IAEAInternational Atomic Energy Agency

Accidents in Radiotherapy

Radiation Sources in Radiotherapy

Day 7 – Lecture 6

Page 2: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 2

Objective

To become familiar with the contributing factors that can led to unwanted exposure in radiotherapy.

To be aware of investigation of accidents; case studies of accidental exposures in radiotherapy and lessons learned.

Role of the Regulatory Body in case of accidents and emergencies in radiotherapy.

Page 3: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA

Contents

• Potential for accidents in radiotherapy.• Deviation from the prescribed dose.• Potential exposures in external beam therapy.• Potential exposures in brachytherapy.• Case examples and lessons learned from accidental

exposures in radiotherapy.• Role of the Regulatory Body.

Page 4: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 4

Potential for accidents in radiotherapy

Radiotherapy is unique from the point of view of radiation safety because:

• it is the only use of radiation sources in which very high radiation doses are given deliberately to the human body.

• not only the tumour, but also normal tissue receives radiation doses comparable to the tumour dose.

Page 5: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA

Potential Exposure

“Prospective exposure that is not expected to be delivered with certainty but that may result from an anticipated operational occurrence, accident at a source or owing to an event or sequence of events of a probabilistic nature, including equipment failures and operating errors.”

[GSR Part 3 Definitions]

Potential for accidents in radiotherapy (cont)

Page 6: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 6

• may involve severe or even fatal consequences;

Deviations from the prescribed dose

Deviations from the prescribed dose:

• require prompt investigation by licensees in the event of an accidental patient exposure.

Page 7: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 7

“Any medical treatment delivered to the wrong individual or the wrong tissue of the patient, or using the wrong radiopharmaceutical, or with an activity, a dose or dose fractionation differing substantially from (over or under) the values prescribed by the radiological medical practitioner, or that could lead to unduly severe secondary affects”

Accidental Medical Exposure

“Any failure of medical radiological equipment, software failure or system failure, or accident error, mishap or other unusual occurrence with the potential for subjecting the patient to a medical exposure that is significantly different from what was intended.”

[GSR Part 3 Requirement 41. 3.179]

Page 8: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 8

For external beam therapy, accidents may be related to:

Causes of accidents

• equipment design;• beam calibration;• maintenance;• treatment planning;• dose calculation;• treatment simulation;• treatment setup and delivery.

Page 9: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 9

An operator of a linear accelerator selected x-ray mode and then quickly changed to electron mode. The machine was unable to complete the first request (to operate in x-ray mode) and subsequently operated with “hybrid” instructions.

Examples – Equipment Design

The same accident occurred in 6 different hospitals and 2 patients died due to doses ranging from 160 -180 Gy

Contributing factors

• The software controlled linear accelerators were not tested for the conditions that occurred in practice at the hospitals.

Page 10: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 10

Examples – Equipment Design (cont)

The same accident occurred in 6 different hospitals and 2 patients died due to doses ranging from 160 -180 Gy.

• The manufacturer took too long to identify the problem and to disseminate the information. By then the 6 hospitals had experienced the same failure and 2 patients had died from their radiation exposure.

Contributing factors (cont)

An operator of a linear accelerator selected x-ray mode and then quickly changed to electron mode. The machine was unable to complete the first request (to operate in x-ray mode) and subsequently operated with “hybrid” instructions.

Page 11: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 11

Accidents due to mistakes in properly determining the radiation dose rate have led to patient over-exposure, in some cases by as much as 60%.

Other reported accidents have concerned misinterpretation of a calibration certificate and an atmospheric pressure correction factor; incorrect use of a plane-parallel ionization chamber; inadequate briefing of a replacement medical physicist.

Examples – Beam Calibration

Contributing factors • Lack of understanding of beam calibration procedures,

calibration certificates, conversion factors and dosimetry instruments.

