Dedicated to successful NIV

Hospital respiratory care

V60 ventilator

Specifi cations

The print quality of this copy is not an accurate representation of the original.

Page 2: Dedicated to successful NIV - Microsoft€¦ · Dedicated to successful NIV Hospital respiratory care V60 ventilator Specifi cations The print quality of this copy is not an accurate

Specifi cations

3. Settings

Settings Range

C-Flex OFF, 1 – 3

CPAP 4 – 25cmH2O

EPAP 4 – 25cmH2O

IPAP 4 – 40cmH2O

I-time (inspiratory time) 0.30 – 3.00sec

Max P (AVAPS maximum IPAP) 6 – 40cmH2O

Min P (AVAPS minimum IPAP) 5 – 30cmH2O

O2 (oxygen percent) 21 – 100%

Ramp time OFF, 5 – 45min

Rate (respiratory rate) 4 – 60bpm

Rise (rise time) 1 – 5

Triggering and cycling Auto-adaptive (Auto-Trak)

AVAPS target tidal volume 200 – 2,000ml btps

Max E 0 – 100cmH2O/I

Max R 0 – 50cmH2O/l/s

PPV% 0 – 100%

Max P (PPV maximum pressure limit) 5 – 40cmH2O

Max V (PPV maximum volume limit) 200 – 3,500ml

The Philips Respironics V60 ventilator combines Respironics’ ventilation expertise with Philips focus on simplifying advanced health care. The result is exceptional noninvasive ventilation with an invasive ventilation fallback and an interactive display that helps simplify patient management.

* May not be available in all markets

1. Patient types

Adult

Pediatric (≥20kg)

2. Modes

Standard

CPAP (continuous positive airway pressure)

S/T (spontaneous with timed backup)

PCV (pressure control ventilation)

AVAPS (average volume assured pressure support)

Optional

PPV (proportional pressure ventilation)*

The print quality of this copy is not an accurate representation of the original.

Page 3: Dedicated to successful NIV - Microsoft€¦ · Dedicated to successful NIV Hospital respiratory care V60 ventilator Specifi cations The print quality of this copy is not an accurate

4. Modes with settings

CPAP S/T PCV AVAPS PPV

Rate

I-time

CPAP

EPAP

IPAP

Rise

Min P

Max P

Max V

Max E

Max R

PPV%

O2

VT (tidal volume)

C-Flex

Ramp time

5. Monitored parameters

Patient data window

Breath phase/trigger indicator Spont, timed, exhale

PIP 0 – 50cmH2O

Patient/total leak 0 – 200l/min btps

Patient trigger 0 – 100%

Respiratory rate 0 – 90bpm

Ti/Ttot 0 – 91%

Minute volume 0 – 99.0l/min btps

Tidal volume 0 – 3,500ml btps

Waveform window

Pressure waveform 0 – 50cmH2O

Flow waveform -240 – 240l/min btps

Volume waveform 0 – 3,500ml btps

6. Alarms

Alarm Adjustable range

Hi Rate (high respiratory rate alarm) 5 – 90bpm

Lo Rate (low respiratory rate alarm) 1 – 89bpm

Hi VT (high tidal volume alarm) 200 – 3,500ml

Lo VT (low tidal volume alarm) Off, 5 – 1,500ml

HIP (high inspiratory pressure alarm) 5 – 50cmH2O

LIP (low inspiratory pressure alarm) OFF, 1 – 40cmH2O

Lo VE (low minute ventilation alarm) OFF, 0.1 – 99l/min

LIP T (low inspiratory pressure delay time)

5 – 60sec

7. Other settings

Alarm volume 1 – 10 (relative scale)

Brightness 1 – 5 (relative scale)

Exhalation port selection • DEP (disposable exhalation port)

• Whisper Swivel• PEV (plateau exhalation

valve)• Other• None (no inline circuit

exhalation port)

Interface selection ET/Trach, 1, 2, 3, Other

Screen lock Off, On

Auto-Trak Plus Optional*

Trigger* Normal, 1 – 7

E-cycle* -2, -1, Normal, 1 – 6

8. Environmental

Temperature

Operating conditions +5 – +40ºC

Storage conditions -20 – +50ºC

Relative humidity

Operating conditions 15 – 95% (non-condensing)

Storage conditions 10 – 95% (non-condensing)

Barometric pressure

Operation and storage 79.9 – 101.1kPa(600 – 765mmHg)

Altitude

Operation and storage 600 to 765 mmHg (approximately -61 to 1951m (-200 to 6400 ft) relative to sea level)

* May not be available in all markets

The print quality of this copy is not an accurate representation of the original.

