Laparoscopic Placement of the BardPort Intraperitoneal ... · PDF fileLaparoscopic Placement of the BardPort Intraperitoneal Catheter and Reservoir. ... Laparoscopic Placement of the
28
Laparoscopic Placement of the BardPort Intraperitoneal Catheter and Reservoir
Laparoscopic Placement of the BardPort Intraperitoneal Catheter and Reservoir
Choice of Catheter and Reservoir is the IV Device
Titanium Port with 9.6F attached catheter
Pre-placement Procedure
Perform laparoscopy in the usual manner with umbilical placement of laparoscope trocar and survey of the abdominal cavity for adhesionsPlace a 5 mm operative trocar in a position contralateral to the side of proposed catheter placementIdentify the prospective site in the lower quadrant for entry of the catheter into the peritoneal cavity
Mark the site for reservoir placement above the right costal margin
Incise the skin and develop a pocket for the reservoir
Incise the skin and develop a pocket for the reservoir
Check that the reservoir fits snugly into the pocket
Place the tunneler on the end of the catheter and start the
passage deep to Scarpa’s fascia
Place a 5 mm transverse incision at the target site
Bluntly with a hemostat extend the incision beneath Scarpa’s fascia to contact the tunneler
Pass the tunneler out through the target incision
Pass the tunneler out through the target incision
Pass the catheter through the tunnel and the target incision and cut
catheter to remove tunneler
Suture the reservoir in place to fascia and confirm catheter function
Place introducer needle through target incision adjacent to catheter
Visualize needle tip at the target site in the peritoneal cavity
Pass the guidewire with “J” tip through the introducer needle
Pass the guidewire with “J” tip through the introducer needle
Visualize “J” tip of the wire at the target site in the peritoneal cavity
Withdraw the introducer needle leaving the guidewire in place
Insert the introducer with the 10F dilator over the guidewire into the
peritoneal cavity
Visualize the dilator AND the peel-away introducer in the peritoneal cavity
Dilator
Dilator
Introducer
Withdraw both the wire and the dilator from the introducer
Pass the catheter through the introducer and assist
laparoscopically in advancing the catheter into the peritoneal cavity
Peel away the introducer while holding the catheter in the
peritoneal cavity
Advance all the remaining catheter into the peritoneal cavity
Adjust position of catheter tip and confirm function by injection