liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in...

7
Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters Patrick Albéric* Université dOrléans CNRS/INSU UMR 6113, Institut des Sciences de la Terre dOrléans (ISTO) 1A rue de la férollerie, 45071, Orléans cedex2, France A commercial interface coupling liquid chromatography (LC) to a continuous-ow isotope ratio mass spectrometry (CF-IRMS) instrument was used to determine the d 13 C of dissolved organic carbon (DOC) in natural waters. Stream and soil waters from a farmland plot in a hedgerow landscape were studied. Based on wet chemical oxidation of dissolved organics the LC/IRMS interface allows the on-line injection of small volumes of water samples, an oxida- tion reaction to produce CO 2 and gas transfer to the isotope ratio mass spectrometer. In ow injection analysis (FIA) mode, bulk DOC d 13 C analysis was performed on aqueous samples of up to 100 mL in volume in the range of DOC concentration in fresh waters (110 mg C.L 1 ). Mapping the DOC d 13 C spatial distribution at the plot scale was made possible by this fairly quick method (10 min for triplicate analyses) with little sample manipulation. The relative contributions of different plot sectors to the DOC pool in the stream draining the plot were tentatively inferred on the basis of d 13 C differences between the hydrophilic and hydrophobic components. Copyright © 2011 John Wiley & Sons, Ltd. Several methods of coupling a dissolved organic carbon (DOC) analyzer to a continuous ow isotope ratio mass spec- trometry (CF-IRMS) system have been developed. [16] As for quantitative DOC analysis, two types of methodologies can be employed to achieve organic carbon oxidation to CO 2 . [1,2] (i) High-temperature combustion (HTC) systems with cryo- genic trapping before transfer of the CO 2 to the IRMS system. [46] Refractory organic carbon may be better oxidized by HTC systems but the isotopic values have to be corrected for a signicant instrument blank depending on the volume of water injected and time allowed for cryotrapping. [46] (ii) Wet chemical oxidation (WCO) methods are applied to larger sample volumes with variable persulphate reagent concentra- tion in a closed reactor. [13] Blank corrections also need to be accounted for, mostly depending on reagent volumes. [2,3] Both types of system can use reduction ovens and gas chro- matographic separation before transfer to the IRMS system. The necessary complete purging of dissolved inorganic carbon (DIC) before DOC (more exactly non-purgeable organic carbon NPOC) analysis has to be carried out off-line by acid introduction into the sample before it is injected into HTC systems. [5,6] This can, however, be conducted on-line before introducing the oxidation reagent into the reactor for WCO systems, allowing in this case both DIC and DOC carbon isotopic analysis. [13] In this study a commercial liquid chromatography (LC)/ IRMS interface (LC IsoLink, Thermo Scientic, Bremen, Germany) was used to analyse the variation in DOC d 13 C in stream and soil waters. The device, which was specially developed for on-line compound-specic isotope analysis of water-soluble samples after LC separation, [7] can also be used without a column in bulk mode analysis for simple [7] and macromolecular compounds. [8] In the eld of complex raw samples the LC IsoLink interface has been already used for bulk carbon isotope ratio determination in alcoholic beverages. [9] The main advantage of this method is that the water sample is injected into a continuous ow of mobile phase (ultrapure deionised water) and reagents mixed at the entry of a heated capillary reactor connected on-line to an open-split coupling and to the IRMS system via a liquid/ gas (water/He) membrane exchanger and a gas-drying unit (see Krummen et al. [7] for a detailed description). The eld selected as an example for this study was a farm- land plot in which the different characteristics (redox conditions, ionic strength, dissolved metals, water-table depths) of several sectors had previously been determined. [10] The factors controlling DOC in soils are complex and they can be inuenced by numerous parameters and processes such as substrate quality, decomposer community, adsorption in mineral soil horizons, pH, ionic strength, temperature, soil moisture, water uxes and redox conditions. [11,12] Some of these factors may also control the variation in the isotope composition of organic carbon in soil proles. As evidence is lacking for any direct isotope fractionation mechanisms, [13,14] the effects of selective sorption and/or preferential preserva- tion of isotopically and chemically heterogeneous complex soil organic matter have been investigated, mostly through * Correspondence to: P. Albéric, Université dOrléans CNRS/INSU UMR 6113, Institut des Sciences de la Terre dOrléans (ISTO) 1A rue de la férollerie, 45071 Orléans cedex2, France. E-mail: [email protected] Presented at the Liquid Chromatography-Isotope Ratio Mass Spectrometry Users Meeting, held 2324 November 2010 at the University of Oxford, Oxford, UK. Copyright © 2011 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2011, 25, 3012 3018 Research Article Received: 18 April 2011 Revised: 4 August 2011 Accepted: 6 August 2011 Published online in Wiley Online Library Rapid Commun. Mass Spectrom. 2011, 25, 3012 3018 (wileyonlinelibrary.com) DOI: 10.1002/rcm.5212 3012

