ABSTRACT Development of Sample Rotator for High-field, Ultra-low Temperature NMR Experiments Nathaniel J. Kan Yale University The purpose of this project was to lay the foundation for the construction of a platform tilt-angle rotator for high-field, ultra-low temperature NMR experiments. The sample rotator will be used in a continuation of NMR experiments being performed in Sean Barrett’s lab to gather more information about the many-electron “particles” known as skyrmions. The induced magnetic field contribution for a rotating piezo-electric actuator made by attocube Systems was measured, and several models were built to determine contribution across the sample at for different physical orientations of the actuator. The actuator produced an induced magnetic field that was comparable to the induced field in a similar sized block of titanium. From this data it appears likely that the actuator will be usable in a sample rotator for future skyrmion experiments. 1

Upload: justin-kan

Post on 18-Nov-2014

3.835 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: My Senior Thesis Project

ABSTRACT

Development of Sample Rotator for High-field, Ultra-low Temperature NMR

Experiments

Nathaniel J. Kan

Yale University

The purpose of this project was to lay the foundation for the construction of a

platform tilt-angle rotator for high-field, ultra-low temperature NMR experiments. The

sample rotator will be used in a continuation of NMR experiments being performed in

Sean Barrett’s lab to gather more information about the many-electron “particles” known

as skyrmions. The induced magnetic field contribution for a rotating piezo-electric

actuator made by attocube Systems was measured, and several models were built to

determine contribution across the sample at for different physical orientations of the

actuator. The actuator produced an induced magnetic field that was comparable to the

induced field in a similar sized block of titanium. From this data it appears likely that the

actuator will be usable in a sample rotator for future skyrmion experiments.

1

Page 2: My Senior Thesis Project

Development of Sample Rotator for High-field, Ultra-low

Temperature NMR Experiments

By

Nathaniel J. Kan

Physics 472 Final Report

Advisor: Sean E. Barrett

2

Page 3: My Senior Thesis Project

Contents 1 Background 4

1.1 Introduction: Skyrmions 4 1.2 Present Experimental Setup 5 1.3 New Experimental Setup 6

2 Sample Rotator 7 2.1 Constraints for a Sample Rotator 7

3 Experimental Setup 9 3.1 Probe 9 3.2 Physical Setup 10

4 Data 12 4.1 Converting into B-field 12

5 Analysis 14 5.1 Sphere Model 14 5.2 Octagonal Prism Model 16 5.3 Field vs Actuator Distance Simulation 18

6 Conclusion 20 6.1 Future Work 20 6.2 Final Comments 21

Appendix A: Igor 5.0 Octagonal Prism Model Procedure 22 Appendix B: Field vs Actuator Distance Simulation 26 Bibliography 29

3

Page 4: My Senior Thesis Project

1 Background

1.1 Introduction: Skyrmions

Optically Pumped Nuclear Magnetic Resonance (OPNMR) has been used to

investigate the behaviors of conduction electrons in two-dimensional electron systems

(2DES) under high field and low temperature conditions. These OPNMR experiments,

particularly those performed in the Barrett OPNMR lab at Yale University, are of

particular relevance to this project. Such experiments have so far provided us with

incomplete descriptions of the many-electron spin-textures known as skyrmions.

The experiments conducted in the Barrett lab use 2DES samples composed of thin

(300Å) layers of GaAs between two layers of Al0.1Ga0.9As. These samples maintain a

structure that behaves as many quantum wells. The physical states of the system are

determined by the Landau filling factor [1]:

θν

cosc

toteBnh

= (1.01)

where n is the 2D density of electrons in the well, h is Planck’s constant, c is the speed of

light, and e is the charge of an electron. If we consider a 2DES system with at its ground

state at ν = 1 (Figure 1.1.1a), and then add a single electron, we bring the system to ν = 1

+ ε. Due to the Pauli Exclusion principal the added electron will be spin oriented opposite

the polarization, and in order to minimize electrostatic repulsion nearby electrons will

have their spins oriented away from the added electron (Figure 1.1.1b). This spin-texture

is referred to as a skyrmion. By controlling the filling factor, the number of skyrmions in

a 2DES can be changed.

