digital es modulevel w/hart communication and · pdf filemechanical design and construction....
TRANSCRIPT
®
Digital ES and Analog EZ Modulevels are advanced, intrin-sically safe two-wire instruments, utilizing simple buoyancyprinciple to detect and convert liquid level changes into astable 4–20 mA output signal. The linkage between the levelsensing element and output electronics greatly simplifiesmechanical design and construction. The in-line verticaldesign of the transmitter reduces instrument weight andsimplifies installation.
The Digital ES Modulevel has microprocessor-based elec-tronics with HART compatible output, in addition to thestandard 4–20 mA DC output. The Analog EZ Modulevelprovides a 4–20 mA DC output only.
• HART network compatible.• Remote calibration with-
out moving level viaHART compatible handheld unit.
• Change specific gravitywithout stopping processvia HART compatiblehand held unit.
• Push button programlocal calibration.
• Continuous self-test with22 mA DC fault signal anddigital warning message.
• Signal damping may bechanged without stop-ping process via HARTcompatible hand heldunit.
• May be recalibrated with-out moving the level intank.
• Two-wire, loop-powered,intrinsically safe.
• Cast iron, NEMA 4X/7/9housing.
• Field selectable direct orreverse output.
• Field wiring in isolatedjunction box.
• Range spring suppressesthe effect of turbulence togive stable output signal.
• Special conduit seals pro-hibit moisture andcondensation seepageinto electronics.
• Specific gravity to as lowas 0.11.
• Level ranges from 14 to 120inches (356 to 3048 mm).
• Non-interacting zero andspan.
• Interface option available.
• 4–20 mA output signal.
• Analog meter available.
• Calibration stand avail-able for bench calibration.
• CSA and FM approved forhazardous locations.
• Process temperatures to600° F (315° C) for non-steam applications.
• Flanged top mounting ore x t e r n a l c a g e w i t hside/side or side/bottomconnections.
• Emission and immunitycompliance to EN50 081-2and EN50 082-2.
• 316 SS displacer andinternal trim on standardmodels.
Applications
• Feedwater heaters
• Scrubbers
• Receivers
• Knock-out drums
• Condensate drip pots
• Separators
• Flash tanks
• Boilers
Worldwide level and flow solutions
Digital ES Modulevelw/HART Communicationand Analog EZ ModulevelElectronic Liquid LevelTransmitters
Digital ES Modulevel Features
Digital ES and Analog EZ Modulevel Features
Principle of operation
Liquid level change acts upon the spring supported displac-er causing vertical motion of a core within a linear variabledifferential transformer (LVDT).
The enclosing tube acts as a static isolation barrier betweenthe LVDT and the process media. As the core positionchanges with liquid level, voltages are induced in the sec-ondary windings of the LVDT. These signals are processedin the electronic circuitry and used to control the current inthe 4–20 mA output current loop.
Digital ES transmitters add the HART output in addition tothe analog 4–20 mA DC output.
2
���
���
����
Electrical enclosure
Printed circuit boardLVDT
Moving LVDT core
External span and zero adjustment (EZ only)
Potted entrance seal
Terminal junction box
Enclosing tube
Range spring
External cage
Displacer
Hand held option
Digital ES models are compatible with Rosemount, Inc.model 275 D9E (English version) Hand Held Communicator.Programming must be added to the HART program for:
• Dry set point• Set specific gravity• 20 mA by percentage• Damping
Contact authorized HART distributors for ES programaddition. New hand held communicators which include theES program are available through Magnetrol.
