15 nanoseconds which corresponds well to the 12ns stated [33]. (The errors were found using the

smallest measurable scale on the oscilloscope printouts).

Figure 4-32: The onset times of all three SPMs using pulsed LED circuits to show that the pulses

are indeed produced in around 12ns.

Page 85: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

72

Figure 4-33: A pulse caused by the pulsing of a red square LED, with black tape around the four

exposed sides on the 1mm SPM.

The SPM pulses caused by the LED photons are much slower (as they are on for µs which is similar to

the scintillator crystal decay time) compared to the dark noise photons which are present for

approximately 50ns. The amplitude of the pulses are also quite different, as the dark noise photons are

approximately 18mV at the 30V bias, whilst the LED pulse causes the SPM to produce voltages of

over 1.5V where the noise is only 1%. The SPM pulse output was measured for LED pulse durations

of 0.5, 1, 2, 4 and 12.5μs using a 3mm round red LED.

Figure 4-34: The SPM response to an LED pulse not fully recorded.

From these LED measurements, an important though unexpected result was found (Figure 4-34). The

SPM response to the LED pulse does produce the maximum pulse voltage for the whole duration of the

LED pulse. After discussions with the manufacturer this was found to be due to the Alternating

Current (AC) coupling, a feature of the pulse pre-amplifier board, which is more evident for the larger

pulse durations (>1μs), as these are not fully recorded. Using an alternative board, a „transimpedance

amplifier‟, would allow much longer light signals to be recorded. (One was received on 25/11/07.)

Page 86: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

73

This was a useful experiment to conduct, as;

It was confirmed that the SPM produces pulses in response to optical photons;

The size and duration of the pulses was identified which helped locating the output from the

scintillators;

The AC coupling due to the preamp board was identified where the full light pulse is not

completely recorded by the SPM.

Components were chosen to allow the LED to pulse at the maximum speed of the IC, with the LED on

for 0.2µs and off for 5µs. As each LED pulse was clearly registered by the SPM, by producing a large

negative pulse, it is seen that the count rate of the detector is at least 2x105Hz. However, this is

severely limited by the maximum rate the circuit could pulse the LED at due to the IC used, and the

detector is able to quoted to be able to count at 10MHz [22].

Figure 4-35: The counting of the SPM to be at least 2x105Hz.

SPM Pulse Linearity

As the resistance to an LED was increased, the power was reduced, decreasing the number of photons

emitted. Figure 4-36 shows the linear (with a high goodness of fit) decrease of the SPM voltage pulse

with the increasing resistance in series with the LED.

Page 87: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

74

Figure 4-36: The effect of resistance (and therefore LED power from the 555IC timer) on the size

of the SPM pulses produced.

This confirms that the SPM response is linear with resistance (directly affecting the power which is

directly proportional to resistance) applied to the LED, which means for any future measurements,

principally spectroscopy, the SPM response the amount of light incident (energy) is directly

proportional to the SPM voltage produced.

SPM Pulse Observations

The output pulses from the SPMs were viewed on an oscilloscope and printed for many SPM, crystal

and source combinations to provide the DAQ with the correct settings required for spectroscopy.

Figure 4-37: A scintillation pulse from the 3mm 20µm SPM (at 32V bias) using CsI(Tl) with the 22

Na source.

Figure 4-37 shows the pulse from a CsI(Tl) crystal using the energy from the

22Na source with a very

clear 1μs decay time as expected from the crystal details value in Table 2-1. The SPM (for these times)

Page 88: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

75

responds for the correct duration. There is also a trend for the voltage of the SPM pulses to reduce with

the decreasing source energy as expected from the decreased light yield produced. The exception to

this are the pulses from 22

Na, which appear higher than the other sources. The cause of this is so far

unknown.

Figure 4-38: A comparison of the voltage pulses from scintillator crystals and sources using the

3mm 20μm SPM at 32V.

LYSO pulses are not present in Figure 4-38 as the pulses from the natural radioactivity were not

distinguishable from those caused by scintillation flashes on the oscilloscope. (Background

measurements were made to confirm that the spectra obtained were from the γ-sources and not

background from Lutetium). The pulses from the CdWO4 crystal were found to have a shorter decay

time than those stated when looking at the light pulses produced by the SPM. From the previous LED

measurements, it is know that longer light pulses over about 1μs are not fully recorded due to the pulse

amplifier board used for pulses, hence the incorrect display of the longer scintillation pulses.

Figure 4-39: The CdWO4 response to 22

Na on the 3mm 20μm SPM, the decay lasting much less

than the 20µs expected.

2µs

200mV

Page 89: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

76

A transimpedance amplifier will reduce this effect, and the longer pulses will be retained. A

measurement was made using the 3mm 20μm SPM with a pulsed LED using this new amplifier (Figure

4-40).

Figure 4-40: Comparing the response of the pulse preamplifier which cuts off the full pulse

duration (left) with the transimpedance amplifier (right).

The SPM responses to the pulsed LED using the transimpedance amplifier allow the full duration of a

long (18μs) pulse to be recorded (Figure 4-40 right), where the negative „well‟ created by the SPM

does not stop until the signal from the LED pulse stops. For comparison, the original board (Figure

4-40 left) shows the cut off effect on much shorter pulses. The use of different boards will therefore

affect the energy resolution possible from the spectroscopic measurements for scintillator crystals with

a longer decay time, as only a fraction of the light will be recorded.

By measuring the size of the voltage pulses produced by the SPMs at two voltages (30 and 32V), it was

found that an increase in SNR (Equation 33) was achieved in pulses from the 1mm to the 3mm (20μm)

SPM and a further increase from the 3mm (20μm) to the 3mm (35μm) SPM. The SNR (Figure 4-41) is

not as high with the bias increase, although the much larger signal pulses (~1V) arise from the photons

from the scintillator crystal which are more easily detected by the acquisition system. (A threshold was

set such that the system records values greater than or equal to this value, which can be observed by

viewing the event rate and increasing the threshold to suppress the noise).