4

Page 5: My Senior Thesis Project

Figure 1.1.1.

Currently many unanswered questions exist about skyrmion behavior. It remains

unproven what formation skyrmions will arrange themselves in (whether skyrmions will

form a crystal). Further measurements of skyrmions at lower temperatures (<0.4 K) will

yield more answers about skyrmion localization and other properties.

1.2 Present Experimental Setup

NMR measurements of the 2DES sample in the ~300 mK – 20 K regime were

conducted in an Oxford Instruments 12Tesla/89mm wide bore Teslatron superconducting

magnet [2]. In experiments conducted near ν = 1, data was recorded while varying the

sample temperature T and Landau filling factor ν. ~300 mK – 20 K temperatures were

achieved by pumping liquid 3He with an Oxford Instruments 3He sorption-pumped

Heliox cryostat made by Oxford Instruments and varying the pumping speed.

To control the filling factor ν, the angle between the sample and the lab field was

adjusted manually between every measurement where the filling factor changed. This

5

Page 6: My Senior Thesis Project

was accomplished by a rotator assembly that could adjust the angle θ between the growth

axis of the sample and the applied field Btot. As the other factors in the filling factor

equation were held constant while manipulating θ it was possible to directly control the

filling factor.

1.3 New Experimental Setup

In order to take further data on skyrmion behavior, a more powerful cooling

system is needed. The project of building a sample rotator for ultra-low temperature and

high-field NMR experiments was originally conceived to fulfill the design needs of an

experiment utilizing a Magnex Scientific 15.3 Tesla/123mm wide bore superconducting

high resolution magnet with a custom ordered dilution refrigerator that would allow

sample temperatures nearing ~15 mK.

The solution used to adjust Landau filling factor in the old experimental setup is

impractical for the <100 mK temperature range. In the previous solution, manual

adjustment of the rotator assembly required a recooling of the sample for every

adjustment of θ. Such rotation would require mechanical access from the exterior of the

dilution refrigerator, which would significantly heat the sample during rotation. Because

of the long cooling time required by the dilution refrigerator to reach the ~15 mK

temperature, it is desirable to develop a rotation mechanism that does not significantly

heat the system. Such a system could cut down wait time between measurements by days.

6

Page 7: My Senior Thesis Project

2 Sample Rotator

2.1 Constraints for a Sample Rotator

In the next phase of Barrett lab’s skyrmion experiments, the goal is to be able to

take measurements at ultra-low temperatures nearing 15 mK. This will allow more

accurate measurements, as lower temperatures results in less movement in the system and

thus clearer resolution in the data. Because the present experimental setup uses a slow

method of ramping up current in the inductor, it is much less difficult to devise a method

of changing the tilt angle of the platform than the lab field. The primary concern of this

project was to lay the foundation for the design of a sample rotation device. This sample

rotator will need to sit inside the dilution fridge that is responsible for cooling the sample.

There are several challenges in designing the sample rotation device. The rotator

mechanism will need to function in the ultra-low temperature regime, and generate as

little heat as possible to minimize heat conduction from the exterior of the dilution fridge

to the sample. The rotator will also need to be composed primarily of non-magnetic

materials, so that emissions from the rotator do not interfere with the measurements from

the sample. Also, the sample rotator design will be confined by the size of the cavity in

the dilution refrigerator.

Currently we are considering using a piezo-electric rotator designed by attocube

Systems as the rotating mechanism in the experiment. This rotator has the advantage of

being controlled in very small precise steps (0.001 deg.), which will allow us to much

more easily devise a method of independently determining θ and minimize heat

7

Page 8: My Senior Thesis Project

conduction by controlling the rate that steps are sent to the actuator. The actuator works

on the slipstick principal—voltage is sent to the actuator in sawtooth waves with a very

long ramp-up time and a very short ramp-down time. Due to friction, this results in the

ramp-up causing rotation in both the top and bottom plates, so that the relative position of

the two plates does not change, while the ramp-down is quickly contracts the piezo and

moves the bottom plate back without changing the location of the top plate.

