No./DMER/Tender Notice/Equip.Pur/15-16/Corri.(06)/7M Date :- 28/08/2015 E-Tender Notice CORRIGENDUM REGARDING SPECIFICATIONS FOR 1. EQ-C-15/15-16 Central Pipe Line for Oxygen, suction & Nitrous Oxide for 250 bedded hospital with Liquid Oxygen Plant 2. EQ-C-40/15-16 Endoscopy Ultrasono (Linear and Radial Probe) 3. EQ-C-41/15-16 Video Endoscopy (Gastroscope) 4. EQ-C-46/15-16 Hemodialysis Machine 5. EQ-C-48/15-16 Central Oxygen System Vacum E Suction system with Essential Part 6. EQ-C-67/15-16 Operative Laparoscope 7. EQ-C-69/15-16 Operative Laproscopy complete with HD Camera, LED Monitor with Hand Instrument with Recording System ISO & CE Certified 1. Directorate of Medical Education & Research, Mumbai, invited e-tender in two envelope systems from the manufacturers / authorized dealers and suppliers to supply instruments/ equipments for the Government Medical and Dental Colleges. This notice is available on the e-tender site i.e http://dmer.maharashtra.etenders.in & http://maharashtra govt. in and also notified in Maharashtra Govt. Gazette supplement - II 2. All the tenderers are hereby informed that modifications made in the specifications after the Pre-Bid meeting held on 20 th August 2015 at 11.00 a.m are as follows. 01. e-tender-15/15-16/ Central Pipe Line for Oxygen, suction & Nitrous Oxide for 250 bedded hospital with Liquid Oxygen Plant (REVISED SPECIFICATION) GENERAL DESCRIPTION : 1) This is required for Operation Theatre Complexes, Surgical I.C.U. and Wards in newly constructed 250 bedded hospital. 2) QUANTITIES given along with of all items are approximate, may vary actually. To be calculated and finalised after ACTUAL SITE VISIT and and AS PER USER DEPARTMENT’S REQUIREMENT. 3) QUALITY Certification from Authentic Authority is MANDATORY where ever required. All the items should comply with ISO 7396-1 / HTM / NFPA with CE-certification (device class – II b) in accordance with the European Directive 93/42/EEC (medical devices). OXYGEN MANIFOLD SYSTEM : Sr. No. Unit & Parameters Specifications 1 Service As secondary/ reserve source, the primary source being the liquid oxygen, bulk cylinders arranged through other contracts. 2 Main manifold configuration 2 x 12 cylinders 3 Pressure rating 145 kg/sq.cm. 4 Flow rate Approx 1300 LPM 5 Format Wall-mounting straight line. 6 Top Frame High pressure copper pipes of size ½ ” I.D. x 15swg, high pressure brass fittings made of high tensile brass, non-return valves & high pressure copper tail pipes, made of high pressure copper pipes of size ¼ ” I.D. x 15swg

Upload: duongkhanh

Post on 06-Sep-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

No./DMER/Tender Notice/Equip.Pur/15-16/Corri.(06)/7M Date :- 28/08/2015

E-Tender Notice CORRIGENDUM REGARDING SPECIFICATIONS FOR

1. EQ-C-15/15-16 Central Pipe Line for Oxygen, suction & Nitrous Oxide for 250 bedded hospital with Liquid Oxygen Plant

2. EQ-C-40/15-16 Endoscopy Ultrasono (Linear and Radial Probe) 3. EQ-C-41/15-16 Video Endoscopy (Gastroscope) 4. EQ-C-46/15-16 Hemodialysis Machine 5. EQ-C-48/15-16 Central Oxygen System Vacum E Suction system

with Essential Part 6. EQ-C-67/15-16 Operative Laparoscope 7. EQ-C-69/15-16 Operative Laproscopy complete with HD Camera,

LED Monitor with Hand Instrument with Recording System ISO & CE Certified

1. Directorate of Medical Education & Research, Mumbai, invited e-tender in two envelope systems from the manufacturers / authorized dealers and suppliers to supply instruments/ equipments for the Government Medical and Dental Colleges. This notice is available on the e-tender site i.e http://dmer.maharashtra.etenders.in & http://maharashtra govt. in and also notified in Maharashtra Govt. Gazette supplement - II

2. All the tenderers are hereby informed that modifications made in the specifications after the Pre-Bid meeting held on 20th August 2015 at 11.00 a.m are as follows.

01. e-tender-15/15-16/ Central Pipe Line for Oxygen, suction & Nitrous Oxide for 250

bedded hospital with Liquid Oxygen Plant (REVISED SPECIFICATION)

GENERAL DESCRIPTION: 1) This is required for Operation Theatre Complexes, Surgical I.C.U. and Wards in newly constructed

250 bedded hospital. 2) QUANTITIES given along with of all items are approximate, may vary actually. To be calculated and

finalised after ACTUAL SITE VISIT and and AS PER USER DEPARTMENT’S REQUIREMENT. 3) QUALITY Certification from Authentic Authority is MANDATORY where ever required. All the items

should comply with ISO 7396-1 / HTM / NFPA with CE-certification (device class – II b) in accordance with the European Directive 93/42/EEC (medical devices).

