DEPARTMENT OF BIOMEDICAL ENGINEERING. University of Wisconsin-Madison. Anna Karas Meghan Olson Blake Hondl Zachary Cohen. Client: Dr. Christopher Green, MD. Advisor: Mitch Tyler. Problem Statement. PTT is not calculated with the current Sleep Apnea Equipment (Flaga system) - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: DEPARTMENT  OF BIOMEDICAL ENGINEERING

DEPARTMENT OFBIOMEDICAL ENGINEERING

Advisor: Mitch TylerClient: Dr. Christopher Green, MD

University of Wisconsin-Madison

Anna KarasAnna KarasMeghan Olson Meghan Olson Blake HondlBlake HondlZachary CohenZachary Cohen

Page 2: DEPARTMENT  OF BIOMEDICAL ENGINEERING

• PTT is not calculated with the current Sleep Apnea Equipment (Flaga system)

• Design a solution which interfaces with the current system and is capable of detecting and measuring changes in pulse transit time which occur due to vasoconstriction during a micro arousal

Problem Statement

Page 3: DEPARTMENT  OF BIOMEDICAL ENGINEERING

• Must interface with existing Sleep Apnea Study Equipment– Cannot void current system’s warranty

• Accurately detect the ECG and Pulse Oximetry peaks– Must reject signals due to noise or motion artifact

• Calculate PTT• Correlate PTT with other sleep study

measurements

Design Constraints

Page 4: DEPARTMENT  OF BIOMEDICAL ENGINEERING

• Affects 12 million Americans• Obstructive Sleep Apnea

– Is caused by obstruction of the trachea (windpipe)

• From tongue, tonsils, uvula, fatty tissue in the throat, and/or relaxed muscles

– Individual can stop breathing for 10-30 sec.– Low blood oxygen level activates “flight or

fight” response • releases adrenaline• causes vasoconstriction

– Individual brought out of REM sleep– Can occur every 1-4 minutes (~400 times a

night)• Results in health problems and poor sleep

Sleep Apnea: Biological Aspect

Page 5: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Time will decrease with a micro arousal Measure from peak

Signal from ECG machine

Pulse Transit Time Theory

• Obstruction of breathing causes release of endorphins monitored by cerebral cortex

• The blood vessels constrict causing higher blood pressure (blood flows faster)

• The amount of time from heart pumping blood for the pressure wave to reach an extremity decreases

• PTT is correlated with micro arousals

Signal from Pulse Oximeter

t

Page 6: DEPARTMENT  OF BIOMEDICAL ENGINEERING

PTT Correlates to Micro Arousals

• Blood pressure rises imply arousals (through PTT) are as good as EEG micro-arousals at predicting subjective sleepiness

• PTT detects subcortical/autonomic arousals, currently undetected by EEG monitoring

• ΔPTT is a sensitive indicator of inspiratory effort variance due to obstructions

• PTT detects events of a respiratory obstruction, but cannot differentiate types of events

» v = K*√{V*ΔP/ΔV} «

v=Pulse Wave Velocity V= Initial Vessel Volume K= Constant ΔP= Pressure Delta ΔV= Vessel Volume Delta

Page 7: DEPARTMENT  OF BIOMEDICAL ENGINEERING

FLAGA / Somnologica System

• Incorporates a variety of sleep study signals: EOG, EEG, ECG, Pulse Oximeter, Respiratory CO2 , Noise/Vibration (snoring), Visual monitoring

– Signals can be sampled at 2-2000 Hz

• Somnologica program process signals and displays information in GUI format

• Data can be extracted in European Data Format (EDF)

Page 8: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design 1: Spaced Transducer Technique

• Original method to test PTT• Pressure sensors placed on

skin consecutively on an artery

• Sensors measure the pulse moving in the artery below

• Sensors placed known distance apart (d)

• Analog to Digital conversion of signal

• Calculate the time between when the first sensor feels the pulse and the second

• Real time PTT calculation• Store information for later

reference

Pressure Sensor

Pressure Sensor

Pressure Sensor

d

d

Page 9: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design 1: Spaced Transducer Technique

• Advantages– Simple design– Low cost– Independent of Flaga system

• Disadvantages– Noisy signals– Difficult to use with sleeping patients– Inaccurate compared with other techniques

Page 10: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design 2: Automated Ambulatory Blood Pressure

• Blood pressure can be measured and correlated with PTT

• Measure blood pressure with piezoelectric microphones• Measures systolic and diastolic blood pressure• Portable, lightweight, can run on batteries• Used from 24-48 hr. periods

– measurements taken at set intervals (~every 5-60 min.)

• Tiba Medical – ~$1800 for a unit– $225/month to rent– Connects to PC– 512 MB storage capability

Page 11: DEPARTMENT  OF BIOMEDICAL ENGINEERING

• Advantages– Lightweight– Portable– Interface with computer for

post analysis– Past studies show it does

not effect sleep of child– Another way to measure

micro arousals (correlated to blood pressure increase)

• Disadvantages– Costly– Dependent on cuff inflation– Data can only be taken at

intervals of 5- 60 min.– Not a continuous

measurement– Micro arousals may not be

detected

Design 2: Automated Ambulatory Blood Pressure

Page 12: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design 3: PTT Software Solution

• Interface with existing Flaga System• Stored data interfaces with Matlab• Programming

– Signal analysis• Rejection of motion artifact• Detection of QRS and Pulse Oximeter wave

– PTT calculation

• Correlation of PTT with collected data from Flaga– Via Epic Number

Page 13: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design 3: PTT Software Solution

• Advantages– Feasibile– Cost– Meets client requirements– Interfaces with existing system

• No additional signal measurements required

– Provides accurate PTT

• Disadvantages– Not real time

– Not portable

Page 14: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Design Matrix

Criteria Transducer Blood Pressure SoftwareCost 2 3 1

Feasibility (time restraints)

2 1 1

Satisfies Client Requirements

3 3 2

Ease of Use 2 2 1

Interfaces with Flaga system

3 3 1

Average Score 2.4 2.4 1.2

Good Good Best

Rated 1-3: 1=Best, 3=Worst

Page 15: DEPARTMENT  OF BIOMEDICAL ENGINEERING

Potential Problems/Future Work

• Obtaining real-time signal from FLAGA system to output PTT in real-time

• Motion artifacts are introduced into ECG, influencing PTT– Need processing algorithms to account for

this

Convert EDF to text file

Using Matlab develop algorithms

to detect wave peaks

Create rejection algorithm for

motion artifact

Calculate PTT & display

Page 16: DEPARTMENT  OF BIOMEDICAL ENGINEERING