In order to determine whether the attocube actuator is a viable solution, we

needed to create a model of the magnetic field profile of the actuator. While attocube

actuators are designed with some non-magnetic components, the body of the actuator is

largely titanium (which is paramagnetic). In the presence of an external magnetic field

(such as the field from an NMR magnet), a paramagnetic field will gain a magnetization

and cause an induced magnetic field. Before we could commit to using the attocube

actuator, it was necessary to measure the induced field and extrapolate the possible

interference across a 2DES sample from a rotator built with the actuator. While the

sample rotator would ideally use the attocube piezo-rotator, at this time the only available

objects for testing is an attocube linear piezo-actuator, which is constructed from the

same materials and works according to the same principals.

8

Page 9: My Senior Thesis Project

3 Experimental Setup

3.1 Probe

To isolate the field produced by the actuator, we repeated measurements with and

without the actuator in the probe. Measurements were done by recording the NMR signal

from a D2O droplet approximately 1 mm3. The Larmor frequency of D2O is 78.4 MHz.

We used the low frequency circuit in the probe (Figure 3.1.1).

Figure 3.1.1.

For a given L, R (where in our circuit R was the resistivity of copper wire

contained by the circuit), and resonant frequency ω, the capacitances of the tuning

capacitor CT and the matching capacitor CM are given by the following equations:

)LR(

LR50R50/RLC 222

222

T ωωωω

++−−

= (3.01)

9

Page 10: My Senior Thesis Project

222M

LR50R50/RC

ωω +−= (3.02)

The matching capacitor matches the impedance of the probe to the impedance of

the 50Ω cable to the NMR spectrometer, while the tuning capacitor is used to adjust the

resonant frequency of the probe’s RF coil. The tuning and matching capacitors in our

probe were both variable capacitors which could be adjusted manually between 2-12 pF.

Additional capacitors were added in parallel to get the circuit within the appropriate

range to allow the variance in the tuning and matching capacitors to bring the resonant

frequency of the circuit to the Larmor frequency of D2O. The probe circuit was then

tuned by adjusting the variable capacitors in order to minimize power from a

spectrometer reflected by the probe at 78.4 MHz.

3.2 Physical Setup

When planning the physical location of the attocube actuator in the probe can, we

wanted to maintain degrees of freedom in two axes. The actuator, which is shaped

roughly like an octagon prism, was attached with non-magnetic material to the end of a

long plastic rod (~5’) close to the center of one of its octagon faces. The rod was placed

in the hollow shaft of the probe, so that when the probe was placed in the magnet bore,

the rod extended out the top of the probe shaft and the magnet. This extended end was

then rigidly affixed to the lab frame. The probe’s RF coil below the actuator with a slight

r-axis offset. In this setup (Figure 3.2.1) the probe (and the D20 sample) could rotate

freely 360 degrees around the actuator. The plastic rod could be unfixed from and

reattached to the lab frame in order to take measurements at different z values. Figure

3.2.2 demonstrates the different positions of the RF coil relative to the actuator.

10

Page 11: My Senior Thesis Project

Figure 3.2.1.

Figure 3.2.2.

11

Page 12: My Senior Thesis Project

4 Data

4.1 Converting into B-field

For any applied field B a particle with a net spin has a Larmor frequency of

Bγν = [3] (4.01)

Our measurements recorded the frequency shift between the resonant frequency

for D2O from to the 12T lab field that was due to the additional field caused by the

induced magnetization in the actuator. The frequency difference (∆ν) between the

measurements with the actuator and without the actuator at a given z,r is related to the

magnetic field due solely to the induced field in the actuator by the gyromagnetic ratio γ:

attocubeBγν =∆ (4.02)

In our measurement coordinate system, z is the z-distance between the center of

the attocube actuator and the sample, r is the radial distance between the sample and the

axis of rotation, and θ is the angular offset of the coil. We recorded data at three values of

z z = 2.6 cm, z = 3.1 cm, z = 3.6 cm, r = 1.18 cm, and θ from 0 to 360 degrees (Figure

4.1.1). To isolate the field due to the actuator, we also first recorded data without the

actuator in the probe. The field due to the actuator for a given z, r, θ is then given by:

γνν controlattocube

attocubeB−

= (4.03)

The gyromagnetic ratio of D2O is 6.65 * 106 s-1 T-1.