OXYGEN MANIFOLD SYSTEM : Sr. No.

Unit & Parameters Specifications

1 Service As secondary/ reserve source, the primary source being the liquid oxygen, bulk cylinders arranged through other contracts.

2 Main manifold configuration

2 x 12 cylinders

3 Pressure rating 145 kg/sq.cm. 4 Flow rate Approx 1300 LPM 5 Format Wall-mounting straight line. 6 Top Frame High pressure copper pipes of size ½ ” I.D. x 15swg, high pressure

brass fittings made of high tensile brass, non-return valves & high pressure copper tail pipes, made of high pressure copper pipes of size ¼ ” I.D. x 15swg

7 Middle & Bottom frames To suit both round and flat bottom cylinders. 8 High pressure regulators For both manifold systems for reducing the cylinder pressure

suitable to the line pressure. 9 Source shut-off Valves For ease of changing and positioning, without closing the bank. 10 Filter Pore size of not exceeding 100um, between the cylinder(s) and the

first pressure regulators. 11 Fixity Cylinder to be set in place using cylinder brackets and fixing chains,

all zinc plated. 12 Painting All works to be powder coated, the color codification complying with

ISO 5359/ IS 2379. 13 Specific requirement Non-halogenated polymer materials only are permitted to be used in

the non-return valves. 14 Delivery to site In degreased condition, plugged or capped at all ends and supplied

sealed in protective polythene bags, accompanied by pre-delivery test certificate.

15 Testing Pre-delivery and post installation pressure testing to 1.5 times the working pressure.

B) OXYGEN and NITROUS OXIDE CONTROL PANEL (Imported): Sr. No

Unit & Parameters Specifications

1 Service As sole controller for supply from right bank & left bank reserve of Oxygen and nitrous oxide:(two source)

2 Model Fully automatic 3 Control Unit/ Alarm on

panel

Electronic based. Left manifold

Line pressure

Right manifold

Normal Main line Normal Standby Normal Standby Low Low Low Power on Warning

buzzer Buzzer mute

Push to Test

4 Make Imported 5 Capacity Upto 130m3/hr at 4 bar 6 Power compatibility 220V, 50Hz. 7 Operation Fully automatic. 8 Changeover solenoid valves. In accordance with ISO 7396-1 / HTM / NFPA, an

alarm signal, which is integrated into the control panel 9 Components • Control panel should provide following displays.

• display of system pressure • display of gas flow • display of currently active source • online checklist for the cylinder change • range calculation for the active source • Line pressure regulator-3nos, pressure gauges, pressure

relief mechanism 10 Bypass in case of

malfunction Bypass Valves

11 Display alarm On panel 12 Output connection for

local alarm Two sets of dry contacts NO/NC.

13 Body Sheet steel, epoxy coated 14 Delivery to site In degreased condition, supplied sealed in protective polythene bags. 15 Testing Pre-delivery and post installation. Post installation testing to be

carried out in the presence of Hospital Engineer. C) NITROUS OXIDE MANIFOLD SYSTEM : Sr. No.

Unit & Parameters Specifications

1 Service As primary source 2 Main manifold

configuration 2 x 4 cylinders

3 Pressure rating 145 kg/sq.cm.

4 Flow rate To,[specify] attach calculation 5 Format Wall-mounting straight line. 6 Top Frame High pressure copper pipes of size ½ ” I.D. x 15swg., high pressure

brass fittings made of high tensile brass, non-return valves, high pressure copper annealed tail pipes, made of high pressure copper pipes of size ¼ ” I.D. x 15swg & brass end adapters to suit N2O cylinders.

7 Middle & Bottom frames To suit both round and flat bottom cylinders.

8 High pressure regulators

For both manifold systems for reducing the cylinder pressure suitable to the line pressure.

9 Source shut-off Valves For ease of changing and positioning, without closing the bank. 10 Filter Pore size of not exceeding 100um, between the cylinder(s) and the

first pressure regulators. 11 Fixity Cylinder should be set in place using cylinder brackets and fixing

chains which should be zinc plated. 12 Painting All works to be powder coated, the color codification complying with

ISO 5359/ IS 2379. 13 Specific requirement Non-halogenated polymer materials only are permitted to be used in

the non-return valves. 14 Delivery to site In degreased condition, plugged or capped at all ends and supplied

sealed in protective polythene bags, accompanied by pre-delivery test certificate.

15 Testing Pre-delivery and post installation pressure testing to 1.5 times the working pressure.

D) VACUUM PUMP 15V: Sr.No. Unit & Parameters Specifications

1 Vacuum Pump INGERSOLL- RAND or equivalent

2 Model 15V 3 Capacity 149.60 cfm or 4235 ltrs/ min 4 Service Air cooled 5 Maximum Vacuum with Closed

Intake 29.20”of hg

6 Vacuum pump model 15V 3 nos 7 Motor:

8 Rating 7.46kw/10HP 9 10Hp Motor 3nos 10 1500 ltrs receiver 2nos 11 Y strainer (2inch) 3nos 12 Bacterial Filter along with by pass 1no 13 Electrical Control panel Fully automatic changeover control along with single

phase protector, Phase indicator, trip indicator, amp meter, Volt meter etc.

