goldmann applanation tonometry in patients with regular corneal astigmatism

4
GOLDMANN APPLANATION TONOMETRY IN PATIENTS WITH REGULAR CORNEAL ASTIGMATISM JACK T. HOLLADAY, M.D., MARK E. ALLISON, M.D., AND THOMAS C. PRAGER, PH.D. Houston, Texas By averaging vertical and horizontal applanation readings from the Goldmann tonometer, we developed a simple technique that accurately measures intraocular pressure in patients with regular corneal astigma- tism. This procedure eliminates the need for keratometry and oblique alignment of the applanator. When the applanator is oriented normally, with the mires displaced horizontally, intraocular pressure is underestimated for with-the-rule corneal astigmatism and overestimated for against-the-rule corneal astigmatism. The error is approximately 1 mm Hg for every 4 diopters of astigmatism. For oblique axis corneal astigmatism, the error is smaller and approaches zero when the axis is either 45 or 135 degrees. The Goldmann applanation tonometer has served as the standard for intraocular pressure measurements since 1955. x A circular applanation area of 7.36 mm 2 , which Goldmann determined was opti- mal to neutralize the corneal structural resistance, creates capillary attraction during applanation, by surface tension of the corneal tear film. This results in an accurate measurement of intraocular pressure independent of corneal rigidity and capillary attraction. 2,3 When regular corneal astigmatism is present, the applanated area is elliptical, not circular. To eliminate error resulting from the elliptical applanation area, the applanator must be oriented at approxi- mately 43 degrees from the flattest corne- al meridian. 4 We have developed a second method Accepted for publication April 11, 1983. From the University of Texas Medical School, Hermann Eye Center, Department of Ophthalmolo- gy, Houston, Texas. Reprint requests to Jack T. Holladay, M.D., Hermann Eye Center, P.O. Box 20420, Houston, TX 77030. that does not require determination of the flattest corneal meridian, is as accu- rate as Goldmann's procedure, and is technically easier to perform. Our tech- nique requires measuring the intraocular pressure with the applanator oriented horizontally, then vertically, and deter- mining the average value. SUBJECTS AND METHODS We selected a consecutive series of 11 patients (12 eyes) with regular corneal astigmatism of more than 3 diopters. Each patient underwent keratometric measurements once and intraocular pres- sure measurements four times. The first measurement was done in the standard manner with the applanation mires dis- placed horizontally (180 degrees). We then measured the intraocular pressure three more times: first, after rotating the applanation mires vertically (90 degrees); second, after rotating the applanation mires to 43 degrees from the flattest corneal meridian (Goldmann's proce- dure); and, finally, as a control, repeat- ing the initial measurement with the applanator oriented horizontally to assure 90 © AMERICAN JOURNAL OF OPHTHALMOLOGY 96:90-93, 1983

Upload: thomas-c

Post on 16-Feb-2017

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Goldmann Applanation Tonometry in Patients with Regular Corneal Astigmatism

GOLDMANN APPLANATION TONOMETRY IN PATIENTS W I T H REGULAR CORNEAL ASTIGMATISM

JACK T. HOLLADAY, M.D. , MARK E. A L L ISO N , M.D. , AND THOMAS C. PRAGER, P H . D .

Houston, Texas

By averaging vertical and horizontal applanation readings from the Goldmann tonometer, we developed a simple technique that accurately measures intraocular pressure in patients with regular corneal astigma­tism. This procedure eliminates the need for keratometry and oblique alignment of the applanator.

When the applanator is oriented normally, with the mires displaced horizontally, intraocular pressure is underestimated for with-the-rule corneal astigmatism and overestimated for against-the-rule corneal astigmatism. The error is approximately 1 mm Hg for every 4 diopters of astigmatism. For oblique axis corneal astigmatism, the error is smaller and approaches zero when the axis is either 45 or 135 degrees.

The Goldmann applanation tonometer has served as the standard for intraocular pressure measurements since 1955.x A circular applanation area of 7.36 mm2, which Goldmann determined was opti­mal to neutralize the corneal structural resistance, creates capillary attraction during applanation, by surface tension of the corneal tear film. This results in an accurate measurement of intraocular pressure independent of corneal rigidity and capillary attraction.2,3

When regular corneal astigmatism is present, the applanated area is elliptical, not circular. To eliminate error resulting from the elliptical applanation area, the applanator must be oriented at approxi­mately 43 degrees from the flattest corne­al meridian.4

We have developed a second method

Accepted for publication April 11, 1983. From the University of Texas Medical School,

Hermann Eye Center, Department of Ophthalmolo­gy, Houston, Texas.

Reprint requests to Jack T. Holladay, M.D., Hermann Eye Center, P.O. Box 20420, Houston, TX 77030.

that does not require determination of the flattest corneal meridian, is as accu­rate as Goldmann's procedure, and is technically easier to perform. Our tech­nique requires measuring the intraocular pressure with the applanator oriented horizontally, then vertically, and deter­mining the average value.