Page 4: Dedicated to successful NIV - Microsoft€¦ · Dedicated to successful NIV Hospital respiratory care V60 ventilator Specifi cations The print quality of this copy is not an accurate

© 2016 Koninklijke Philips N.V. All rights reserved.Specifications are subject to change without notice.

Bernoulli is a registered trademark of Cardiopulmonary Corp.DataCaptor is a trademark of CapsuleTech.CareAware, and iBusare trademarks of Cerner Corp.

www.philips.com

4522 991 22601 * OCT 2016

9. Communication

Philips IntelliBridge EC10

Philips IntelliBridge EC40/80

Philips VueLink Open Interface

Respi-Link remote diagnostic system

Bernoulli® management system

Capsule DataCaptorTM device interface driver

GE Healthcare (Centricty Critical Care)

Cerner CareAware® iBus™

Other monitoring and patient information systems

RS232 digital and analog

10. Electrical

External

AC voltage 100 – 240 VAC

AC frequency 50/60Hz

AC power 300 VA

Battery (optional)

Nominal voltage 14.4V

Capacity 11.0Ah

Battery chemistry Lithium-ion

Operating time 6 hours in normal conditions

12. Regulatory compliance

2nd edition standards

EN 60601-1-2 Electromagnetic Compatibility Requirements and Tests

EN 55011 Radiated and Conducted RF Disturbance Characteristics--Limits and Methods of Measurement (Level A)

EN 55014-1 Electromagnetic Compatibility Requirements. Part 1: Emissions

EN 61000-3-2 Limits for Harmonic Current Emissions

EN 61000-3-3 Limitation of Voltage Changes, Fluctuations, and Flicker Emissions

EN 61000-4-2 Electrostatic Discharge Immunity Test (8/15KV)

11. Physical

Weight 11.7kg (25.7lb) with optional battery10.6kg (23.3lb) without optional battery

Dimensions 33.7cm (13.3in) height39.4cm (15.5in) width42.9cm (16.5in) depth

Please visit www.philips.com/V60

EN 61000-4-3 Radiated Electromagnetic Field Immunity Test (10V/M)

EN 61000-4-4 Electrical Fast Transient/Burst Immunity Test

EN 61000-4-5 Surge Immunity Test

EN 61000-4-6 Immunity to Conducted RF Disturbances (10V)

EN 61000-4-8 Power Frequency Magnetic Field Immunity Test

EN 61000-4-11 Voltage Dips, Short Interruptions, and Voltage Variations Immunity Tests

MIL-STD 461E RE101

Electromagnetic Field Generation (Army Level)

ANSI/AAMI ES 60601-1

Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

WEEE recycling directive

Compliant with the WEEE recycling directive

3rd edition standards

IEC 60601-1; Ed. 3.1

Medical electrical equipment – Part 1: General requirements form basic safety and essential performance

IEC 60601-1-2; Ed. 3.0

Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances

IEC 60601-1-6; 2013

Medical electrical equipment – Part 1-6: General requirements for safety

IEC 60601-1-8; Ed. 2.1

Medical electrical equipment – Part 1-8: General requirements for safety

IEC 62366; 2007 + A1: 2004

Medical devices - Application of usability engineering to medical devices

ISO 14971; 2007

Medical devices – Application of risk management to medical devices

EN ISO 14971; 2012

Medical devices – Application of risk management to medical devices

ISO 80601-2-12; 2011

Medical electrical equipment – Particular requirements for basic safety and essential performance of critical care ventilators

ISO 60529; Ed. 2.1

Degrees of protection provided by enclosures (IPX1 @ 0° tilt)

IEC 62304; Ed. 1.0

Medical device software - Software life cycle processes

ANSI/AAMI ES 60601-1

Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

WEEE recycling directive

Compliant with the WEEE recycling directive