12

Page 13: My Senior Thesis Project

Figure 4.1.1.

We later introduced several changes to correct various quirks observed in the

data. In order to explain a mysterious second period in the data, we believe it was very

likely that our θ = 0 changed with respect to the coil location and that we reversed the θ

direction we were recording in between the trials with and without the actuator. Also, in

order to make our B-field data better fit an expected 1/r3 decay, we adjusted our z values

by ±5 mm. This offset appears to be due to a regular error measurement in our z-

coordinate.

Figure 4.1.2.

13

Page 14: My Senior Thesis Project

5 Analysis

5.1 Sphere Model

Our first attempt to model the actuator was to assume that the behaved like a

uniformly magnetized titanium sphere (Figure 5.1.1).

Figure 5.1.1.

The induced magnetization of M of a paramagnetic object in a uniform field Bz is

given by:

0

zBµ

K=M (5.01)

where K is the magnetic volume susceptibility, and µ0 is the Permeability of Free Space.

14

Page 15: My Senior Thesis Project

K can be expressed in terms of the molar susceptibility χm and the molar volume Vm [4]:

m

m

=K (5.02)

For titanium, χm = 3709.5 * 10-6 cm3 mol -1 and Vm = 10.64 cm3 mol -1. In the 12 T lab

field this yields a magnetization M = 1.7 * 103 T.

The field due to a sphere with a uniformly magnetization M is a textbook

problem. The potential of the sphere for a position x = φ, θ, r is given by [1]:

( ) ( )θµπφ sin

rraM

34xA 2

200 ⎟

⎟⎠

⎞⎜⎜⎝

⎛=

>

< (5.03)

where a is the radius of the sphere and r< (r>) is the smaller (larger) of a and r (Figure

5.1.2).

Figure 5.1.2.

0B =φ (5.04)

( )θθθ φ sinA

sinr1Br ∂

∂= (5.05)

( )rArr

1B φθ ∂∂

−= (5.06)

We are only interested in the cases where r > a, so (5.03) becomes:

( ) ( )θµπφ sin

raM

34xA 2

3

00= (5.07)

Plugging this into (5.05) and (5.06) gives:

θµπ cosraM

38B 3

3

00r = (5.08)

θµπθ sin

raM

34B 3

3

00= (5.09)

15

Page 16: My Senior Thesis Project

We can transform this to give us the z-component of the field:

⎟⎠⎞

⎜⎝⎛ −= θµ 2

3

3

00z sin32

raMB (5.10)

We did a Х2 fit allowing the XY location (because the actuator’s center of mass

was slightly offset from the axis of rotation and center of volume) and the radius of the

sphere to vary. The best fit to the data was

Figure 5.1.2.

5.2 Octagonal Prism Model

The next model we did was an octagonal prism with a rectangular slit removed

from it, the dimensions of which were the actual figures measured from the actuator. The

magnetic potential from any uniformly charged object is:

( ) Ar

nM4

r 20 ∂

⋅=Φ ∫

rrr

πµ

(5.11)

16

Page 17: My Senior Thesis Project

The field is given by:

∫ Φ∂⋅∇=r

B (5.12)

For the octagonal prism model we wrote a procedure in the program Igor 5.0 that

modeled the surface shell of the prism as an array of points, and then numerically

summed the B-field contributions of each point to each point in a sample ring (which

approximated our ring of sample points).

Due to the nMrr⋅ term in (5.11), the only surfaces that give non-zero contributions

are the surfaces parallel to the XY plane. To account for the missing rectangular box we

summed the contributions from points of an overlaid box with an opposite magnetization.

The total field in the z-direction at a sample point is then

∑ ⎟⎠⎞

⎜⎝⎛ ×

∆∆=

rz

ryx

4M

B 200

z πµ

(5.13)

where r is the distance between the nth point and the observation point. See Appendix A

for a complete copy of the procedure.

