gc and gc-ms. gas chromatography function components common uses chromatographic resolution...

45
GC and GC-MS

Upload: carlton-babcock

Post on 01-Apr-2015

247 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

GC and GC-MS

Page 2: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Gas Chromatography

• Function

• Components

• Common uses

• Chromatographic resolution

• Sensitivity

Page 3: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Function

• Separation of volatile organic compounds• Volatile – when heated, VOCs undergo a

phase transition into intact gas-phase species

• Separation occurs as a result of unique equilibria established between the solutes and the stationary phase (the GC column)

• An inert carrier gas carries the solutes through the column

Page 4: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Components

• Carrier Gas, N2 or He, 1-2 mL/min

• Injector

• Oven

• Column

• Detector

Page 5: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Gas tank

Oven

Column

Injector

Syringe

Detector

Page 6: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Injector

• A GC syringe penetrates a septum to inject sample into the vaporization camber

• Instant vaporization of the sample, 280 C• Carrier gas transports the sample into the

head of the column

• Purge valve controls the fraction of sample that enters the column

Page 7: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Splitless (100:90) vs. Split (100:1)

Injector

Syringe

Injector

Syringe

Purge valveopen

Purge valveclosed

GC column GC column

HeHe

Page 8: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Split or splitless

• Usually operated in split mode unless sample limited

• Chromatographic resolution depends upon the width of the sample plug

• In splitless mode the purge valve is close for 30-60 s, which means the sample plug is 30-60 seconds

• As we will see, refocusing to a more narrow sample plug is possible with temperature programming

Page 9: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

0.32 mm ID

Liquid Stationary phase

Mobile phase (Helium) flowing at 1 mL/min

Open Tubular Capillary Column

15-60 m in length

0.1-5 m

Page 10: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

FSOT columns

• Coated with polymer, crosslinked– Polydimethyl soloxane (non-polar)– Poly(phenylmethyldimethyl) siloxane (10%

phenyl)– Poly(phenylmethyl) siloxane (50% phenyl)– Polyethylene glycol (polar)– Poly(dicyanoallyldimethyl) siloxane– Ploy(trifluoropropyldimethyl) siloxane

Page 11: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Polar vs. nonpolar

• Separation is based on the vapor pressure and polarity of the components.

• Within a homologous series (alkanes, alcohol, olefins, fatty acids) retention time increases with chain length (or molecular weight)

• Polar columns retain polar compounds to a greater extent than non-polar– C18 saturated vs. C18 saturated methyl ester

Page 12: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

C16:0

C18:0

C18:1C18:2

C16:1

C16:0

C18:0

C18:1

C18:2

C16:1

RT (min)

RT (min)

Polar column

Non-polar column

Page 13: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Oven• Programmable• Isothermal- run at one constant

temperature• Temperature programming - Start at low

temperature and gradually ramp to higher temperature– More constant peak width– Better sensitivity for components that are

retained longer– Much better chromatographic resolution– Peak refocusing at head of column

Page 14: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Typical Temperature Program

Time (min)0 60

50C

220C

160C

Page 15: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Detectors

• Flame Ionization Detectors (FID)

• Electron Capture Detectors (ECD)

• Electron impact/chemical ionization (EI/CI) Mass spectrometry

Page 16: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

FIDs

• Effluent exits column and enters an air/hydrogen flame

• The gas-phase solute is pyrolized to form electrons and ions

• All carbon species are reduced to CH2+

ions• These ions collected at an electrode held

above the flame• The current reaching the electrode is

amplified to give the signal

Page 17: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

FID

• A general detector for organic compounds

• Very sensitive (10-13 g/s)

• Linear response (107)

• Rugged

• Disadvantage: specificity

Page 18: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

ECD

• Ultra-sensitive detection of halogen-containing species

• Pesticide analysis

• Other detectors besides MS– IR– AE

Page 19: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Mass Spectrometry

Page 20: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

What kind of info can mass spec give you?

• Molecular weight

• Elemental composition (low MW with high resolution instrument)

• Structural info (hard ionization or CID)

Page 21: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

How does it work?