Page 12: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 12

• Lack of training and expertise in radiotherapy physics;

Examples – Beam Calibration (cont)

Contributing factors (cont)

• Lack of redundant and independent determination of absorbed dose (mistakes were not detected);

• Lack of formal procedures for communication and information exchange with new personnel.

Page 13: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 13

In one case verification of a 60Co teletherapy beam had not been undertaken for 22 months.

• The physicist responsible was fully engaged with a new linear accelerator and “ignored” the 60Co unit.

Examples – Beam Calibration (cont)

Contributing factors

• Management did not review staffing needs when the new accelerator was installed.

Page 14: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 14

Incorrect adjustment of the electron energy of a linear accelerator resulted in 27 patients being over exposed. Doses were between 3 and 10 times greater than intended, causing several deaths.

• A repair was made by a maintenance technician who did not understand the accelerator circuitry nor the consequences of incorrectly adjusting the electron energy (i.e. insufficient or incorrect training and experience).

• The repair was not communicated to the medical physicist and patient treatments resumed without beam verification.

Examples – Maintenance problems

Contributing factors

Page 15: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 15

• It was possible to operate the machine with the energy selector disabled.

• Conflicting displays and signals on the control panel were misinterpreted (the energy selector indicated a different energy to the instrument on the control panel).

Examples – Maintenance problems (cont)

Contributing factors (cont)

Page 16: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 16

Repeated, intermittent and unresolved equipment faults were interrupting treatments. The radiation oncologist took a decision to operate the accelerator in “physics” mode which bypassed safety interlocks.

Operation of a linear accelerator in “physics” mode for treatments resulted in one death.

Examples – Maintenance problems (cont)

Contributing factors

Page 17: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 17

Accident relating to the commissioning of the Treatment Planning System (TPS) leading to wrong doses to many patients.

Examples – Treatment planning

• A distance correction was applied manually but the TPS already had the data in the computer plan. This error resulted in underexposure of over 1000 patients by as much as 30%.

Contributing factors

• Insufficient understanding of the TPS (no staff training on the new equipment for treatment planning);

• Other accidents were related to entering incorrect data into the TPS computer.

Page 18: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 18

Examples – Treatment planning (cont)

Contributing factors (cont)

• Lack of formal commissioning of the TPS (no test before using the TPS for treating patients).

• Lack of an independent check of the dose planning (either by manual calculations to selected points or by measurement on a phantom, or by “in vivo” dosimetry);

Page 19: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 19

Anatomical error with identification of the patient’s left and right sides interchanged.

Examples – Treatment Simulation

• the treatment was simulated in an unusual position;

Contributing factors

• there was no check of the anatomical site relative to the check film.

Page 20: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 20

Examples – Treatment set up and delivery

Treatment of the wrong patient, wrong site, giving the wrong treatment, applying the wrong treatment schedule.

Contributing factors

• Lack of procedures for proper identification of the patient, the anatomical site or the treatment schedule;

• procedures not followed rigorously.

Page 21: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 21

Brachytherapy accidents may be related to:

Causes of accidents

• Equipment design;• Source order and delivery;• Source calibration and acceptance;• Source preparation for the treatment;• Treatment planning and dose

calculation;• Source removal.

Page 22: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 22

Sources were dislodged from the drive mechanism of a high dose rate (HDR) remote after loading device and remained in the patient. The patient died from the radiation dose.

Examples – Equipment Design and testing

• The HDR device apparently was not adequately tested for conditions such as a source train jammed in the catheter or a kinked catheter.

Contributing factors

Page 23: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 23

• Conflicting signals - the HDR device indicator was showing “source shielded” while an area monitor indicated the presence of radiation. The wrong signal was accepted.

Examples – Equipment Design and testing (cont)

Contributing factors (cont)

• The external radiation monitor had a history of malfunctions and was not trusted.

• A portable survey instrument was not used to confirm the return of the sources to the HDR device after the treatment.

Page 24: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 24

Different units of activity were used by the hospital (mCi) and the manufacturer (mg-Ra-equivalent). This led to a patient receiving a radiation dose 74% greater than prescribed.