E) DISTRIBUTION FOR COPPER PIPING : Sr. No. Unit & Parameters Specifications

1 Pipe material Copper Pipes will be Solid drawn, seamless, deoxidized , non arsenical, half hard tempered and degreased, materials conforming to BS EN: 13348 Lloyd’s certified.

2 Fittings End-feed type, made from the same grade of copper as the pipes, and in accordance with the requirements of BS-EN 1254-1: 1998 Part 1.

3 Sizes 12MM OD * 0.7MM THICKNESS 15MM OD * 0.9MM THICKNESS 22MM OD * 0.9MM THICKNESS 28MM OD * 0.9MM THICKNESS 35MM OD * 1.2MM THICKNESS 42MM OD * 1.2MM THICKNESS 54MM OD * 1.2MM THICKNESS 76MM OD * 1.5MM THICKNESS 108MM OD as per required

4 Delivery to site In degreased condition, plugged or capped at both ends and Supplied sealed in protective polythene bags, accompanied by certificate from Lloyd’s.

5 Joints, on site To be brazed, except for mechanical joints used for components, employing a method that Permits the joints to maintain their mechanical characteristics up to 600 degC.

6 Brazing system Fluxless brazing using a silver-copper-phosphorous brazing alloy to BS-1845, employing an oxygen-free nitrogen inert gas shield and Cadmium-free (less than 0.025% mass fraction)

7 Pipe clamps Shall be non-reactive to copper and be of non-ferrous or non-deteriorating plastic suitable for the diameter of the pipe

8 Pipeline supports Spacing not to exceed 1.5M, Irrespective of the pipe diameter. 9 Passage through

walls & floors Through suitable sleeves.

10 Earthing To be connected to one or more earth terminals. 11 Painting All exposed pipes to be painted with two coats of synthetic enamel paint,

the color codification complying with ISO 5359/IS 2379 12 Marking and colour

coding To mark with the name and / or symbol adjacent to shut-off valves, at the junctions and the changes of direction, before and after walls and partitions, etc. at the intervals of no more that 10M and adjacent terminal units.

13 Post-installation After erection, the pipelines are to be flushed and purged clean with dry nitrogen gas.

14 Testing Pressure testing using dry nitrogen gas at 1.5 times the working pressure for at least 24 hours. System should exhibit its integrity and no-leakage. All contraindications, if any to be carefully examined, problems diagnosed and necessary examined, problems diagnosed and necessary rectifications / Substitutions carried out, and test repeated after re-purging the lines.

15 Test documentation and witnessing

All test, including failed tests, to be documented covering all the details. The Engineer shall witness the tests and receive the documents.

Providing and laying of Pipelines. 1) The successful tenderer shall be responsible for designing and laying of medical gas pipeline. The

design shall be in accordance with HTM 0102 and such that all the outlet points shall have required pressure and flow rate. Further, care shall be taken that layout of pipelines shall not disturb the existing hospital systems working.

2) The quantities mentioned in the Schedule of Quantities are approximate hence it is mandatory to quote per unit /running foot price for each item.

3) Certification of the pipes shall be attached along with tender. 4) Each Copper Pipe shall have its respective thicknesses and diameters mentioned/marked on the

pipe itself and also the lot/batch of testing. (e.g. 22mm x 0.9mm, EN13348 and batch /lot testing number).

5) Typical chemical composition of the copper shall be Copper Not less than 99.85%. Arsenic not more than – 0.05%, Phosphorous not less than 0.015% and not more than 0.080%. The successful tenderer shall submit the test certificate of chemical composition of the Copper pipes.

Installation: Pipes, fittings and valves shall be degreased and shall be made free of particulate matter and toxic residues before their use. The pipes shall be individually capped at both ends and delivered to the site. P.V.C saddles shall be used for diameters smaller than 28mm OD pipes and powder coated metal saddles shall be used for diameters larger than 35mm OD pipes. The saddles shall be properly mounted on the wall with the help of Screws, brackets or such suitable method. Joints of Pipes shall be made using inert gas flux less brazing method. Joints shall be of copper to copper and shall be brazed by silver brazing filler material without flux. During the process of brazing, the joints will be continuously purged with oil free dry nitrogen to prevent the formation of oxides on the inner surface of the pipe and fittings While laying a copper pipeline from floor to floor, all the pipes shall be encased with suitable size G.I. Pipe. On installation, pipes shall be tested with dry nitrogen at a pressure of 10 Bar for 48 hours along with testing for cross-connection. Painting of the pipes shall be carried out as per IS 2379:1963. All exposed piping shall be properly painted with two coats of Synthetic enamel Paint over a coat of primer after installation. Colour Code for Medical Gases as per IS 2379: 1963 Service Ground Colour First Band Second Band Oxygen Canary Yellow White

Nitrous Oxide Canary Yellow French Blue Vacuum Sky Blue Black Comp. Air Sky Blue Black White F) ISOLATION VALVE Sr. No.

