organic transformations in ionic liquidsshodhganga.inflibnet.ac.in/bitstream/10603/4058/7/07_chapter...
TRANSCRIPT
Chapter 1
Introduction
Chapter 1
1.1 Luminescence :
1.1.1 Brief history of luminescence:
Our heritage from ancient times has been knowledge of luminescence in
the form of fireflies, glowworms, the lantern fish (Lucerna piscis), Nile fish
(dilyxnos), bacteria, rotten wood, fungus, mushrooms, mollusc, squid, etc. In
addition, the ancients had noted luminescence connected with electrical
phenomena such as the aurora borealis which are northern and southern polar
lights displays in sky usually observed at night, ignis lambens and pollux. The
fairly wide knowledge of luminescence begin with Aristotle (384-322 B.C.) in
Greece [1]. He observed luminescence of flesh, fish and wooden materials. The
Greek geographer Strabo (63 B.C. to 24 A.D.) mentioned dilyxons as a Nile
fish. Nicolas Monardes (1565) observed the light emission from an infusion of
plant Lignum Nephriticum. The emission of light in solids was discovered by
Bolognian Vincenzo Cascariolo (1603) in the form of Bolognese stone which
glows after exposure to light. Robert Boyle (1667) was the first to make
chemical experiments on bioluminescence. He worked on shining wood, fish,
certain, bacteria, fungi as well as on thermoluminescence [2]. Benjamin
Franklin (1752) identified the luminescence of lightning. The red emission of
chlorophyll was noted by Brewster in 1883. Besides of these, Arthanasius
Kricher (Germany), Francesco Grimaldi (Italy), Issac Newton (England), all
have observed the phenomenon of luminescence but they could not explain the
reasons behind it. The first detailed paper on luminescence was in 1852 by Sir.
G. G. Stokes (England) described the theoretical basis for the technique by
giving a mechanism of the absorption and emission process and termed the
phenomenon as fluorescence because the specimens used were of the fluorspar
mineral [3,4].
The term ‘Luminescence’ was first used in 1888 by German physicist
and historian of science Eilhardt Wiedemann. He defined luminescence as ‘all
those phenomena of light which are not solely conditioned by the rise in
temperature’. Wiedemann recognized luminescence as the contrast of
1
Chapter 1
incandescence where luminescence refers to cold light and incandescence
refers to hot light.
Wiedemann basically classified luminescence into six classes-depending
upon method of excitation. By using modern terminology this classification can
be extended into following classes as given in Table 1.1
Table 1.1: Types of luminescence:
Sr. No.
Type of Luminescence
Mode of Excitation
1 Photoluminescence (Fluorescence, Phosphorescence)
UV-Visible light
2 Thermoluminescence Heating after prior storage of energy.
3 Electroluminescence Electrical field
4 Crystalloluminescence Crystallization from solutions
5
Triboluminescence (Piezoluminescence )
Frictional and electrostatic forces
6
Chemiluminescence Chemical reactions
7
Galvanoluminescence Passage of electric current through aqueous solutions
8
Sonoluminescence Intense sound waves
9
Lyoluminescence Dissolution of crystals
10
Bioluminescence Biochemical reactions
11
Radioluminescence Particles emitted from radioactive material
12
Roentgenoluminescence High energy x-rays
13
Cathodoluminescence Cathode rays
14
Ionoluminescence Positive or negative ions
15
Anodoluminescence Anode rays
2
Chapter 1
Luminescent compounds are of very different kinds. Some of them can be
listed as follows:
a) Organic compounds: Aromatic hydrocarbons, coumarins, polyenes.
oxazines, fluorescein, amino acids, etc. b) Inorganic compounds: Uranyl ion (UO2
+), lanthanide ions, glasses with Nd,
Mn, Ce, Sn, Cu, Ag, etc, crystals like ZnSe ZnS,
CdS, CdSe, GaS, GaP, Al2O3, etc.
c) Organometallic compounds: Lanthanide ion complexes, Ruthenium
complexes, complexes with fluorogenic
chelating agents like 8-hydroxyquinoline, etc.
1.1.2 Origin of Photoluminescence:
Photoluminescence is one of the major classes of luminescence in which
substance absorbs photons and attains excited state. Consequently it re-emits
photons to return ground state. The emission of photons accompanying de-
excitation is called photoluminescence. It comprises mainly fluorescence and
phosphorescence.
The first reported observation of fluorescence was made by Spanish
physician Nicolas Monardes in 1565. He described wonderful blue color of an
extract of a wood called Lignum Nepriticum. This phenomenon was further
studied by G. G. Stokes by performing experiments with the solution of quinine
sulphate. Initially he called this phenomenon as ‘dispersive reflexion’ but after
words renamed as fluorescence [5, 6].The name ‘fluorescence’ originates from
‘fluorspar’ or ‘flurospath’ which are minerals containing calcium fluoride and
exhibit the phenomenon of fluorescence.
The term phosphorescence comes from Greek word ‘phosphor’ meaning
‘which bears light’. The term phosphor has been assigned from ancient periods
for those materials which glow in the dark after exposure to light. This property
was earlier reported by V. Cascariolo (1602) from Bologna for the bolognian
phosphor. In early times fluorescence and phosphorescence seems to be
identical as both are relevant to photoluminescence. The distinction between
fluorescence and phosphorescence based on experimental facts was made in
3
Chapter 1
nineteenth century. Fluorescence is an emission of the light which disappears
with end of excitation and in phosphorescence the emission persists after the
end of excitation. The first theoretical distinction between these two
phenomena was provided by Francis Perrin [7].
The twentieth century and specially the period of 1918-48 was prolific
for development of the major experimental and theoretical concepts of
fluorescence and phosphorescence. J.Perrin, Stern, Volmer, F.Weigen, S. J.
Vavilov, W. L. Levshin, F. Perrin, E. Gaviola, E. Jette, W. West, A. Jablonski,
Th. Förster are some of the names who made their efforts to make the concept
of photoluminescence more and more clear [8].
Nowadays there has been tremendous growth in the use of fluorescence
in various branches of science. Fluorescence spectroscopy and time resolved
fluorescence are considered important research tools in biochemistry and
biophysics. Fluorescence is leading technology used extensively in medical
diagnostics, biotechnology, drug analysis, flow cytometry, DNA sequencing,
forensics, genetic analysis, etc. The biochemical application of fluorescence
generally includes anisotropy measurements, resonance energy transfer, etc.
Anisotropy measurements provide information on the size and shape of the
proteins. It has been used to measure protein-protein associations and fluidity
of membranes. Resonance energy transfer has been used to investigate the
binding interactions and to measure molecular distances. The measurements
can provide information on a wide range of molecular process. Fluorescence
spectroscopy will continue to contribute to advancements in biology,
biotechnology and nanotechnology.
1.2 Light and its interaction with matter:
1.2.1 Nature of light:
An electromagnetic radiation, which generally spoken as light, is
characterized by a wavelength (λ) and frequency (υ) which are interrelated as ,
λυ c= (1.1)
where c is the velocity of light = 3 x 1010 cm/s
4
Chapter 1
After striking with matter, light can either pass through matter without
any absorption or it can be absorbed by matter entirely or partially. The energy
from light is absorbed in integral units, called quanta or photon. The
relationship between quanta and energy is given by,
λυ hchE == (1.2)
where E is the energy associated with photon and h is Planck’s constant.
Each molecule has a series of closely spaced energy levels and
absorption of a discrete quantum of light results the jump from lower to higher
energy level.
The mechanism of photoluminescence is based upon absorption and
emission of photon from electromagnetic radiation. It is divided mainly into
two categories fluorescence and phosphorescence depending on the nature of
the excited state. There are two possible arrangements in excited state as shown
in following Figure 1.1
Figure 1.1: Distinction between singlet and triplet excited states.
In ground state, two electrons are paired with each other and according
to equation, M = 2S + 1, where M is multiplicity and S is total spin quantum
number, the multiplicity of ground state is singlet. After absorption of photon,
two probable excitation states can be observed.
5
Chapter 1
(a) the electron in the excited orbital is paired to second electron in the ground
state orbital resulting into singlet multiplicity and hence called as excited
singlet states which are denoted as S0, S1, S2 , etc.
(b) the electron in the excited orbital is parallel to second electron in the
ground state orbital resulting into triplet multiplicity and hence called as
excited triplet states which are denoted as T1, T2, T3, etc.
The absorption of radiation is highly specific and radiation of particular
energy is absorbed only by a characteristic structure.
1.2.2 Fluorescence and Phosphorescence:
On the basis of excited state involved in absorption and emission
process, fluorescence and phosphorescence can be distinguished as follows:
a) Fluorescence
When molecule is in excited state, some energy, in excess, of the lowest
vibrational energy level is rapidly dissipated. If all the excess energy is not
further dissipated by collisions with other molecules the electron returns to the
ground state, with emission of energy. This phenomenon is called as
fluorescence. It involves the transition from lowest excited singlet to ground
singlet state. Because some energy is lost in the short period before the
emission, the fluorescence is of longer wavelength than the energy that was
absorbed. Generally the time period of fluorescence is 10-8 sec.
b) Phosphorescence
The phosphorescence involves the transition from excited singlet state to
excited triplet state and then from excited triplet state to ground singlet state.
This process is highly improbable as it is forbidden process because it involves
electron spin reversal. The characteristic transition times of phosphorescence
are 10-4 to 10 sec. It involves afterglow i.e. emission continues even after the
excitation source is removed. This is because of the relatively long life time of
the triplet state.