Upload: patrick-alberic

Post on 06-Jul-2016

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

Research Article

Received: 18 April 2011 Revised: 4 August 2011 Accepted: 6 August 2011 Published online in Wiley Online Library

Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

3012

Liquid chromatography/mass spectrometry stable isotopeanalysis of dissolved organic carbon in stream and soil waters†

Patrick Albéric*Université d’Orléans – CNRS/INSU UMR 6113, Institut des Sciences de la Terre d’Orléans (ISTO) 1A rue de la férollerie, 45071,Orléans cedex2, France

A commercial interface coupling liquid chromatography (LC) to a continuous-flow isotope ratio mass spectrometry(CF-IRMS) instrument was used to determine the d13C of dissolved organic carbon (DOC) in natural waters. Streamand soil waters from a farmland plot in a hedgerow landscape were studied. Based on wet chemical oxidation ofdissolved organics the LC/IRMS interface allows the on-line injection of small volumes of water samples, an oxida-tion reaction to produce CO2 and gas transfer to the isotope ratio mass spectrometer. In flow injection analysis (FIA)mode, bulk DOC d13C analysis was performed on aqueous samples of up to 100 mL in volume in the range of DOCconcentration in fresh waters (1–10 mg C.L–1). Mapping the DOC d13C spatial distribution at the plot scale was madepossible by this fairly quick method (10 min for triplicate analyses) with little sample manipulation. The relativecontributions of different plot sectors to the DOC pool in the stream draining the plot were tentatively inferred onthe basis of d13C differences between the hydrophilic and hydrophobic components. Copyright © 2011 John Wiley& Sons, Ltd.

(wileyonlinelibrary.com) DOI: 10.1002/rcm.5212

Several methods of coupling a dissolved organic carbon(DOC) analyzer to a continuous flow isotope ratio mass spec-trometry (CF-IRMS) system have been developed.[1–6] As forquantitative DOC analysis, two types of methodologies canbe employed to achieve organic carbon oxidation to CO2.

[1,2]

(i) High-temperature combustion (HTC) systems with cryo-genic trapping before transfer of the CO2 to the IRMSsystem.[4–6]Refractory organic carbon may be better oxidizedby HTC systems but the isotopic values have to be correctedfor a significant instrument blank depending on the volumeof water injected and time allowed for cryotrapping.[4–6] (ii)Wet chemical oxidation (WCO) methods are applied to largersample volumes with variable persulphate reagent concentra-tion in a closed reactor.[1–3] Blank corrections also need to beaccounted for, mostly depending on reagent volumes.[2,3]

Both types of system can use reduction ovens and gas chro-matographic separation before transfer to the IRMS system.The necessary complete purging of dissolved inorganiccarbon (DIC) before DOC (more exactly non-purgeableorganic carbon NPOC) analysis has to be carried out off-lineby acid introduction into the sample before it is injected intoHTC systems.[5,6] This can, however, be conducted on-linebefore introducing the oxidation reagent into the reactor for

* Correspondence to: P. Albéric, Université d’Orléans –CNRS/INSU UMR 6113, Institut des Sciences de la Terred’Orléans (ISTO) 1A rue de la férollerie, 45071 Orléanscedex2, France.E-mail: [email protected]

† Presented at the Liquid Chromatography-Isotope RatioMass Spectrometry Users Meeting, held 23–24 November2010 at the University of Oxford, Oxford, UK.

Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

WCO systems, allowing in this case both DIC and DOCcarbon isotopic analysis.[1–3]

In this study a commercial liquid chromatography (LC)/IRMS interface (LC IsoLink, Thermo Scientific, Bremen,Germany) was used to analyse the variation in DOC d13C instream and soil waters. The device, which was speciallydeveloped for on-line compound-specific isotope analysis ofwater-soluble samples after LC separation,[7] can also be usedwithout a column in bulk mode analysis for simple[7] andmacromolecular compounds.[8] In the field of complex rawsamples the LC IsoLink interface has been already used forbulk carbon isotope ratio determination in alcoholicbeverages.[9] The main advantage of this method is that thewater sample is injected into a continuous flow of mobilephase (ultrapure deionised water) and reagents mixed at theentry of a heated capillary reactor connected on-line to anopen-split coupling and to the IRMS system via a liquid/gas (water/He) membrane exchanger and a gas-drying unit(see Krummen et al.[7] for a detailed description).

The field selected as an example for this study was a farm-land plot in which the different characteristics (redoxconditions, ionic strength, dissolved metals, water-tabledepths) of several sectors had previously been determined.[10]

The factors controlling DOC in soils are complex and they canbe influenced by numerous parameters and processes such assubstrate quality, decomposer community, adsorption inmineral soil horizons, pH, ionic strength, temperature, soilmoisture, water fluxes and redox conditions.[11,12] Someof these factors may also control the variation in the isotopecomposition of organic carbon in soil profiles. As evidence islacking for any direct isotope fractionation mechanisms,[13,14]

the effects of selective sorption and/or preferential preserva-tion of isotopically and chemically heterogeneous complexsoil organic matter have been investigated, mostly through

Copyright © 2011 John Wiley & Sons, Ltd.

