gilson guide to pipetting - gilsoncn.comgilson guide to pipetting third edition introduction. ......

4
More information available at www.gilson.com 1 Gilson Guide to Pipetting ird edition

Upload: others

Post on 09-Mar-2020

22 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Gilson Guide to Pipetting - gilsoncn.comGilson Guide to Pipetting Third edition Introduction. ... The Pasteur pipette is still in use today. The next significant improvement on pipettes

More information available at www.gilson.com 1

Gilson Guide to PipettingThird edition

Introduction

Page 2: Gilson Guide to Pipetting - gilsoncn.comGilson Guide to Pipetting Third edition Introduction. ... The Pasteur pipette is still in use today. The next significant improvement on pipettes

More information available at www.gilson.comGuide to Pipetting

How the pipette story began…

Over a century ago, Louis Pasteur invented the glass Pasteur pipette to reduce contamination when transferring samples. The Pasteur pipette is still in use today.

The next significant improvement on pipettes occurred in the late 1950s with the introduction of a handheld, piston-operated pipette as a safe alternative to potentially dangerous mouth pipetting. The first handheld pipettes had pre-established volume setting (fixed volume pipettes). Further improvement was provided with the more flexible stepper volume setting (variable volume pipettes).

In 1972, Dr. Warren Gilson invented the first continuously adjustable pipette. As an original error-preventing feature, the selected volume was now clearly displayed on the Gilson pipette (direct digital readout). Today, Gilson precision pipettes are still the world standard for accuracy, precision, and reliability.

Today, the company still maintains its continuous effort of creativity and offers innovative, robust, and reliable pipettes to help scientists in their daily work.

The pipetting system is our core expertise, and we truly enjoy sharing this knowledge and experience with pipette users so that they may achieve their goals.

This guide will provide you with keys to understand how pipettes should be used and to get the most out of them. Examples given are based on our own pipette range, but the techniques described are equally applicable to other brands of pipettes and tips.

Gilson is one of the winners of the Red Dot Award: Product Design for MICROMAN E.

Page 3: Gilson Guide to Pipetting - gilsoncn.comGilson Guide to Pipetting Third edition Introduction. ... The Pasteur pipette is still in use today. The next significant improvement on pipettes

Guide to Pipetting More information available at www.gilson.com

table

of co

nten

tChapter 1 - Selecting the right pipette

1.1 Working principle of pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

1.2 The right choice for your application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

1.3 Specific pipettes for specific vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14

1.4 High throughput & repetitive pipetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15

Chapter 2 - Pipetting techniques2.1 Adjust the volume display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17

2.2 Forward or reverse mode pipetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18

2.3 Tips for mistake-free pipetting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

2.4 Pipetting ergonomics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

Chapter 3 - Selecting the right tip3.1 Fitting a disposable pipette tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27

3.2 Choosing the best tip for your application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29

3.3 Evaluating tip quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31

Chapter 4 - Preventing contamination4.1 Types of contamination and how to prevent them . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

4.2 Decontaminating your pipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

Chapter 5 - Pipette service & maintenance

5.1 Pipette specifications according to ISO 8655 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

5.2 Repair in the lab or return for service? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

5.3 Quick pipette diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

5.4 How to calculate volumetric accuracy and precision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

5.5 Pipette calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40

5.6 Calibration with the gravimetric method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

5.7 Performance check procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44

Appendix

Pipetting terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46

Example of a performance check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48

Z factor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49

Evaporation loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50

Chemical resistance of plastics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

FAQ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52

Page 4: Gilson Guide to Pipetting - gilsoncn.comGilson Guide to Pipetting Third edition Introduction. ... The Pasteur pipette is still in use today. The next significant improvement on pipettes

Guide to Pipetting More information available at www.gilson.com1 2

Positive-displacement pipetteAir-displacement pipette

150

150

T051401

• Push-button

• Volume adjustment knob

• Ejector of capillary and piston

Volumeter display

Shaft(also known as

capillary holder)

Disposablecapillary and piston

Serial numberSerial number

• Push-button

• Large volume adjustment knob

Tip ejector button

Tip ejector

Thumbwheel(also known as fine

volume adjustment ring)

Shaft(also known as tip holder)

Disposable tip

Volumeter display

Sealing ringIdentification

mark

Filter

Levelmark

150

150

T051401

GLOS

SARY

Pipettes with different designs are available. For more information, visit www.gilson.com