In three cases, patients received less than the prescribed dose due to the use of sources without a check on the source activity.

Examples – Source orders, delivery, etc

Contributing factors

• Different activity units were used by the hospital and manufacturer;

• Insufficient check of documents (order and delivery);• A source calibration check was not undertaken before use (source

strength determination).

Page 25: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 25

Errors in treatment time calculations ranged from - 59% to + 49%.

Examples – Treatment planning

Contributing factors

• Copies of an obsolete form were still available for clinical use;

• Records did not include the treatment distance relevant to the treatment time calculations;

• Miscommunication between the radiation oncologist, physicist and others (a treatment plan was modified but an unmodified plan was used);

• Lack of independent verification of treatment time calculations.

Page 26: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 26

• Wrong sources were used resulting, in one instance, of treatment with less than 50% of the prescribed dose;

Examples – Source preparation

• In one case the manufacturer delivered a source with essentially no activity;

• Two 192Ir sources were lost;

• A leaking 125I source was re-used;

• Sources withdrawn from clinical use were used with incompatible applicators.

Page 27: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 27

Contributing factors

Examples – Source preparation (cont)

• Personnel handling sources and applicators lacked proper training;

• Source activity was not verified;

• Failure to properly audit the inventory of radiation sources;

• Source leakage was not detected during preparation. A similar incident in another hospital had not triggered a check action;

• Sources withdrawn from clinical use were re-used by mistake.

Page 28: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 28

• a resident physician did not implant all prescribed sources;

Examples – Delivery of treatment

• a wrong patient was treated;

• a source ribbon was dislodged from the catheter and was taped by a nurse on the face of the patient;

• sources became displaced;

• sources were removed by the patient;

• a source that did not match the applicator was loose and fell out of the applicator.

Page 29: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 29

Contributing factors

Examples – Delivery of treatment (cont)

• untrained physician working without supervision;

• nurses without appropriate training were caring for brachytherapy patients;

• no written safety procedures;

• poor communication of instructions that were not understood.

Page 30: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 30

Examples – Source removal

• Sources lost due to lack of appropriate checks after removal.

• Sources were not accounted for after removal;

Contributing factors

• The patient, their clothes, the room and waste from the treatment room were not monitored.

Page 31: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 31

Studies have shown that most accidents could have been prevented by consistent application of the Requirements 34 – 42 (3.144 – 3.184) of the GSR Part 3.

Lessons learned

Page 32: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 32

Accidents in radiotherapy

Human error resulting in a therapeutic treatment delivered:

Equipment malfunction

• to the wrong treatment site

• with a substantially different dose, or dose fraction, to that prescribed by the medical practitioner

• to the wrong patient

Page 33: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 33

The Panama accident

This accident, investigated by the IAEA, occurred at a facility that provided radiotherapy for cancer patients.

• In keeping with common practice, the facility used blocks of shielding material to modify the shape of the radiation beams to protect normal tissue, including critical structures, during treatment.

Page 34: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 34

The Panama accident (cont)

• A computerized treatment planning system (TPS) was used to calculate the dose distribution and determine treatment times.

• The TPS allows a maximum of four shielding blocks per field to be taken into account when calculating treatment times and dose distributions.

• The data for each shielding block should have been entered into the TPS separately.

Page 35: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 35

The Panama accident (cont)

• In August 2000, to satisfy the request of a radiation oncologist to include five blocks in the field, the method for digitizing the shielding blocks was changed.

• It was found that it was possible to enter data into the TPS for multiple shielding blocks together, as if they were a single block, thereby apparently overcoming the limitation of four blocks per field.

Page 36: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 36

The Panama accident (cont)

• Although the TPS accepted entry of the data for multiple shielding blocks as if they were a single block, the computer output indicated a treatment time substantially longer than it should have been.

• The result was that patients received a proportionately higher dose than that prescribed.