Page 2: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

LC/MS stable isotope analysis of DOC d13C in natural water

experiment.[15–17] Mapping DOC d13C in soil water at the plotscale can be a complementary approach to decomposition orsorption experiments but it involves managing a large numberof water samples. The aim of the present work was first totest the applicability of the LC IsoLink interface to sucha soil water field study. As soils account for the principalsource of DOC in streams,[18] the isotope composition ofseveral organic fractions extracted from the water of thesmall creek draining the plot were compared with thoseof the soil water DOC obtained from the different sectorsin the plot. Descriptions of the site, soil horizons, and pro-cedures of sampling or isolation of specific organic frac-tions (with XAD resins) prior to d13C analysis will not bedetailed in this paper. The objective is to describe the fea-sibility of analysing a large number of small volumessamples for their DOC d13C values at fresh water naturalconcentrations with a commercial LC/IRMS interface.

0

500

1000

1500

2000

2500

3000

3500

0 100 200 300 400 500 600Time (sec)

Am

plitu

de (m

V)

Soil water W7 : 13CDOC = -27.99‰ ± 0.02‰Total area = 74 Vs ± 0.3 Vs

m/z46

m/z45

m/z44

Figure 1. Triplicate DOC d13C flow injection analysis (FIA).Run starts with three monitor gas pulses for d13C calculationand ends with two for control. Traces of m/z 44, 45 and 46(two later amplified) are indicated. Full loop injection 100 mL.Sample preparation: filtration, acidification, gas purge.Mobilephase: MilliQ water pH 2 with H3PO4, 300 mL/min; T reactor100 �C; reagents: H3PO4 (1.5 M 50 mL/min); Na2S2O8 (0.4 M40 mL/min). Triplicate analysis in 10 min.

301

EXPERIMENTAL

Site and soil description and water sampling

The site and field work have been detailed elsewhere.[10,19]

Briefly, the plot selected for the study is located on a hillsideused as pasture in a hedgerow landscape. The soils arePlanosols developed on altered gneisses and exhibit a stonyhorizon rich in Fe–Mn oxides overlaying a clayey B horizonsupporting a seasonal water table in winter. Soil waters werecollected below the water table from 80 cm depth to near thesurface, free-flowing in auger-drilled holes. The volumesof water filling the auger holes were between 0.1 and 2 L.Landscapesmarked by hedgerows (bocage landscape) are typi-cally non-uniform areas in which the 3D distribution of soilhorizons and anthropogenic structures (e.g. hedges, ditchesor wheel ruts) affects the composition and circulation ofwater.[20] Based on water solute composition the plot could besubdivided into four geochemical sectors:[10] (i) a sector linedby hedges characterised by a deeper water table, higherdissolved oxygen content and ionic strength; (ii) a sector withhigher DOC concentrations (spatially correlated with shallowsoil water levels enhancing contact between soil solutionand humic top soil horizons); (iii) a sector influenced by theupward infiltration of underground waters rich in reducedmanganese in locations where the deep clayey B horizon isthinner or missing;[21] and, finally, (iv) two anoxic Fe2+ andMn2+ rich spots close to the bank of the stream. Each sectormay contribute differently to the chemical composition ofthe stream water.

Sample preparation

Soil water subsamples for DOC analysis were filtered(0.45 mm) and acidified to pH 1–2 with either concentratednitric or phosphoric acid (but not HCl so as to preservepersulphate oxidation efficiency, see below).Separation of the different DOC fractions was carried out

on larger volumes of water (about 7 L) sampled in the streamdraining the plot. Hydrophobic (HPO) and transphilic (TPH)acids were recovered by tandem XAD8/XAD4 adsorptionchromatography.[22] HPOacid and TPHacid fractions wereadsorbed at pH 2 and then eluted in a small volume

Copyright © 2011Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

(50 mL) of basic solution (pH 12), respectively, from theXAD8 and the XAD4 resins (80 mL resin bed each): TPHacidsare more hydrophilic than HPOacids. Other classicallydefined fractions[22] either remained in solution in the waterpassing through the tandem columns (i.e. the most hydro-philic compounds, HPI) or were retained on the XAD8 resin(i.e. the more hydrophobic neutral compounds, HPOneutral).In addition, analytical separation using only a small XAD8column[23] yielded 50% of the HPOacid and 30% of theHPI + TPH.

