pahs in surface water by pda and fluorescence detection

6
Introduction Heightened awareness of polycyclic aromatic hydrocarbons (PAHs) has become prevalent due to urban background levels found in surface water, soil, air, cosmetics and food. They are generated by the combustion of fossil fuels and are always found as a mixture of individual compounds that differ in behavior, environmental distribution, and their effect on biological systems. PAHs encompass a wide molecular weight range, differing based on their physical, chemical, and biological characteristics. Lower molecular weight PAHs, in general, are quite water soluble and have significant acute toxicity to invertebrate aquatic organisms. Naphthalene is structurally the simplest PAH and is especially water soluble. It is the most abundant component of road, driveway and parking lot sealcoating. The U.S. Environmental Protection Agency (EPA) has classified naphthalene as a Group C possible human carcinogen. 1 PAHs in Surface Water by PDA and Fluorescence Detection APPLICATION NOTE Authors: Catharine Layton Wilhad M. Reuter PerkinElmer, Inc. Shelton, CT Liquid Chromatography

Upload: perkinelmer-inc

Post on 16-Aug-2015

26 views

Category:

Science


4 download

TRANSCRIPT

Page 1: PAHs in Surface Water by PDA and Fluorescence Detection

Introduction Heightened awareness of polycyclic aromatic hydrocarbons (PAHs) has become prevalent due to urban background levels found in surface water, soil, air, cosmetics and food. They are

generated by the combustion of fossil fuels and are always found as a mixture of individual compounds that differ in behavior, environmental distribution, and their effect on biological systems. PAHs encompass a wide molecular weight range, differing based on their physical, chemical, and biological characteristics.

Lower molecular weight PAHs, in general, are quite water soluble and have significant acute toxicity to invertebrate aquatic organisms. Naphthalene is structurally the simplest PAH and is especially water soluble. It is the most abundant component of road, driveway and parking lot sealcoating. The U.S. Environmental Protection Agency (EPA) has classified naphthalene as a Group C possible human carcinogen.1

PAHs in Surface Water by PDA and Fluorescence Detection

A P P L I C A T I O N N O T E

Authors:

Catharine Layton

Wilhad M. Reuter

PerkinElmer, Inc. Shelton, CT

Liquid Chromatography

Page 2: PAHs in Surface Water by PDA and Fluorescence Detection

2

High molecular weight PAHs are expected to bio-accumulate in aquatic organisms such as oysters, rainbow trout, bluegills and zooplankton.2 They have low solubility in water and bind very strongly to soils, sediments and particulate matter. Benzo(a)pyrene, a by-product of incomplete combustion or burning of organic material, such as cigarettes, gasoline and wood, is considered to be one of the most carcinogenic high molecular weight PAHs.3

PAHs in surface water result from a variety of sources including residential, industrial and commercial outlets, streets and parking lots, and atmospheric fallout. In general, most surface waters contain individual PAHs at levels up to 0.05 µg/L (ppb), but highly polluted rivers can have concentrations of up to 6 ppb. The PAH levels in groundwater and drinking water are typically within the range of 0.0002 to 0.0018 ppb. Levels of individual PAHs in rainwater range from 0.01 to 0.2 ppb, whereas, in highly urban areas, levels of up to 1 ppb have been detected in snow and fog.4 The EPA has developed ambient water quality criteria to protect human health from the carcinogenic effects of PAH exposure. The current recommendation is a maximum contaminant level (MCL) of no greater than 0.2 ppb in drinking water.5

A pre-concentration step, often solid phase extraction, is necessary to concentrate PAHs in aqueous environmental samples to detectable levels by PDA and fluorescence. In this application, via a spiking experiment, we explore the levels at which PAHs in surface water can be monitored by UHPLC with a sub-2 µm particle sized column combined with photo diode array (PDA) and fluorescence (FL) detection.