Unit & Parameters Specifications

1 Type non-lubricated ball valve brass full bore 2 Make Indian/Imported to be quoted separately) 3 Fitting arrangement Threaded nut nipple type 4 Sizes 15mm/22mm/28mm/35mm/42mm/54mm/76mm 5 Handle Butterfly handle 6 Working pressure 50kgs cm2 7 Testing Pressure testing using dry nitrogen gas at 1.5 times the working

pressure for at least 24 hours. System should exhibit its integrity and no-leakage. All contraindications, if any to be carefully examined, problems diagnosed and necessary rectifications / Substitutions carried out, and test repeated after re-purging the lines.

8 Test documentation and witnessing

All test, including failed tests, to be documented covering all the details. The Engineer shall witness the tests and receive the documents.

G) DOUBLE LOCKING TYPE OUTLET FOR O2, N2O, AIR & VACUUM CONNECTIONS: Sr. No.

Unit & Parameters Specifications

1 Type Conforms to BS EN type

2 Make Imported 3 Type of outlet Double locking type (front loaded) 4 Valves Primary & Secondary check valves for all positive gas & negative

services. 5 Inlet Copper stubs around 175mm long, 12mm OD copper pipe. 6 Front plate Colour coded 7 Delivery to site In degreased condition, plugged or capped and supplied sealed in

protective polythene bags, accompanied by certificates from third-party inspection agencies.

8 Post- installation purging

After erection, to be flushed and purged clean with dry nitrogen.

9 Delivery to site In air-tight polythene hood and suitable crate. 10 Testing All contraindications, if any, to be carefully examined, Problems

diagnosed and necessary replacement carried out, and test repeated after re-purging the lines. Testing to be carried out in the presence of Hospital Engineer.

H. FLOWMETER WITH HUMIDIFIER Sr. No Unit & Parameters Specifications 1 Type Back Pressure compensated ( Brass bodied), with humidifier. 2 Capacity 1 to 12 l pm with inlet pressure of 50 to 60 psi. 3 Fitting

arrangement Wall-mounted.

4 Body Brass 5 Humidifier Polycarbonate, unbreakable, reusable, autoclavable (134 deg C) 6 Delivery to site In degreased condition, plugged or capped and supplied sealed in

protective polythene bags. 7 Post-installation

purging After erection, to be flushed and purged clean with dry nitrogen.

8 Delivery to site In degreased condition, supplied sealed in protective polythene bags, accompanied by pre-delivery test certificates

I) WARD VACUUM UNIT: No Unit & Parameters Specifications 1 Type Suction regulator type with jar 2 Capacity 2000 ml./ 600ml 3 Fitting arrangement Wall-mounted, with bracket / universal back plate. 4 Suction regulator Wide membrane continuous suction controller 5 Suction bottle Cap Unbreakable material (Nylon).

6 Regulator body Metal; colour coated 7 Collection 2000 ml/ 600ml autoclavable, totally transparent jar, with overflow

safety device. 8 Delivery to site In degreased condition, plugged or capped and supplied sealed in

protective carton box. J)THEATRE SUCTION UNIT: Sr. No.

Unit & Parameters Specifications

1 Type Suction regulator type with twin jar 2 Capacity 2000ltrs x 2nos 3 Fitting arrangement Trolley-mounted 4 Suction regulator Wide membrane continuous suction controller 5 Suction bottle Cap Unbreakable material (Nylon). 6 Regulator body Metal; colour coated 7 Collection 2000 cc autoclavable, totally transparent polysulphone, with overflow

safety device. 8 Delivery to site In degreased condition, plugged or capped and supplied sealed in

protective carton box. K) AREA LINE PRESSURE ALARM (ALPA): No. Unit & Parameters Specifications 1 Make Indian/Imported (to be quoted separately) 2 Type Micro processor control 3 Capacity For OT : To monitor 4 gases (Oxygen, N2O, Medical air (4kg/cm2),

and Vacuum For ICU: To monitor 3 gases (Oxygen, Medical Air, (4kgs/cm2) and Vacuum. Each with normal & two fault (high & low) conditions plus annunciator

4 Audio-visual output True digital LED display read pit with red “alarm”, yellow “caution” and green “normal” trend indication for each service

5 Pressure indicator PSI or Bar read out 6 Manual intervention Only audio alarm mute & test button 7 Body Gasketed hinged doors, dust & vermin proof. 8 Mounting Wall-mounted. 9 Delivery to site Sealed in protective polythene bags. L) SINGLE ARM PENDANT: Indian/Imported (to be quoted separately) (Should comply with IEC/CE standards) Single arm Moveable pendant arm length 1mtr having Oxygen points- 2nos, Medical Air4-1no, surgical air7-1no, N2o-1no & Vacuum-2nos. electrical plug points 6 nos of 16 amps. Infusion pump pole and IV stand combined. Pendants having heavy platform shelf for keeping monitor. (Adjustable type capable to take weight of 50kg) along with flexible hose for connecting the medical gas pipeline. M) BED HEAD PANELS for Surgical Intensive Care Unit:(10 +4 HDU beds) Sr. No.