Page 90: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

77

Figure 4-41: The effect of bias on the SNR for each SPM.

SPM Spectroscopy

Following the modelling and characterisation experiments using LEDs, the arrival of scintillator

crystals allowed spectroscopic experiments to begin leading to dozens of measurements, the results of

which are summarised graphically and in Table 4-5. Due to the ambitious number of measurements

designed to fully test the SPM system, and the scintillator crystals and 3mm SPMs not arriving until

mid-October, spectra were limited to 30 minute integrations to allow for as many tests as possible to be

conducted. (This was in addition to the time required to trial a range of rise times and gains, and

physically setting up each experiment by coupling the detector to the power boards, coupling the

scintillator to the SPM and adjusting the bias where necessary).

The acquisition system used was primarily a way of collecting the data, rather than for complex

analysis of the spectrum. The energy of only one peak can be entered at a time (to provide the energy

resolution of, or the number of counts in, the energy peak). However, when selecting a peak as a

region of interest (ROI) it can be seen that the peak shifts left when decreasing energies are applied and

spectra collected, and approximate values for the centre of the peak are displayed based on the energy

of the one ROI.

Page 91: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

78

Figure 4-42: The energy spectrum using CsI(Tl) with the 22

Na source, providing an energy

resolution of 14% using the 3mm 35μm SPM.

An energy spectrum from the SPMs is shown in Figure 4-42 for the

22Na source and the CsI(Tl) crystal.

Following a well defined peak at 511keV, the Compton continuum continues due to portions of the

1274keV peak not being fully recorded in the scintillator until 1274keV. Better energy resolutions

were occasionally found using the 22

Na source in place of the 137

Cs source despite its lower energy,

which corresponds with the larger pulses seen when using the oscilloscope in earlier measurements.

Some of the experiments matched very well with the model, which predicted several outcomes:

One of the best energy resolutions possible was found using the 3mm 35μm SPM (with the

higher PDE) and the scintillator crystals with the highest light yield. This was measured and

an energy resolution of 11.78 0.03% was found for CsI(Tl) (predicted value 10.98%);

The best energy resolution comes from the 3mm 35μm SPM for all measurements (due to the

higher PDE), which was found to be the case for most measurements (Table 4-5);

Page 92: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

79

Figure 4-43: The modelled energy resolutions (top) and measured average energy resolutions

using CsI(Tl) for each SPM (bottom).

The best energy resolution occurs at 662keV.

This was found to be the case for many measurements (see Table 4-5) as more charges are created per

event reducing the statistical broadening of the energy peak, increasing the energy resolution.

Some combinations produce values for energy resolution of over 100%.

The model predicted that there would be some experimental combinations using BGO and CdWO4 and

the lower energy γ-sources that would provide such a low light output that energy resolutions would be

as high as 145%. Poor energy resolutions were experimentally verified using these crystals such that

BGO on the 3mm 20μm SPM gave an energy resolution of 126 2% when using the 241

Am (59.keV)

source. Having an energy resolution of more than 126% at 59.5keV means that an energy range of

76.2keV either side of the energy peak is required to resolve additional peaks.

Figure 4-44: An energy spectrum from the 3mm 20μm SPM using BGO and 137

Cs.

Page 93: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

80

The linearity of the detector, in terms of number of counts received with increasing time, was explored

for the SPMs by measuring the maximum number of counts of the energy peak in two separate

experiments; using the 3mm 20μm SPM with CsI(Tl) and the 137

Cs source, and the 3mm 35μm SPM

with BGO and the 22

Na source.

Figure 4-45: Two separate SPM linearity experiments; using the 3mm 20μm SPM with 137

Cs and

CsI(Tl), and the 3mm 35μm SPM with 22

Na and BGO.

As can be seen in Figure 4-45, there is a linear increase with a high goodness of fit for the increase in

the maximum number of counts with time for two different sources and SPMs. This shows the stability

of the detector over time and that there is no measured random counting drift in the detector in this

range.

Energy (keV) 59.5 122 511 662

Average Measured Energy

Resolution (%) 126 2 82.7 0.2 25.0 0.1 19.9 0.5

Root Energy Result 118.4 88.0 40.4 22.0

Modelled 122.5 85.9 42 36.9

Table 4-4: A close match for the measured values to that expected from the model for energy

resolution measurements.

In many cases, the measured energy resolutions were a close match to those modelled and those

predicted using Equation 20, for the measured energy resolution varying with root of the energy, as

seen in Table 4-4. In this example, at 511 and 662keV the energy resolutions were better than

predicted, so the scintillation losses may not have been as poor as initially predicted.

Page 94: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

81

Plotting Equation 21 (Figure 4-46) shows a linear relationship is present when selected energy

resolutions are chosen, and therefore Equation 20, stating that energy resolution is proportional to root

energy is verified.

Figure 4-46: A ln-ln plot of energy vs. energy resolution using BGO on the 3mm 35μm SPM.

The very poor energy resolutions obtained for the CdWO4 crystal have been attributed to the pulse

amplifier which has been used in place of a transimpedance amplifier (which would be more

appropriate for the longer sized pulses). The decay time of CdWO4 is approximately 20μs where

nearly all of the light pulse from the crystal is not recorded using the pulse amplifier.

Figure 4-47: The average measured energy resolution results using BGO, LYSO and CsI(Tl)

scintillator crystals on the 3mm 20µm SPM.