Experimental

Hardware/SoftwareA PerkinElmer Altus™ UPLC® system was used, including the A-30 Solvent Module (quaternary pump), Sample Module and Column Module (PerkinElmer, Shelton, CT, USA). A PerkinElmer Brownlee™ HRes DB PAH 1.9 μm 50 x 2.1 mm column (PerkinElmer, Shelton, CT, USA) was used for all separations. Detection was accomplished with PerkinElmer Altus A-30 PDA and FL detectors (PerkinElmer, Shelton, CT, USA). All instrument control, analysis and data processing was performed via Waters® Empower® 3 CDS software.

Method ParametersThe HPLC method parameters are shown in Table 1.

Solvents, Standards and SamplesAll solvents and diluents used were HPLC grade.

A stock PAH standard mixture was obtained from PerkinElmer (Part # 00891543; PerkinElmer, Shelton, CT). The stock mixture contained: naphthalene (1000 ppm), acenaphthylene (1000 ppm), 1-methylnaphthalene (1000 pm), 2-methylnaphthalene (1000 ppm), acenaphthene (1000 ppm), fluorene (500 ppm), phenanthrene (500 ppm), anthracene (1000 ppm), fluoranthene (500 ppm), pyrene (500 ppm), benz(a)anthracene (500 ppm),

chrysene (500 ppm), benzo(j)fluoranthene (500 ppm), benzo(b)fluoranthene (500 ppm), benzo(k)fluoranthene (500 ppm), benzo(a)pyrene (500 ppm), dibenz(a,h)anthracene (500 ppm), benzo(g,h,i)perylene (500 ppm) and ideno(1,2,3-cd)pyrene (500 ppm) in methylene chloride.

In a 25-mL volumetric flask, a stock PAH standard solution was prepared in methanol to equal the final standard concentration of 1 ppm/2 ppm, depending on the analyte. The solution was further diluted in methanol to 0.5 ppm/1 ppm for PDA detection and to 10 ppb/20 ppb for fluorescence detection. These solutions were used to determine linearity at ppm and ppb levels for the PDA and FL detectors, respectively.

Surface water was collected from: 1) the Housatonic River, downstream of a water treatment plant; 2) a suburban driveway with sealcoating; and 3) the surface surrounding a gas station pump area. Each water sample was spiked 1:1 with the PAH standard solution mix. Recoveries for two PAHs, naphthalene and benzo(a)pyrene, were calculated using both PDA and fluorescence detection.

Prior to injection all samples were filtered through a 0.22-µm filter to remove insoluble particles.

Results and Discussion

Figure 1 shows the peak profile of the 19 PAHs at a concentration of 1 ppm/2 ppm. The upper chromatogram, (a), was collected by PDA, while the lower chromatogram, (b), was collected by fluorescence. Separation for all of the 19 PAHs was achieved in less than 4.5 minutes and most were quantifiable down to ppb levels, as specified in Table 1. For fluorescence detection, fluoranthene and indeno(1,2,3,-cd)pyrene are not detected using the excitation and emission wavelengths specified in the method.

Table 1. HPLC Method Parameters.

HPLC Conditions

Column: PerkinElmer Brownlee HRes 1.9 µm 50 x 2.1 mm DB PAH, Part # N9303995

Mobile Phase Gradient:

Solvent A: waterSolvent B: acetonitrile

PDA Detector: Wavelength: 254 nm

FL Detector: Step 1: Ex 280 nm / Em 340 nm Step 2: Ex 280 nm / Em 390 nm at 1.52 mins

Oven Temp.: 35 ºC

Injection Volume: 1 µL

Sampling (Data) Rate: 10 pts./sec

Time (min)

Flow Rate (mL/min) %A %B %C %D Curve

1 Initial 0.700 50.0 50.0 0.0 0.0 Initial

2 4.00 0.700 5.0 95.0 0.0 0.0 6

3 4.50 0.700 5.0 95.0 0.0 0.0 6

4 8.00 0.700 50.0 50.0 0.0 0.0 6

Page 3: PAHs in Surface Water by PDA and Fluorescence Detection

3

As shown in Figure 2, chromatographic repeatability was confirmed via ten injections of the 1 ppm/2 ppm PAH standard mixture, (PDA at 254 nm).