SUBJECTS AND M E T H O D S

We selected a consecutive series of 11 patients (12 eyes) with regular corneal astigmatism of more than 3 diopters. Each patient underwent keratometric measurements once and intraocular pres­sure measurements four times. The first measurement was done in the standard manner with the applanation mires dis­placed horizontally (180 degrees). We then measured the intraocular pressure three more times: first, after rotating the applanation mires vertically (90 degrees); second, after rotating the applanation mires to 43 degrees from the flattest corneal meridian (Goldmann's proce­dure); and, finally, as a control, repeat­ing the initial measurement with the applanator oriented horizontally to assure

90 © AMERICAN JOURNAL OF OPHTHALMOLOGY 96:90-93, 1983

Page 2: Goldmann Applanation Tonometry in Patients with Regular Corneal Astigmatism

VOL. 96, NO. 1 APPLANATION TONOMETRY 91

that none of the measurements were arti­ficially low as a result of repeated appla-nation.5 If the first and last intraocular pressure measurements were not identi­cal, we repeated all four measurements until this criterion was satisfied. Meas­urements were done independently by two observers and the results were statis­tically evaluated by analysis of variance and Pearson product-moment correla­tion.

R E S U L T S

A one-way analysis of variance per­formed after collapsing data across both sets of observed data showed no signifi­cant difference (P>.05) between Gold-mann's method of measuring intraoc­ular pressure and our averaging tech­nique (Goldmann: X = 13.11 mm Hg, S.D. = 2.36; averaging technique: X = 13.08 mm Hg, S.D. = 2.59). A Pearson product-moment correlation between the two methods was almost linear, r = +0.9936 (Table).

Our data showed that an error of 1 mm Hg for every 4 diopters of with- or against-the-rule astigmatism was present when we used only the normal, horizon­tal measurement. For oblique axis corne-al astigmatism, the difference was less than a ratio of 1:4 and approached zero as the axis neared 45 or 135 degrees.

D I S C U S S I O N

Goldmann and Schmidt2,3 were aware of the applanation error induced by regu­lar astigmatism when the mires were not aligned 43 degrees from the flattest cor-neal meridian. To our knowledge, the specific amount of the applanation error has never been reported, because for astigmatism of more than 3 diopters Goldmann's method is used.

Goldmann's determination of 43 de­grees as the proper applanator alignment is an average value of an angle that varies with the amount of astigmatism.4 Gold­

mann's intraocular pressure error is roughly 0.3% for 5 diopters, 1.8% for 10 diopters, and 2 .1% for 20 diopters of regular corneal astigmatism. In compari­son, our averaging technique theoretical­ly underestimates the intraocular pres­sure by 0.0%, 0.5%, and 0.7%, respectively. This underestimation is a result of the difference between the true geometric mean and our arithmetic mean. Although these differences are of theoretic interest, they are not significant clinically.

The horizontal-vertical averaging tech­nique offers three advantages that Gold­mann's original method does not: (1) Ker-atometric measurements are not neces­sary. (2) Goldmann's method of aligning the applanator at 43 degrees from the flattest corneal meridian usually requires oblique orientation of the mires. It is difficult to perform this technique, which requires simultaneous vertical and hori­zontal movement of the slit lamp. Using the averaging technique obviates the need for oblique alignment. (3) The verti­cal and horizontal measurements deter­mine the precise applanation intraocular pressure for that patient's cornea, and thus determine the error of the routine horizontal applanation.

The averaging technique is clinically useful after cataract surgery, because a significant incidence of with-the-rule astigmatism results from the use of 9-0 and 10-0 nonabsorbable sutures that com­press the corneoscleral wound.6,7 Com­pressing sutures, combined with wound swelling, often produce 3 to 10 diopters of corneal astigmatism during the early postoperative period.6 Measuring intra­ocular pressure with applanation mires in the normal, horizontal orientation leads to an underestimation of approximately 1 mm Hg for every 4 diopters of corneal astigmatism. For example, with +12 di­opters of with-the-rule astigmatism, in­traocular pressure is underestimated by 3

Page 3: Goldmann Applanation Tonometry in Patients with Regular Corneal Astigmatism

CD

to

TAB

LE

APP

LA

NA

TIO

N I

NT

RA

OC

UL

AR

PR

ESS

UR

ES

IN P

AT

IEN

TS

WIT

H R

EG

UL

AR

CO

RN

EA

L A

STIG

MA

TIS

M

Eye

No.