Unit & Parameters Specifications

1 Type Modular, compartmentalized, multi-function.. 2 Orientation Horizontal / vertical, as required. ( to be quoted separately) 3 Construction Extruded aluminium section, wall thickness min 3mm. 4 Painting Epoxy powder coated 5 Medical gas points 2 oxygen outlets, 1 medical air outlet & 1 suction outlet. 6 Electrical points Modular 6/15 Amp sockets, with switches : 8 nos 7 Delivery to site In air-tight polythene hood and suitable crate. N Compressed Air system

complete with 2nos 2

5

Each compressor of 50 CFM capacity . Non- lubricated ,non noisy. Reciprocating Air cooled type Base Frame mounted Air Compressors ( Anest Iwata or Equivalent Model) BFT150 C-9 each with 15 Hp motors and automatic changeover electrical panel Duplex dryer assy

57.18 cfm each and Receiver of 500 liter water capacity along with duplex 4 stage bacterial filter assy. (Domnick hunter /imported or equivalent) Flow rate 1600 ltrs/min 1. Air Cooled Oil free Air Compressor Anest Iwata Motherson Make Model BFT150C-9 2nos. 2. 15hp Motor 2nos. 3. Air Receiver 0.5 m3 (500lts ) 2nos. 4.Air cooler with Moisture Seperator 2nos. 5. Air Suction Filter 2nos 6. Safety Valve 2nos 7. Belt Guard 2nos 8. Automatic Drain Trap 4Nos 9. Heatless Air Dryer Delair make (DC126) 2Nos 10. Domnick hunter 4 stage filtration System(imported)2 Sets 11. Pressure Reducing Regulator 4Nos. 12. Electrical Panel (automatic changeover) 1No

O-1 Copper Piping as per BS EN: 13348 standards complete with fittings and installation material (Approximate, May vary actually during site visit and installation process)

A 54 mm OD Copper pipes 100 Meter B 42 mm OD Copper pipes 300 Meter C 35 mm OD Copper pipes 800 Meter D 28 mm OD Copper pipes 2000 Meter E 22 mm OD Copper pipes 2000 Meter F 15 mm OD Copper pipes 2800 Meter G 12mm OD Copper pipes 800 Meter O-2 Isolation Valve: -Ball valve Imported A 15 mm Isolation Valve 82 Nos. B 22 mm Isolation Valve 30 Nos. C 28 mm Isolation Valve 1 Nos. D 35 mm Isolation Valve 1 Nos. E 54 mm Isolation Valve 1 Nos. O-3 Outlet Point: A Double locking outlet points for oxygen 196 Nos. B Double locking outlet points for Air4 43 Nos. C Double locking outlet points for Air7 4 Nos. D Double locking outlet points for N2o 1 Nos. E Double locking outlet points for vacuum 171 Nos. O-4 Oxygen flow meter (0-15 LPM) with humidifier bottle 43 Nos. O-5 Twin flow meter 126 Nos. O-6 Vacuum unit 600ml 142 Nos. O-7 Vacuum unit 2000ml 28 Nos. O-8 Indigenous Theatre Suction Unit with trolley with caster wheel, with 2

nos of 2000ml Collection Jars with suction regulator 22 Nos.

O-9 Valve Box powder coated complete with front glass cover, pressure gauges, Lock complete

O-10 Valve Box unit for Five gas services. 12 Nos. O-11 Valve Box unit for three gas services. 9 Nos. O-12 Valve Box unit for Two services. 9 Nos. O-13 Medical gas Digital Gas Alarm system with pressure indication displaying

the measured levels on the panel in PSI and Bar

A Area Alarm Panel for five gas services. 11 Nos. B Area Alarm Panel for three gas services. 5 Nos. C Area Alarm Panel for two gas services. 9 Nos. P Single arm Moveable pendant arm length 1mtr having

Oxygen points- 2nos, Medical Air4-1no, surgical air7-1no, N2o-1no & Vacuum-2nos. electrical plug points 6 nos of 16 amps. Infusion pump pole and IV stand combined. Pendants having heavy platform shelf for keeping monitor. (Adjustable type capable to take weight of 50kg) along with flexible hose for connecting the medical gas pipeline.

11 Nos.

21 ICU Type Bed Head Panel Exclusive aluminium powder coated 14 Nos.

- Panel with provision for 2 oxygen outlets, 1 medical air outlet + 1 suction outlet. -Provision of 8socket for electrical 6 x 16mm -Dimensions -1500mm (59 inches)

22 Liquid Medical Oxygen Foundation for and Installation of Liquid tank 6000ltrs along with fencing, digging for trench for oxygen, N2O, Air & vacuum pipeline from plant room to building along with GI pipe, Cement pipe etc and covering it back. This job is to be carried out as per standards of CCOE on Turn key project

1 Job

Product : Liquid Medical Oxygen (Inox or Equivalent)

Fixed Facility Charges :for the refilling 6 KL storage tank per month or as per requirement for atleast 5 yrs.