Figure 2. Chromatographic repeatability of the 19 PAH standard mixture by PDA at 254 nm.

Overlay Report

Injection Id 2468, 2464, 2480, 2455, 2492, 2484, 2476, 2472, 2488, 2460

Sample Set Id 2412

Chromatogram Overlay with Z Axis Offset

AU

0.000

0.005

0.010

0.015

0.020

0.025

0.030

0.035

0.040

0.045

0.050

0.055

0.060

0.065

0.070

0.075

Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50

Figure 1. UHPLC chromatogram showing a standard mixture of 19 PAHs at 1 ppm/2 ppm concentrations by a) PDA at 254 nm; b) fluorescence: Ex/Em wavelengths as specified in method.

Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50

12 3

4

56

7

8

910

11

12

14

15

16

1718

19

a)

b)

1. Naphthalene 2. Acenaphthylene 3. 1-Methylnaphthalene 4. 2- Methylnaphthalene 5. Acenaphthene

6. Fluorene 7. Phenanthrene 8. Anthracene 9. Fluoranthene 10. Pyrene

11. Benzo(k)antharcene 12. Chrysene 13. Benzo(j)fluoranthene 14. Benzo(b)fluoranthene 15. Benzo(k)fluoranthene

16. Benzo(a)pyrene 17. Dibenzo(a,h)anthracene 18. Benzo(g,h,i)perylene 19. Indeno(1,2,3-cd)pyrene

13

Page 4: PAHs in Surface Water by PDA and Fluorescence Detection

4

Linearity, at ppm levels by PDA detection and ppb levels by fluorescence, was determined for each of the 19 PAHs using a 1:1 water-diluted stock solution. Representative of the 19 PAHs, the linearity plots for both naphthalene and benzo(a)pyrene are shown below.

In Figure 4, linearity for benzo(a)pyrene between 0.25-2.5 ppm is shown for PDA at 254 nm. Linearity is also demonstrated from 5-50 ppb by fluorescence detection (Ex/Em wavelengths as specified in method).

In Figure 3, we demonstrate linearity for naphthalene between 0.5-5 ppm by PDA detection at 254 nm. Similarly, linearity is demonstrated from 20-100 ppb by fluorescence detection (Ex/Em wavelengths as specified in method).

Figure 4. Linearity plots of benz(a)pyrene a) 0.25-2.5 ppm by PDA; b) 5-50 ppb by fluorescence.

CLCalReportIndivPUBLICATION

Injection Id ****

Sample Set Id ****

Processing Method: PAH_FLR_PM Project Name: PAHProcessing Method ID: 3977 System: Acquity with FLR

5030Calibration ID: ACQUITY FLR ChA Channel:Proc. Chnl. Descr.: ****

1/23/2015 11:53:10 AM ESTDate Calibrated:

Cal Curve Id 5044

Are

a

0

50000

100000

150000

200000

Amount0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00

B

Benzo(a)pyrene FL, ppb

CL Cal Report Individuals

Injection Id ****

Sample Set Id ****

Processing Method: CEL PAHs ppm Project Name: PAHProcessing Method ID: 5925 System: Acquity with FLR

6193Calibration ID: PDA Ch1 [email protected]:Proc. Chnl. Descr.: ****

1/30/2015 8:36:57 AM ESTDate Calibrated:

Area

-10000.0

0.0

10000.0

20000.0

30000.0

40000.0

50000.0

60000.0

Amount0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

A

Benzo(a)pyrene PDA, ppm

R2 = 0.99995 R2 = 0.99988

Figure 3. Linearity plots of naphthalene a) 0.5-5 ppm by PDA; b) 20-100 ppb by fluorescence.