Figure 5.2.1.

The results of the octagonal prism were interesting. The octagonal prism model is

slightly less accurate in matching the data than the spherical approximation. This is due

to the fact that there was no Х2 fit to maximize various input parameters. The octagonal

model much more accurate depicts the true volume and location of the actuator, however.

17

Page 18: My Senior Thesis Project

Figure 5.2.2.

5.3 Field vs Actuator Distance Simulation

For the actuator is usable for the purposes of a sample rotator, we estimate that the

induced field across the sample should be no greater than 5 ppm (for our models less than

60 x 10-6 T). Once we were fairly confident that our models of a paramagnetic uniformly

magnetized object were reasonably good approximations, we constructed a final model

that measured the induced magnetic field against the distance from a magnetized block of

comparable volume to the actuator. The block in the sample has its large face normal to

the XY plane (Figure 5.3.1), which simulates a possible orientation of the actuator in a

sample rotator. See Appendix B for the full Igor 5.0 procedure.

18

Page 19: My Senior Thesis Project

Figure 5.3.1.

According to the model, the sample must be at least 2 cm away from the center of

actuator to reduce the force from the field to below 5 ppm (Figure 5.3.2). In the

orientation of components for the new dilution refrigerator probe, the sample rotator will

most probably be 3-5 cm away from the sample. Because of these constraints we believe

the attocube actuator is a workable solution for the purpose of a sample rotator.

Figure 5.3.2.

19

Page 20: My Senior Thesis Project

6 Conclusion

6.1 Future Work

There are additional projects that need to be completed before a sample rotator

can be built. The heat conduction due to the electrical connectors of the piezo-actuator

needs to be tested to determine the necessary cooling time between θ rotations and

measurements. Also, it would be desirable to develop a system of measuring the platform

angle θ independent of counting the number of ramping steps sent into the actuator. By

using an independent method of tracking θ the experimenter can be sure that noise or

distortions in the voltage signal have not caused a smaller or greater angular step per

wave, and therefore a miscalculation of θ. At this point there is no definite “best method”

for determining absolute θ; several methods being considered are building a mechanical

stop into the rotator and stepping from that angle in each of our measurements, or a

similar solution using an attocube fiber-optic system that tracks location by reflecting

light off a grating on the rotating platform.

Finally, the piezo-actuator sample rotator needs to be constructed. The rotator

could be as simple as the attocube rotating actuator with a non-magnetic plate extending

from its rotating face for the sample to rest on. The location of the actuator will be limited

by the physical constraints of the dilution refrigerator. It may also be desirable to retest

the magnetic field profile due to the completed working rotator to ensure that it is within

acceptable levels across the sample.

20

Page 21: My Senior Thesis Project

6.2 Final Comments

Throughout the process of testing the actuator to develop a field profile we have

learned several important lessons in experimental procedure. For future experimenters

working on this project it may be helpful to review these notes.

First, in our tests there was a significant time lapse (~several weeks) between

the NMR measurements without the actuator and the measurements with the actuator. As

the field due to the NMR magnet drifts over time, this resulted in a larger drift of the lab

field between our two data sets, which has to be approximated and accounted for. By

cutting down the wait time between measurements (for example, doing the measurements

on consecutive days) we feel this would drastically reduce error due to drift estimations.

Second, it is important to build the circuit so that the test object (the actuator)

can fit inside. While this may seem obvious, we initially planned to build our circuit to

certain physical dimensions that would allow the actuator free rotation at several different

z-values. However, through the course of building the circuit, components had to be

moved or added. This resulted in a final circuit that would not accommodate the actuator,

which was not discovered until we had taken the test data for the control profile.

Third, the experimenter should make sure to use a consistent angular convention

in every test. Four, in order to maximize the effect of the induced field, the applied lab

field should be as large as possible. This will make it much easier to see the effects of the

object on field structure.