1.2.3 Mechanism of fluorescence with Jablonski diagram:
The mechanism of fluorescence can be well explained with the classical
Jablonski diagram, proposed by Professor Alexander Jablonski in 1935 to
6
Chapter 1
describe absorption and emission of light. Prof. A. Jablonski is known as
‘Father of Fluorescence Spectroscopy’ because of his versatile contribution to
the branch of fluorescence spectroscopy studies including descriptions of
concentration depolarization and defining the term ‘anisotropy’ to describe the
polarized emission from solutions [9,10]. The Jablonski diagram has many
forms, one such classical diagram is as shown in Figure 1.2.
The various singlet ground and excited are denoted by S0, S1, S2,........... etc. At each of these electronic energy levels the molecule can exist in a
number of vibrational energy sublevels. Similarly various triplet excited states
are denoted by T1, T2, T3,…. etc.
Following light absorption, which take place usually in about 10-15 sec,
several processes can occur. A molecule is excited to some higher vibrational
level either S1 or S2. The molecule in higher excited state rapidly relax to the
lowest vibrational level of S1. This process is called as internal conversion (IC)
and generally occurs in 10-12 sec. Consequently transition from S1 to S0 is called
fluorescence. Since fluorescence life times are near 10-8 sec, internal
conversion is generally complete prior to emission. Hence fluorescence
emission generally results from the thermally equilibrated excited state. As for
absorption, the electronic transition down to the lowest electronic level also
results in an excited vibrational state. This state will also reach thermal
equilibrium in about 10-12 sec.
The molecule in S1 state can undergo conversion to the first triplet state
T1 which is termed as intersystem crossing (ISC). The emission, following by
transition from T1 to S0 , is called as phosphorescence. It is generally shifted to
longer wavelengths relative to fluorescence. The transition from T1 to ground
state is forbidden and as a result the rate constant for such emission is smaller
than those of fluorescence.
1.2.4 Types of fluorescence:
The fluorescence emission from S1 state is referred to as prompt or
steady state fluorescence and it persists until the excitation is in process. As
soon as the excitation is stopped, the fluorescence emission cuts off. The
7
Chapter 1
phosphorescence is long lived delayed emission having spectral characteristic
very different from fluorescence. However, there are delayed emission whose
Figure 1.2: Jablonski diagram
spectra coincides exactly with prompt fluorescence from lowest singlet state
with only difference being in their lifetimes. These processes having lifetime
property of phosphorescence and spectral properties of prompt fluorescence are
known as delayed fluorescence [11].
1.2.4.1 E-type delayed fluorescence:
When excitation source is cut off, the excited triplet molecules do not
emits immediately due to its longer lifetime. If the energy gap between first
excited singlet and triplet ( ) is comparably smaller as in the case of dye
molecules, then the back energy transfer from triplet to singlet can occur. In
this sequence, the triplet excited molecule in the lowest vibrational level
acquire some thermal energy to go into the vibrational level of isoenergetic
with lowest vibrational level of first singlet state (S
11 TSE −Δ
1). This is followed by
energy transfer between isothermal vibrational level of triplet and first excited
singlet state. The first excited singlet state, consequently deactivates with the
8
Chapter 1
emission of light. This type of fluorescence was first observed in deoxygenated
solutions of eosin in glycerol and ethanol and hence it is referred as E-type
delayed fluorescence. Subsequently similar type of fluorescence was observed
from dyestuffs in fluid solutions [12].
1.2.4.2 P-type delayed fluorescence:
If energy gap between singlet and triplet state (ΔES ـT) is large, the
population of excited singlet state through back energy transfer is not possible.
In such cases, lowest excited singlets are formed in triplet-triplet annihilation
process. The emission occurred from lowest excited singlet state during
deactivation is termed as P-type delayed fluorescence as it was first observed in
pyrene and phenanthrene solutions.
1.3 Photophysical processes:
A photophysical process is defined as a physical process resulting from
the electronic excitation of a molecule or system of molecules by non ionizing
electromagnetic radiation. There are number of photophysical processes to
occur with interaction between radiation and molecule which are mainly
classified as:
1) Unimolecular processes and 2) Bimolecular processes.
1.3.1 Unimolecular processes:
The unimolecular photophysical processes that can occur in an isolated
molecule in the vapour phase at low pressure or in dilute solution in a
transparent medium can be divided into following categories:
a) Radiative excitation (absorption) transitions-
In this process, the molecule is excited from a lower to higher electronic
state by the absorption of photon.
These transitions are sub divided as
i) S0-S1and S0-Sp (P>1) i.e. the absorption from ground singlet state to
any excited singlet state is spin allowed and corresponds to main
electronic absorption spectrum.
9
Chapter 1
ii) S0-T1 and S0-Tq (q>1) i.e. the absorption from ground singlet state to
any excited triplet state is spin forbidden and can be observed with
intense light sources or perturbation methods.
iii) T1-Tq (q>1) i.e. the absorption from triplet state to higher triplet state
which commonly observed by flash photolysis. T1 is populated by
intersystem crossing from S1 which initially excited by an intense
flash light.
iv) S1-Sp (p>1) i.e. the absorption from first excited singlet state to
higher excited singlet state which observed by nanosecond flash
photolysis [13].
b) Radiative de-excitation (luminescence) transitions-
In this process, the molecule is deactivated from a higher to a lower
electronic state by emission of a photon. A radiative transition between states
of the same multiplicity is described as fluorescence while that between states
of different multiplicity is termed as phosphorescence.
These transitions are subdivided as
i) S1-S0 The fluorescence of short duration (~1-103 ns) corresponds to
the normal fluorescence emission.
ii) T1-S0 The phosphorescence which generally occurs with long
duration as it is spin forbidden.
iii) Sp-S0 Such fluorescence is observed in few compounds like azulene
[14].
iv) Tq-S0 This phosphorescence is very improbable process reported in
fluoranthene .
v) Tq-T1 The fluorescence corresponding to the inverse of T1-Tq
absorption reported in azulene and naphthalene [15].
c) Radiationless transitions
These involves transitions between isoenergetic vibrational levels of
different electronic state. Such transitions are normally proceeded by
radiationless thermal activation of the initial electronic state and/or followed by
raditaionless thermal deactivation of the final electronic state. A radiationless
10
Chapter 1
transition between states of the same multiplicity is termed as internal
conversion while that between state of different multiplicity is termed as
intersystem crossing. These transitions are subdivided as[16]
i) S2-S1 and Sp-Sp-1 The internal conversion process usually occurs
rapidly.
ii) T2-T1 and Tq-Tq-1 The internal conversion process usually occurs
rapidly.
iii) S1-S0 The internal conversion to the ground state.
iv) S1-T1 and S1-Tq The intersystem crossing constitutes the internal
quenching of S1 which competes with the normal fluorescence.
v) T1-S0 The intersystem crossing competes with the normal
phosphorescence.
vi) T1-S1 The intersystem crossing may occur by thermal activation of
T1 during its excitation lifetime to a vibrational level isoenergetic
with S1. This leads to E-type delayed fluorescence.
vii) Sp-Tq The intersystem crossing from higher excited singlet states.
1.3.2 Bimolecular processes:
The processes which occur in concentrated or aggregated systems due to
interactions with molecules of the same species (homopolar) or in mixed
molecular systems due to interactions with molecules of different species
(heteropolar) are called bimolecular processes.
The various bimolecular processes are divided into following categories:
a) Perturbation processes:
The interaction with an adjacent molecule may perturb the energy levels
of the excited molecule and modify its photophysical properties and behaviour.
b) Excitation migration:
The interaction between excited and unexcited molecules can lead to the
transfer of its excitation energy either by a radiative process or by a
radiationless process.
11
Chapter 1
c) Complex formation by aromatic molecules:
The heteropolar donar–acceptor complexes are formed between
aromatic hydrocarbons and other appropriate molecules in the ground state and
photophysical properties of these complexes which may be fluorescent or
phosphorescent ,differ from those of their constituent molecules.
d) Complex formation by an excited molecule and an unexcited molecule:
Many molecules in their first excited singlet state ( 1M*) interact with
unexcited molecules(1M) of same species to produce excited dimer (1D*)
1M* + 1M → 1D*
These homopolar excited dimeric complexes are known as excimers.
The excimers are distinct molecular species which exhibits its own
characteristic photophysical properties .
Similarly an excited molecular complex of definite stoichiometry which
is dissociated in the ground state is described as an exciplex. The exciplex
(1E*) is formed by the interaction of molecules in excited singlet state of one
species (1M*) with unexcited molecules of another species( 1Q) 1M* + 1Q → 1E*
e) Interaction between two excited molecules:
This type of interaction is to be observed between two identical
molecules each in the excited triplet state ( 3M*) 3M* + 3M* → 1D*
or 3M* + 3M* → 1M* + 1M
This process yields both excited molecules( 1M*) and excimer ( 1D*) and the
fluorescence of these entities involves P-type delayed fluorescence.
1.4 Processes competing with fluorescence:
The number of photophysical processes occurs as a result of interaction
between matter and radiation as discussed earlier. Though these processes
seems to be quite easily occurring, there are non-radiative processes which
precede or compete with fluorescence. Hence in accordance with the study of
fluorescence, it is important to consider the processes which competes with
fluorescence.