DOC d13C LC/IRMS analysis

A system composed of an LC unit (Surveyor MS pump withautosampler) coupled to a DeltaV-Advantage isotope ratiomass spectrometer via a LC-IsoLink interface (all suppliedby Thermo Scientific, Bremen, Germany) was used. Bulkanalyses of DOC d13C were run in flow injection analysis(FIA)[8] mode (i.e. without column). Per mil (%) deviationsfrom the international standard Vienna Peedee Belemnite(VPDB) were calculated by the Isodat software (ThermoScientific). d13C standardization was carried out using simplemolecules from IAEA (International Atomic Energy Agency,Vienna, Austria) and USGS (US Geological Survey, Reston,VA, USA) and complex aquatic fulvic acids from IHSS (Inter-national Humic Substances Society, St Paul, MN, USA).

The mobile phase (MilliQ water: Millipore, Billerica, MA,USA; brought to pH 2 with H3PO4) and reagents (see Fig. 1caption) were degassed under vacuum in a sonic bath andthen continuously purged with He to prevent air re-gassing.[8,9] The LC flow rate was fixed at 300 mL/min, basedon its influence on the signal intensity and the d13C values.[8]

The flow rates were fixed at 50 mL/min for the acid reagentand between 35 and 40 mL/min for the oxidant reagent tomaintain the signal for O2 (m/z 32) at about 10 V which is indi-cative of a sufficient level of oxidant through the reactor. Thereactor temperature was set at 100 �C. A sample volume up to

wileyonlinelibrary.com/journal/rcmJohn Wiley & Sons, Ltd.

3

Page 3: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

P. Albéric

3014

100 mL was routinely used to lower the detection limit to0.5 mg C L–1. Acidified samples were manually purged withHe (for DIC-CO2 elimination) in the 1.8 mL autosampler vials(during 3–4 min) prior to being placed in the autosamplerrack and subsequent injection. For organic concentratedfractions (HPOacid, TPHacid), 10 mL of sample diluted ten-fold with water were used. Each run started with threegas pulses for d13C determination and ended with twofurther pulses for control purposes; analysis of the samplein triplicate was completed in 10 min (Fig. 1).

Precision and accuracy of the system

The d13C values were calculated relative to the valuesobtained from pulses of CO2 analysed at the beginning ofeach run. The standard deviation (SD, 1s) for triplicateanalysis of standard solutions or water samples was less than0.1%. The d13C value of the CO2 monitor gas was repeatedlycalibrated using a set of reference standard compoundsdissolved in water. The set comprising both simple moleculesand natural complex macromolecules ranged from �10.45%(sucrose IAEA-CH-6) to �27.6% (Suwannee River fulvic acidIHSS-1S101F). Intermediate d13C values were determinedwith sucrose IAEA-C6, –10.8%; oxalic acid IAEA-C8, –18.3%; and glutamic acid USGS-40, –26.39%. The stabilityover time of the system was controlled by plotting thereference standard d13C values versus the measured values.In the example shown in Fig. 2, the maximum differencebetween the targeted and measured values did not exceed

IHSS-1S101F

IAEA-CH-6

IAEA-C6

USGS-40

IAEA-C8

-30

-25

-20

-15

-10

-30 -25 -20 -15 -10

δ13C measured

δ13C

std

Figure 2. Accuracy of the system. Calibration of the d13Cvalue of the CO2 monitor gas over a set of reference standardcompounds dissolved in water. Stability over time of the sys-tem was controlled by plotting the reference standard d13Cvalues versus measured values. The set of reference materialsranged from �10.45% (sucrose IAEA-CH-6) to �27.6%(Suwannee River fulvic acid IHSS-1S101F). Intermediated13C values correspond to sucrose IAEA-C6, –10.8%; oxalicacid IAEA-C8, –18.3%; and glutamic acid USGS-40, –26.39%. Standard deviations for triplicate were less than themark size. The difference between targeted and measuredvalues did not exceed 0.5%. Least-squares linear regressionof the data can be used to correct measured values within a0.3% difference.

wileyonlinelibrary.com/journal/rcm Copyright © 2011 John Wile

0.5%. Least-squares linear regression of the data was usedto correct the measured values within a 0.3% difference fromthe standard values.