1 2 3 4 5 6 7 8 9 10

11

12

Ker

atom

etric

Rea

ding

s (D

iopt

ers)

Flat

test

40.0

0 at

170

° 41

.62

at

41.7

5 at

170

° 41

.12

at

42.5

0 at

30

° 41

.12

at

41.5

0 at

17

° 41

.12

at 1

80°

40.6

2 at

41

.00

at

41.2

5 at

42

.00

at 1

80°

Stee

pest

50.2

5 at

80

° 48

.12

at

90°

47.7

5 at

80

° 48

.50

at

92°

47.2

5 at

118

° 46

.00

at

95°

49.3

7 at

107

° 46

.25

at

85°

47.5

0 at

98

° 47

.50

at

90°

49.0

0 at

92

° 47

.00

at

90°

Asti

gmat

ism

10.2

5 6.

50

5.00

6.

38

4.75

4.

88

7.88

5.

13

6.88

7.

50

8.75

5.

00

Hor

izon

tal

15.0

11

.0

12.0

9.

5 12

.0

11.5

14

.5

15.0

8.

0 12

.0

12.5

12

.5

Obs

erve

r 1

(mm

Hg)

Ver

tical

20.0

13

.0

15.0

11

.5

13.0

14

.0

17.0

19

.0

10.5

15

.0

15.5

14

.5

Ave

rage

17.5

12

.0

13.5

10

.5

12.5

12

.8

15.8

17

.0

9.8

13.5

14

.0

13.5

43

Deg

rees

*

17.0

12

.0

13.0

10

.5

12.5

13

.0

15.5

17

.0

9.0

14.0

14

.5

13.5

Diff

eren

ce

+0.5

0

+0.5

0 0

-0.2

+0

.3

0 +0

.8

-0.5

-0

.5

0

Hor

izon

tal

12.0

7.

5 9.

0 9.

0 11

.0

10.5

12

.0

13.0

8.

0 11

.0

15.0

15

.5

Obs

erve

r 2

(mm

Hg)

Ver

tical

18.0

10

.0

11.0

11

.0

13.0

13

.5

16.0

18

.0

11.0

14

.0

19.0

18

.0

Ave

rage

15.0

8.

8 10

.0

10.0

12

.0

12.0

14

.0

15.0

9.

5 12

.5

17.0

16

.8

43

Deg

rees

*

15.0

9.

0 9.

5 10

.0

12.0

11

.5

14.0

15

.5

10.0

12

.5

17.0

16

.5

Diff

eren

ce

0 -0

.2

+0.5

0 0

+0.5

0

-0.5

-0

.5

0 0 +0

.3

*Gol

dman

n's

met

hod

of ta

king

the

mea

sure

men

t at

43

degr

ees

to t

he f

latte

st c

orne

al m

erid

ian.

D

iffe

renc

es i

n in

trao

cula

r pr

essu

re b

etw

een

the

two

tech

niqu

es.

Page 4: Goldmann Applanation Tonometry in Patients with Regular Corneal Astigmatism

VOL. 96, NO. 1 APPLANATION TONOMETRY 93

mm Hg. Thus, using the averaging tech­nique after surgery prevents underesti­mates of the intraocular pressure without keratometry and oblique alignment of the applanator.

The averaging technique may be clini­cally important in measuring the intraoc­ular pressure of an older patient. Older individuals have an increased incidence of against-the-rule astigmatism,6 result­ing in an overestimation of the intraocular pressure if the. applanation mires are ori­entated horizontally. Unnecessary treat­ment may be an unfortunate consequence of this error.

We have found that using this simple averaging technique prevents significant applanation intraocular pressure errors in patients with regular corneal astigma­tism. We suggest that whenever the ap­planation mires do not appear to be circu­

lar they should be rotated to the vertical orientation and a second reading taken.

REFERENCES 1. Goldmann, H.: Applanation tonometry. In

Newell, F. W. (ed.): Glaucoma. Transactions of the Second Conference, December 3, 4, and 5, 1956. New York, Josiah Macy Foundation, 1957, pp. 167-220.

2. Goldmann, H., and Schmidt, T.: Über Applanationstonometrie. Ophthalmologica 134:221, 1957.

3. : Weiterer beitrag zur Applanations­tonometrie. Ophthalmologica 141:441, 1961.

4. Schmidt, T.: Zur Applanationstonometrie an der Spaltlampe. Ophthalmologica 133:337, 1957.

5. Moses, R. A.: Repeated applanation tonome­try. Ophthalmologica 142:663, 1961.

6. Jaffe, N. S., and Clayman, H. M.: The patho-physiology of corneal astigmatism after cataract ex­traction. Trans. Am. Acad. Ophthalmol. Otolaryngol. 79:615, 1975.

7. Emery, J. M., and Paton, D. (eds.): Current Concepts in Cataract Surgery. Selected Proceedings of the Fourth Biennial Cataract Surgical Congress, 1975. St. Louis, C. V. Mosby, 1976, pp. 179-193.