12
Chapter 1
1.4.1 Vibrational relaxation:
It is assumed that at room temperature, before excitation, all molecules
are in the lowest vibrational levels of the ground electronic state. By absorbing
radiation, a molecule is excited to one of the vibrational level of excited
electronic state. After arriving in the excited state, the excited molecule may be
in vibrationally excited state. Then the molecule will start to vibrate with a
characteristic frequency of that state, loosing its excess vibrational energy in
the form of infrared quanta or in the form of kinetic energy imparting to other
colliding molecules. Thus the excited molecule gets relaxed thermally to the
lowest vibrational level of the electronically excited singlet state. In gaseous
state, the deactivation of molecule from same vibrational level to which it is
excited occurs but in the solids and solutions, the excited molecule have to fall
into lowest vibrational level of the excited state before to deactivate. This
process of dissipation of energy in the form of heat and vibrations is known as
vibrational relaxation having life time of 10-14 to 10-12 sec. In this non-radiative
process, the molecule fall into the lowest vibrational level of an excited state
and then emission occurs as stated by Kasha’s rule. Hence when fluorescence
from solution occurs, it involves a transition from the lowest vibrational level
of an excited state [17]. Due to vibrational relaxation, the fluorescence band for
given electronic transition is shifted towards longer wavelengths form the
absorption bands.
1.4.2 Internal conversion (IC):
Internal conversion is a non-radiative transition between two electronic
states of the same spin multiplicity. In solution, this process is followed by a
vibrational relaxation towards, the lowest vibrational level of the final
electronic state.
Internal conversion can be achieved in one of three possible ways:
a) If there is considerable overlap between the lower vibrational level of
the higher electronic state and higher vibrational level of the lower
electronic state then the upper and lower electronic states will be in
13
Chapter 1
transient thermal equilibrium. Then the molecule crossover from higher
to a lower excited singlet state by this vibrational coupling.
b) If there is no considerable overlap but they separated by a small gap,
the molecule in the upper electronic state will convert to the lower
electronic state by tunneling mechanism.
c) If the energy separation of the upper and lower electronic states are
relatively large, the radiative transition takes place to any one of a
number of vibrational levels of the lower electronic state. This radiative
transition is nothing but the fluorescence .
Internal conversion is very rapid process taking about 10-12 sec. The
average lifetime of the lowest excited singlet state is of the order of 10-8 sec .
Therefore even if a molecule can not pass efficiently from its lowest excited
singlet state to the ground state, it may undergo other processes which may
compete with fluorescence [18].
1.4.3 Intersystem crossing(ISC):
Intersystem crossing is a non-radiative transition between two
isoenergetic vibrational levels belonging to electronic states of different
multiplicities. This is spin dependent internal conversion which may be fast
enough taking about 10-8 sec.
For efficient transfer to triplet state , molecule should have to satisfy
following conditions:
a) the energy difference between the lowest singlet state and the triplet
state, just below it, must be small.
b) vibrational coupling should be more between the excited singlet state
and triplet state.
In aromatic hydrocarbons where the singlet-triplet splitting is large, the
ISC is less efficient than in certain dye molecules where triplet-triplet
splitting is small [19]. As ISC occurs, subsequently the molecule undergoes
the IC process and falls to the lowest vibrational level of the first excited
triplet state. Therefore ISC can compete with fluorescence and this it
decreases the quantum efficiency of fluorescence. The population of triplet
14
Chapter 1
state has a significance in producing delayed fluorescence and
phosphorescence, which is radiative decay of triplet state molecule to the
ground state.
The possible de-excitation pathways are summarized in following Figure 1.3
Figure 1.3: Possible de-excitati
on pathways of excited molecules.
1.5 Ch racteristics of fluorescence:
he phenomenon of fluorescence has a number of general characteristics.
The fluorescence spectrum recorded is a fluorescence intensity as a function of
wavelength. The wavelength at which maximum emission takes place is
referred as emission wavelength (λem) while the height of emission peak at λem
gives intensity of fluorescence (F).
1.5.1 Fluorescence intensity:
he intensity of fluorescent light (F) is directly related with the
concentration of fluorescent solute in solution as given by the relation,
F =( I – I ) Φ (1.3)
is
fficient ε of the molecule [20,21].
1.5.2 F
a
T
T
0 F
If ΦF is constant then the shape of the fluorescence spectrum
determined solely by the extinction coe
actors influencing the fluorescence intensity:
Fluorescence intensity of a compound is altered by following factors-
a) Structure of the compound:
15
Chapter 1
It is observed that all organic compounds are not fluorescent but those
which exhibit fluorescence are usually aromatic or contain conjugated double
electronic phenomenon, the molecules having
readily
ne-pair’
electrons.
these
d compound is likely to
be mo if
e the π-electrons, there will be a diminution of
fluores
re fluorescent than benzene and its derivatives. Naphthalene,
anthrac
of the compound to be assayed is very important consideration in
quantit
at high dilutions where the
l enough to make the extent of re-
bonds. As fluorescence is an
available electrons for energy transitions are capable to show
fluorescence. Such electrons are π or delocalized electrons and ‘lo
If a compound contains π-electrons, there is good possibility that it will
show fluorescence and if a substituent, increasing the freedom of
electrons, is added to the compound, then the substitute
re fluorescent than the parent compound [21,22]. On the other hand ,
the substituent tends to localiz
cence.
For example, cyclohexane having no conjugated double bands is non –
fluorescent while benzene, an aromatic compound is weakly fluorescent. In
polycyclic aromatic systems, the number of π-electrons available is greater than
in benzene and therefore these compounds and their derivatives are usually
much mo
ene and biphenyl derivatives [23,24] are much more fluorescent than
the corresponding fluorescent benzene derivatives.
b) Concentration of fluorescent solute:
The intensity of fluorescence is proportional to the concentration of the
fluorescent compound only in highly dilute solutions and therefore the
concentration
ative work [25]. In most fluorimeters, the fluorescence emitted from the
cell holding the solution is measured at right angles to the path of exciting light.
The fluorescence emitted has therefore to pass through the solution to the
detector and during this passage some of it is re-absorbed by other molecules of
the compound under study. Higher the concentration of the compound, greater
is the proportion of the re-absorption. Therefore, linearity between fluorescence
intensity and concentration can only be expected
number of molecules present is smal
16
Chapter 1
absorption trifling compared with the amount of fluorescence emitted.
However, the effect of concentration is dependent to some extent upon
instrumental parameters such as slit width, intensity of the exciting light and
the type of detector. To get reproducible results, slit width and light intensity
need to keep constant during fluorimetric assay.
c) Effect of solvent:
The solvent used for fluorimetric analysis can affect the intensity and
wavelength of fluorescence. The solvent effect can be discussed into three
aspects-
i) Purity of solvent:
Since fluorimetry is a highly sensitive technique, it is important
that the solvent used should themselves be non-fluorescent and free
from fluorescent impurities. These solvents may be used either for
extracting the desired materials or for the actual fluorescence
measurements. Apart from water, number of solvents including
methanol, butanol, ether, hexane, heptane etc may be used for
elated work.
sorbs ultraviolet light and hence it is preferable to clean cuvettes in
chromic acid.
ii) Non
:
of the buffer affect the fluorescence. For example, in case of phosphate
fluorescence r
All solvents should be free from contaminants which may enter
through cleansing agents of glassware. For example, chromic acid
ab
nitric acid rather than
-aqueous solvent:
The fluorescence wavelength can alter depending upon the
physical properties of solvent such as dielectric constant, the association
of solvent and solute by hydrogen bonding, etc. For example, the reports
on indole shown that, the fluorescence wavelength increases with the
dielectric constant of the solvent due to an effect on the π-electrons .
iii)Aqueous buffer solutions
The fluorescence measurements are carried out in aqueous buffer
solutions. In such cases, it is important to know whether the constituent
17
Chapter 1
buffer, an increase of phosphate concentration frequently leads to
diminution in fluorescence intensity.
d) pH
The ef
import
may b
fluores
range of
the pH-fluorescence change. In such
indicators, flu
e) Tem
decrea
molecu ndency for collisions. This would
some of the energy which might have radiated as
same time cause
decomposition. However, even though photo-decomposition
e two characteristic spectra: the excitation
ission spectrum. The large molecule have large number of
electrons and nuclei hence the absorbed energy can be readily distributed
of the solution:
fect of pH upon the fluorescence of a compound is of more
ance. A compound may be fluorescent over limited range of pH and it
e practically non–fluorescent in remaining pH regions. Again it may be
cent over a considerable range of pH, but over a certain section of that
it may be much more fluorescent than over the rest, the working
fluorescent indicator is based on
orescence is visible only in specific pH ranges.
perature:
Fluorescence intensity tends to increase with fall in temperature and to
se to zero at high temperatures. When temperature rises, the motion of
les increases and there is greater te
result in the loss of
fluorescence. With most compounds, a change at 1oC may cause an intensity
change of about 1% [20, 22, 26].
f) Irradiation effect:
The stability of compound when it is radiated by ultraviolet light is an
important consideration in fluorimetry. The extent of photo-decomposition
depends upon the intensity of the light source and as a very intense light source
may enhance the sensitivity of an instrument, it may at the
increased photo-
does occurs, rapid measurements can be carried out before much of the
compound is decomposed. It should be noted that the photo-decomposition
does not always lead to loss of fluorescence and in some cases it leads to the
enhancement of fluorescence [18, 26, 27].