Due to very poor energy resolutions measured using CdWO4, the values have been removed from

Figure 4-47 to avoid distorting the values recorded using the other scintillator crystals. For the 3mm

20µm SPM, the best energy resolutions come from the LYSO crystal providing better energy

Page 95: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

82

resolution at higher energies. This is thought to be due to the much faster decay time being better

suited to the amplifier used with the SPM for these measurements. With the transimpedance amplifier,

it is expected that the energy resolution for the CsI(Tl) crystal will provide the best energy resolutions

overall based on the better light yield, as previously modelled.

Figure 4-48: The average measured energy resolution results from several scintillator crystals

coupled to the 3mm 35µm SPM.

Figure 4-48 shows how the overall trend for the improvement in energy resolution (%) is followed with

increasing energy. LYSO provides the better energy resolution for the 3mm 35µm SPM at the two

lower energies, however, as predicted by the model, the best energy resolution on the 3mm 35µm SPM

is found using CsI(Tl) crystal at 662keV.

Comparing the 3mm 20μm and 3mm 35μm SPMs (to identify the effect of the number of microcells

changing the PDE), shows that in nearly all of the measurements the average energy resolution (from

two measurements) is improved for the 35μm SPM, which is due to the higher PDE in this SPM. The

LYSO crystal is able to offer comparable energy resolutions to CsI(Tl) at higher energies even with its

lower light yield, which has been attributed to LYSO‟s much quicker decay time.

Page 96: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

83

Figure 4-49: The average measured energy resolution results for the CsI(Tl) scintillator crystal

compared to the modelled result for the 1mm SPM.

Figure 4-49 shows the energy resolution improving with source energy for the 1mm 20µm SPM for the

modelled values and those found experimentally. The energy resolutions achieved are the poorest of

all three SPMs using the 1mm 20µm SPM (as predicted by the model). This is likely to be a

combination of the crystal size mismatch causing approximately 1 in 9 of the photons to be incident on

the SPM area, the lower PDE and lower active area. It was found that the 3mm 20μm SPM provided

better energy resolutions than the 1mm 20µm SPM (for all but the 122keV source energy).

Due to the mismatch in scintillator and SPM area for the 1mm SPM, the model estimating the energy

resolution was modified, allowing for only 1 in 9 scintillation photons reaching the SPM for the CsI(Tl)

crystal. It was found that the measured energy resolution fits between both the modelled energy

resolutions, where it is assumed that areas match completely, and the other model (labelled modelled

1/9th ER) where only 1 in 9 photons reach the SPM. As the model has been proved to work well for

other combinations of SPM and crystals, it is reasonable to assume that the mismatch has caused the

loss of some, but not as much as 1/9th of the light emitted by the crystal as seen in Figure 4-50.

Figure 4-50: The measured energy resolutions with the 1mm SPM and the modelled results based

on a complete match in areas and at a 1/9 area match.

Page 97: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

84

One of the project aims was to determine the difference in the energy resolution between the three

SPMs, therefore comparing the different pixel sizes and the two fill factors. Figure 4-51 shows how

the measured energy resolution varies for the same scintillator crystal (CsI(Tl)) using each of the three

SPMs. It is clear for each SPM that the increase in energy resolution is gained with increased energy.

Additionally, the best energy resolution is found using the 3mm 35µm SPM, followed by the 3mm

20µm and finally the 1mm 20µm SPM (in all but 1 case) matching with the predictions.

Figure 4-51: A comparison of the measured energy resolutions for the three SPMs using all of the

sources on the CsI(Tl) crystal.

Using Equation 20, the energy resolutions of various combinations of scintillator and SPMs were

extrapolated to provide the expected energy resolution at 140keV, the energy used in SPECT imaging.

The results in Figure 4-52 show that the energy resolutions at this energy are unlikely to be better than

30% for any combination, which means that other energy peaks can be resolved above 180keV or

below 100keV. With this large dynamic range, it is seen that using SPM technology at two important

energies, 140 and 511keV, can be both simultaneously and separately recorded at a very high photon

counting rate.

Page 98: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

85

Figure 4-52: The extrapolated energy resolution (based on measured results) to 140keV for eight

scintillator and SPM combinations.

The measured SPM energy resolution never reached the current performance possible using PMTs, or

those experimentally found using a different semiconductor detector (CZT). However, the testing of

this technology has proved that spectroscopy is possible. Their size and gain make them a serious

contender for low power, light weight, fast counting passive γ-ray detectors.

3mm 20µm 59.5keV 122keV 511keV 662keV

CdWO4 343 0 175 37 38 5 36 2

BGO 126 2 82.7 0.2 25.0 0.1 19.9 0.5

LYSO 62.3 0.7 57 3 10.6 0.1 12.9 0.5

CsI(Tl) 104 1 64 3 23 2 17.3 0.8

3mm 35µm

CdWO4 0 0 0 0 25.75 0.02 35.3 0.1

BGO 0 0 72.9 0.1 34.9 0.4 18 2

LYSO 54.6 0.3 31 3 61.6 0.3 42 2

CsI(Tl) 62 16 46.7 0.4 14.0 0.2 11.78 0.03

1mm 20µm

CsI(Tl) 0 0 51.8 0.8 35 2 28 2

Table 4-5: The average measured energy resolutions (%). (Errors at the 95% confidence level).

A summary of the measured energy resolutions obtained and the corresponding error at the 95%

confidence level is displayed in Table 4-5. (A value of 0 indicates that no measurement (or repeat) was

possible for this experimental arrangement.)

Page 99: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

86

Observing Complex Spectra using SPMs

To observe complex spectra, the same experimental setup was used for the SPM spectroscopic

measurements but with an additional source near the scintillator when required. A series of

experiments was conducted to identify the effect on the energy spectrum produced when simultaneous

sources are incident on the scintillator crystals coupled to the SPMs (CsI(Tl) on the 3mm 35μm was

initially tested as it provided among the best energy resolutions in previous measurements). Integrating

briefly for two minutes using 137

Cs in the standard experimental setup (Figure 3-17) produces the

spectra in Figure 4-53 (left) and adding 241

Am (placed several cm further back from the crystal to avoid

saturation) provides Figure 4-53 (right).