B

Naphthalene FL, ppb

CEL Cal Curve

Project Name: PAHs\PAHReported by User: SystemReport Method: CEL Cal Curve Date Printed:

5361 2/2/2015Report Method ID: 53613:47:55 PM US/EasternPage: 1 of 1

Name: Napthalene; Processing Method: PAH_FLR_PM; Fit Type: Linear (1st Order); Cal Curve Id: 5031; A: -1.474686e+003; B: 3.617692e+003; R^2: 0.999819; Equation Y = 3.62e+003 X - 1.47e+003; Units ppb

Are

a

-50000

0

50000

100000

150000

200000

250000

300000

350000

400000

Amount0.00 10.00 20.00 30.00 40.00 50.00 60.00 70.00 80.00 90.00 100.00

Error LogAll Peaks Table group contains information that doesn't match the data being reported.

A

Naphthalene PDA, ppm

R2 = 0.99988 R2 = 0.99982

Area

-2000.0

0.0

2000.0

4000.0

6000.0

8000.0

10000.0

12000.0

14000.0

16000.0

Amount0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00

Page 5: PAHs in Surface Water by PDA and Fluorescence Detection

5

Table 2. PAHs and their calculated LOQs and LODs, by PDA and FL detection.

Polyaromatic Hydrocarbon RT (min) PDA LOQ

(ppb)PDA LOD

(ppb)FL LOQ (ppb)

FL LOD (ppb)

Naphthalene 0.81 207 62 17 5

Acenaphthylene 0.99 343 103 ND ND

1-Methylnaphthalene 1.12 376 113 8 2

2-Methylnaphthalene 1.18 286 86 9 3

Acenaphthene 1.26 598 179 4 1

Fluorene 1.33 56 17 10 3

Phenanthrene 1.52 26 8 38 11

Anthracene 1.70 13 4 227 68

Fluoranthene 1.94 110 33 ND ND

Pyrene 2.08 146 44 6 2

Benzo(a)anthracene 2.65 52 16 2 1

Chrysene 2.75 33 10 11 3

Benzo(j)fluoranthene 3.12 134 40 11 3

Benzo(b)fluoranthene 3.24 48 14 64 19

Benzo(k)fluoranthene 3.44 74 22 45 13

Benzo(a)pyrene 3.61 57 17 15 5

Dibenzo(a,h)anthracene 3.99 232 69 7 2

Benzo(g,h,i)perylene 4.12 146 44 26 8

Indeno(1,2,3-cd)pyrene 4.25 67 20 ND ND

ND = Not Detected

As listed in Table 2, LOQ and LOD levels were established for each of the PAHs, based upon a s/n of > 10/1 for LOQ and > 3/1 for LOD. It should be noted that for all PAHs, the LOQs established by PDA were below 1 ppm.

As can be observed in Figure 5, no quantifiable PAHs were found in any of the three unspiked, undiluted surface waters analyzed. In order to visualize the amount of PAHs recoverable by PDA and FL detection, surface water samples were spiked with the PAH standard mixture, as shown in Figure 6.

Figure 5. Fluorescence overlay of 10 ppb/20-ppb PAH standard mixture (black) with undiluted surface water from the Housatonic River (blue), a suburban driveway with sealcoating (green), and a gas station pump area (red).

Overlay Report FLR

Injection Id 6461, 5359, 5341, 5353 Sample Set Id 6433, 5328

Chromatogram Overlay with Z Axis Offset

Sample Name: Sunoco puddle filtered 5ul; Date Acquired: 1/26/2015 1:17:57 PM ESTSample Name: driveway filter 5ul; Date Acquired: 1/26/2015 1:43:21 PM ESTSample Name: stratford filter 5ul; Date Acquired: 1/26/2015 1:56:04 PM ESTSample Name: Std2 10ppb + 20ppb; Date Acquired: 2/1/2015 10:46:50 PM EST