1.6 Experimental observables:
Fluorescent molecules hav
spectrum and the em
18
Chapter 1
am many vibrational and rotational modes. Consequently, their
fluorescence properties are distinctly different from those of small molecules.
Following experimental observables are used to measure the properties
of any luminescent system.
1.6.1 Absorption spectrum:
It shows the dependence of the degree of light absorption by the
compound on the wavelength of light. The quantization condition for the
absorption or emission of light by an atom or by molecule is given by Einstein
relation as given by equation (1.2)
ong
12 EEhc−= h =υ
λ
where E2 and E1 are the electronic energy levels.
The absorption of energy by a molecule is governed by the Beer-Lambert’s law.
According to this relationship,
lcI
..log 010 ε=⎟⎞⎜⎛ (1.4)
I ⎠⎝
where I0 – intensity of incident light,
ted light, I – intensity of transmit
ε – molecular extinction coefficient,
c – concentration of the path length,
l – path length of the absorbing system through which light passes,
and ⎟⎞
⎜⎛
II 0
10log - optical density or ab⎠⎝
sorbance of the material.
In general, the absorption spectrum s plotted in terms of molecular
avelength. The probability of
ound state S00 and the wave function of the
vibrational level of the first excited singlet state S →S .
The positions of the absorption peaks
ution, the broad absorption band is an
indicat cules in the ground state while the
i
extinction coefficient (ε) against frequency or w
the absorption depends upon the degree of overlap of the wave function of the
lowest vibrational level of gr
10 1n
and its nature are of significance
in the spectroscopic studies. In sol
ion of dimeric nature of mole
19
Chapter 1
structu of monomolecular species[28]. But
in soli tructured as in solution. The nature
of abs ular
under study.
1.6.2 Emission Spectrum:
ission spectra. It is observed that the absorption spectra
onal levels of the excited state and the emission
spectra
est vibrational level of S1. This relaxation occur
in abou
y independent of the excitation wavelength [29].
1.6.3 E
red spectrum indicates the existence
ds the absorption spectra are not as s
orption band also give an idea about the lattice structures of molec
system
Emission spectrum defines the relative intensity of radiation emitted at
various wavelengths [18]. The emitted light comprises fluorescence, delayed
fluorescence and phosphorescence, thereby yielding three types of emission
spectra. The Fluorescence emission spectrum shows almost mirror like
symmetry with its absorption.
The delayed fluorescence is specially identical with prompt fluorescence
while the phosphorescence spectrum, although similar in shape is red shifted as
a whole.
The fluorescence emission spectrum is obtained by irradiating the
sample by a wavelength of maximum absorption as given by absorption
spectrum of the sample. The ground state and excited state are associated with
the absorption and em
gives data about the vibrati
yield data about the vibrational levels of the ground state.
The same fluorescence emission spectrum is generally observed
irrespective of the excitation wavelength. Upon excitation into higher
electronic and vibrational levels, the excess energy is quickly dissipated,
leaving the molecule in the low
t 10-12 sec and is presumable a result of strong overlap among numerous
states of nearly equal energy. Because of this rapid relaxation, emission spectra
are usuall
xcitation spectrum:
It defines as the relative efficiency of different wavelengths of exciting
radiation to induce fluorescence. The excitation spectrum is obtained by
measuring the fluorescence intensity at a fixed emission wavelength while the
20
Chapter 1
excitation wavelength is scanned. For large, complex molecules, the excitation
spectrum is quite stable, independent of the emission wavelength.
The excitation spectrum will be identical to the absorption spectrum
where ε.c.l << 1. The measurement of quantum intensity is limited by the
sensitivity of the spectrofluorimeter and that depends upon the intensity of the
excitation source. Parker (1968) estimated that concentrations as low as 10-12
mol dm-3 can be detected by excitation spectroscopy compared with a
minimum concentration of 10-8 mol dm-3 by absorption spectroscopy [30].
Excitation spectroscopy is also used to determine the quantum efficiency of
r and acceptor molecules.
1.6.4 M
e takes place
almost
energy transfer between dono
irror symmetry:
Mirror image symmetry exists between the absorption or excitation
spectrum and the fluorescence emission spectrum as shown in Figure1.4. The
red side of the absorption and excitation spectrum forms a mirror image of the
blue side of the fluorescence spectrum. This is because fluorescenc
exclusively from the lowest vibrational level of the excited state.
Consequently, the absorption spectrum reflects the vibrational levels of excited
states while the emission spectrum reflects those of the ground state. The
mirror symmetry occurs because the vibrational structures seen in the
absorption and emission spectra are similar, as the spacing of the vibrational
energy levels is not significantly altered by the excitation. The absence of
mirror symmetry indicates a strong interaction in the excited state. For example,
excimers have no mirror symmetry [18].
1.6.5 Stokes shift:
Fluorescence radiation always occurs at wavelengths longer than the
exciting wavelength by a wavelength interval depending on the energy loss in
the excited state due to vibrational relaxation. This phenomenon was first
observed by Stokes in 1852 [5]. This separation between the excitation and
21
Chapter 1
Figure 1.4: Characteristic mirror symmetry of excitation and emission
spectra.
Figure 1.5 : Stokes shift.
emission band maxima is known as the Stokes shift as depicted in Figure 1.5. It
is the characteristic of all complex molecules and usually greater than 10 nm.
22
Chapter 1
When the emission band lies within 30 to 50 nm of excitation wavelength,
measurement problems can arise due to difficulty in separating the Rayleigh
scatter of the excitation light from the emission band. The interactions of solu
molecules with the solvent usually introduce large spectral red shifts of
fluorescence. These shifts are occasionally solvent specific and are also called
Stokes shift [18]. The Stokes shift is of interest to analytical chemists since the
emission wavelength can be greatly shifted by varying the form of the molecule
being excited.
1.6.6 Fluorescence quantum yield:
The quantum efficiency Φ denotes the ratio of the total energy emitted
by any molecule per quantum of energy absorbed. Higher the value of
greater the fluorescence observed of a compound. A nonfluorescent molecule is
one whose quantum efficiency is zero or close to zero that the fluorescence is
te
Φ,
rgy absorbed by such a molecule is rapidly lost by not measurable i.e. all ene
collisional deactivation.
The value of Φ can be determined by measuring the fluorescence of
dilute solution of a standard, such as quinine sulphate, whose quantum
efficiency is known. The fluorescence of the new compound is then measured
and the quantum efficiency is calculated as follows-
unk
std
std
unkstdunknown A
AFF .
.. ×Φ=Φ (1.5)
where F is the relative fluorescence determined by integrating the area beneath
the corrected florescence spectrum, Φ is the respective quantum yields and A is
the absorbance. Quantum yield is characteristic for each fluorescent compound
nd emission wavelengths [31].
1.6.7 F
I and the lifetime τ is
and is independent of the excitation a
luorescence lifetime:
The fluorescence lifetime of most organic molecules is in the
nanosecond region. The fluorescence life time refers to the mean lifetime of the
excited state i.e. the probability of finding a given molecule that has been
excited still in the excited state after time t is e-t/τ. The general equation relating
the fluorescence intensity
23
Chapter 1
I = I0 e-t /τ (1.6)
where I is the fluorescence intensity at time t, I0 is the maximum fluorescence
intensity during excitation, t is the time after removal of the excitating radiation
and τ is the average lifetime of the excited state.
The precise measurement of the observe
be used to calculate the natural lifetime τo or the absolute quantum efficiency,
ective than
e includes two wavelengths in the form of
excitat
us sample constituents.
The information regarding excitati
ng the excitation and emission
mono
d bandwidths. When working with a mixture
of fluorescent components, t s to
extent of spectral overlaps.
1.7 Flu
fluorescence.
d lifetime is important since it can
Фo ,if one or the other is known.
τ = Фo τo (1.7)
1.6.8 Synchronous fluorescence spectrum:
The fluorescence and phosphorescence methods are more sel
absorptiometry because thes
ion and emission wavelengths. However, conventional emission scans at
fixed excitation wavelength or excitation scan at fixed emission wavelength do
not fully utilize this advantage. These scans provide useful analytical
selectivity only if there are substantial differences in the absorption and/or
luminescence spectral characteristics of the vario
on and emission spectra can be used
more efficiently if synchronous scanning techniques are used. It is possible to
scan the excitation and emission monochromators simultaneously. Often,
synchronous fluorimetry is carried out by scanni
chromators at the same rate while keeping the wavelength difference
between them constant. The main purpose of synchronous scanning is to
generate spectra having decrease
he synchronous scanning is more advantageou
greatly simplify the spectrum and decrease the
orescence quenching:
The lowering of fluorescence intensity by a competing deactivating
process resulting from the specific interaction between fluorophor and another
substance present in the system is called fluorescence quenching [32]. When
one compound diminishes the fluorescence of another, it is said to quench the
24
Chapter 1
A variety of interactions can result in quenching which include excited
state reactions, molecular rearrangements, energy transfer, ground state
hers.
nching, concentration quenching are the other
main t
Dep
n the fluorophore and quencher.
complex formation and collisional quenching.