Influence of acidification and purging – reproducibility

Purging DIC before analysis is an obligatory step when mea-suring DOC d13C in natural waters.[1] For the in-flow injectionof the water sample in the mobile phase on-line to the reactorand the gas exchanger, the simple method employed consistedof acidifying and purging samples before injection (NPOCtreatment). Depending on the alkalinity of the samples, 1 or 2drops of concentrated phosphoric acid or nitric acid (MerckSuprapurW; Merck, Dartmstadt, Germany) were used per5mL ofwater sample. The influence of purging timewas testedon MilliQ solutions of Suwannee River fulvic acid andK-biphthalate (Merck standard compound for elementary ana-lysis used for DOC quantification) and a natural soil watersample, all acidified with phosphoric acid (Fig. 3). Purgingwas carried out manually in the 1.8 mL autosampler vials witha flux of He (derived from the reagents and mobile phasepurging lines) introduced with a stainless steel syringe needlethrough the vial septum. As shown in Fig. 3, the influence ofdissolved CO2 was smaller in the case of the MilliQ solutionsthan for natural water but still remained substantial. A purgingtime of 3 to 4 min was considered sufficient. The mean valuesand SDs for the ≥3 min purging time range (≥1 min forK-biphthalate) are also given in Fig. 3. The mean d13C value(�27.58 � 0.05%, 1s) found for the river fulvic acid solutionalong the 3 to 10 min purge time range conformed to the valuegiven by the IHSS (�27.6%). Assays with unacidified, oracidified but unpurged, fulvic acid solution, in addition toyielding a less negative mean d13C value, also had a largerSD, probably due to variable levels of contamination by CO2.Duplicate tests with nitric acid did not significantly changethe mean values and the SDs. On the contrary, an alteration inthe peak shape and more negative values were found whenHCl was used, due to an incomplete oxidation of organic car-bon in the reactor. The interference caused by halides in WCOmethods has been overcome in large volume reactor systemsby adding larger quantities of persulphate.[2,3] This interferenceseems, however, to be an obstacle when analysing low DOCsaline waters with the LC interface used in this work. Thepotential effects of adding acid in samples were commentedon by De Troyer et al.[6] who concluded that the NPOC treat-ment was acceptable for HTC d13C analysis. The long-termstorage of acidified DOC-rich samples should, however, beavoided because of the possible flocculation of macromolecularsubstances. This will increase the risk of plugging the on-linefilter set at the entry of the LC interface reactor and of losing afraction of the sample DOC pool, potentially resulting in biasedd13C values.

Linearity – Background and blank levels – Quantification of DOC

Using benzoic acid solutions, Krummen et al.[7] in the origi-nal publication on the LC IsoLink interface found the line-arity of the system to range from 200 to 2000 ng C. Dilutionseries of the same solutions as used above (Fig. 3) wereanalysed to confirm the compatibility of the method withthe concentration range and the type of organic matterfound in natural waters. As shown in Fig. 4, the meand13C values were within about �0.1 SD for the range

y & Sons, Ltd. Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

Page 4: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

Figure 4. Linearity. d13C values against DOC concentrationsfor dilution series of the three different solutions as in Fig. 3.

-27.8

-27.7

-27.6

-27.5

-27.4

-27.3

-27.2

-27.1

-27.0

0 5 10

He purge time (min)

13C

(‰)

IHSS -1S101F (5 mg C/L)

-27.58 ± 0.05‰ (n = 6)

-27.0

-26.9

-26.8

-26.7

-26.6

-26.5

-26.4

-26.3

-26.2

0 1 2 3 4 5

He purge time (min)

13C

(‰)

K-Phtalate (6 mg C/L)

-26.83 ± 0.06‰ (n = 4)

-30

-28

-26

-24

-22

-20

-18

-16

0 5 10

He purge time (min)

13C

(‰)

Soil water J36 (DOC 6.5 mg C/L)

-28.38 ± 0.04‰ (n = 3)

Figure 3. d13C values as a function of He purge time for threedifferent solutions: Suwannee River fulvic acid IHSS-1S101F,K-biphthalate and a natural soil water sample. Symbols anderror bars represent mean � standard deviation (SD) fortriplicates. The mean d13C values for the ≥3 min purging timerange (≥1 min for K-biphthalate) are also given.

LC/MS stable isotope analysis of DOC d13C in natural water

301

1–10 mg DOC/L (i.e. 100–1000 ng C). The mean values forthe soil water sample and the stream fulvic acid were notdistinguishable from those obtained in the runs shown inFig. 3. The value calculated for K-biphthalate by this runwas less negative (�26.3% vs. –26.8%) but the mean valuewas still within a 0.3% SD.With the water mobile phase used in this study a low m/z

44 background of between 60 and 70 mV was maintained.The pulse intensity of the CO2 monitor gas was set around2 V (1750 mV in Fig. 1) by fixing the gas pressure on the LCIsoLink interface entry gauge. Over time, many factors cancause variation in the intensity of the system signal[8] (e. g.new filament, source focusing, adjustment of the open splitunits and He flows) which can be monitored by fixing thereference gas entry pressure. Unlike with the HTC systems,the mobile phase and reagents used in the continuous-flow

Copyright © 2011Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

LC interface generate no instrument blank. Injection of 100mL of degassed acidified MilliQ water gave, however, a blanksignal of around 1 � 0.5 Vs (equal to about 0.1 mg DOC/L),mostly attributable to the acid introduced into the samples.No blank correction of the d13C values was considerednecessary from a concentration limit of 1 mg DOC/L. Underthat limit the signal intensity (area in Vs) appears too small togive a correct calculation of d13C values in accordance withlinearity evaluations.