HART Hand Held CommunicatorMagnetrol Part Number 089-5213-01X
3
Agency Approval Models
E5X-XXXX,XE5X-XXXX,
EXP/I/1/BCD,E6X-XXXX or XE6X-XXX withTransmitter Codes EZA, B, C,
DIP/II, III/1/EFG G, L, N, Q, R, S, T, U, V, 1, 2,FM ➂ 3, 4, 5, 6; ESA, G, 1, 4, Q, T
99-5031 – EntityE5X-XXXX or E6X-XXXX
IS/I, II, III/1/BCDEFGwith Transmitter Codes EZA, G, Q, T, 1, 4
Class I, Div. 1, EXX-XXXX with TransmitterGroups B, C, & D Codes EZA, B, C, G, L, N, Q,Class II, Div. 1 R, S, T, U, V, 1, 2, 3, 4, 5, 6Groups E, F, & G
Class I, Div. 1,
CSA ➂ Groups A, B, C, & DClass II, Div. 1,
Groups E, F, & G E6X-1XXX, E6X-2XXX,Intrinsically safe E6X-3XXX with Transmitterwhen connected Codes EZA, G, 1, 4through a certifiedsafety barrier rated≤28 V and ≥300 ohms
EEx d IIC T6EXX-XXX with transmitter codes EZA, B, C, G, L, N
EEx ib IIC T4EXX-XXX with transmitter codes EZD or EZI
Cenelec EEx ia IIC T6 (ambient EXX-XXX with transmitter temp. to 40° C/104° F) codes ESD,or ESIEEx ia IIC T5 (ambient
EXX-XXX with transmittertemperature to 55° C/codes ESD,or ESI131° F)
Agency approvals
➂ FM and CSA approvals pending on Digital ES models.
These units have been tested to EN 50081-2 and EN 50082-2 andare in compliance with the EMC Directive 89/336/EEC.
➀ These performance characteristics appy to units measuring liquid level; interface measurement units may not exhibit the same perfor-mance.
➁ Minimum temperatures given are for carbon steel construction materials. Complete stainless steel is recommended for lower processtemperatures. Maximum temperatures given are based on ambient temperatures up to +120° F (+49° C) on models without cooling fins.Higher ambient temperatures require reduced process temperatures.
Description Digital ES Modulevel Specifications Analog EZ Modulevel Specifications
12 to 36 VDCPower supply at transmitter 14 to 36 VDC 120 VAC, 50/60 Hz
240 VAC, 50/60 Hz
Specific gravity range 0.11 to 2.20 SGU
Process Connections NPT, socket weld or flanged
Output signal4–20 mA DC direct or reverse acting 4–20 mA DC field selectable HART compatible direct or reverse acting
Max. loop resistance 420 ohms at 24 VDC 480 ohms at 24 VDC
Level range Limited only to displaced length. From fractional inches to 10 feet (3 m) or more.
Materials of constructionChambers: carbon steel or 316/316L stainless steel
Wetted components: 304 or 316 stainless steel.
Linearity (independent) 0.25% of full span ➀
Repeatability 0.20% of full span ➀
Ambient temperature limits Electronics: -40° F to +160° F (-40° C to +71° C)
Process temperature ➁Steam applications: -20° F to +500° F (-29° C to +260° C)
Non-steam applications: -20° F to +600° F (-29° C to +315° C)
Temperature effect Maximum zero shift is 0.031% of range per ° F from 0° F to +160° F.
Specifications
Intrinsic safety
FM Entity parametersfor transmitter codes EZA,EZG, EZQ, EZT, EZ1, and EZ4
Vmax = 40 VImax = 150 mACi = 0.005 uFLi = 50 uH
CENELEC Entity parametersfor transmitter codes EZD and EZI
Ui = 28.4 V; li = 80 mA; Pi = 0.57 W; orUi = 28 V; li = 93 mA; Pi = 0.65 W; orUi = 26.4 V; li = 99 mA; Pi = 0.66 W; orUi = 25 V; li = 125 mA; Pi = 0.78 WCi = 0.05 uFLi = 200 uH
CENELEC Entity parametersfor transmitter codes ESD and ESI
Ui = 28.4 V; li = 94 mA; Pi = 0.67 WCi = 72 uFLi = 550 uH
Analog EZ Transmitter Codes
Use w/Specific Gravity and Temperature
Input PowerCodes 1, 2, or 3 and Maximum Temperatures
of 300° F (149° C) Steam; 400° F (204° C) Non-Steam
Blind w/Analog Meter
24 VDC EZA EZG
120 VAC EZB EZL
240 VAC EZC EZN
24 VDC EZD➆ EZI➆
Digital ES Transmitter Codes
24 VDC ESA ESG
24 VDC ESD➇ ESI➇
Integral mount
ProcessCode
Model Connection Head Flange RatingSize 150 lb. 300 lb. 600 lb.
3 inch G3 G4 G5
E61, E62 4 inch H3 H4 H5
6 inch K3 K4 K5
Liquid SpecificCode
Gravity Range 400° F max. 500° F max.➃ 600° F max.➃➄(204° C max.) (260° C max.) (315° C max.)