EU

-2.00

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

18.00

20.00

22.00

24.00

26.00

28.00

30.00

M inutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50

10 ppb/20 ppb PAH Standard Mixture by Flourescence Housatonic River Suburban Driveway with Sealcoating Gas Station Pump Area

Page 6: PAHs in Surface Water by PDA and Fluorescence Detection

For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs

Copyright ©2015, PerkinElmer, Inc. All rights reserved. PerkinElmer® is a registered trademark of PerkinElmer, Inc. All other trademarks are the property of their respective owners. 012102_01 PKI

PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA P: (800) 762-4000 or (+1) 203-925-4602www.perkinelmer.com

Conclusion

This work has demonstrated the fast, effective chromatographic separation of 19 PAHs using a PerkinElmer Altus UPLC® system with A-30 PDA and FL detectors. The results exhibited exceptional linearity for each PAH over the tested concentration ranges.

Though none of the analyzed surface waters showed any detectable amount of PAHs, the naphthalene and benzo(a)pyrene spike recovery analysis demonstrated the ability of the A-30 PDA and A-30 FL detectors to detect PAHs at ppb levels.

References

1. U.S. Environmental Protection Agency, Hazard Summary – Naphthalene, 2000

2. U.S. Centers for Disease Control Agency for Toxic Substances and Disease Registry (ATSDR) “Toxic Profile for Polyaromatic Hydrocarbons (PAHs)”. 1995.

3. U.S. Environmental Protection Agency, “IRIS Chemical assessment Tracking System: Benzo(a)pyrene”, 2006

4. Prabhukumar, Giridhar; et. al.; “Polycyclic Aromatic Hydrocarbons in Urban Runoff – Sources, Sinks and Treatment: A Review”, 2010

5. U.S. Environmental Protection Agency, Nation Primary Drinking Water Regulations, 2015

Figure 6. Fluorescence overlay of surface water from the Housatonic River (blue), a suburban driveway with sealcoating (green), and a gas station pump area (red), each spiked 1:1 with the 10 ppb/20-ppb PAH standard mixture (black).

Overlay Report FLR

Injection Id 1097, 1088, 1106, 1021 Sample Set Id

Chromatogram Overlay with Z Axis Offset

Sample Name: Acc Sunoco1 10+20ppb oncol; Date Acquired: 2/2/2015 4:05:20 PM ESTSample Name: Acc Drive 1, 10+20ppb oncol; Date Acquired: 2/2/2015 3:02:23 PM ESTSample Name: Acc Stratford 1, 10+20ppb oncol; Date Acquired: 2/2/2015 2:11:52 PM ESTSample Name: Std1 10ppb + 20ppb; Date Acquired: 2/2/2015 12:30:58 PM EST

EU

-2.00

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

18.00

20.00

22.00

24.00

26.00

28.00

30.00

Minutes0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50

10 ppb/20 ppb PAH Standard Mixture by Fluorescence Housatonic River Suburban Driveway with Sealcoating Gas Station Pump Area

Recoveries for naphthalene and benzo(a)pyrene, in each spiked sample, are shown in Table 3. Recovery was determined at both ppm and ppb levels, using both PDA and fluorescence detection.

Naphthalene and benzo(a)pyrene are reported for each water sample, using the PDA and fluorescence detectors at ppm and ppb levels. Recoveries were between 89-107 %.

Table 3. Recoveries of naphthalene and benzo(a)pyrene in spiked surface water collected from the Housatonic River, a suburban driveway with sealcoating, and gas station pump area, each diluted 1:1 with methanol.

PAH of Interest Housatonic River

% Recovery from Surface Water

Suburban Driveway with Sealcoating

Gas Station Pump Area

PDA Detection

1-ppm benzo(a)pyrene 107 104 107

2-ppm naphthalene 102 101 98

FL Detection

10-ppb benzo(a)pyrene 102 100 102

20-ppb naphthalene 89 92 89