• The modified treatment protocol was used for 28 patients, treated between August 2000 and March 2001 for prostate cancer and cancer of the cervix.

Page 37: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 37

The Panama accident (cont)

• In November 2000, a radiation oncologist started to observe unusually prolonged diarrhoea in some patients. In December 2000, the effect was observed in other patients.

• The Department’s physicists reviewed the treatment plans but did not find any anomaly.

• The computer outputs (dose distribution and treatment time) calculations were not checked.

Page 38: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA

The outcome at 30 May 2000

• 8 of the 28 patients dead.

The Panama accident (cont)

5 deaths - radiation related.

2 deaths - unknown if radiation related.

1 death due to metastatic cancer.

• 20 patients surviving.

Page 39: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 39

The Panama accident (cont)

• “Additional radiation effects will become apparent in the affected patients over the next months and years and, given the radiation doses received, the morbidity and mortality can be expected to increase.”

Investigation of an Accidental Exposure of Radiotherapy Patients in PanamaIAEA Vienna 2001

• “Most of the surviving patients already have serious medical problems related mainly to bowel and bladder overexposure. Most of the untoward bowel and bladder effects cannot be remedied.”

Page 40: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 40

Compliance with legislationThe radiotherapy department of the facility applied for authorization in 1997 but this had not been granted by the time of the accidental exposure, pending the resolution of several outstanding issues.

The Panama accident (cont)

Records showed that the Regulatory Body had issued a number of reminders to the department since 1997:

• requesting information, principally the manuals of procedures for radiation protection and quality assurance.

Page 41: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 41

Compliance with legislation (cont)

The Panama accident (cont)

• A number of brachytherapy sources from the facility were found to be in use in a different hospital without authorization;

• An incident had occurred in which a cobalt therapy source had not returned to the ‘OFF’ (shielded) position, risking excessive radiation exposure of staff and patients. No radiation oncologist was present at the hospital at the time.

The Regulatory Body sent a letter to the facility reminding it of its obligation to have at least one radiation oncologist always present when patients are being treated.

Page 42: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 42

The event was triggered by:

• the search for a way to overcome the limitation to four shielding blocks.

The Panama accident (cont)

• There was no warning on the computer screen when the blocks were digitized in an unacceptable way i.e. in a way different to that prescribed in the manual;

Contributory factors

• Several different ways of digitizing the shielding blocks were accepted by the computer;

Page 43: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 43

• When blocks were digitized incorrectly, the TPS produced a diagram which was the same as that produced when data were entered correctly, thereby giving the impression that the calculated results were correct.

The Panama accident (cont)

• The modified protocol was used without a verification test i.e. without a manual calculation of the treatment time for comparison with the computer calculated treatment time, or a simulation of treatment by irradiating a water phantom and measuring the dose delivered.

• Although the treatment times were about twice those required for correct treatment, the error went unnoticed.

Page 44: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 44

Not an isolated event

• More than 90 cases of accidental exposures in radiotherapy have been documented.

• Accidents have occurred in both brachytherapy and external beam radiotherapy.

• Accidents occur in both developed and developing countries

Page 45: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 45

Consequences in practice

• is too low - reduction of tumor control probability. There is no second chance!

If the dose:

There is less likelihood that an accident will occur where a country has a functioning regulatory system. However, these countries may be over-represented in accident summaries possibly because all (?) accidents are actually reported.

• is too high – risk of acute complications and late complications.

Page 46: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 46

Lessons learned

“Registrants and licensees shall promptly investigate any of the following unintended or accidental medical exposures:

[GSR Part 3 Requirement 41. 3.179]

(a) any therapeutic treatment delivered to either the wrong patient or the wrong tissue, or using the wrong pharmaceutical, or with a dose or dose fractionation differing substantially from the values prescribed by the medical practitioner or which may lead to undue acute secondary effects;”

Page 47: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 47

• Treatment planning is a critical aspect of treatments.

Lessons learned from the Panama accident

• Written procedures are required.