Last, for any experimenters pursuing this project, it may be helpful to retake the

data. When doing so, it would be a good idea to take data at different orientations of the

actuator in order to better test the accuracy of the model. Also, a more accurate method of

fixing the test object in the lab frame while rotating the probe would yield much more

accurate measurements along the z-axis.

21

Page 22: My Senior Thesis Project

Appendix A: Igor 5.0 Octagonal Prism Model Procedure Function RectangleBvariationNeg(cx,cy,cz,cs,sx,sy,sz,sr,ss,ca,cb,ch,thetastart) // Calculate the Bfield for an array of points defined by the circle with center at (sx,sy) and radius sr. Variable cx,cy,cz,cs,sx,sy,sz,sr,ss,ca,cb,ch,thetastart //Variable cx,cy,cz,cnpnts,sx,sy,sz,sr,snpnts //ca = start hole x, cb = end hole x, ch = height hole, thetastart = phase shift Variable m m = 1.703*10^3 // m = total magnetization of the object //Variable cs,ss // cs = step size for x and y dim of rectangle, ss = step size in radians along circle //cs = cx*cy/cnpnts //cnpnts number of points on 'charged rectangle' over which to integrate. //snpnts number of points in sample to be evaluated. //ngpnts number of points in absent rectangle to be evaluated Variable cnpnts,snpnts,ngpnts,grdpnts grdpnts = 2*(cx/cs)*(cy/cs)*2*pi/ss+2*((cb-ca)/cs)*(cy/cs)*2*pi/ss print "start time",time() DoWindow/K graph0 DoWindow/K graph1 DoWindow/K graph2 DoWindow/K table0 DoWindow/K table1 //kills leftover graphs and tables created by this macro from previous run. Make/N=(grdpnts)/D/O xc, yc, xn, yn, xs, ys // generates longer waves than needed. waves will be redimensioned later. Variable xj, yj // looping variables for tracking current x,y within charged plate Variable k, i, theta // loop variable xj=0 yj=0 k=0 i=0 theta=thetastart ngpnts=0 // first generate array of points for the charged rectangular plate do // loop through x if(xj>=(ca-0.0001)&&xj<=(cb+.0001)) do // loop through y xc[i]=xj yc[i]=yj xn[ngpnts]=xj yn[ngpnts]=yj ngpnts+=1 i+=1 yj+=cs while(yj<=(cy+0.0001)) else do // loop through y xc[i]=xj

22

Page 23: My Senior Thesis Project

yc[i]=yj i+=1 yj+=cs while(yj<=(cy+0.0001)) endif yj=0 // reset yj to -cr and step to next xj value. xj+=cs while(xj<=(cx+0.0001)) // now generate array of points for the sample circle do // loop through theta xs[k]=sx-sr*sin(theta) ys[k]=sy+sr*cos(theta) theta+=ss k+=1 while(theta<=2*Pi+thetastart) snpnts=k //use i and k to redeminsion waves xs, ys, zs and xc, yc. cnpnts=i ReDimension/N=(snpnts) xs, ys ReDimension/N=(cnpnts) xc, yc ReDimension/N=(ngpnts) xn, yn print "number of points for 'charged rectangle'", cnpnts print "number of points for sample", snpnts print "number of points for negative block", ngpnts Display yc vs xc ModifyGraph mode=3,marker=19 AppendToGraph ys vs xs ModifyGraph mode=3,marker=19,rgb(ys)=(0,0,65535) AppendToGraph yn vs xn ModifyGraph mode=3,marker=19,rgb(yn)=(0,65535,0) Make/N=(snpnts)/D/O Bext //this is the field wave evaluated at points (xs,ys) Bext=0// initialize wave to zero print "start integration loop 1 at", time() i=0 // index for points on rectangle plate k=0 // index for Bext points in sample Variable xdelta, ydelta, zdelta, rsquare, multiplyconst // multiplyconst = the constants in front of the whole thing multiplyconst = 10^-7*m print "multiply factor", multiplyconst //Variable height = cz/2 do // loop for different points within the sample do // integrating over lower & upper charged rectangles at same time. xdelta=xc[i]-xs[k] ydelta=yc[i]-ys[k] // Top plate has positive charge and located at 0 zdelta = sz rsquare = (zdelta^2+xdelta^2+ydelta^2) Bext[k]+= (+1)*zdelta*cs*cs/rsquare

