Nurse Call System
Standard packages:
Nurse Console Patient Panel Hand Unit
Room Indicator Bathroom Reset Panel
Pull Cord Emergency Call
Excellence in Medical Engineering
Introduction
WTK Nurse Call System is designed to provide nursing services to modern application.
Description
It is important communication link between patients and nursing staff and being designed to meet various kind of visual and tone signal needed for patient call. It provides a two way nurse to patient speech for instant communication between nurse and patient or other staff. With easy to use numeric key pad, the nursing staff can access any call point inside the system to open up a two way communication with the calling point. It is a smart system where the attendant will automatically answer the highest priority call or any call intended to make by inputting the room and bed number through key pad. Main Supporting System will be patient panel, hand unit, emergency panel, reset panel, overdoor indicator (Tri-Color) and pull cord system
Patient Panel Hand Unit Emergency Panel Reset Panel Overdoor Indicator WT07002 WT07003 WT07005 WT07007 WT07004
WTK Technologies (M) Sdn. Bhd. 6, Jalan BA/13, Kawasan Perusahaan Bukit Angkat 43000 Kajang. Selangor. Malaysia Tel. : +603-87332880 Fax. : +603-87332535 E-mail : [email protected] http://www.wtktechnologies.com.my
TÜV SÜD certified company for ISO 13485 & CE Marking
0123
Typical Wiring Layout of Nurse Call System
* Wiring is loop type using 2 x 0.5mm (12V) & 4 x 0.2mm (data) cable from interface board to patient panels,
toilet reset panels and light indicator. Emergency panel is linked to reset panel via 4 x 0.2mm cable
WTK Technologies (M) Sdn. Bhd. 6, Jalan BA/13, Kawasan Perusahaan Bukit Angkat 43000 Kajang. Selangor. Malaysia Tel. : +603-87332880 Fax. : +603-87332535 E-mail : [email protected] http://www.wtktechnologies.com.my
TÜV SÜD certified company for ISO 13485 & CE Marking
0123
Typical Layout of Nurse Call Console
TECHNICAL SPECIFICATION:
Part List:
Unique & Reliable Design
Designation Nurse Call System, 2 Way Speech Type
Power Type: Input 230 VAC. Low voltage design 12 VDC
Room & Bed: Unlimited assignment up to 255 locations per console
Configurations: Default bed locations (ex. 001) or bed + room assignment (ex. ICU 001)
Auto Detection: Broken Links (loose wiring, handset dis-engaged, wiring/jumper disconnects)
Handset Disable: Included in design
WT07001 Nurse Console
WT07002 Patient Panel
WT07003 Hand Unit
WT07004 Room Indicator
WT07005 Emergency Call
WT07006 Pull Cord
WT07007 Bathroom Reset Panel