köhler illumination with the inverted microscope - mikroskop · köhler illumination with the...

1
Köhler Illumination with the Inverted Microscope EN_41_050_005 | CZ-01/2014 | Design, scope of delivery and technical progress subject to change without notice. | © Carl Zeiss Microscopy GmbH Brightfield and Phase Contrast Set condenser to brightfield position. Move condenser vertically to a +/- middle position using the condenser focusing knob. Open both the luminous field stop and aperture stop fully. 1. Focus specimen using the focusing knob. Keep specimen in focus during the whole alignment procedure. 2. Close luminous field stop until it appears in the field of view. 3. Focus the luminous field stop image by changing the height of the condenser with the condenser focusing knob. Now, the edges of the luminous field stop appear with maxi- mum sharpness. 4. Center the luminous field stop image with the condenser centering screws. 5. Open the luminous field stop until its image just disappears behind the field of view edge. For alignment of phase contrast please go to step 8. 6. To adjust the aperture stop diameter remove an eyepiece and look into the tube (Auxiliary microscope recommended). Close the aperture stop... 7a. ...until the visible objective opening is illuminated at least by 2/3 of its diameter. Insert eyepiece. With each objective change realign luminous field and aperture stop. Now, the microscope is optimally aligned for brightfield examinations. 7b. Insert specimen and adjust Köhler illumination according to steps 1-6. 8. Make sure the condenser phase stop matches the chosen phase contrast objective. 9. Replace one eyepiece with the auxilliary microscope (alternative diopter) and focus on the dark objective phase ring and the bright condenser phase stop images. 10. Insert centering keys into the phase alignment openings of the condenser. Rotate the keys until the images of phase ring and phase stop match as shown. Replace the auxilliary microscope with the eyepiece. With each objective change, realign luminous field stop. If necessary, change to the condenser phase stop indicated on the phase contrast objective. Now, the microscope is opti- mally aligned for phase contrast examinations. 11. Carl Zeiss Microscopy [email protected] www.zeiss.com/microscopy

Upload: others

Post on 05-Jun-2020

13 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Köhler Illumination with the Inverted Microscope - mikroskop · Köhler Illumination with the Inverted Microscope EN_41_050_005 | CZ-01/2014 | Design, scope of delivery and technical

Köhler Illumination with the Inverted Microscope

EN_4

1_05

0_00

5 | C

Z-01

/201

4 | D

esig

n, s

cope

of

deliv

ery

and

tech

nica

l pro

gres

s su

bjec

t to

cha

nge

wit

hout

not

ice.

| ©

Car

l Zei

ss M

icro

scop

y G

mbH

Brightfield and Phase Contrast

Set condenser to brightfield position. Move condenser vertically to a +/- middle position using the condenser focusing knob. Open both the luminous field stop and aperture stop fully.

1.

Focus specimen using the focusing knob. Keep specimen in focus during the whole alignment procedure.

2.

Close luminous field stop until it appears in the field of view.

3.

Focus the luminous field stop image by changing the height of the condenser with the condenser focusing knob. Now, the edges of the luminous field stop appear with maxi-mum sharpness.

4.

Center the luminous field stop imagewith the condenser centering screws.

5.

Open the luminous field stop until its image just disappears behind the field of view edge. For alignment of phase contrast please go to step 8.

6.

To adjust the aperture stop diameterremove an eyepiece and look into the tube (Auxiliary microscope recommended).

Close the aperture stop...

7a.

...until the visible objective opening is illuminated at least by 2/3 of its diameter. Insert eyepiece. With each objective change realign luminous field and aperture stop. Now, the microscope is optimally aligned for brightfield examinations.

7b.

Insert specimen and adjust Köhlerillumination according to steps 1-6.

8.

Make sure the condenser phase stop matches the chosen phase contrast objective.

9.

Replace one eyepiece with the auxilliarymicroscope (alternative diopter) and focus on the dark objective phase ring and the bright condenser phase stop images.

10.

Insert centering keys into the phase alignment openings of the condenser. Rotate the keys until the images of phase ring and phase stop match as shown. Replace the auxilliary microscope with the eyepiece. With each objective change, realign luminous field stop. If necessary, change to the condenser phase stop indicated on the phase contrast objective. Now, the microscope is opti-mally aligned for phase contrast examinations.

11.

Carl Zeiss Microscopy [email protected] www.zeiss.com/microscopy