CCOE License should be procured Mobilization of the storage tank and allied equipment’s up to Nagpur

site should be tenderers responsibility

Approximate Area available for Liquid oxygen unit(SMPV area) and CPL

plant room is 25m x 10 m as per CCOE requirements. Necessary map for area, approach road and civil work is attached herewith Gas Manifold room & Vaccum Pump room should be Constructed near liquid oxygen plant as per the standards laid down and considering all safety protocols.

02. e-tender-40/15-16/ Endoscopy Ultrasono (Linear and Radial Probe) Sr.No. Previous Specifications Revised Specifications Endoscopic Ultrasound (EUS) System consisting

of Video Linear Ultrasound Endoscope, Video Radial Endoscope, Video Processor with inbuilt Xenon Light Source (Xenon) and standalone Color Doppler, Medical Grade Monitor and latest configuration computer with accessories

Endoscopic Ultrasound (EUS) System consisting of Video Linear Ultrasound Endoscope, Video Radial Endoscope, Video Processor and Xenon Light Source (Xenon) and standalone Color Doppler, Medical Grade Monitor and latest configuration computer with accessories

A 3 VIDEO LINEAR ULTRASOUND ENDOSCOPE VIDEO LINEAR ULTRASOUND ENDOSCOPE

b. It should have depth of field of 5- 100mm b. It should have depth of field of 3- 100mm

c. It should have 100- 120° field of view with a forward oblique view of 45-55 Degree

c. It should have 100- 140° field of view

d. It should have minimum Tip Deflection of Up: 130° Down : 130°, Right: 120° Left :120°

d. It should have minimum Tip Deflection of Up: 130°or more, Down : 130° or more, Right: 120° or more, Left :120° or more

e. Distal End Diameter of probe should be less than 12.5

e. Distal End Diameter of probe should not be less than 12.5 mm.

j. It should have a ultrasound scan angle of 120 degree

j. It should have a ultrasound scan angle of 120 degree or more

B b. It should have depth of field of 5- 100mm b. It should have depth of field of 3-100mm or more

t. 3D imaging option for radial scanning probe To be deleted

C i. Processor should have in-built light source To be deleted

D f. Real-time Tissue Elastography facility to visualize tissue strain in real time, offering information of its stiffness (optional)

f. Real-time Tissue Elastography facility to visualize tissue strain in real time, offering information of its stiffness.

g. It should have facility for Tissue Harmonic Echo imaging (optional)

g. It should have facility for Tissue Harmonic Echo imaging.

h. It should have ability to measure blood flow (Color flow mode) using Pulse wave Doppler and optionally H-FLOW (High resolution flow).

h. It should have ability to measure blood flow (Color flow mode) using Pulse wave Doppler and H-FLOW (High resolution flow).

To Be Added Equipment should be USFDA or European Notified body CE certified.

To be Added For all Imported Equipment the bidder has to submit bill of entry copy for equipment imported in last 02 years

To be added Successful bidder will be required to submit Custom duty clearance copy alongwith supplies of equipment or else payment will not be released.

03. e-tender-41/15-16/ Video Endoscopy (Gastroscope) To Be Added Equipment should be USFDA or European Notified

body CE certified. To be Added For all Imported Equipment the bidder has to

submit bill of entry copy for equipment imported in last 02 years

To be added Successful bidder will be required to submit Custom duty clearance copy alongwith supplies of equipment or else payment will not be released.

04. e-tender-46/15-16/ Hemodialysis Machine Sr.No. Previous Specifications Revised Specifications 2 Operational Requirements To be added 2.4Dialysis Machine should be "open system".

Machine should run on the consumable from other make as well.

3.2 Machine should have two bacterial filter (Pyrogen filters) one at water inlet and one before water going to dialyser

To be deleted

3.6 Variable conductivity setting between 12 to 15 Variable conductivity setting between 12 to 16 3.16 Should have accurate UF control by flow

measurement technique Should have accurate UF control by balanced chember

3.21 Should have touch button screen Should have touch screen or button 3.24 Blood pump rate from 0-600 ml/min adaptable

to all standard A-V blood lines Blood pump rate from 50-600 ml/min adaptable to all standard A-V blood lines

7.4 Comprehensive warranty for 5 years and 5 years AMC ofter warranty supported by the principal manufactures

Comprehensive warranty for 2 years and 5 years AMC ofter warranty supported by the principal manufactures (as per tender terms conditions)

To be added Cost of the consumables should be quote separately and will be frozen for 05 years.

To Be Added Equipment should be USFDA or European Notified body CE certified.

To be Added For all Imported Equipment the bidder has to submit bill of entry copy for equipment imported in last 02 years

To be added Successful bidder will be required to submit Custom duty clearance copy alongwith supplies of equipment or else payment will not be released.