1.7.1 Quenchers of fluorescence:
A number of substances act as quenchers of fluorescence. One of the
well known collisional quencher is molecular oxygen [33], which quenches
almost all known fluorophors. Therefore it is frequently necessary to remove
dissolved oxygen to obtain reliable fluorescence yields. Aromatic and aliphatic
amines are also efficient quenchers of most unsubstituted aromatic
hydrocarbons [34]. Heavy atoms such as iodide and bromide can also act as
quenchers. The examples stated above are some of the impurity quenc
Besides these, temperature que
ypes of quenching. In temperature quenching, usually with rise in 10C in
temperature, the intensity decreases by 1%. The concentration quenching
includes the decrease in fluorescence intensity with increase in concentration of
fluorescent solution. For this reason in fluorimetric analysis, dilute solutions
are preferred.
ending upon the mechanism of quenching there are two types:
1) Collisional quenching and 2) Static quenching.
Both of these, require molecular contact betwee
1.7.1.1 Collisional quenching:
The quenching resulting from collisional encounters between the
fluorophore and quencher is called collisional or dynamic quenching. In this
type, the quencher must diffuse to the fluorophor during the life time of the
excited state. Upon contact, the fluorophore returns to the ground state, without
emission of a photon. This is a time dependent process.
Collisional quenching of fluorescence is described by the Stern-Volmer
equation-
[ ] [ ]QKQkFF
Dq +=+= 11 00 τ (1.8)
25
Chapter 1
In this equation F0 and F are the fluorescence intensities in the absence and
presence of quencher respectively, qk is the bimolecular quenching constant, τ0
is the lifetime of fluorophor in abs of quencher andence [ ]Q is the concentration
kq τ0. If
the olmer constant will be
of quencher. The Stern-Volmer quenching constant is given by KD =
quenching is known to be dynamic, the Stern-V
represented by KD, otherwise this constant will be described by KSV.
The quenching data are usually presented as plot of F
0 versus [ ]Q
which is known as Stern-Volmer plot.
F
FF0 is expected to be linearly dependent
upon the concentration of quencher. A plot of FF0 versus [ ]Q yields an intercept
of one -1
oncentration at which
on the Y-axis and a slope equal to KD. It is observed that KD is the
quencher cFF0 = 2 or 50% of the intensity is quenche
linear Stern-Volmer plot is generally indicative
The quenching resulting from ground state complex formation between
rns to the
ground
For static quenching, the dependence of the fluorescence intensity upon
ed
constant for complex formation. This constant is given by,
d. A
of a single class of fluorophors,
all equally accessible to quencher.
1.7.1.2 Static quenching:
fluorophor and quencher is called static quenching. The static quenching
provides valuable data regarding the binding between these two molecules.
The ground state complex formed in static quenching process is non-
fluorescent. When this complex absorbs light, it immediately retu
state without emission of a photon.
quencher concentration is easily deriv by consideration of the association
[ ][ ][ ]QF
QFS .
K −= (1.9)
where [F-Q] is the concentration of the complex, [F] is the concentration of
Q] is the concentrati
complexed species are non-fluorescent then the fraction of the fluorescence that
uncomplexed fluorophore, and [ on of quencher. If the
26
Chapter 1
remains (F/Fo) is given by the fraction of the total fluorophores that are not
complexed.
The total concentration of fluorophore [ ] 0.F is given by
[ ] [ ] [ ]QFFF −+=0. (1.10)
by substituting equation ( 1.10) into equation (1.9) we get,
[ ] [ ] [ ][ ][ ] [ ][ ] [ ]QQFQF
K S ..−== (1.11)
We can substitute the fluorophore concentration for fluorescence intensities
and by rearranging equation (1.11)
FFF 10.0 −
[ ]QKFF
S+=0
The dependence of
1 (1.12)
FF0 on [ ]Q is linear as the case in dynamic quenching.
1.7.1.3 Comparison of static and dyna
crease in temperature, the stability of complexes decreases and
conseq
.8 Fluorescence spectrometry:
The fluorescence and fluorescence excitation spectra of the solutions of
arious drug samples as donor and acceptor m
1.6. It has following specifications:
mic quenching:
Dynamic and static quenching constants have different dependency on
temperature. Dynamic quenching depends upon diffusion. With rise in
temperature, diffusion coefficients increases and hence bimolecular quenching
constants are expected to increase with increasing temperature. In contrast to
this, with in
uently it decreases the static quenching constants.
1
1.8.1 Instrumentation:
v olecules in different solvents
were recorded on P. C. based spectrofluorimeter. The experimental set up is
shown in the photograph i.e. Figure
27
Chapter 1
During recording the fluorescence and fluorescence excitation spectra
the parameters like spectral bandwidth (10 nm), data pitch (1 nm) and
wavelength scanning speed (250 nm/min) were kept constant. The other
parameters such as excitation wavelength, emission wavelength were varied as
per the requirement of the experiment.
1.8.2 Optical system of FP-750 spectrofluorimeter:
The optical system of the instrument is given in Figure 1.7. The light
from the source (Xenon lamp) is focused on the entrance slit of the excitation
monochromator by the ellipsoidal mirror M1 and spherical mirror M0. The light
Holographic grating with 1200 lines/mm
: 220 nm to 730 nm
Spectr
avelength scanning speed : 60, 250, 1000, 4000 nm/min
esponse : Fast, Medium, Slow, Auto
ensitivity : Signal to noise ratio of Raman band of
water is higher than 300:1
hotometric display : -999 to +999
ample chamber : Single cell holder (standard)
Detector : Silicon photodiode for Ex.
monochromator and Photomultiplier tube
for Em. monochromator
Instrument : PC based spectrofluorophotometer
Make : JASCO, Japan
Model : FP-750
Light source : 150 W xenon lamp with shielded
lamphouse
Monochromator :
Wavelength range
al bandwidth : 10, 20 nm on both Ex. and Em
monochromator
Wavelength accuracy : ±3 nm
Wavelength threw speed : 30,000 nm/min
W
R
S
P
S
28
Chapter 1
from the slit is dispersed by the diffraction grating G1 and monochromatic light
by the exit slit. A a d to the
ring silicon photodiode, P of splitter, BS, while the
omatic light that has transmitted the beam splitter is led to the sample
e plane mirror M an s focused
on the centre of the sample cell. Th focused on to
f the emission o 4
ors M5 and M . am is taken out from the
he diffractio
after going through the exit slit an etric photomultiplier tube
irror M7
is taken out p rt of the monochromatic light is le
monito S , by the beam
monochr
chamber by th 2 d ellipsoidal mirror M3 where it i
e emission from the sample is
the entrance slit o m nochromator (Em) by ellipsoidal mirror M
and two plane mirr 6 Monochromatic be
light dispersed by t n grating G2 of the emission monochromator
d is led to photom
PMT by the spherical m .
29
Chapter 1
30
Figure 1.6: The experimental setup of the fluorimeter
Chapter 1
Figure 1.7: Optical system of FP-750 Spectrofluorimeter
31
Chapter 1
1.8.3 Detecting and recording system:
The schematic diagram for the FP-750 system is shown in Figure1.8.
The light incident on the monitoring detector (silicon photodiode) and the
emission detector (PMT) is converted into an electrical signal and then
converted into a digital signal by the A/D converter and is introduced to the
microcomputer. The signal subjected to arithmetic operation by the
microcomputer is outputted to the display unit as digital data or spectrum. Both
wavelength as well as slit drives were controlled by the microcomputer.
1.8.4 Operating procedure:
The steps involved during recording of fluorescence and fluorescence
excitation spectra of samples under study are explained as follows-
i) Visual fluorescence color was observed by exciting the sample at
365nm (Hg line) excitation wavelength.
ii) The emission monochromator was set at the approximate wavelength
of visually observed color.
iii) The excitation monochormator was scanned from 250 nm to a
wavelength of emission monochromator.
iv) The excitation spectrum was recorded and the λex was noted.
v) The excitation monochromator was set at λex observed in excitation
spectrum.
vi) The emission monochromator was allowed to scan in the range 300
nm to 750 nm.
vii) The fluorescence emission spectrum was recorded and the λem was
noted.
viii d excitation
spectrum monochromator was scanned and thus the excitation
spectrum is recorded
ix) Finally the fluorescence spectrum was obtained by setting the
excitation monochromator at λex obtained in step (viii). Similarly
fluorescence excitation spectrum was obtained by setting λem
observed in the final emission spectrum.
) The emission monochromator was then set at the λem an
32
Chapter 1
Figure 1.8
: System diagram
33
Chapter 1
1.8.5 Characteristics of an ideal spectrofluorimeter:
To achieve correct analysis by spectrofluorimetric method, the
components of instrument must posses following characteristics-
1.8.5.1: The light source must yield a constant radiation output at all
wavelengths. At present the most versatile light sources are the high pressure
xenon arc lamps. These lamps provide relatively continuous light output from
270 to 700 nm. In addition, the operation of these lamps does not generate
ozone in the surroundings. Xe lamp have useful life of about 2000 hrs. The
lamphouse provides more safety to lamp and analyzer. The high pressure
mercury lamps have higher intensities than Xe lamps, but the intensity is
concentrated in lines. Mercury lamps are only useful if the Hg lines are at
suitable wavelengths for excitation of the fluorophore. Xe-Hg arc lamps, low
pressure Hg lamps are other examples of light sources but these are less
superior than Xe arc lamps.