Figure 4-53: A 137

Cs Spectrum for two minutes on the 3mm 35μm SPM (left) and the effect seen

when 241

Am (further back at 5cm) from the crystal is added to the experiment (right).

Higher attenuation of photons occurs at lower energies (Figure 2-20), causing many more counts to be

recorded at lower energies (photons are more easily stopped). Counts at higher energies occur less

frequently and are not visible on the energy spectrum if a low energy source is placed at a similar

distance to a higher energy source. This was found experimentally to be the case. By replacing the

241Am source by

57Co a γ-peak should be found at higher energy than the

241Am peak at 59.5keV. This

was successfully found, as is displayed in Figure 4-54.

Page 100: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

87

Figure 4-54: The 122keV peak clearly to the right of the cursors showing the position around the

59.5keV peak.

The 136keV peak from the 57

Co source is not visible in any of the SPM measurements, due to the poor

energy resolutions obtained from the experimental equipment. (An energy resolution of 50% at 122keV

means that another peak could be resolved at 122 61keV, where to resolve a peak at 136keV, an

energy resolution of 11% at 122keV (or better) is required, hence it is visible on the CZT detector).

Peaks corresponding to high and low energy sources are clearly distinguishable in the same energy

spectrum based on Figure 4-54.

When using both the low energy sources (59.5 and 122keV) together, the peaks are not as

distinguishable due to the large energy resolutions achieved (Table 4-5).

Figure 4-55: Illumination with 57

Co (left) and both 241

Am and 57

Co (right) showing a broadening

due to the 59.5keV source from 33% to 52% at 150 counts.

Page 101: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

88

An increase in the broadening of the peak (by 57%) is detected (Figure 4-55) but the energy resolutions

are so large that the peaks at 59.5 and 122keV cannot successfully be resolved. Using both high energy

sources (511 and 662keV) together allows some indication that additional peaks are present due to

better energy resolutions at higher energies (Figure 4-56 and Figure 4-57).

Figure 4-56: Both high energy sources (511 and 662keV) when integrating for two minutes.

Figure 4-57: The energy spectrum for 22

Na (511keV) without the 662keV source, the peak is

missing when integrating for two minutes.

Using BGO with 137

Cs for two minutes on the 3mm 35μm SPM provides a definite spectrum with a

clear peak at 662keV from the 137

Cs source is visible.

Page 102: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

89

Figure 4-58: BGO with 137

Cs for two minutes on the 3mm 35μm SPM.

By adding 22

Na to the experiment, after two minutes neither the 511 nor 662keV peak is well resolved

but the continuum due to the 1274keV peak continues from the 22

Na source in Figure 4-59.

Figure 4-59: The addition of 22

Na to the 137

Cs source for two minutes.

By integrating for 30 minutes, a better defined spectrum is produced (Figure 4-60) which shows the

662keV peak sitting on the 511keV peak and again the continuum present until the peak at 1274keV

from the 22

Na source.

Figure 4-60: Integrating 22

Na and 137

Cs for 30 minutes better defines the spectrum.

Page 103: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

90

Figure 4-61: The beta spectrum from LYSO taken for 30 minutes with no additional radioactive

sources using the 3mm 20μm SPM.

Figure 4-61 is the naturally occurring radioactive energy spectrum from the Lutetium in the LYSO

scintillator showing a low level background always present with this crystal as predicted. The effect is

negligible compared to the peak source energies.

SPM Detection Time

In most detection applications, the faster the detection time the better. The time taken to produce an

energy spectrum using the 3mm 35μm SPM and the CsI(Tl) crystal with 57

Co (closest energy for

SPECT imaging) was explored (Figure 4-62). This verifies the uniform counting of the detector over

the time and energy ranges explored. In just 10 seconds, over one thousand counts are in present the

122keV energy peak allowing identification that the energy is present.

Figure 4-62: The linear increase of counts with integration time for the 3mm 35µm using 57

Co

and CsI(Tl).

Page 104: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

91

Extrapolating the best fit line in Figure 4-62, an energy peak with 1000 counts at 122keV would be

present in less than 10 seconds. The closeness of the energies between 122 and 140keV allows

comparisons to be drawn to SPECT medical detection of 140keV, where rapid detection of this energy

would be possible using SPMs and scintillator crystals.

SPM Efficiency Results

For the 3mm 35μm SPM, the trend was successfully identified that as the source energy increases, the

detection efficiency decreases. This corresponds with Figure 2-20, where fewer photons are attenuated

at higher energies.

Figure 4-63: A comparison of the system efficiencies for each SPM using CsI(Tl).

Figure 4-63 displays the measured detection efficiency of the three SPMs using the CsI(Tl) crystal,

showing how the efficiency decreases with increasing source energy (as expected). The attenuation

data (curve labelled CsI Att) shows how the ideal attenuation in the crystal over the energies should

decrease with increasing energy. It is seen that the experimental observations follow the predicted

outcome where the efficiencies reduce with increasing energy. However, the measured observations

show lower efficiencies than the ideal curve. Possible reasons for this include the loss of photons from

any slight mismatch of the alignment of the scintillator crystal to the SPM, and absorption of photons

in the silicone grease between the scintillator and SPM, which would cause fewer photons to be

detected by the SPM. To improve this setup the scintillators would be coupled directly using superior

optical grease and the crystals would be perfectly matched to the SPM area to maximise the efficiency.