05. e-tender-48/15-16/ Central Oxygen System Vacum E Suction system with Essential

Part Sr.No. Previous Specifications Revised Specifications B Specification for central O2, N2O, Air and

Vacuum

3 Fully automatic digital control panel for oxygen

3 Fully automatic digital control panel for oxygen (Imported)

8 Vacuum System Model: V255 vacuum pump any Indian make, Type : Squirrel cage , Rating 5HP, Cylinders: 2nos, Size & Stroke -5” & 5” x 3-1/2”, Piston displacement: 60lpm, Maximum Vacuum with Closed Intake-736.6mm of hg, V-Belt driven, Air Cooled

8 Vacuum System Model: V255 vacuum pump any Indian make, Type : Squirrel cage , Rating 10HP Duplex, Cylinders: 2nos, Size & Stroke -5” & 5” x 3-1/2”, Piston displacement: 60lpm, Maximum Vacuum with Closed Intake-736.6mm of hg, V-Belt driven, Air Cooled

Vacuum receiver: Vacuum Receiver Tank will have Pressure 600mm of Hg and will be hydraulically tested at 10kg/sq.cm. It will be fabricated by low heat input Electrodes having dimensions 84” x 30” and capacity 1000ltrs approx. The thickness of the Vacuum reservoir tank will be 4mm. The tank will be provided with a drain cork and a gauge

Vacuum receiver: Vacuum Receiver Tank will have Pressure 600mm of Hg and will be hydraulically tested at 10kg/sq.cm. It will be fabricated by low heat input Electrodes having dimensions 84” x 30” and capacity 2000 ltrs approx. The thickness of the Vacuum reservoir tank will be 4mm. The tank will be provided with a drain cork and a gauge

05. e-tender-67/15-16/ Operative Laparoscope Sr.No. Previous Specifications Revised Specifications

1 HD Laparascope 10 mm, 30 degree, High speed III rod lens Medical grade stainless steel body, sapphire tips for more strach resistance, Lazer welding joints, high depth of focus, high resolution uniform brightness wide angle 75 dgree, working length 330 mm, high magnification, autoclavable full C ring elumination, three tube design

High Definition 3 chip Camera System Full HD native output of 1920 X 1080P in DVI-D format Aspect Ratio 16:9, Sensitivity - 2 Lux @ f (1.4) Signal? Noise Ratio-70db, window, anti-moire filter, buttons on camera head for 6 functions such as scop selection [Speciality wise Surgery selection], digital zoom function, white balance, picture freez function, gain,video recording On/Off and still image functions on camera head button. 15 step digital enhancer, Facility to invert or flip image. Progressive Scan CCD-1.3 x 3, buttons on head, detachable C- Mount with optical zoom coupler. U.S.B. port on camera control for still images and video recording or same other HD recording system / either built in or external Digital outputs - HD-SDI-2, Y/C-2, Composite-2,DVI-1, RGB-1,RS-232.

2 HD Laparascope 10mm,0degree, High speed III rod lens, Medical gade stainless steel body, sapphire tips for more strach Resistance, laser welding joints, Wide angle 75 degree, working length 330 mm high magnification autoclavable, full 0 ring, elumination, three tube design, high depth of focus, high resolution, uniform brightness

FULL HD LED Medical Grade Monitor 1. Size of the Monitor should not less then 24" 2. Monitor should have full HD, 1920 x 1080

resolution. 3. It should have digital DVID input. 4.All other on screen menu adjustments like

gamma, colour, contrast, brightness etc. should be provided

5. Aspect ratio can be selectable 16:9 or 4:3

3 HD Laparascope 5 mm, 30 degree, High speed III rod lens, Medical grade stainless steel body,sapphire tips for more Strach Resistance, lazer welding joints, high depth of focus, high resolution Uniform brightness, wide angle 75 degree, working length 330mm, high magnification, autoclavable full C ring, elumination three tube design

Led Light Source Accurate colour temperature, long life of 50000 hours, standby mode, sensor for F.O. cable, step intensity adjustment, consistent lux intensity, equivalent to 300 watts of Xenon bulb

4 CO2 Insufflator flow rute 30 lt/min, microprocessor based, Facility to adjust flow, dedicated verss mode, set flow adjustment, venting system, without high pressure tubing five parameter display for set pressure actual pressure set flow , actual flow, volume light weight, built in self testocclusion test, continnous monitoring of safety parameters.

Fiber Optic Cable 4.8mm x 230cm, metal braded, Autoclavable and heat resistant with CE mark

5 Morcillator with accessories, To be Deleted 6 Laparascopy Hand Instruemnts with

C.E.Mark 5. Suction irrigation system with 5 mm tubes,

two way stop cock

To be Deleted

To Be Added Equipment should be European Notified body CE certified.

To be Added For all Imported Equipment the bidder has to submit bill of entry copy for equipment imported in last 02 years

To be added Successful bidder will be required to submit Custom duty clearance copy alongwith supplies of equipment or else payment will not be released.

05. e-tender-69/15-16/ Operative Laproscopy complete with HD Camera, LED Monitor with Hand Instrument with Recording System ISO & CE Certified Sr.No. Previous Specifications Revised Specifications 01 High Definition 3 chip Camera System 3 buttons on camera head for 6 functions such

as scop selection Buttons on camera head for 6 functions such as scop selection

light source selection such as LED, Xenon, Halogen.