Solutions of either K-biphthalate or benzoic acid (IVAAnalysentechnik, Meerbusch, Germany) were used as DOCstandards for calibration, with no difference being observedin the concentration (mg C/L) versus IRMS peak area (Vs)relationship. The calibration may, however, vary over aperiod of years by more than 50% and it needs to be checkeddaily. DOC quantification down to 0.25 mg C/L was possible

wileyonlinelibrary.com/journal/rcmJohn Wiley & Sons, Ltd.

5

Page 5: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

Figure 5. Comparison between DOC determinations by LC/IRMS flow injection analysis (FIA) and HTC Shimadzu TOCanalyser. The solid line indicating the trend (and the correla-tion coefficient) between the two sets of data did not signifi-cantly diverge from the 1:1 line (dotted). In view of thedelays between quantification runs and the multiplicityof subsamples, the �1 mg C/L standard error (n = 63) indi-cates no systematic error in DOC estimation by LC/IRMS(see text).

Table 1. d13C composition relative to VPDB

Sampled13C(%)

Soil water DOC (average, n = 70) �28.7 � 0.6Soil water DOC (maximum) �27.4Soil water DOC (minimum) �29.9Bulk stream DOC (average, n = 3) �28.8 � 0.4Bulk TPHacid (stream) �28.5Bulk HPOacid (stream) �30.0

P. Albéric

3016

with a precision (SD) of ≤0.02 mg/L and an standard error ofabout 0.1 mg C/L (n = 10). Comparison with quantificationusing a high-temperature catalytic total organic carbon(TOC) analyser (Shimadzu TOC 5000A; Shimadzu, Kyoto,Japan) is shown in Fig. 5 for part of the whole sample set.Considering the delays between the two quantification runsand the multiplicity of subsamples, the 1 mg C/L standarderror concordance indicates that there is no systematic errorin the DOC concentrations obtained by LC/IRMS. A recent

Figure 6. Spatial distribution of soil waternegative values (≤ –29.3%) overlappedplot (green grass symbol), while less nalong hedgerows. Black contour lines reabove sea-level. Abscissa and ordinate t

wileyonlinelibrary.com/journal/rcm Copyright © 2011 John Wile

better-controlled interlaboratory comparison test (LC Iso-Link-IRMS; ISTO versus Shimadzu TOC 5000 LGE/IPGP;Laboratoire de Géochimie des Eaux/Institut de Physique duGlobe de Paris, Paris, France) on stream waters from theAmazonas-Rio Negro River system was conducted with onlya 0.25 mg/L standard error being observed in a 2–8 mg/LDOC range (unpublished results).

APPLICATION TO NATURALWATER

Mapping DOC d13C in soil waters at the plot scale

The results for the study of natural waters are shown in Fig. 6and Table 1. The values fell in a narrow range, between �27.4and �29.9%, averaging as a whole �28.7 � 0.6% (1s, n = 70).The soil water d13C values were only averaged to comparethem with stream water values and with separate domainsin the plot. It appears that the spatial distribution of thevalues (although not large enough to be analysed statistically)is not uniform. A cluster with more negative values is distin-guishable overlapping with the previously defined sector(ii), where higher DOC concentrations were assumed to belinked to the upper levels of the water table and an enhancedcontact between the water and organic top soil layers.[10]

DOC d13C in the plot. Most of the moresector (ii) in the central zone of the

egative values were generally foundfer to soil surface elevation in metresick intervals: 10 metres.

y & Sons, Ltd. Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

Page 6: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

LC/MS stable isotope analysis of DOC d13C in natural water

301

Comparison with the stream waters draining the plot

The d13C values of both the bulk stream DOC (�28.8 � 0.4%,n = 3) and the XAD4 TPHacid fraction (�28.5%) matched theplot average soil water value (�28.7 � 0.6%). However, themore hydrophobic XAD8 HPOacid fraction (�30.0%) wasfound to be 1.5% more negative, being therefore closer tothe set of values found in sector (ii) (green grass symbol inFig. 6). The more negative d13C signature of hydrophobicfractions has already been documented, for example, in forestsoils by Kaiser et al.[16] who found a difference of about 3%.This general characteristic of hydrophobic fractions wasattributed to the larger contribution of compounds derivedfrom lignin that are depleted in 13 C.The comparison between the TPHacid and HPOacid d13C

values in the stream and the spatial DOC d13C distributionin the plot suggests that the different areas in the plot maycontribute differently to the stream DOC pool. Well-aeratedand drained hedgerow banks may produce more hydrophilicand transphilic organic fractions (less negative d13C values)whereas more poorly drained places may supply a morehydrophobic pool (more negative values). This hypothesisagrees with the higher ionic strength and more oxicconditions prevailing in the proximity of hedgerows,[10]

which are factors known to enhance flocculation and sorptionof the more hydrophobic dissolved humic substances[11,12]

and thus to influence DOC d13C evolution through soilprofiles.[15,16,24]