Page 105: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

92

Chapter 5 : Review, Conclusions and Further Work

6.1 Conclusions

Over the course of the M.Phys. placement year, research was performed mainly at the Dstl laboratory,

exploring radiation detection. Modelling the attenuation of transmission X-rays for a range of

materials was undertaken initially, showing the distinct difference between metals and organics. This

followed a review of the different detection techniques which populated a graphical model to provide

information on where the research would be useful, principally in identifying the spectroscopic

performance of radiation detectors. CZT detector preparation and characterisation experiments were

conducted to gain an understanding into detector theory and application.

The main aims of this project were to determine and quantify the performance of the CZT and SPM

detector systems as new detectors for spectroscopy and imaging. This is of great importance as

spectroscopic information provided by γ-detection and could provide the possibility for material

discrimination based on the energy of the material present. This could have an impact in medical and

security applications.

An ICCD detector was unable to provide any spectroscopic information, but was used as a basis of

comparison to the CZT detector for the backscattered X-ray images. When directly comparing the

ICCD detector to the CZT for X-ray backscatter imaging, it was seen (Figure 4-26) that the CZT

detector provides lower noise and better object to background ratios (Figure 4-20). The ICCD detector

is quite long about 0.6m, whereas the CZT detector is only 0.25m thick with the specifically produced

shielding applied. An SPM could further reduce this thickness to less than 5cm including a scintillator

crystal. An array of SPMs to cover the area of the current ICCD detector could provide a thin

lightweight alternative to both the CZT and ICCD detectors, to provide imaging and spectroscopic

information with the ability to be a fast counting passive γ-detector.

The energy resolution of the CZT array, required to identify how well close peaks can be resolved, was

quantified before and after a hardware upgrade, which identified an improvement in the energy

resolution using two different γ-sources. It was found that the CZT detector provides excellent energy

Page 106: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

93

resolution of 4% at 122keV allowing direct application as a γ-ray detector which is very close to the

energy of the γ-line used in SPECT imaging for medical applications. The inverse square law for the

decrease in radiation with increased distance was verified for the CZT detector array when several

distances (15 to 30cm) were tested with a range of integration times (1 to 10 seconds). The linear

increase of the counts with time was also verified, proving the detector‟s stability.

To use the CZT detector array as a pinhole imager, the thickness of different Z materials was calculated

to provide shielding only allow X-rays through the pinhole. Pinhole imaging using the CZT detector to

detect X-ray backscattered photons demonstrated the concept worked for both the ICCD and CZT

detectors. The images produced from the X-ray backscatter experiments show that without image

processing (by flat fielding) the ICCD detector produces images which are initially clearer due to the

large number of smaller pixels. After flat fielding, the CZT images are made clearer, with the effect of

flat fielding being more pronounced for the CZT detector (as seen in the change in uniformity before

and after the flat field corrections were applied). Unfortunately, the graded shielding manufactured,

requires additional lead to prevent the bright spots seen in the CZT images. An optimum configuration

using extra lead required to prevent these spots was used for the measurements and will be used until

the shielding problem is fixed. From the backscatter measurements, the larger pixel size of the CZT

was found to provide better sensitivity, shown as a lower background noise (at least 10% less than the

ICCD detector where no object is present) and a better object to background ratio than the ICCD

detector when using the same experimental setup.

By determining the number of counts detected, the efficiency of the CZT detector was determined.

The highest efficiency of the CZT detector was found to be 70 1 % (at 59.5keV) which is lower than

expected based on the attenuation of the γ-ray photons over the 200keV energy range explored. This is

possibly due to attenuation in the thin carbon window on the aperture, or the 5mm thickness not fully

depleted and able to attenuate and detect radiation photons. Considering the spectroscopic energy

resolution and detection efficiency, this detector was found to have many useful application areas, and

has been shown to be successful as a high energy resolution passive γ-ray detector, and as an active

imager for X-ray backscatter measurements.

Page 107: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

94

Testing on novel SPMs initially involved modelling how the energy resolution varies when using

different sized SPMs with four scintillator crystals and four radioactive sources to activate the crystals,

to explore the possibility of spectroscopy. The model showed that energy resolutions of less than 15%

(and more than 140%) would be possible with some combinations of scintillator crystals, SPMs and

radioactive sources with improving energy resolutions at higher energies.

A vast amount of testing was then performed on novel SPM detector systems providing several key

findings:

The pulse output is directly proportional to the amount of light incident on the SPM

with direct application to radiation spectroscopy, where increased energy will

linearly produce more scintillation light;

The decay constant of the crystals can be accurately seen using the SPM when

exploring the short scintillation (< 1µs) pulses on an oscilloscope;

The pulse preamplifier board used for the SPM measurements is unsuitable for longer

(>1μs) light pulses and at the time of writing a new amplifier arrived which was

proved to work with longer light pulses;

Spectroscopy was found to be possible for various combinations of SPM and

scintillator sources over a wide range of γ-energies. There are direct comparisons

between the model predicting the energy resolutions, and with the spectroscopic

measurements made. Energy resolutions were measured to be from 11% up to

approximately 175%;

When more than one source is used on the scintillator and SPM combination, it was

found that the peaks corresponding to two sources are clearly distinguishable when

large source energy separations are used. The resolution of close peaks is more

difficult, and in some cases impossible with similar energies due to the poor energy

resolutions found previously;

The SPM detectors were shown to count linearly with increasing integration time

using two different SPMs and different scintillator crystals;

Page 108: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

95

The detection efficiency was found to decrease with the increasing source energy as

expected from the modelling of the crystal attenuation modelling. The best SPM

detection efficiency was measured to be 65 1%.

Currently the size limitation of the SPM being only several millimetres, ensures that PMTs are still the

standard for detection of light from scintillator crystals for the majority of applications. However, the

development of arrays of SPMs is underway [17].

According to [6] the energy resolutions achievable using a PMT are 8.5% at 122keV using a NaI

scintillator. Using Equation 20, this extrapolates to an energy resolution of 3.6% at 662keV. The SPM

spectroscopic measurements never matched or improved upon these values however, the best possible

energy resolution did match closely to that modelled proving the SPM detector concept.