0.11–0.54 1 7 7
0.55–1.09 2 8 8
1.10–2.20 3 9 9
Liquid SpecificCode
Gravity Range 300° F max. 400° F max.➃ 500° F max.➃➄(149° C max.) (204° C max.) (260° C max.)
0.11–0.54 1 7 7
0.55–1.09 2 8 8
1.10–2.20 3 9 9
Part number construction
E - -
Mounting & materials of construction
Specific gravity & temperature
Tank connection
Level range
Transmitter electronics
Digital ES and Analog EZtransmitters are identifiedby a ten digit alphanumericpart numbering system. Thepart number describes model
configuration, materials ofconstruction, level ranges,and other options vital tothe performance and func-tion of the instrument.
Selection data
1
2
3
4
5
Analog EZ Transmitter Codes
Use with codes 1, 2, 3, 7, 8, Use with codes 7, 8, oror 9➈ and Maximum 9➉ and Maximum
InputTemperatures of 400° F Temperatures of 500° F
Power(204° C) Steam; (260° C) Steam;500° F (260° C) 600° F (315° C)
Non-Steam Non-Steam
Blind w/Analog Blind w/Analog Meter Meter
24 VDC EZ1 EZ4 EZQ EZT
120 VAC EZ2 EZ5 EZR EZU
240 VAC EZ3 EZ6 EZS EZV
Digital ES Transmitter Codes
24 VDC ES1 ES4 ESQ EST
Remote mount (connecting cable required)
Model MountingMaterials of Construction
Code Arrangement Cage or Flange Displacer && Head Assembly Internal Trim➀
E61Flanged Top➅
Fabricated Steel 316 SS
E62 Fabricated 316 SS 316 SS
E63Side/Bottom
Fabricated Steel 316 SS
E64 Fabricated 316 SS 316 SS
E65Side/Side
Fabricated Steel 316 SS
E66 Fabricated 316 SS 316 SS
C e r t a i n A n a l o g E ZModulevel instruments areavailable for quick shipment,usually within one weekafter factory receipt of apurchase order, through the
Expedite Ship Plan (ESP).Models covered by ESPservice are convenientlycolor coded in the selectiondata chart below. To takeadvantage of ESP, simply
match color coded modelnumber codes (standarddimensions apply). ESPservice may not apply toorders of ten units of more.Contact your local repre-
sentative for lead times onlarger volume orders aswell as other products andoptions.
Expedite ship plan
Non-steam applications➂
Steam applications➁
Tank connection3
Level range➅4
Mounting and materials of construction1
Transmitter electronics5
Specific gravity and temperature2
4
Standard Pressure Models
Connecting cableRequired for remote mount models. Refer to page 5.
CodeModel Process Connection ANSI Displacer Cage Rating
Size Type 150 lb. 300 lb. 600 lb.
NPT C5 C7 C9
E63 11⁄2" Socket weld C6 C8 C0
E64 Flanged P3 P4 P5
E65 NPT D5 D7 D9
E66 2" Socket weld D6 D8 D0
Flanged Q3 Q4 Q5
Code A B C D E F G H I
Inches 14 32 48 60 72 84 96 108 120
mm 356 813 1219 1524 1829 2134 2438 2743 3048
Dia. 2.0 1.25 1.0 .88 .88 .75 .75 .63 .63
mm 50 31 25 22 22 19 19 16 16
Code A B C D
Inches 14 32 48 60
mm 356 813 1219 1524
Dia. 2.0 1.25 1.0 .88
mm 50 31 25 22
Integral mount
5
Connecting cable(Required for remote mount models)
Applications which require a remote mount transmitter alsorequire a connecting cable. Lengths of 1 to 75 feet (30.5 cm to22.9 meters) are available in increments of 1 foot (30.5 cm). Toorder, specify one of the part numbers listed below with thelast two digits representing the cable length required.
For Transmitter codes ES1, ES4, EZ1, 2, 3, 4, 5, and 6specify remote cable 037-3218-0XX.
For Transmitter codes ESQ, EST, EZQ, R, S, T, U, and Vspecify remote cable 037-3220-0XX.
Calibration stand
For bench calibration of transmitter electronics
Part Number: 031-6107-001
Level range➅4
Tank connection3
Mounting and materials of construction1 Transmitter electronics5
Specific gravity and temperature2
Process Code
Model Connection Head Flange RatingSize 900 lb. 1500 lb. 2500 lb.