To ascertain this, analysis of the hydrophobic and hydro-philic pools of the soil waters should be carried out. The off-lineextraction and concentration of organic fractions by XAD resinsare, however, highly time- and sample-consuming procedureswhile on-line LC/IRMS methods are opening up newperspectives. The application of LC/IRMS coupled size-exclusion chromatography (SEC) to the XAD resin fractionsanalysed in this study,[25] although potentially a promisingtool for discriminating dissolved substances on a molecularweight and stable carbon isotope basis, did not answerthe question. The on-line analytic separation of hydrophilicand hydrophobic fractions on small volumes[23] would be abetter approach provided that it is adapted to LC/IRMSconstraints. Temperature-programmable LC capabilities[26]

should preferably be explored.

CONCLUSIONS AND RECOMMENDATIONS

Although the above discussion requires further work to becarried out for the results to be conclusive, the LC IsoLinkinterface was shown to be a useful high-sensitivity tool forthe reproducible and accurate IRMS on-line determinationof d13C values of DOC in fresh waters. The technique allowslarge sampling and a fast field survey in a approach compa-rable with that developed by Jochmann et al.[9] in anotherdomain. Analysis in triplicate takes only 10 min andconsumes small sample volumes (a few mL, 100 mL singleinjection) of natural water that have only been filteredand acidified. Accurate d13C determination was possibledown to 1 mg DOC/L, and DOC quantification down to0.25 mg C/L with a precision (SD) of ≤0.02 mg/L and anstandard error of about 0.1 mg C/L. Comparison with theresults obtained using an HTC TOC analyzer did not revealany systematic error in DOC evaluation by this WCO-based

Copyright © 2011Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018

interface. Compared with other existing HTC or WCOmethods, a further advantage of LC/IRMS is that thereis no need to apply blank corrections for d13C calculations.Extreme care should be taken during the purge step priorto injection since experience has shown that an uncontrolledHe flow deficiency during manual purge is the main causeof failed analyses. The use of an automatic sampler witha gas-sparging option (as in most commercial TOC analysers)would be a significant improvement. The main drawbackof the WCO principle of the LC/IRMS interface concernsthe analysis of low DOC saline solutions owing to their beingonly partial oxidation of the organic compounds in theon-line reactor.

AcknowledgementsSophie Cornu (INRA), Aurélie Vennink and Ary Bruand(ISTO) are warmly thanked for their help during the winterfield work and Dieter Juchelka (Thermo Scientific, Bremen,Germany) for technical support. The referees are sincerelyacknowledged for their valuable comments which have con-tributed to improving the manuscript.

REFERENCES

[1] G. St-Jean. Automated quantitative and isotopic (13C)analysis of dissolved inorganic carbon and dissolvedorganic carbon in continuous-flow using a total organiccarbon analyser. Rapid Commun. Mass Spectrom. 2003,17, 419.

[2] S. Bouillon, M. Korntheuer, W. Baeyens, F. Dehairs. A newautomated setup for stable isotope analysis of dissolvedorganic carbon. Limnol. Oceanogr.: Methods 2006, 4, 216.

[3] C. L. Osburn, G. St-Jean. The use of wet chemical oxidationwith high-amplification isotope ratio mass spectrometry(WCO-IRMS) to measure stable isotope values of dissolvedorganic carbon in seawater. Limnol. Oceanogr.: Methods 2007,5, 296.

[4] S. Q. Lang, M. D. Lilley, J. I. Hedges. A method to measurethe isotopic (13C) composition of dissolved organic carbonusing a high temperature combustion instrument. Mar.Chem. 2007, 103, 318.

[5] R. J. Panetta, M. Ibrahim, Y. Gelinas. Coupling a high-temperature catalytic oxidation total organic carbonanalyzer to an isotope ratio mass spectrometer to measurenatural-abundance d13C-dissolved organic carbon in marineand freshwater samples. Anal. Chem. 2008, 80, 5232.

[6] I. De Troyer, S. Bouillon, S. Barker, C. Perry, K. Coorevits,R. Merckx. Stable isotope analysis of dissolved organiccarbon in soil solutions using a catalytic combustion totalorganic carbon analyzer-isotope ratio mass spectrometerwith a cryofocusing interface. Rapid Commun. Mass Spec-trom. 2010, 24, 365.

[7] M. Krummen, A. W. Hilkert, D. Juchelka, A. Duhr, H. J.Schlüter, R. Pesch. A new concept for isotope ratio monitor-ing liquid chromatography/mass spectrometry. RapidCommun. Mass Spectrom. 2004, 18, 2260.