23

Page 24: My Senior Thesis Project

// Bottom plate has negative charge and located at -cz zdelta = cz+sz rsquare = (zdelta^2+xdelta^2+ydelta^2) Bext[k]+= (-1)*zdelta*cs*cs/rsquare i+=1 // step through points on disk to integrate while (i<=cnpnts-1) i=0 //reset loop variable if(k == round(.25*snpnts)) print "25% complete at", time() elseif(k == round(.50*snpnts)) print "50% complete at", time() elseif(k == round(.75*snpnts)) print "75% complete at", time() elseif(k == round(.90*snpnts)) print "90% complete at", time() endif k+=1 while (k<=snpnts-1) print "start integration loop 2 at", time() i=0 // index for points on rectangle plate k=0 // index for Bext points in sample //Variable height = cz/2 do // loop for different points within the sample do // integrating over lower & upper charged rectangles at same time. xdelta=xn[i]-xs[k] ydelta=yn[i]-ys[k] // Top plate has negative and located at 0 zdelta = sz rsquare = (zdelta^2+xdelta^2+ydelta^2) Bext[k]+= (-1)*zdelta*cs*cs/rsquare // Bottom plate has positive charge and located at -ch zdelta = ch+sz rsquare = (zdelta^2+xdelta^2+ydelta^2) Bext[k]+= (+1)*zdelta*cs*cs/rsquare i+=1 // step through points on disk to integrate while (i<=ngpnts-1) // multiply Bext[k] by all the constants Bext[k] *= multiplyconst i=0 // reset loop variable if(k == round(.25*snpnts)) print "25% complete at", time() elseif(k == round(.50*snpnts)) print "50% complete at", time() elseif(k == round(.75*snpnts)) print "75% complete at", time()

24

Page 25: My Senior Thesis Project

elseif(k == round(.90*snpnts)) print "90% complete at", time() endif k+=1 while (k<=snpnts-1) print "stop integration loop at", time() Edit xc,yc,xn,yn,xs,ys,Bext // displays these waves in a table. Display Bext // plots Bext as a function of kth point. ModifyGraph mode=0 End

25

Page 26: My Senior Thesis Project

Appendix B: Field vs Actuator Distance Simulation Function OctagonBvariationJustin(cx,cy,cz,cs,sx,sy,sz,sr,ss,thetastart) // Calculate the Bfield for an array of points defined by the circle with center at (sx,sy) and radius sr. Variable cx,cy,cz,cs,sx,sy,sz,sr,ss,thetastart //Variable cx,cy,cz,cnpnts,sx,sy,sz,sr,snpnts //ca = start hole x, cb = end hole x, ch = height hole, thetastart = phase shift // a = corner x length, b = corner y height Variable m m = 1.703*10^3 // m = total magnetization of the object (assumes ~180 ppm of dimensionless chi for Titanium, and B=12 Tesla, and mu_0=4Pie-7) //Variable cs,ss // cs = step size for x and y dim of rectangle, ss = step size in radians along circle //cs = cx*cy/cnpnts //cnpnts number of points on 'charged rectangle' over which to integrate. //snpnts number of points in sample to be evaluated. //ngpnts number of points in absent rectangle to be evaluated Variable cnpnts,snpnts,ngpnts,grdpnts,dpnts grdpnts = 2*(cx/cs)*(cy/cs)*2*pi/ss print "start time",time() DoWindow/K graph0 DoWindow/K graph1 DoWindow/K graph2 DoWindow/K table0 DoWindow/K table1 //kills leftover graphs and tables created by this macro from previous run. Make/N=(grdpnts)/D/O xc, yc, xs, zs, ys, angle // generates longer waves than needed. waves will be redimensioned later. Variable xj, yj // looping variables for tracking current x,y within charged plate Variable k, i, theta // loop variable Variable ymin, ymax // hold the max y and min y for calculating corners xj=0 yj=0 k=0 i=0 theta=thetastart ngpnts=0 dpnts=0 // first generate array of points for the charged rectangular plate do // loop through x do // loop through y xc[i]=xj yc[i]=yj i+=1 yj+=cs while(yj<=(cy+0.0001)) yj=0 // reset yj to -cr and step to next xj value. xj+=cs while(xj<=(cx+0.0001)) // now generate array of points for the sample circle // now generate array of points for the sample circle do // loop through theta