Detector Comparison

The pixel pitch of the CZT detector used is 1.6mm which produces coarser images than the ICCD

detector (pitch is 0.5mm). An array of SPMs scaled up to the size of the CZT would have about 870

3mm pixels where the CZT has 6400, and therefore any possible images would be very pixellated

compared to the CZT reducing the spatial resolution. A single 3mm pixel of SPM can count photons at

10MHz which means the 870 pixels able to fit in the area of the CZT would be able to count at about

9GHz, where the same area of CZT is able to currently count at only 100kHz. However, saturation

occurs at fewer counts per pixel on the SPM as this is limited to the number of microcells. Based on

the number of microcells in the 3mm 35μm SPM (providing the best energy resolution), an array of

these SPMs made to the same area as the fully populated CZT detector would be able to detect just

over 3 million simultaneous photon interactions; thirty times more than the CZT detector currently can.

The energy range of the CZT detector used is currently limited from 0-200keV, whereas the SPM has

so far successfully detected from 59.5keV up to 1274keV providing a much larger dynamic range.

However a comparison between the best obtainable energy resolutions (Table 5-1) for two energies

shows that the CZT is far superior (Table 5-1) to the SPMs at both energies.

Page 109: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

96

Detector Best Energy Resolution at

59.5keV (%)

Best Energy Resolution

at 122keV (%)

SPM 55 32

CZT 8 4

Table 5-1: The best obtainable spectroscopic energy resolutions (nearest %) for the two detectors

at the same energies.

The detection efficiency of the CZT detector (Table 5-2) was found to be 6% above the value obtained

for the SPM using CsI(Tl) at the using the same 59.5keV source, showing that both detectors are able

to successfully attenuate photons in the 59.5-122keV energy range. This was predicted from the

attenuation curves, where CsI is able to attenuate slightly less over the energy range with comparable

detection efficiencies.

Detector Best Efficiency at

59.5keV (%)

SPM 65

CZT 71

Table 5-2: The best obtainable detection efficiencies (nearest %) for the two detectors at the same

energies.

Creating an array of 1mm SPMs could provide better spatial resolution that the current CZT provides

are smaller pixels would be used, however, as seen from the spectroscopic results, this would be at the

expense of energy resolution, as the 1mm SPM provided the poorer energy resolutions.

6.2 Further Research

Many measurements were made using both the CZT and SPM detector systems. Due to time

constraints with the equipment, it was not possible to explore as many combinations for spectroscopy

as planned earlier in the year. With more time, a complete comparison with the modelled results

(particularly with the 1mm SPM) would be possible, as with any imaging device, the smaller the pixel

size, used the more coarse the image will be as seen in the images between CZT and ICCD.

Experimentation with a newly available 16 3mm pixel SPM array (Figure 5-1) will allow a small area

imaging device to be tested (to be connected to the 16-channel acquisition system) to determine the

spatial resolution and sensitivity of a new spectroscopic imaging detector. This array would allow a

Page 110: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

97

wider area detector, capable of imaging to be tested for sensitivity for comparison to the CZT and

ICCD for image quality.

Figure 5-1: An array of 16 3mm pixel SPMs [34].

Obtaining an LED pulser should allow the single photoelectron spectrum for each SPM to be collected

to determine the SPM gain. This will provide a direct comparison with PMTs which have well

researched gain of the order of 106.

Repeating all of the spectroscopic experiments with the new transimpedance amplifier (in place of the

pulse preamplifier supplied) should improve the energy resolution for the slower scintillator crystals

(CsI(Tl) and especially CdWO4) and provide another comparison to the PMT energy resolutions.

Further exploration of the lower than expected CZT detector efficiency is required to identify if the full

5mm thickness of the CZT detector is able to detect photons. Additionally, calculating the attenuation

in a carbon window protecting the CZT, to identify if this is the cause of the efficiency loss.

In conclusion to this research project, characterisation experiments of the CZT and the SPM detector

systems was conducted providing key findings to the department. This research project has been

exceptionally interesting to explore over the placement, and it is hoped that the research conducted will

continue to be of use to both the department and the wider physics community.

Page 111: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

98

References The sources, including books, papers and technical data sheets, which have been referred to throughout

the research project are listed below.

1. www.dstl.gov.uk, January 2007

2. Explosive detection systems (EDS) for aviation security, Sameer Singh et al., Maneesha, 2002

3. Introductory Nuclear Physics, Kenneth Krane, 0-471-80553-X

4. Fundamentals of Physics, 6th Edition, Halliday / Resnick / Walker, 0-471-39222-7

5. Oxford Dictionary of Physics, 2003, Oxford University Press, 0-19-860759-8

6. Radiation Detection and Measurement, 3rd Edition, G. F. Knoll, 0-471-07338-5

7. http://bioeng.berkeley.edu/budinger/xraytransmission.html, October 2007

8. Principles of Physics, M. Nelkon, 8th Edition, 0-582-05416-8

9. Properties of Narrow Gap Cadmium-based Compounds, Inspect publication, IEE, 1994, p589 10. Evaluation of a CdZnTe pixel array for X- and γ-ray spectroscopic imaging, F. Quarati et al., 2006

11. Digital Camera Fundamentals, www.andor.com , December 2007

12. http://www.hilger-crystals.co.uk/properties.asp?material=7, December 2007

13. Physical properties of Common Inorganic Scintillators, Saint-Gobain Crystals,

(http://www.detectors.saint-

gobain.com/Media/Documents/S0000000000000001004/SGC_Scintillation_Properties_Chart.

pdf)