1.8.5.2: The monochromator must pass radiation of all wavelengths with equal
efficiency. In most of the spectrofluorimeters the diffraction gratings are used
as monochromators. The performance of a monochromator depends on the
dispersion and the straylight levels. For best results, low stray light levels are
required to avoid problems due to scattered stray light. The grating
monochromators may have planar or concave gratings. Planar gratings are
mechanically produced and may contain imperfections while concave gratings
are usually produced by holographic and photo resist methods and having less
imperfections. Imperfections of the gratings are the major source of stray light
transmission by the monochromators and of ghost images from the grating. For
is reason, the holographic gratings are preferable.
.8.5.3: The monochromator efficiency must be independent of polarization.
f monochromator is dependent upon orientation
f polarizer either vertical or horizontal. The polarization characteristics of
onochromators have important consequences in the measurement of
uorescence. Such measurements must be corrected for the varying efficiencies
f each component.
th
1
The transmission efficiency o
o
m
fl
o
34
Chapter 1
1.8.5.4: The detector must detect radiations of all wavelengths with equal
rtional to the
- light absorption by donor.
hoton by the acceptor. The
efficiency. Almost all fluorimeters use photomultiplier tubes (PMT) as
detectors. It is best regarded as a source of current, which is propo
light intensity.
1.9 Energy transfer phenomenon:
1.9.1 Electronic energy transfer mechanism:
The electronic energy mechanism has become one of the most useful
processes in photochemistry having wide applications as a mechanistic tool and
in photochemical synthesis. It allows photosensitization of physical and
chemical changes in the acceptor molecule by the electronically excited donor
molecule. There are two types of energy transfer mechanism-
1.9.1.1 Non-radiative energy transfer:
The process can be defined by the following two steps-
D + hυ → D*
D* + A → D + A* - energy transfer from donor to acceptor.
The electronically excited donor D* is formed initially by direct light
absorption. This can transfer the electronic energy to a suitable acceptor
molecule A resulting in simultaneous quenching of D* and electronic
excitation of A to A*. The transfer occurs before D* is able to radiate and hence
is known as non-radiative transfer of excitation energy. The A* molecule thus
excited indirectly can undergo various photochemical and photophysical
processes. Such processes are called photosensitized reactions.
1.9.1.2 Radiative energy transfer:
The radiative energy transfer involves the trivial process of emission by
the donor and subsequent absorption of the emitted p
process takes place as-
D* → D + hυ
A + hυ → A*
It is called trivial because it does not require any energetic interaction
between the donor and the acceptor. It is merely reabsorption of the
35
Chapter 1
fluorescence radiation. Though it is called trivial, it causes radiation
imprisonment and may introduce error in fluorescence measurement [11].
1.9.2 Förster (Fluorescence) Resonance Energy Transfer (FRET):
It is a physical phenomenon described over 50 years ago, that is being
l research and drug discovery today. It
r mechanism between two
chrom
rst paper
portance of energy in photosynthesis.
sonance energy
a)
e intensity and excited state lifetime
and increase in the acceptor’s emission
bove, there should be proximity between donor and
acceptor (10-100 Å) and absorption/ excitation of acceptor must overlap with
emission spectrum n in figure 1.9.
used more and more in biomedica
describes a non radiative energy transfe
ophores. It is generally acronymed as FRET.
FRET relies on the proximity/ distance dependent transfer of energy
from a donor molecule to an acceptor molecule. This energy transfer
mechanism termed Förster resonance energy transfer named after the German
scientist Theodor Förster. By 1946, Professor Förster had written his fi
on energy transfer and pointed out the im
When both molecules are fluorescent, the term fluorescence re
transfer is used.
The efficiency of FRET depends on the following parameters-
the distance between donor and acceptor molecules,
b) the extent of overlap of the emission spectrum of the donor with the
absorption spectrum of the acceptor,
c) the relative orientation of the donor and acceptor transition dipoles,
d) the quantum yield of the donor.
FRET is the radiationless transmission of energy from a donor molecule,
which initially absorbs the energy, to acceptor molecule. The transfer of energy
leads to reduction in donor’s fluorescenc
intensity. A pair of molecules that
interacts in such manner that FRET occurs is referred as donor-acceptor pair.
As mentioned a
of the donor as show
36
Chapter 1
with emission of
1.9.3 C
rster
dis 0 nge of 20 to 60 Å [35].
The ra
Figure 1.9: Spectral overlap of excitation of acceptor
donor.
haracteristics of FRET:
The distance at which energy transfer is 50 % efficient is called Fö
tance (R ), which is typically in the ra
te of energy transfer from a donor to an acceptor is given by-
( )6
01⎟⎠⎞
⎜⎝⎛=
rR
rkD
T τ (1.13)
where Dτ is the decay time of the donor in the absence of acceptor, R0 is the
Förster distance and r is the donor to acceptor distance. Hence the rate of
transfer is equal to the decay rate of the donor ⎟⎟⎠
⎞⎜⎜⎝
⎛
Dτ1 when the D-A distance (r)
is equal to the Förster distance (R0) and the transfer efficiency is 50 %. At
37
Chapter 1
r = R0, the donor emission would be decreased to half its intensity in the
absence of acceptor. Thus the rate of FRET depends strongly on distance and is
proportional to r -6 [29].
Förster distances in the range 20-90 Å are convenient for studies of
biological macromolecules, vitamins, drug molecules etc. Any condition that
affects the D-A distance will affect transfer rate, allowing the change in
distance to be quantified. In such applications, the extent of energy transfer
between a fixed donor and acceptor is utilized to calculate the D-A distance
and thus to obtain structural information about the molecule. Such distance
measurement is important aim of FRET and hence it is called as ‘spectroscopic
ruler’ [36, 37]. The use of the energy transfer as a proximity indicator
illustrates an important characteristic of energy transfer. It is observed that
FRET will occur if the spectral properties are suitable and the D-A distance is
comparable to R0. A wide variety of biochemical interactions results in changes
in distance and thus can be calculated using FRET.
1.9.
1.9. r of energy:
energy transfer have been derived from
4 Theoretical aspects of energy transfer:
4.1 Rate of transfe
The theory of resonance
classical and quantum mechanical considerations. The rate of transfer for a
donor and acceptor separated by distance r is given by [29]-
( ) ( ) ( ) ( ) λλλελ dFKrk ADD
T4
456
2 10ln9000∫πτ NnrD 0128
∞
⎠⎝
N = Avogadro’s number
⎟⎞
⎜⎛Φ
= (1.14)
where DΦ = the quantum yield of the donor in absence of acceptor,
n = refractive index of medium,
r = is the distance between the donor and acceptor ,
Dτ = is the lifetime of donor in absence of acceptor ,
( )λDF is the corrected fluorescence intensity of the donor in the wavelength
range λ to λ + ∆λ with the total intensity (area under curve) normalized to unity,
38
Chapter 1
( )λε A is the extinction coefficient of the acceptor at λ. The term K2 is a factor
describing the relative orientation in space of the transition dipoles of the donor
and acceptor. K2 is usually assumed to be equal to 2/3 which is appropriate for
dynamic random averaging of the donor and acceptor.
The overlap integral (J) expresses the degree of spectral overlap
between the donor emission and the acceptor absorption.
( ) ( )( ) ( )
( )∫
∫∫ ∞
∞
∞
==
0
0
4
0
4
λλ
λλλελλλλελ
dF
dFdFJ
D
AD
AD (1.15)
F
D (λ) is dimensionless. If ( )λε A is expressed in units of M-1cm-1 and λ is in
nm, then J is in units of M-1cm-1 (nm)4.
1.9.4.2 Förster distance (Ro):
For biochemical processes it is usually convenient to consider distances
than transfer rates. For this reason equation 1.14 is written in terms of the
ecules decay by energy transfer
with
Förster distance R0 at which half the donor mol
( ) 1−= DT rk τ ,following equation is from equations (1.13) and (1.14)
obtained [29].
( ) ( ) ( ) λλλελπ
dFnNKR AD
D 4
045
26
0 .12810ln.9000
∫∞
expression, Förster distance can be calculated from the spectral
properties of the donor and acceptor and the donor quantum yield.
This expression can be made simpler to calculate R0 in terms of the
g the
ngth is expressed in nm then
Φ= (1.16)
From this
experimentally known values which is accomplished by combinin
constant terms in equation (1.16). If the wavele
( )λDF is in units of M-1cm-1(nm)4 and the Förster distance in Å is given by –
( ) 6/1420 211.0 JnKR DΦ= − (1.17)
( )JnKR DΦ×=∴ −− 4256 1079.8 (1.18) 0
length is expressed in cm and J is in units of M-1cm3 then,
If the wave
( )JnKR −− 42256 1079.8 (in cm6) (1.19) DΦ×=0
39
Chapter 1
The
the transfer rate is much
faster t
otons absorbed by the
donor which are transferred to the acceptor [29]. The fraction is given by
rate of energy transfer can be easily calculated by knowing R0 from
equation (1.13).
1.9.4.3 Efficiency of energy transfer:
The energy transfer will be efficient only when
han the decay rate. If reverse is the case then FRET will be inefficient.
The efficiency of energy transfer (E) is the fraction of ph
( )( )rk
rkETD
T
+= −1τ
which is the ratio of the transfer rate to the total decay rate of the donor in the
(1.20)
presence of acceptor. By substituting equation (1.13), equation (1.20) can be
rearranged as
660 rR
E+
= (1.21)
From this equation it is clear that the efficiency of the energy transfer is
strongly dependent on distance when the D-A distance is near R
60R
nsfer efficiency is typically measured using the relative
fluorescence intensity of the donor, in the absence (F0) and presence (F) of
acceptor.