[8] J.-P. Godin, J. Hau, L.-B. Fay, G. Hopfgartner. Isotope ratiomonitoring of small molecules and macromolecules byliquid chromatography coupled to isotope ratio massspectrometry. Rapid Commun. Mass Spectrom. 2005, 19, 2689.

[9] M. A. Jochmann, D. Steinmann, M. Stephan, T. C. Schmidt.Flow injection analysis-isotope ratio mass spectrometry forbulk carbon stable isotope analysis of alcoholic beverages.J. Agric. Food Chem. 2009, 57, 10489.

wileyonlinelibrary.com/journal/rcmJohn Wiley & Sons, Ltd.

7

Page 7: Liquid chromatography/mass spectrometry stable isotope analysis of dissolved organic carbon in stream and soil waters

P. Albéric

3018

[10] P. Albéric, A. Vennink, S. Cornu, H. Bourennane, A. Bruand.A snapshot of soil water composition as an indicator ofcontrasted redox environments in a hedged farmland plot.Sci. Total. Environ. 2009, 407, 5719.

[11] K. Kalbitz, S. Solinger, J.-H. Park, B. Michalzik, E. Matzner.Controls on the dynamics of dissolved organic matter insoils: A review. Soil Sci. 2000, 165, 277.

[12] S. Fiedler, K. Kalbitz. Concentrations and properties ofdissolved organic matter in forest soils as affected by theredox regime. Soil Sci. 2003, 168, 793.

[13] J. Balesdent, A. Mariotti, in Mass Spectrometry of Soils, (Eds:J. W. Boutton, S.I. Yamosaki). Marcel Dekker, New York,1996, pp. 88–111.

[14] E. M. Galimov. Isotope organic geochemistry. Org. Geochem.2006, 37, 1200.

[15] C. C. Cleveland, J. C. Neff, A.R. Townsend, E. Hood. Com-position, dynamics, and fate of leached dissolved organicmatter in terrestrial ecosystems: Results from a decomposi-tion experiment. Ecosystems 2004, 7, 275.

[16] K. Kaiser, G. Guggenberger, W. Zech. Isotopic fractionationof dissolved organic carbon in shallow forest soils as affectby sorption. Eur. J. Soil Sci. 2001, 52, 585.

[17] B. Ludwig, B. Heil, H. Flessa, F. Beese. Dissolved organiccarbon in seepage water – Production and transformationduring soil passage. Acta Hydrochim. Hydrobiol. 2000,28, 77.

[18] M. Meybeck, in Erosion and Carbon Dynamics, (Eds: E. J.Roose, R. Lal, C. Feller, B. Barthès, B. A. Stewart),Advances in Soil Science, CRC Press, Boca Raton, 2005,pp. 209–238.

wileyonlinelibrary.com/journal/rcm Copyright © 2011 John Wile

[19] S. Cornu, V. Deschatrettes, S. Salvador-Blanes, B. Clozel,M. Hardy, S. Branchut, L. Le Forestier. Trace element accu-mulation in Mn–Fe-oxide nodules of a planosolic horizon.Geoderma 2005, 125, 11.

[20] G. Bourrié, V. Maître, P. Curmi. Mise en évidence de deuxdynamiques saisonnières du fer dans les sols hydromorphesen climat tempéré. C. R. Acad. Sci. – Série II 1994, 318, 87.

[21] F. Delarue, S. Cornu, J. Daroussin, S. Salvador-Blanes, H.Bourennane, P. Albéric, A. Vennink, A. Bruand, D. King.3D representation of soil distribution: An approach forunderstanding pedogenesis. C. R. Geoscience 2009, 341, 486.

[22] G.R. Aiken, D. M. McKnight, K. A. Thorn, E. M. Thurman.Isolation of hydrophilic acids from water using nonionicmacroporous resins. Org. Geochem. 1992, 18, 567.

[23] C. Marchand, P. Albéric, E. Lallier-Vergès, F. Baltzer. Distri-bution and characteristics of dissolved organic matter inmangrove sediment pore waters along the coastline ofFrench Guiana. Biogeochemistry 2006, 81, 59.

[24] E. Lloret, C. Dessert, J. Gaillardet, P. Albéric, O. Crispi,C. Chaduteau, M.F. Benedetti. Comparison of dissolvedinorganic and organic carbon yields and fluxes in thewatersheds of tropical volcanic islands, examples fromGuadeloupe (FrenchWest Indies). Chem. Geol. 2011, 280, 65.

[25] P. Albéric, P. Gautret, D. Juchelka. irm-LC/MS: d13C analy-sis of dissolved organic matter in natural waters. ApplicationNote Thermo Scientific 2010, AN30198.

[26] J.-P. Godin, G. Hopfgartner, L. Fay. Temperature-programmed high-performance liquid chromatographycoupled to isotope ratio mass spectrometry. Anal. Chem.2008, 80, 7144.

y & Sons, Ltd. Rapid Commun. Mass Spectrom. 2011, 25, 3012–3018