Pleur-evac cactus chest drainage system atrium Oasis A-6000 Series 3600 Series Suction Dry suction to 40 cm of water Seal Wet Water Seal needs to be established through the addition of 2 cm of H 2 O Patient is at risk of exposure to atmospheric air until water seal has been established Patient Protection upon Tip Over One-way valve protects the patient from atmospheric air in any orientation Suction ON One-way valve protects the patient from atmospheric air in any orientation when suction is ON Suction OFF Risk of atmospheric air entering the patient around angled fluid when suction is OFF Fluid Migration upon Tip Over Fluid remains in seal when tipped. Drainage can migrate among sections of collection chamber Suction Source Requirements 16 LPM 20 LPM Suction Control Dial Clicks in place at prescribed level Adjustable but does not click into place Suction Indicator Orange float Bellows Suction Indicator Active Distinct “On/Off” Orange Float only visible when suction has reached prescribed level (pre-set -20 cm) Variable bellows. Suction bellows will expand to the mark or beyond when suction is connected and operating at a regulator setting of -20 cmH 2 0 or higher. If the bellows is expanded but less than mark, increase the suction source vacuum to -80 mmHg or higher. For regulator settings less than -20 cmH 2 0, and visible bellows expansion in bellows window will confirm suction operation. Optional Stop Cock Suction Control No Yes Air Leak Meter Air leak meter must be filled to detect air leak Meter Labeled LOW (1) to HIGH (7) LOW (1) to HIGH (5) Dry Suction/Wet Seal Feature Comparison Overview PLEUR-EVAC ® CACTUS SERIES AND ATRIUM OASIS SERIES

Upload: doanmien

Post on 26-Jul-2019

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dry Suction/Wet Seal Feature Comparison Overview cactus chest drainage system atrium Oasis A-6000 Series 3600 Series Suction Dry suction to 40 cm of water Seal Wet Water Seal needs

Pleur-evac cactus chest drainage system atrium OasisA-6000 Series 3600 Series

Suction Dry suction to 40 cm of water

Seal WetWater Seal needs to be established through the addition of 2 cm of H2O

Patient is at risk of exposure to atmospheric air until water seal has been established

Patient Protection upon Tip Over One-way valve protects the patient from atmospheric air in any orientation

Suction ON One-way valve protects the patient from atmospheric air in any orientation when suction is ON

Suction OFF Risk of atmospheric air entering the patient around angled fluid when suction is OFF

Fluid Migration upon Tip Over Fluid remains in seal when tipped. Drainage can migrate among sections of collection chamber

Suction Source Requirements 16 LPM 20 LPM

Suction Control Dial

Clicks in place at prescribed level Adjustable but does not click into place

Suction Indicator Orange float Bellows

Suction Indicator Active

Distinct “On/Off” Orange Float only visible when suction has reached prescribed level (pre-set -20 cm)

Variable bellows. Suction bellows will expand to the mark or beyond when suction is connected and operating at a regulator setting of -20 cmH20 or higher. If the bellows is expanded but less than mark, increase the suction source vacuum to -80 mmHg or higher. For regulator settings less than -20 cmH20, and visible bellows expansion in bellows window will confirm suction operation.

Optional Stop Cock Suction Control No Yes

Air Leak Meter Air leak meter must be filled to detect air leak

Meter Labeled

LOW (1) to HIGH (7) LOW (1) to HIGH (5)

Dry Suction/Wet Seal Feature Comparison OverviewPLEUR-EVAC® CACTUS™ SERIES AND ATRIUM OASIS™ SERIES

Page 2: Dry Suction/Wet Seal Feature Comparison Overview cactus chest drainage system atrium Oasis A-6000 Series 3600 Series Suction Dry suction to 40 cm of water Seal Wet Water Seal needs

Teleflex, Mini Sahara, Pleur-evac and Sahara are trademarks or registered trademarks of Teleflex Incorporated or its affiliates. © 2015 Teleflex Incorporated. All rights reserved. MC-001244

TELEFLEX 3015 Carrington Mill Boulevard, Morrisville, NC 27560 Toll Free: 866.246.6990 Phone: +1.919.544.8000 TELEFLEX.COM

Teleflex, Pleur-evac and Cactus are trademarks or registered trademarks of Teleflex Incorporated or its affiliates.© 2015 Teleflex Incorporated. All rights reserved. MC-001391

Oasis is a registered trademark of Atrium Medical Corporation.

Pleur-evac® cactus™ chest drainage system atrium Oasis™

A-6000 Series 3600 Series

Collection Chamber 2500 ml• 1 ml increments up to 100 ml, then 2 ml up to 200 ml

• Over 200 ml calibrations are in 5 ml increments

2100 ml• 1ml increments up to 100 ml, then 2 ml up to 200 ml

• Over 200 ml calibrations are in 10 ml increments

Specimen Sampling

Needleless Sample Port in patient tube

Needle Not recommended 20 gauge needle or smaller

In-Line ATS Bag Yes (A-1500-08LF) Yes (2550)

Capacity 1400 cc gravity 700 cc

Internal Spring No Yes

Dedicated filter Yes (4") No

Possible to pressure cuff Yes No

Suspension Side Front

Infant Unit A-6020-08LF (1/2 ml increments up to 10 ml/1 ml to 150 ml) 3612 (1 ml up to 100/2 ml up to 200)

Dual A-6002-08LF (950/1900 ml) 3620 (1090/900 ml)

Continuous Reinfusion Models A-9250LF 3650/3652

Rapid Transfer Bag Yes (S-250-08) Yes (2450)

Spring External - can be pressure cuff reinfused Internal - cannot be pressure cuff reinfused

Capacity 700 cc 700 cc