0. The
efficiency quickly increases to 1 as the D-A distance decreases below R0.
The tra
0
1FFE −= (1.22)
1.9.4.4
that o
yield of the donor
Various terms involved in calculations:
To calculate the D-A distance, it is necessary to know R0, which in turn
depends upon K2, n, DΦ and J. The refractive index is often assumed to be near
f water (n =1.33) or small organic molecules (n =1.39). The quantum
DΦ is determined by comparis
The overlap integral must be evaluated for each D-A pair. The greater the
overlap of the emission spectrum of the donor with the
tinction
coefficients result in larger R0 values .The orie 2 is nt
on with standard fluorophors.
absorption spectrum of
the acceptor, the higher the value of R0. Acceptors with larger ex
ntation factor K depende
40
Chapter 1
upon geometrical considerations of emission transition dipole of the donor and
the absorption transition dipole of the acceptor. It is generally assumed equal to
acceptors that randomize by rotational
mon applicati
between two sites on a macromolecules.
d to study macromolecular system when there is more than a
ule near a d
d) It can be used to measure binding interactions between molecules in
these
acellular probes and for making the surfaces functional. FRET can be
such inte
f) FRET has been used to develop a number of sensors. The use of donor
because the measurements
2/3, which is the value for donors and
diffusion prior to energy transfer [29].
1.9.4.5 Applications of FRET:
FRET has number of applications in various areas of science. Some of
these are summarized as follows:
a) The most com on of FRET is to measure the distances
b) It can be used to measure the extent of binding. The steady state
measurements are often used to measure binding interactions.
c) It is use
single acceptor molec onor molecule.
solution [38] or in microscopy [39-40].
e) Quantum dots [QDs], semiconductor nanoparticles have high quantum
yields, narrow emission spectra and good photostability and hence
are widely used as fluorescent probes. The knowledge of the
interactions of QDs with proteins is important for their use as
intr
used in understanding ractions [41].
to acceptor intensity ratios is valuable
become mostly independent of the overall intensity. This independence
is important in fluorescence microscopy where it is not possible to
control the local fluorophore concentration. This property has been used
to develop sensors for a variety of analytes [42].
g) FRET is used extensively in DNA analysis. The donor-acceptor pairs
can be designed which display distinct emission spectra and similar
intensities with a single excitation wavelength. Such pairs with their
FRET mechanism can be used for DNA sequencing [43,44].
41
Chapter 1
h) FRET is to be used with fluorescence life time imaging microscopy
(FLIM) to study association reactions in cells [45-46].
1.10 Binding Mechanism:
When the interaction occurs between two different molecules, the
constant and number of binding sites can be
determi
binding parameters like binding
ned by following equation [47].
[ ]QnKF
F
FFloglog 0 =
−log+ (1.23)
where ssed earlier, K is the binding
con
interac
easily
0 and F have same meaning as discu
stant, n is the number of binding sites for that particular molecular
tion and [Q] is the concentration of quencher.
The binding constant (K) and the number of binding sites (n) can be
determined by plotting the graph of F
FF −0log versus [ ]Qlog . The nature
of this plot will be a straight lin Y-axis. The slope
determ
constant of that interaction.
1.11 Thermodynamic parameters:
(ΔS) and the enthalpy change (
fluorescenc o
determ
quenchi
recorded at various temperatures by
binding constants are related with
equation,
e with intercept on
ines the number of binding sites while intercept gains the binding
One can determine the change in free energy (ΔG), the entropy change
ΔH) for any particular interaction by using
e quenching data. By applying equation 1.23, it is possible t
ine binding constants at various temperatures, with fluorescence
ng measurements. The fluorescence of the system under study can be
keeping it in thermostat. The various
different temperatures by van’t Hoff
RS
RTHK Δ
+Δ
−=ln (1.24)
temper
accom
here, K is binding constant, R is the gas constant, T is absolute
ature, ΔH is change in enthalpy and ΔS is change in entropy
panying the molecular interaction.
42
Chapter 1
obtain
G = ΔH –TΔS (1.25)
n the different molecules essentially
includes, hydrogen bond, van der Waals’ force
hydrophobic interaction [48]. Ross summed up the thermodynamic laws for
force while if both are
ion mainl py dr
ΔH ≈ 0 and ΔS have positive value then there will be electrostatic force
sfer was introduced and developed
ET [54].
Förster calculated the critical distanc
hodamine B in glycerol,2-5 diphenyl
oxazole, 9-10 dibromoanthracene in cyclohexane, etc [55]. Förster himself
Subsequently, free energy change at different temperatures can be
ed by Gibb, s equation,
Δ
Thus, fluorescence study enables to get idea about thermodynamic
parameters. The nature of binding forces can be predicted by observing these
parameters. The acting forces betwee
s, electrostatic interactions and
estimating the type of the binding force between organic micromolecule and
biological macromolecule. If ΔH and ΔS are both positive or having some
higher values the main force will be hydrophobic
negative or some lower values then hydrogen bond and van der Waals’ forces
will be key forces of interaction and react y enthal iven. Also if
between the acting molecules. The negative value of ΔG will indicate
spontaneity of reaction and vice-versa[49-52].
1.12 Literature survey on FRET studies:
The concept of resonance energy tran
by Professor Theodor Förster. J.Perrin was the first person to realize the
interaction from one molecule to another through interactions between
oscillating dipoles of closely placed molecules [53]. Förster proposed a non-
radiative mechanism for FRET which includes long range dipole-dipole
interactions and energy transfer between donor and acceptor which arises from
mutual resonance dipole perturbation. He explained the energy transfer
phenomenon in photosynthesis in his first paper and developed a correct
theoretical basis of FR
e between donor and acceptor at
which the efficiency of energy transfer will be 50 % which is called as Förster
distance [35]. A.Kawasaki et al confirmed the Förster’s theory of RET with
various donor-acceptor pairs containing r
43
Chapter 1
confirmed the validity of energy transfer theory with fluorescein and
rhodamine B in glycerol and with fluorescein and rhodamine 6G [56].
J.Hung et al [57] st
-diethylaniline.
he interaction of BSA with 6-
mercap
ct was utilized by L.Matyus et al [63].The
interac
udied the interaction between antileishmanial drug,
Boldine, with the promastigotes of leishmania mexicana and applied the FRET
to determine the mechanism of biological action of the drug. Y.Li et al [58]
applied FRET for studying the molecular interaction between dibucaine and
acriflavine in aqueous as well as in polystyrene latex dispersions. Dibucaine is
local anesthetic drug which was donor and acriflavine, having role of sensitizer
in photodynamic therapy, was acceptor molecule. The studies shown that the
molecular interaction involved electrostatic and hydrophobic forces.
The binding of antimicrobial drug, Gatifloxacin with BSA molecule was
studied by M.Guo et al [59] by using fluorescence quenching mechanism. The
binding parameters as well as effect of drug on conformation of BSA has been
analysed by synchronous fluorescence spectroscopy. M.H.Zhang et al [60]
reported the fluorescence study between Chinese herbal medicine Hypocrellin,
having activities against skin diseases, and N-N
Y.J. Hu et al [61] investigated t
topurine (6-MP), an antiviral agent of purine series. The change in
conformation of BSA in presence of 6-MP, dynamic quenching of BSA
fluorescence by 6-MP and thermodynamic parameters were studied. S.
Chatterjee et al [62] observed the interaction between anionic dye fluorescein
and cationic dye safranine T in aqueous as well as micellar medium which
includes the effect of viscosity on the FRET by fluorescence anisotropy study.
Fluorescence quenching studies used to obtain adequate information
about the structure and dynamics of biologically important macromolecular
systems like proteins. This fa
tion between well known dye methylene blue and BSA was studied by
Y.J.Hu et al [64] with respect to FRET and all corresponding parameters. The
fluorescence quenching of HSA by antidepressant drug doxepine hydrochloride
involves static quenching mechanism was concluded by P.B.Kandagal et al
[65]. The study included determination of intermolecular distance between
44
Chapter 1
donor and acceptor, critical distance, thermodynamic parameters of binding
process, etc. Further more it pointed at the possible binding site between
protein
studied by
X.Che
nor. From this
interac
,
interfe
and drug.
C.Wang et al [66] studied the interaction of carbamazepine with BSA
and determined the binding, thermodynamic parameters. The fluorescence
quenching study between riboflavin and serum albumins was carried out by
Z.Hongwei et al [67]. The binding constants and binding sites were obtained at
various temperatures.
An antibiotic ,lomefloxacin is a drug belonging to fluoroquinolones. The
energy transfer between bovine lactoferrin and lomefloxacin was
n et al [68].The change in conformation of bovine lactoferrin is
determined by this study. The binding mechanism of anti inflammatory drug
cromolyn sodium to BSA was investigated by Y.Liu et al [69] with respect to
nature and magnitude of interaction.
The flavonoid nevadensin acts as quencher of lysozyme fluorescence
intensity. Lysozyme is small monomeric protein acts as do
tion, critical distance of energy transfer, the distance between donor and
acceptor, etc were calculated by D.Li et al [70].An alkaloid class drug brucine
interacted with HSA which was investigated by Y.Q.Wang et al [71]
spectrofluorimetrically from which binding parameters has been calculated.
From this review, we planned to utilize fluorescence properties of some
drug molecules to investigate drug-drug, drug-vitamin, drug-BSA molecular
interactions from which the binding parameters, drug absorption, distribution
rence of coexisting species can be determined.