26

Page 27: My Senior Thesis Project

zs[k]=sz-sr*sin(theta) ys[k]=sy+sr*cos(theta) xs[k]=sx theta+=ss angle[k]=360-(theta-thetastart)*180/Pi k+=1 while(theta<=2*Pi+thetastart) snpnts=k //use i and k to redeminsion waves xs, ys, zs and xc, yc. cnpnts=i ReDimension/N=(snpnts) xs, ys, zs, angle ReDimension/N=(cnpnts) xc, yc print "number of points for 'charged rectangle'", cnpnts print "number of points for sample", snpnts Display/W=(1,1,500,500) yc vs xc ModifyGraph mode=3,marker=19,rgb(yc)=(65535,0,0) AppendToGraph ys vs xs ModifyGraph mode=3,marker=19,rgb(ys)=(0,0,65535) Make/N=(snpnts)/D/O Bext //this is the field wave evaluated at points (xs,ys) Bext=0// initialize wave to zero print "start integration loop 1 at", time() i=0 // index for points on rectangle plate k=0 // index for Bext points in sample Variable xdelta, ydelta, zdelta, rsquare, multiplyconst // multiplyconst = the constants in front of the whole thing multiplyconst = 10^-7*m print "multiply factor", multiplyconst //Variable height = cz/2 do // loop for different points within the sample do // integrating over upper charged rectangle first xdelta=xc[i]-sx ydelta=yc[i]-ys[k] zdelta=cz-zs[k] // Top plate has positive charge rsquare = (zdelta^2+xdelta^2+ydelta^2)^(3/2) Bext[k]+= (+1)*zdelta*cs*cs/rsquare i+=1 // step through points on disk to integrate while (i<=cnpnts-1) i=0 //reset loop variable do // integrating over lower charged rectangle second xdelta=xc[i]-sx ydelta=yc[i]-ys[k] zdelta=-zs[k]

27

Page 28: My Senior Thesis Project

// Bottom plate has negative charge rsquare = (zdelta^2+xdelta^2+ydelta^2)^(3/2) Bext[k]+= (-1)*zdelta*cs*cs/rsquare i+=1 // step through points on disk to integrate while (i<=cnpnts-1) // multiply Bext[k] by all the constants Bext[k] *= multiplyconst i=0 //reset loop variable if(k == round(.25*snpnts)) print "25% complete at", time() elseif(k == round(.50*snpnts)) print "50% complete at", time() elseif(k == round(.75*snpnts)) print "75% complete at", time() elseif(k == round(.90*snpnts)) print "90% complete at", time() endif k+=1 while (k<=snpnts-1) print "start integration loop 2 at", time() Edit xc,yc,zs,ys,angle,Bext // displays these waves in a table. Display Bext vs angle // plots Bext as a function of kth point. ModifyGraph mode=0 End

28

Page 29: My Senior Thesis Project

Bibliography

[1] S. E. Barrett, R. Tycko, “Optically Pumped NMR of Semiconductors and Two-

dimensional Electron Systems,” Encyclopedia of Nuclear Magnetic Resonance,

Supplementary Volume.

[2] A. E. Dementyev, “Spin in the Quantum Hall Effect: OPNMR Studies of Skyrmions

and Composite Fermions,” Dissertation, Yale University.

[3] J.P. Hornak, “The Basics of NMR,” http://www.cis.rit.edu/htbooks/nmr/inside.htm

[4] “Magnetic Susceptibility of the Elements and Inorganic Compounds,” http://www-

d0.fnal.gov/hardware/cal/lvps_info/engineering/elementmagn.pdf

29