UV Detectors

Fixed Wavelength

Variable Wavelength

Diode array

Page 3: UV Detectors Fixed Wavelength Variable Wavelength Diode array

High PressureMercuryLamp

Fixed Wavelength

Mercury vapor lamps are probably the most common and emit intense light at 253.7 nm (and certain other wavelengths). Because of the limitedemission spectra, of the lamp wavelengthsare not adjustable.

Because of the intensity of the radiation, fixed wavelength detectors can be up to 20 times more sensitive than variable wavelength detectors.

Page 4: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Variable WavelengthThe variable types use a deuterium or similar lamp that produces a broad spectrum of wavelengths that are separated by a defraction grating. The grating works like a prism but generally the resolving power is higher for gratings than for prisms.

Page 5: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Diode ArrayThe photo diode array detector passes a wide spectrum of light through the sample. The spectrum of light is directed to an array of photosensitive diodes. Each diode can measure a different wavelength which allows for

the monitoring of many wavelengths at once.

- Peak Purity

- Quantify a peak with an interfering peak

Compound Identification

Monitor compounds withdifferent UV max.

Page 6: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Intensity of Lamp Energy

Page 7: UV Detectors Fixed Wavelength Variable Wavelength Diode array

As the lamps age, the light intensity drops significantly.

Therefore it is important to monitor lamp usage.

Arc lamps have an average lifetime of about 200

hours and most detectors are equipped with a timer that

keeps track of lamp usage.

Lamp Life

Page 8: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Fluorescence DetectorsFluorescence detectors are among the most sensitive of HPLC detectors, and depending on the compound, can be from 10-1000 times more sensitive than UV detectors analyzing strong UV absorbing compounds.

The fluorescence detector is based on the principle that some compounds fluoresce when bombarded with UV light. If the compound of interest fluoresces this is a very sensitive detector.

The analyte is excited by light commonly at 253.7 nm from a low pressure mercury lamp. The light is absorbed and the molecule is excited and it gives off light of a different wavelength. This wavelength is monitored by a detector which sits at right angles to the UV light source used to excite the analyte.

Page 9: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Electrical Conductivity

The electrical conductivity detector was one of the first in-line detectors developed and is still in use for detecting electrically charged ions. It is based on the principle that the electrical conductivity of a liquid is proportional to the ionic concentration in the mobile phase. The design is simple, and the detector can be made small enough for an effective sensing volume of a few nanoliters, making this detector very useful for capillary electrophesis.

Page 10: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Refractive Index Detectors

In contrast to the fluorescence detector, the refractive index detector is one of the least sensitive (about 1000 times less sensitive than UV detectors).

Refractive index detectors are often used for the analysis of carbohydrates. Carbohydrates are generally not ionic do not fluoresce, and do not absorb UV or visible light to any great extent, so the refractive index is one of the few detectors that can detect them.

Page 11: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Light Scattering Detectors

The flow from the column is converted into a fine spray or mist. The mobile phase is evaporated, leaving tiny particles of the analytesThe particles are passed through alaser beam and they scatter the laser light. The scattered lightis measured at right angles to the laser beam by a photodiode detector.

Page 12: UV Detectors Fixed Wavelength Variable Wavelength Diode array

Mass Spectrometric Detectors

For all mass spectrometer techniques, the analytes are first ionized since the detector only measures charged particles. The charged particles then enter the separator that separatesions on the basis of mass using a magnetic field. The masses of different fragments hit the detector wherethey give up their charge creating an electric signal which is recorded.

It is difficult to use a mass spectrometer with liquid chromatography because the mobile phase must be removed before the analytes can enter the detector. A number of interfaces have been developed to solve this problem (ie., electrospray, ion spray, and thermospray) and LC-MS is becoming more common.