Page 1: Laboratory QA/QC

An Overview

Page 2: Laboratory QA/QC

Quality Assurance: QA is defined as the overall program that ensures the final results reported by the laboratory are correct.QA is a broad plan for maintaining quality in all aspects of a program.QA establishes the need for quality control.

Page 3: Laboratory QA/QC

Quality Control: The measures that must be included during each laboratory procedure to verify that the test is working properly.QC refers to routine technical activities with the purpose to control error.

QC can be considered as the “HOW” of the QA process.

QC is applicable to field, lab and office procedures (administration).

Page 4: Laboratory QA/QC

Quality Assessment - quality assessment (also known as proficiency testing) is a means to determine the quality of the results generated by the laboratory. Quality assessment is a challenge to the effectiveness of the QA and QC programs. Quality Assessment may be external or internal.

Page 5: Laboratory QA/QC

“QC aims at simply ensuring that the results generated by the test are correct. However, QA is concerned with much more. It checks whether the right test is carried out on the right sample, and that the right result and right interpretation is delivered to the right person at the right time”

Page 6: Laboratory QA/QC

We need QA to: Understand data reliability; Quantify areas of analytical uncertainty; and Standardize measurement to allow for

repeatable and comparable data across time and place.

Page 7: Laboratory QA/QC

Quality Assurance (QA)• broad program plan • establishes the need

for QC

Quality Controls (QC) • individual checks

and balances• the “How to” of QA

Page 8: Laboratory QA/QC

Quality control is applicable in all aspects of a soil, plant and water sampling project including: Field data collection and sampling Laboratory analysis and processing Data evaluation and assessment Reporting and project documentation

QC provides steps to ensure lab data will meet defined standards of quality with a standard level of confidenceQC provides steps to ensure lab data will meet defined standards of quality with a standard level of confidence

Page 9: Laboratory QA/QC

In most cases field QC (soil, water and plant sampling) is out of laboratory control;

QC is particularly critical in field data collection; Often the most costly aspect of any project and the

most limiting factor is field sampling; Data is never reproducible under the exact same

condition or setting; Therefore, field sampling QA is also needed to

assure that best possible (most reliable) set of data is obtained.

Page 10: Laboratory QA/QC

7.5 cm core Irrigation projectRepresenting 10 ha

Page 11: Laboratory QA/QC

Laboratory data analysis, data measurement, and data acquisition:Chain of custody formsEquipment calibrationStorage practicesAnalytical methodsHolding timesLimit of detection (LODs),

previously known as MDLs.

www.odc.gov/noeh/dls

Page 12: Laboratory QA/QC

Educational background and training of personnel;Condition of the samples;Controls used in the test runs;Reagents quality;Maintenance status of equipment;Interpretation of the results;Recording of results; andReporting of results.

Page 13: Laboratory QA/QC

True value: This is an ideal concept which practically cannot be achieved.Accepted true value: The value approximating the true value, the difference between the two values should be negligible (not statistically significant).Error: The discrepancy between the result of a measurement and the true (or accepted true value).

Page 14: Laboratory QA/QC

Input data required: Such as standards used, calibration values, and values of physical constants;Inherent characteristics of the quantity being measured;Instruments used: Accuracy, repeatability;Observer unreliability: Reading errors, blunders, equipment selection, analysis and computation errors;

Page 15: Laboratory QA/QC

Environment: Any external influences affecting the measurement; andTheory assumed: Validity of mathematical methods and approximations.

Page 16: Laboratory QA/QC

An error that varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions. Random errors create a characteristic spread of results for any test method and cannot be accounted for by applying corrections. Random errors are difficult to eliminate, but repetition reduces the influences of random errors.

Page 17: Laboratory QA/QC

Examples of random errors include: errors in pipetting;

changes in incubation period; or

the time used for extraction/centrifuging.

Random errors can be minimized by training,

supervision and adherence to standard operating

procedures (SOPs).

Page 18: Laboratory QA/QC

x

x x

x x

True x x x x

Value x x x

x x x

x

x

x

Page 19: Laboratory QA/QC

An error that, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions, or varies according to a definite law when conditions change. Systematic errors create a characteristic bias in the test results and can be accounted for by applying a correction. Systematic errors may be induced by factors such as variations in incubation temperature, change in the reagent batch or modifications in testing methodology.

Page 20: Laboratory QA/QC

x

x x x x x x x

True x

Value

Page 21: Laboratory QA/QC

Internal Quality Control: “Controllable” by those responsible for

performing the laboratory analysis. External Quality Control:

A “set of measures” established for and conducted by people outside the analytical laboratory (lab auditors, regional or national laboratories, accreditation process, etc).

Page 22: Laboratory QA/QC

Internal Quality Control:Equipment calibrationProper training and

certification of practitionersProper sampling techniques Proper data documentation

Page 23: Laboratory QA/QC

IQC samples comprises either In-house prepared aliquot of known values, orInternational standards with values within significant ranges for the element to be measured.