• New procedures must be tested - including documentation.

• Computer calculations should be verified manually.

• Treatment planning software should include:

an instruction manual;

on-screen warnings;

fool-proof tests.

Page 48: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 48

• Ongoing communication with the treatment device manufacturer is important for both service and staff training.

Lessons learned from the Panama accident (cont)

• Radiotherapy facilities must be appropriately resourced (staffing, equipment etc) in keeping with the patient workload.

• The licensee, responsible medical practitioner, RPO and senior staff must encourage a strong safety culture in all personnel through appropriate supervision and interaction.

• Directions from the Regulatory Body must be complied with promptly.

Page 49: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 49

Investigation of accidents

• Minor accidents should be investigated by the licensee and the results, along with the corrective actions, reported to the Regulatory Body.

• The Regulatory Body should prescribe the types of accidents and incidents that must be reported.

• For more serious, or potentially serious accidents and emergencies, an independent investigation by the Regulatory Body (and sometimes also by other government authorities) should be conducted in addition to the investigation conducted by the user.

Page 50: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 50

• Determine the root causes and contributing factors.

Investigation of accidents (cont)

• Assess the consequences in terms of exposure and the likelihood of exposure of:

exposure of patient, staff (or public)

• Identify corrective actions.

• Derive the lessons to be learned and the significance of the findings for other radiotherapy facilities.

Page 51: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 51

What action should the Regulatory Body take in response to a lost radioactive source?

If a therapy radioactive source is no longer on-site and no longer under the licensee’s control (through poor inventory

control, improper transfer or disposal, theft etc) and that loss is confirmed, the Regulatory Body should immediately inform government and initiate the appropriate emergency

response plan.

Consider the need for wider national and international notification

Page 52: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 52

What action should the Regulatory Body take in response to a reported accident or incident?

Determine if an inspection is necessary to obtain first-hand information of the

cause of the accident or incident.

The licensee’s report should be reviewed (in collaboration with other authorities and government

agencies, if appropriate to the circumstances)

Page 53: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 53

Determine how similar accidents or incidents can be best avoided in future and recommend corrective measures.

Accidents and

Incidents

What action should the Regulatory Body take in response to a reported accident or incident?

Determine what enforcement action, if any,

may be appropriate

Page 54: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 54

Particularly in cases where injury (or the perceived risk of injury) may lead to litigation, determine the potential health

outcomes, responsibility and liability (under the radiation safety legislation)

for the accident or incident.

What action should the Regulatory Body take in response to a reported accident or incident?

Page 55: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 55

Disseminate the findings of the investigation to all relevant clients, organizations and professional associations.

What action should the Regulatory Body take in response to a reported accident or incident?

Page 56: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 56

References

• IAEA Human Health Series No. 14 Planning National Radiotherapy Services. IAEA, Vienna 2010

• Setting Up a Radiotherapy Programme: Clinical, Medical Physics, Radiation Protection and Safety Aspects. IAEA, Vienna 2008

• IAEA-TECDOC-1040 Design and implementation of a radiotherapy programme: Clinical, medical physics, radiation protection and safety aspects

Page 57: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 57

• Medical Handling of Accidentally Exposed Individuals. IAEA Safety Series No. 88

• Event No. 8: “Calibration error after a source change in a 60Co unit.”

• Event No. 20: “Design error in accelerator control software.”

• Event No. 70: “Inadequate implantation and loss of sources.”

References (cont)

Page 58: IAEA International Atomic Energy Agency Accidents in Radiotherapy Radiation Sources in Radiotherapy Day 7 – Lecture 6

IAEA 58

• Radiation Protection and Safety of Radiation Sources: International Basic Safety Standards. Generic Safety Requirements. GSR Part 3 (Interim) Vienna (2011)

• IAEA Safety Guide: Radiological Protection for Medical Exposure to Ionizing Radiation (Safety Standards Series No. RS-G-1.5, 2002)

References (cont)