E514 inch H6 H7 H8
6 inch K6 K7 K8
Code
ModelProcess Connection
ANSI Displacer Cage RatingSize Type 900 lb. 1500 lb. 2500 lb.
NPT L5 L7 L9
11⁄2" Socket Weld M5 M7 M9
E53 Flanged P6 P7 P8
E55 NPT L6 L8 L0
2" Socket Weld M6 M8 M0
Flanged Q6 Q7 Q8
Non-steam applications➂
Liquid SpecificCode
Gravity Range 400° F max. 500° F max.➃ 600° F max.➃➄(204° C max.) (260° C max.) (315° C max.)
0.55–1.09 2 8 8
Steam applications➁
Liquid SpecificCode
Gravity Range 300° F max. 400° F max.➃ 500° F max.➃➄(149° C max.) (204° C max.) (260° C max.)
0.55–1.09 2 8 8
Materials of Construction
Model Mounting Cage or Flange Displacer andCode Arrangement and Internal
Head Assembly Trim➀
E51 Flanged Top➅ 304 SSE53 Side/Bottom Fabricated Steel and
E55 Side/Side 316 SS
High Pressure Models
➀ Inconel range spring.➁ Applications measuring water with steam and steam condensate
present at high temperatures.➂ Applications such as hydrocarbons. Do not use ES/EZ Modulevel
on Dowtherm® or other heat transfer media.➃ Codes 7, 8, and 9 are supplied with remote mount electronics
and two sets of cooling fins.➄ Codes 7, 8, and 9 for high temperature applications (500° F or
600° F) utilize transmitter codes ESQ, EST, EZQ, EZR, EZS, EZT,EZU, and EZV only.
➅ Top mounting models require an adjustable hanger, part number032-3110-001, when the span is more than 9.31" (236 mm) belowthe mounting flange.
➆ Transmitter codes with Cenelec approval EEx ib IIC T4.➇ Transmitter codes with Cenelec➈ Specify remote cable assembly 037-3218-0XX.➉ Specify remote cable assembly 037-3220-0XX.
Analog EZ Transmitter Codes
Use w/Specific Gravity and Temperature
Input PowerCode 2 Only and Maximum Temperatures of
300° F (149° C) Steam; 400° F (204° C) Non-Steam
Blind w/Analog Meter
24 VDC EZA EZG
120 VAC EZB EZL
240 VAC EZC EZN
24 VDC EZD➆ EZI➆
Digital ES Transmitter Codes
24 VDC ESA ESG
24 VDC ESD➇ ESI➇
Analog EZ Transmitter Codes
Use with codes 2 or 8➈ and Use with code 8➉ andMaximum Temperatures Maximum Temperatures
Inputof 400° F (204° C) of 500° F (260° C)
PowerSteam; Steam;
500° F (260° C) 600° F (315° C)Non-Steam Non-Steam
Blind w/Analog Blind w/Analog Meter Meter
24 VDC EZ1 EZ4 EZQ EZT
120 VAC EZ2 EZ5 EZR EZU
240 VAC EZ3 EZ6 EZS EZV
Digital ES Transmitter Codes
24 VDC ES1 ES4 ESQ EST
Remote mount (connecting cable required)
Range SpringSpecific Gravity A B C
Rating
.11 to .54 14.06 19.06 23.25(357) (484) (591)
.55 to 1.09 12.12 17.12 21.311.10 to 2.20 (308) (435) (541)
Dimensional specifications inches (mm)
Pipe D ESize
11⁄2" 3.19 5.94(81) (151)
2" 3.19 6.25(81) (159)
Integral Side/Side MountExternal Cage Model Codes 1, 2, 3
Remote Side/Bottom MountExternal Cage Model Codes 7, 8, 9
Integral Top Mount
10.75(273)
8.62(219)
± .19 (5)
Up to 75' (23 m)of Connecting Cable, part number037-3218-0XXor037-3220-0XX,run in conduit.Cable must beordered separately.