To be deleted

Progressive Scan CCD-1.3 x 3, 3 buttons on head, detachable C- Mount

Progressive Scan CCD-1.3 x 3, buttons on head, detachable C- Mount with optical zoom coupler

U.S.B. port on camera control for still images and video recording

U.S.B. port on camera control for still images and video recording or same other HD recording system / either built in or external

02 FULL HD LCD MONITOR FULL HD LED Medical Grade Monitor 03 Led Light Source

Accurate colour temperature, long life of 50000 hours, standby mode, sensor for F.O. cable, step intensity adjustment, consistent lux intensity

Led Light Source Accurate colour temperature, long life of 50000 hours, standby mode, sensor for F.O. cable, step intensity adjustment, consistent lux intensity, equivalent to 300 watts of Xenon bulb.

04 Fiber Optic Cable 4.8mm x 230cm, metal braded, Autoclavable, with CE mark

Fiber Optic Cable 4.8mm x 300 cm, metal braded, Autoclavable and heat resistant with CE mark

05 H.D. Laparascop 10mm, 30 degree,High speed III rod lens, HD Optics medical grade stainless steel body, sapphire tips for more starch Resistance, lazer welding joints, uniform brightness, wide angle 75 degree, working length 330mm, minimum vignetting, Autoclavable, full c ring elumination three tube design, high magnification, Full Screen image on monitor, glass cone at cable end, high depth of focus, High resolution with CE mark

H.D. Laparoscope 10mm, 30 degree, HD Optics medical grade stainless steel body, sapphire tips for more starch Resistance, lazer welding joints, uniform brightness, wide angle 75 degree, working length 300- 330mm, minimum vignetting, Autoclavable, full c ring elumination three tube design, high magnification, Full Screen image on monitor, high depth of focus, High resolution with CE mark

06 CO2 Insufflator flow rate 40lts / min, Microprocessor based Low flow & high flow mode, venting system, with warmer, Functions to set actual flow, set flow, actual pressure, set pressure, volume veriss mode [3 liters/min] Withb high pressure tubing, light weight, continuous monitoring of safety parameters. eith CE mark

CO2 Insufflator flow rate 40lts / min, Microprocessor based Low flow & high flow mode, venting system, with warmer, Functions to set actual flow, set flow, actual pressure, set pressure, volume veriss mode With high pressure tubing, light weight, continuous monitoring of safety parameters with CE mark.

07 Laparoscopy Hand Instruments with CE Mark Laparoscopy Hand Instruments should be of European Notified body CE certified

Traumatic Forcep Detachable, rotatable grasping & dissecting forceps with carbon reinforced handles 5mm dia, removable inserts with or without ratchet & HF connector

Traumatic Forcep Detachable, rotatable grasping & dissecting forceps with peek material handles 5mm dia, removable inserts with or without ratchet & HF connector

Bowel Forcep Detachable, rotatable grasping & dissecting forceps with carbon reinforced handles 5mm dia, removable inserts with or without ratchet & HF connector

Bowel Forcep Detachable, rotatable grasping & dissecting forceps with peek material handles 5mm dia, removable inserts with or without ratchet & HF connector

Babcock Forcep Detachable, rotatable grasping & dissecting forceps with carbon reinforced handles 5mm dia, removable inserts with or without ratchet & HF connector

Babcock Forcep Detachable, rotatable grasping & dissecting forceps with peek material handles 5mm dia, removable inserts with or without ratchet & HF connector

Tinaculam Forcep Detachable, rotatable grasping & dissecting forceps with carbon reinforced handles 5mm dia, removable inserts with or without ratchet & HF connector

Tinaculam Forcep Detachable, rotatable grasping & dissecting forceps with peek material handles 5mm dia, removable inserts with or without ratchet & HF connector

Metzembam curved Scissor Detachable, rotatable grasping & dissecting forceps with carbon reinforced handles 5mm dia, removable inserts with or without ratchet & HF connector

Metzembam curved Scissor Detachable, rotatable grasping & dissecting forceps with peek material handles 5mm dia, removable inserts with or without ratchet & HF connector

High end Detacheble Bipolar forcep rotatable & HF cable

To be deleted

(All the telescopes, equipments and hand instruments should be of same single parant company with CE & I.S.O. Certified except accessories)

(All the telescopes, equipments and hand instruments should be of European Notified body CE & I.S.O. Certified except accessories)

To Be Added Equipment should be European Notified body CE certified.

To be Added For all Imported Equipment the bidder has to submit bill of entry copy for equipment imported in last 02 years

To be added Successful bidder will be required to submit Custom duty clearance copy alongwith supplies of equipment or else payment will not be released.

To be Added Every Item in specification should be quote separately.

5. All the tenderer should note the above modifications in the Specification and do

accordingly. 6. The Directorate of medical Education & Research reserves the right to increase or decrease

the quantity of tendered equipments and reserves the right to reject the tender without any reason thereto.

For e-tendering help/information , please contact. Website:-http://dmer.maharashtra.etenders.in Email :[email protected] Contact No.:- 91-022-2620361-65 (Ext. 338 / 339 / 340 ) Sd/-

Director Medical Education & Research, Mumbai