14. High-Energy Photon Detection with LYSO Crystals, R.W. Novotny et al., 2006

15. Scintillation: mechanisms and new crystals, M.J. Weber, 2004

16. The quest for the ideal inorganic scintillator, S.E. Derenzo et al., 2003

17. Tiled Silicon Photomultipliers for large area, low light sensing applications, P J Hughes et al.,

SensL, 2007

18. First Results of Scintillator Readout With Silicon Photomultiplier, Deborah J. Herbert et al.,

2006

19. SPM Photon Detection Efficiency Technical Note Rev 1.3, SensL, August 2007

20. http://www.sensl.com/Products/03Silicon_Photomultipliers_-_High_Gain_APDs--

SPMScint_High_Performance_SPM_for_Radiation_Detection.html, November 2007

21. Study of the Properties of New SPM Detectors, A G Stewart et al., 2006

22. Introduction to the Silicon Photomultiplier Technical Note, SensL, Rev1.0 August 2007

23. The Silicon Photomultiplier for application to high-resolution Positron Emission Tomography,

D.J. Herbert et al, 2006

24. The Expression of Uncertainty in Testing UKAS Publication ref: LAB 12, Edition 1, October

2000

25. TCT characterization of different semiconductor materials for particle detection, J. Fink et al.,

2006

26. Email, 04/10/07

27. Email, 09/10/07

28. http://www.britannica.com/eb/topic-529015/scintillation-efficiency, November 2007

29. Email, 20/11/07

30. Industrial Electronics, Noel Morris, 2nd Edition, 0070842256

31. NE555 precision timer datasheet, Texas Instruments, SLFS022, Revised Feb 1992,

http://docs-europe.electrocomponents.com/webdocs/002b/0900766b8002b5d8.pdf, October

2007

32. http://www.kpsec.freeuk.com/555timer.htm, October 2007

33. SPM Pulse Preamplifier Technical Note, SensL, Rev 1.3, August 2007

34. http://www.sensl.com/Products/05Silicon_Photomultipliers_-_High_Gain_APDs--

SPMArray_Position_Sensitive_Multi_Anode_High_Gain_APD.html, October 2007

35. http://www.shinpoly.co.jp/business/connector/english/product/category/detail/af.html, April

2007

36. http://www.xia.com/DGF_Pixie-4.html, November 2007

37. Email, 22/10/07

38. Study of Cadmium Zinc Telluride (CZT) Radiation Detector Modules under Moderate and

Long-Term Variations of Temperature and Humidity, Gunnar Mæhlum, November 2007

Page 112: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

99

Chapter 6 : Appendices

APPENDIX I: OBTAINING A SPECTRUM

Due to the novelty of the acquisition system, a certain amount of “feet-finding” was required, and after

much manual reading, contact with the product manufacturer and general trial and error, the best

procedure for setting up the system was found and can be seen below.

One ‘Quick Start’ method is to load a pre-existing experiment, such as those on the desktop and load the settings (from the ‘Load’ menu on the ‘Settings’ tab). The spectrum can then be refined for the new device by adjusting key parameters. When starting a new series of experiments from a new SPM: 1. Explore the pulses from the detector using an oscilloscope and note the rise time and the decay time.

2. Once logged into the computer, open the "Shortcut to Pixie4" from the desktop. The following screen will appear, and click on 'Start Up System'.

Figure 6-1: The system start-up screen.

3. Once started, the Pixie4 Run Control menu will be displayed, which contains the four tabs (circled in Figure 6-2) to setup the system, and to take and analyse the data.

Page 113: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

100

Figure 6-2: The Pixie4 Run Control menu.

Settings Tab

Note: If these options are not all visible, you should press the 'More' button on the bottom of the menu. As the SPM produces negative pulses, it is necessary to ensure that the 'Trigger positive button' in the channel menu of the settings tab is unselected. The trigger filter is particularly important and should be set a value such that the MCA is not triggered on the noise pulses. (From the oscilloscope the number of real light pulses can be estimated (<50 per second) whilst the noise pulses are in the MHz region). Calibrate Tab Here, the gain can be adjusted to utilise the full 14-bit dynamic range of the instrument. Values of 1.1-1.3 seem best for SPM spectra. Run Tab

The integration time is set here and after a few minutes, a good indication of the general shape of the spectrum should be produced. Whilst running, you can periodically ‘Update’ the MCA view to view the event rate (in the Analyse tab) and see how the energy spectrum progresses. Analyse Tab Here the number of counts per second is displayed, if this is in the hundreds or higher, for a detector which is known not to produce this number of pulses, the trigger and threshold need to be increased. The rise time value and gain should be adjusted and short (10 minute) runs in MCA mode should be obtained to ensure the peak can be distinguished above the noise. Once these values are optimised, take a longer run (30 minutes) to reduce statistical fluctuations, as more counts will be received. Peak Fitting

On a peak of the observed spectrum, a Gaussian curve can be fitted to an energy peak, from which the energy resolution can be found, select a cursor (a square or a circle) and drag it to

Page 114: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

101

the left of the peak, and drag the remaining cursor to the right of the peak. Click on the 'Fit’ down arrow and select the channel holding the data (0 or 1 usually). This will provide values for FWHM% and FWHM abs. The FWHM% is the energy resolution. (An automatic should be available in a newer version of the software.)

Figure 6-3: The positions of the cursors to find the energy resolution.

Figure 6-4: The cursors around the peak to provide the energy resolution and value of the peak.

To calibrate the system, the value for this peak can be entered into the ‘Peak’ box (shown in black) for that channel. Attenuation

Jumper settings on the each channel can be used to provide an attenuation of 7.5x if required (for larger voltage pulses) by changing jumper combinations seen in Figure 6-5.

Page 115: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

102

Figure 6-5: The attenuation jumpers for each channel of the acquisition system.

A jumper on the 50 pins

keeps the input at 50Ohms

(else 5kOhms).