45
Chapter 1
1.13 References:
1] E.N. Harvey, A History of Luminescence: From the Earliest Times
Until 1900,The American Philosophical Society, Philadelphia, (1957).
2]
w York, (1968), 90.
4]
6), 9-38.
9]
4),
., 23(8), (1955), 1390-
1391.
5] H. Schuler,G. Arnold, Z. Naturforsch.,16a, (1961), 1091-1092.
6] J.B. Birks, Photophysics of Aromatic molecules, John Wiley and sons
Ltd., London, (1970).
7] C.A. Parker, Photoluminescence of Solutions, Elsevier Publication,
Amsterdam, (1988).
D.M. Needham, Machina Carnis: The biochemistry of muscular
contraction in its historical development, Cambridge University Press,
(1971).
3] H.V. Laverenz, An Introduction to Luminescence of Solids, Dover
Publications, Ne
T.C. O’Haver, J. Chem. Educ., 55, (1978), 423-428.
5] G.G. Stokes, Phil.Trans.,142, (1852), 463-562.
6] G.G. Stokes, Phil.Trans., 143, (1853), 385-396.
7] F. Perrin, Doctoral thesis, Annales de Physique, Paris,12, (1929), 2252-
2254.
8] B. Nickel, Newsletter, 58, (199
A. Jablonski, Z.Phys., 94, (1935), 38-46.
10] J. Szudy, Born 100 years ago: Aleksander Jablonski (1898-1980),
Uniwersytet Mikolaja Kopernika,Torun, Poland (1998).
11] K.K. Rohatgi-Mukherjee, Fundamentals of Photochemistry, New Age
International (P) Ltd., New Delhi, (2006).
12] C.A. Parker, C.G. Hatchard, T.A. Joyce, J.Mol.Spectro., 14, (196
311-319.
13] (a) R.G.W. Norrish, G. Porter, Nature, 164, (1949), 658;
(b) G. Porter, Proc. Roy. Soc., A200, (1950), 284-300.
14] M. Beer, H.C. Longuet-Higgins, J.Chem.Phys
1
1
1
46
Chapter 1
18] S.S. Deshpande, critical reviews in food science and nutrition, 41(3),
19]
21]
22] illiams, Nature, 196, (1962), 59-61.
lliams, Biochem. J., 96(3), (1965),
(2005), 1367-1377.
iology and Medicine,
27] l, H.E. Feigl, D. Goldstein, J. Am. Chem. Soc., 77(15), (1955),
(1954), 275-276.
1] 1), 53-88.
, (1990).
34] hem, 72, (1968),
35] ter, Ann Phys, 2, (1948), 55-75.
38] , Biophys J., 84, (2003), 3992-4010.
40] G.W. Gordon, G. Berry, X.H. Liang, B. Levine, B. Herman, Biophys
(2001), 155-224.
M. Kasha, Radiation Res. Suppl., 2, (1960), 243.
20] R.T. Williams, J.W. Birdges, J. Clin. Pathol., 17(4), (1964), 371-394.
J.W. Bridges, Ph.D. Thesis, University of London, (1963).
J.W. Bridges, R.T. W
23] C.A. Parker, W.T. Rees, Ibid, 87, (1962), 83.
24] J.W. Bridges, P.J. Creavan, R.T. Wi
872-878.
25] B. Liu, Z. Liu, G. Jing, Analytical letters, 38(9),
26] S. Undenfriend, Fluorescence Assay in B
Academic Press, New York, (1962).
F. Feig
4162-4163.
28] T. Förster, K. Kasper, Z. Electrochem., 1,
29] J.R. Lakowicz, Principles of fluorescence spectroscopy, 3rd Ed.,
Plenum Press, New York, (2006).
30] J.B. Birks, J.E. Dyson, Proc. Roy. Soc., 275A, (1963), 135.
3 C.D. Geddes, Meas. Sci. Technol., 12, (200
32] G.G. Guilbault, Practical Fluorescence, 2nd Ed., Marcel Dekker Inc.,
New York
33] H. Kautsky, Trans. Faraday Soc., 35, (1939), 216-219.
H. Knibbe, D. Rehm, A. Weller, Ber Bunsenges Phys C
257-263.
T. Förs
36] L. Stryer, Annu Rev Biochem, 47, (1978), 819-846.
37] I.Z. Steinberg, Annu Rev Biochem, 40, (1971), 83-114.
C. Berney, G. Danuser
39] A. Hoppe, K. Christensen, J.A. Swanson, Biophys J., 83, (2002), 3652-
3664.
47
Chapter 1
J., 74, (1998), 2702-2713.
I.L. Medintz, J.H.41] Konnert, A.R. Clapp, I. Stanish et al, Proc Natl Acad
4),
2, (1997), 78-
69.
4.
004), 915-919.
F .Photochem.
C.X. Zhou, X.L. Ding, Y. Liu, J.
), 1773-1789.
logical fluids, Pergamon Press,Oxford, (1972), 511-
53] The vital contributions of Perrin and Förster, Biophotonics
A lec. Phys., 6(9),
ht,Göttingen, (1951), 109-113.
Sci USA, 101(26), (2004), 9612-9617.
42] M. Awais, M. Sato, K. Sasaki, Y. Umezawa, Anal Chem, 76, (200
2181-2186.
43] S.C. Hung, R.A. Mathies, A.N. Glazer, Anal Biochem, 25
88.
44] G.S. Jiao, L.H. Thoresen, K. Burgess, J. Am. Chem. Soc., 125, (2003),
14668-146
45] M. Elangovan, R.N. Day, A. Pariasamy, J. Microsc., 205, (2002), 3-1
46] B.J. Bacskai, J. Skoch, G.A. Hickey, R. Allen, B.T. Hyman, J. Biomed.
Opt., 8(3), (2003), 368-375.
47] Y.J. Hu, Y. Liu, J.B. Wang, X.H. Xiao, S.S Qu, J. Pharm. Biomed.
Anal., 36, (2
48] . Cui, X. Kong, L. Qin, G. Zhang, Q. Liu, B. Lei, X. Yao, J
Photobiol. B, 95, (2009), 162-169.
49] Y.Z. Zhang, B. Zhou, Y.X. Liu,
Fluorescence, 18(1), (2008), 109-118.
50] G. Neméthy, H.A. Scheraga, J. Phys .Chem., 66, (1962
51] S.N. Timaseff, Thermodynamics of protein interactions, in:H. Peters
(Ed.) Proteins of bio
519.
52] P.D. Ross, S. Subramanian, Biochemistry, 20, (1981), 3096-3102.
R. Clegg,
International, (2004).
54] T. Förster, Naturwissenschaften, 6, (1946), 166-175.
55] . Kawsaki, E. Kuten, J. Kaminski, J. Phys. B:Atom. Mo
(1973), 1907-1916.
56] T. Förster, Fluoreszenz Organischer Verbindungen, Vandenhoeck and
Ruprec
48
Chapter 1
57] Jimenez, M. Hasegawa, M. Rodriguez,
58] kaki, K.Nakashima, Photochem.
59] . Zou, P.G. Yi, Z.C. Shang, G.X. Hu, Q.S. Yu, Anal Sci.,
60] . Zhang, M. Wong, S. Chen, W.L. Xia, L.J. Jiang, D.W. Chen,
61] , W. Jiang, R.M. Zhao, S.S. Qu, Photochem. Photobiol.B,
61-166.
.
68] en, J.C. Fan, Y. Wang, C.P. Fan, Z.C. Shang, Anal.Sci., 22, (2006),
l.,
(2007), 114-120.
-45.
J. Hung, J. Castillo, G.
Spectrochim. Acta.A, 59(13), (2003), 3177-3183.
Y. Li, H. Kuwabara, Y.K. Gong, Y. Ta
Photobiol.B, 70, (2003), 171-176.
M. Guo, J.W
20, (2004), 465-470.
M.H
Photochem. Photobiol.A, 96, (1996), 57-63.
Y.J. Hu, Y. Liu
80, (2005), 235-242.
62] S.Chatterjee, S. Nandi, S.C. Bhattacharya, J.Photochem. Photobiol.A,
173, (2005), 221-227.
63] L. Matyus, J. Szollosi, A. Jenei, Photochem. Photobiol.B, 83, (2006),
223-226.
64] Y.J. Hu, Y. Liu, R.M. Zhao, J.X. Dong, S.S,Qu, Photochem.
Photobiol.A, 179, (2006), 324-329.
65] P.B. Kandagal, S. Ashoka, J. Seetharamappa, V. Vani, S.M.T. Shaikh,
Photochem. Photobiol.A, 179, (2006), 1
66] C. Wang, Q.H. Wu, Z. Wang, J.Zhao, Anal. Sci., 22, (2006), 435-438
67] Z. Hongwei, G.M.Z. Zhaoxia, W. Wenfeng, W. Guozhong, Spectrochim.
Acta.A, 65, (2006), 811-817.
X. Ch
427-430.
69] Y.J. Hu, Y. Liu, T.Q. Sun, A.M. Bai, J.Q. Lu, Z.B. Pi, Biol. Macromo
39, (2006), 280-285.
70] D. Li, J. Zhu, J. Jin, Photochem. Photobiol.A, 189(1),
71] Y.Q. Wang, H.M. Zhang, G.C. Zhang, W.H. Tao, S.H. Tang, J. Mol.
Stru., 830, (2007), 40
49