• Gas-phase ions are separated according to mass/charge ratio and sequentially detected

Page 22: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Parts of a Mass Spec

• Sample introduction

• Source (ion formation)

• Mass analyzer (ion sep.) - high vac

• Detector (electron multiplier tube)

Page 23: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Sample Introduction/Sources

Volatiles• Probe/electron impact (EI),Chemical ionization (CI)• GC/EI,CIInvolatiles• Direct infusion/electrospray (ESI)• HPLC/ESI• Matrix Assisted Laser Adsorption (MALDI)Elemental mass spec• Inductively coupled plasma (ICP)• Secondary Ion Mass Spectrometry (SIMS)

– surfaces

Page 24: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

EI, CI• EI (hard ionization)

– Gas-phase molecules enter source through heated probe or GC column

– 70 eV electrons bombard molecules forming M+* ions that fragment in unique reproducible way to form a collection of fragment ions

– EI spectra can be matched to library stds

• CI (soft ionization)– Higher pressure of methane leaked into the

source (mtorr)– Reagent ions transfer proton to analyte

Page 25: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

To massanalyzer

filament

70 eV e-

anoderepeller Acceleration

slits

GC column

EI SourceUnder high vacuum

Page 26: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

EI process

• M + e- M+*

f1 f2 f3f4

This is a remarkably reproducible process. M will fragment in the same pattern every time using a 70 eV electron beam

Page 27: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity
Page 28: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Ion Chromatogram of Safflower Oil

RT: 14.48 - 24.30

15 16 17 18 19 20 21 22 23 24

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Re

lative

Ab

un

da

nce

RT: 20.82AA: 3547389BP: 67

RT: 21.04AA: 665791BP: 55

RT: 16.04AA: 304398BP: 74

RT: 21.90AA: 291543BP: 28RT: 16.84

AA: 78898BP: 28

NL: 9.69E5

TIC F: {0,0} + c EI det=350.00 Full ms [ 25.00-510.00] MS ICIS evanssaf

Page 29: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity
Page 30: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

CI/ ion-molecule reaction

• 2CH4 + e- CH5+ and C2H5

+

• CH5+ + M MH+ + CH4

• The excess energy in MH+ is the difference in proton affinities between methane and M, usually not enough to give extensive fragmentation

Page 31: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

EI spectrum of phenyl acetate

Page 32: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity
Page 33: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Mass Analyzers• Low resolution

– Quadrupole– Ion trap

• High resolution– TOF time of flight– Sector instruments (magnet)

• Ultra high resolution– ICR ion cyclotron resonance

Page 34: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Resolution

• R = m/z/m/z

• Unit resolution for quad and trap

• TOF up to 15000

• FT-ICR over 30000– MALDI, Resolve 13C isotope for a protein that

weighs 30000– Resolve charge states 29 and 30 for a protein

that weighs 30000

Page 35: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

High vs low Res ESI

• Q-TOF, ICR– complete separation of the isotope peaks of a

+3 charge state peptide– Ion abundances are predictable– Interferences can be recognized and

sometimes eliminated

• Ion trap, Quad– Unit resolution

Page 36: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

MVVTLIHPIAMDDGLR594.3594.7

595.0

601.3

595.3601.0

601.7

602.0

m/z

C78H135N21O22S2+3

Q-TOF

901.4

891.7

902.3

900.6

891.2

892.6

LCQ

R = 0.88

m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Page 37: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Quadrupole Mass Ion Filter

Page 38: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Ion Trap

Page 39: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Time of Flight -TOF

Page 40: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

                                                      

Where:

•mi = mass of analyte ion

•zi = charge on analyte ion •E = extraction field •ti = time-of-flight of ion

•ls = length of the source

•ld = length of the field-free drift region •e = electronic charge (1.6022x10-19 C)

Page 41: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

TOF with reflectronhttp://www.rmjordan.com/tt1.html

Page 42: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Sector instrumentshttp://www.chem.harvard.edu/mass/tutorials/magnetmovie.html

Page 43: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

FT-ICRMS

• http://www.colorado.edu/chemistry/chem5181/MS_FT-ICR_Huffman_Abraham.pdf

Page 44: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Mass accuracy

• Mass Error = (5 ppm)(201.1001)/106 =

0.0010 amu

• 201.0991 to 201.1011 (only 1 possibility)

• Sector instruments, TOF mass analyzers

• How many possibilities with MA = 50 ppm?

with 100 ppm?

Page 45: GC and GC-MS. Gas Chromatography Function Components Common uses Chromatographic resolution Sensitivity

Exact Mass Determination

• Need Mass Spectrometer with a high mass accuracy – 5 ppm (sector or TOF)

• C9H15NO4, FM 201.1001 (mono-isotopic)

• Mass accuracy = {(Mass Error)/FM}*106

• Mass Error = (5 ppm)(201.1001)/106 =

0.0010 amu