A jumper on the „Attn‟

pins removes 7.5x

attenuation (else there is

7.5x attenuation).

[39]

Page 116: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

103

APPENDIX II: GMS

GMS allows easily identifiable links between „nodes‟ so the items that are linked can be instantly

identified. When the mouse pointer is held over a box (or node), the box and any links with that box

are illuminated easily showing the links. This makes things easy to see when the screen gets busy with

lots of information. Figure 6-6 and Figure 6-7 are screenshots from a GMS tutorial highlighting the

effect of linking, showing nothing highlighted and then with the mathematics students shown.

Similarly, it can be seen which subjects Mark takes by moving the mouse over his name.

Figure 6-6: GMS with no object selected.

Figure 6-7: Identifying who takes Mathematics by moving the mouse over Mathematics.

Weld View is a piece of software which can easily show the relationships between the quantities in the

GMS model such that they can be read from a grid (Figure 6-8).

Page 117: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

104

Figure 6-8: The ‘Weld View’ of the Students Tutorial.

Page 118: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

105

APPENDIX III: Initial Research: Introduction and CZT Resistivity

The following section is based on reports produced during the initial part of the research program.

Gold contacting a CZT sample is performed to connect the sample to a circuit to determine the

characteristics and for practical use of the device (applied voltage required for detector operation). An

experiment explored the use of conducting foam in place of / before gold contacting CZT. It would be

very useful to be able to find the resistivity of a detector sample before the contacting process as

contacting is a costly procedure in both time (taking hours to obtain the correct vacuum for each

contact, where two or three contacts are required in total) and financial benefits (cost of pure gold).

Finding if a sample has a good enough resistivity to continue with experiments before contacting would

therefore be very useful.

Figure 6-9: A schematic of the conducting foam used [35].

A non-contacted CZT sample (about 5x5x1mm) was sandwiched between two separate pieces of

conducting foam, and the current through the sample was measured by applying four voltage ranges in

varying step sizes (Figure 6-10). Resistance was found using the inverse gradient of an IV graph.

With known sample dimensions, the resistivity can be found from Equation 16 to find values shown in

Table 6-1.

Page 119: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

106

Figure 6-10: The experimental arrangement to find the CZT sample resistivity.

Voltage Range

(V)

Step Size

(V)

Readings

/ Sample

Gradient

(1/Ω)

Resistivity

(Ωcm)

-1010 1 Default 7.44E-12 3.05 0.02E+11

-1010 0.1 Default 8.85E-12 2.57 0.02E+11

-11 0.01 10 1.69E-12 1.34 0.01E+12

-11 0.01 40 1.76E-12 1.29 0.01E+12

Table 6-1: Resistivity values and results from the conducting foam experiment.

Higher than expected values of resistivity were found using the non-contacted CZT sample and

conducting foam. The increase in the resistivity is thought to come from a larger than true value being

used for the area in contact with the needle, as the area of the sample was used in calculations, when

only the area in contact with the foam should have been used. Using a smaller value for the area based

on the size of the needle rather than the whole detector sample in Equation 16 will reduce the

resistivity. The experiments are being performed again based on a known value for contact area of a

larger needle, and a more reasonable (lower than previously found) value for resistivity should be

found. This looks like it may be a promising way to find the resistivity of a sample without the need

for gold contacting providing several key benefits, including the potential for money saving, by making

faster resistivity measurements.

Page 120: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

107

APPENDIX IV: Experimental Equipment List

The research project required many of the items to be procured by the author taking time for the

ordering (obtaining the specific requirements, specifications and quotations where necessary), in

waiting for the equipment to arrive, and checking the correct equipment arrived. The testing required

different SPMs, radioactive sources, scintillator crystals and an acquisition system to record the pulses

produced.

Silicon Photomultipliers: These were the main emphasis of the testing and three were ordered to

compare both the physical area and the effect of the fill factor in the detector. These were available as

single pixel detectors, in 1mm SPM 20µm fill factor, 3mm SPM 20µm fill factor and 3mm SPM 35µm

fill factor. The SPM detectors of 1mm pixel size provide an area of 1mm2 and for the 3mm the area is

9mm2.

Scintillator Crystals: Four different scintillator crystals were provided as cubes (3x3x3mm) to provide a

range of light yields from 9,000 to 52,000 photons/Mev in decay times from 0.04 to 1µs to fully test the

response of the SPMs.

Data Acquisition System / MCA: The Digital Gamma Finder (DGF) Pixie-4 is a 14-bit digitising

module allowing 4 simultaneous acquisition and analysis of pulse height for each channel. The system

fits in a standard PCI crate and more modules were added together for acquisition of 16 inputs. “Pixie-

4 Viewer is a graphical user interface written using IGOR from Wavemetrics” operating in the

Windows XP (TM) environment makes the easier to use [36].

There are various settings required so an MCA run can be started, which include the signal rise time,

decay time, gain, run time and signal thresholds. Some of these can be found by first exploring the

pulses on a standard oscilloscope. APPENDIX I provides the details to allow an energy spectrum to be

found.

Page 121: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

108

APPENDIX V: Varying Spectra with Energy

The following figures are the spectra produced from the CsI(Tl) crystal using the 3mm 35µm SPM

using decreasing energies from 662keV (as the main source energy) down to 59.5keV.

Figure 6-11: The change of source energy from 662keV (top) to 59.5keV (bottom).

Page 122: PERFORMANCE TESTING OF X-RAY AND GAMMA … TESTING OF X-RAY AND GAMMA-RAY DETECTORS FOR IMAGING AND SPECTROSCOPY Richard Giordmaina A dissertation submitted to the Physics Department

109

These figures clearly show the energy spectra changing with the source energy. As the peak γ-energy

is reduced, the energy peak moves left to reflect this, whilst the decrease in energy causes the peak to

broaden, decreasing the energy resolution.