Page 24: Laboratory QA/QC

External quality control:Performance auditsSplit sample analysisReplicate (duplicate)

sample analysis

Page 25: Laboratory QA/QC

Successful data collection and analysis is dependant upon “The PARCC Parameters”: Precision Accuracy Representativeness Completeness Comparability

The key concepts of QA/QC are the “PARCC” Parameters – the WHY of the QA

The key concepts of QA/QC are the “PARCC” Parameters – the WHY of the QA

Page 26: Laboratory QA/QC

Precision - degree of agreement there is between repeated

measurements of the same characteristic can be biased – meaning there is a consistent

error in the results Accuracy -

measures how close data results are to a true or expected value – does not allow for bias

Page 27: Laboratory QA/QC

accuracy = (average value) – (true value) precision represents

repeatability bias represents

amount of error low bias and high

precision = statistical accuracy

http://www.epa.gov/owow/monitoring/volunteer/qappexec.html

Page 28: Laboratory QA/QC

Representativeness - extent to which measurements actually

represent the true environmental condition or population at the time a sample was collected.

Representative data should result in repeatable data

Does this

represent this?? http://pubs.usgs.gov/fs/fs-0058-99

Page 29: Laboratory QA/QC

Comparability -the extent to which data can be compared

between sample locations or periods of time within a project, or between projects

Will similar data from these sites be Comparable ??

Page 30: Laboratory QA/QC

Quality Assurance (QA) broad program plan establishes the need

for QC

Quality Controls (QC) standardized tests

and methods the “HOW” of QA

Page 31: Laboratory QA/QC

An internal quality control program depends on the use of internal quality control (IQC) samples, and using statistical analysis methods for interpretation.

Page 32: Laboratory QA/QC

A Shewhart Control Chart depend on the use of IQC samples and is developed in the following manner:Put up the IQC specimen for at least 20 or more sample runs and record down the readings;Calculate the mean (x) and standard deviations (Sd);Make a plot with the sample run on the x-axis, and concentration readings on the y axis.

Page 33: Laboratory QA/QC

Draw the following lines across the y-axis: mean, -3, -2, -1, 1, 2, and 3 Sd;Plot concentration reading obtained for the IQC specimen for subsequent sample runs.Major events such as changes in the reagent batch and/or instruments used should also be recorded on the chart.

Page 34: Laboratory QA/QC

A Shewhart control chart consists of:Points representing a statistic (e.g., a mean, range, or proportion) of measurements of a quality characteristic in samples analyzed at different times [the data];The mean of this statistic using all the samples is calculated (e.g., the mean of the means, mean of the ranges, or mean of the proportions);

Page 35: Laboratory QA/QC

A center line is drawn at the value of the mean of the statistic;The standard error (e.g., standard deviation) of the statistic is also calculated using all the samples; andUpper and lower control limits (sometimes called "natural process limits"), indicating the threshold at which the process output is considered statistically 'unlikely' are drawn typically at 3 Sd from the center line.

Page 36: Laboratory QA/QC

The formulation of Westgard rules were based on statistical methods. Westgard rules are commonly used to analyze data in Shewhart Control charts. Westgard rules are used to define specific performance limits for a particular analysis and can be use to detect both random and systematic errors.

Page 37: Laboratory QA/QC

There are six commonly used Westgard rules of which three are warning rules and the other three are mandatory rules. The violation of warning rules should trigger a review of test procedures, reagent performance and equipment calibration. The violation of mandatory rules should result in the rejection of the obtained results.

Page 38: Laboratory QA/QC

0102030405060708090100

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

+3 sd

-3 sd

+2 sd

-2 sd

-1 sd

+1 sd

Sam

ple reading

Target value

Sample run

Page 39: Laboratory QA/QC

Warning 12SD : It is violated if the IQC value exceeds the mean by 2SD. It is an event likely to occur normally in less than 5% of cases.Warning 22SD : It detects systematic errors and is violated when two consecutive IQC values exceed the mean on the same side of the mean by 2SD.Warning 41SD : It is violated if four consecutive IQC values exceed the same limit (mean 1SD) and this may indicate the need to perform instrument maintenance or reagent calibration.

Page 40: Laboratory QA/QC

Mandatory 13SD : It is violated when the IQC value exceeds the mean by 3SD. The test is regarded as out of control.Mandatory R4SD : It is only applied when the IQC is tested in duplicate. This rule is violated when the difference in Sd between the duplicates exceeds 4Sd.Mandatory 10x : This rule is violated when the last 10 consecutive IQC values are on the same side of the mean or target value.

Page 41: Laboratory QA/QC

0102030405060708090100

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

+3 sd

-3 sd

+2 sd

-2 sd

-1 sd

+1 sd

Sam

ple reading

Target value

Assay Run

Page 42: Laboratory QA/QC

0102030405060708090100

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

+3 sd

-3 sd

+2 sd

-2 sd

-1 sd

+1 sd

Sam

ple reading

Target value

Assay Run

Page 43: Laboratory QA/QC

There are three options as to the action to be taken in the event of a violation of a Westgard rule:

1) Accept the test run in its entirety: This usually applies when only a warning rule is violated.

2) Reject the whole test run: This applies only when a mandatory rule is violated.

3) Enlarge the grey zone and thus re-test range for that particular test run: This option can be considered in the event of a violation of either a warning or mandatory rule.