C
1" NPT
5.00(127)
3.00(76)
2.00 (51)
3.50 (89)2.37 (60)
E
D
Level rangeplus 3.50
(89)
Level rangeplus 8.62
(219)
Remote MountBracket
1/2" NPT
1" NPTconduitconnector
10.75(273)
8.62(219)
5.00 (127) Adjustable hangercable available(specify 032-3110-001)
A
Levelrange
1" NPTconduitconnector
10.75(273)
8.62(219)
E
D
Level range
3.00 (76)
1" NPT drain(plug not provided)
B
1" NPTconduitconnector
6
Standard pressure models: E61, E62, E63, E64, E65, E66
Dimensional specifications inches (mm)
E51 Series with Top Mounting900, 1500, 2500 lb. Class
Levelrange
Dia.
ANSI standard flangeC
10.75(273)
8.62(219)
B
1" NPTconduitconnector
E55 Series with Side/Side Connections900, 1500, 2500 lb. Class
1" NPTdrain
Levelrange
(2) 11/2" or 2"NPT or
socket weld
(2) 11/2" or 2"ANSI standard
flange(Refer to chart for
class and size.)
F
J
G
L
10.75(273)
8.62(219) 1" NPT
conduitconnector
MidRange
B
E
E53 Series with Upper Side/Bottom Connections900, 1500, 2500 lb. Class
10.75(273)
8.62(219)
B
Levelrange
(2) 11/2" or 2" ANSIstandard flange(Refer to chart forclass and size.)
MidRange
(2) 11/2" or 2" NPTor socket weld
G
MH
L
F
KE
1" NPTconduitconnector
Outlet SizeDimensions
L M
11⁄2" NPT or Socket Weld4.00 3.44(102) (87)
2" NPT or Socket Weld4.38 3.50(111) (89)
Flange Size Dimensions& Class B C E F G H J K
11⁄2" – 900 lb. 7.25 8.4316.38 9.31 25.81 9.31 (184) (214) 3.44 11.50(416) (236) (655) (236) 8.18 9.38 (87) (292)
2" – 900 lb. (208) (238)
11⁄2" – 1500 lb.7.93 9.13
16.75 9.31 26.19 9.31 (201) (232) 3.44 12.25(425) (236) (665) (236) 8.87 10.13 (87) (311)
2" – 1500 lb.(225) (257)
11⁄2" – 2500 lb. 9.06 10.2517.62 9.31 27.06 9.31 (230) (267) 3.44 14.00(447) (236) (687) (236) 9.87 11.13 (87) (356)
2" – 2500 lb.(251) (283)
7
High pressure models: E51, E53, E55
BULLETIN: 48-105.14EFFECTIVE: December 1997SUPERSEDES: August 1996
ASSURED QUALITY & SERVICE COST LESS
All Magnetrol electroniclevel and flow controls arewarranted free of defects inmaterials or workmanshipone full year from the dateof original factory ship-ment.
If returned within the war-ranty period; and, uponfactory inspection of the
control, the cause of theclaim is determined to becovered under the warranty;then, Magnetrol will repairor replace the control at nocost to the purchaser (orowner) other than trans-portation.
Magnetrol shall not beliable for misapplication,
labor claims, direct or con-sequential damage orexpense arising from theinstallation or use of equip-ment. There are no otherwarranties expressed orimplied, except specialwritten warranties coveringsome Magnetrol products.
The quality assurance sys-tem in place at Magnetrolguarantees the highestlevel of quality throughoutthe company. Magnetrol iscommitted to providing fullcustomer satisfaction both
in quality products andquality service.
Magnetrol's quality assur-ance system is registeredto ISO 9001 and Z299.1affirming its commitment to
known international qualitystandards providing thestrongest assurance ofproduct/service qualityavailable.
Quality assurance
5300 Belmont Road • Downers Grove, Illinois 60515-4499 • 630-969-4000 • Fax 630-969-9489 • Website http://www.magnetrol.com6291 Dorman Road • Mississauga, Ontario L4V-1H2 • 905-678-2720 • Fax 905-678-7407Heikenstraat 6 • B 9240 Zele, Belgium • 052 45.11.11 • Telex 25944 • Fax 052 45.09.93Regent Business Ctr., Jubilee Rd. • Burgess Hill, Sussex RH15 9TL U.K. • 01444-871313 • Fax 01444-871317
Product warranty
©1997 Magnetrol International, Incorporated. All rights reserved. Printed in the USA.Magnetrol and Magnetrol logotype are registered trademarks of Magnetrol International.
The brand and product names contained within this document are trademarks or registered trademarksof their respective holders.