choosing the proper power for the iol brannon aden, md miles h. friedlander, md, facs

24
Choosing the Proper Choosing the Proper Power for the IOL Power for the IOL Brannon Aden, MD Brannon Aden, MD Miles H. Friedlander, MD, Miles H. Friedlander, MD, FACS FACS

Upload: brook-mcbride

Post on 23-Dec-2015

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Choosing the Proper Power Choosing the Proper Power for the IOLfor the IOL

Brannon Aden, MDBrannon Aden, MD

Miles H. Friedlander, MD, FACSMiles H. Friedlander, MD, FACS

Page 2: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Goal’s of Surgery Have Goal’s of Surgery Have Changed.Changed.

In past the goal was good In past the goal was good visual visual outcomeoutcome

Now an equal goal is a good Now an equal goal is a good refractiverefractive outcome outcome• Central to that is an accurate Central to that is an accurate

calculation of the correct IOL powercalculation of the correct IOL power• Next came a variety of formulas Next came a variety of formulas

aimed at achieving that accuracyaimed at achieving that accuracy

Page 3: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Possible Sources of Error Possible Sources of Error in IOL Calculationin IOL Calculation

Systematic error-weakness in formula or Systematic error-weakness in formula or weakness in a measurement techniqueweakness in a measurement technique• Example of technique is altering the axial Example of technique is altering the axial

length of the eye by using a contact type length of the eye by using a contact type probeprobe

Random errorRandom error• Not common but tend to produce larger Not common but tend to produce larger

errorserrors– Example is presence of a staphylomaExample is presence of a staphyloma

Page 4: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

FormulasFormulas

What is the current standard of What is the current standard of care for accuracy?care for accuracy?• 50% +/- 0.5D50% +/- 0.5D• 90% +/- 1.00D90% +/- 1.00D• 99.9% +/- 2.00D99.9% +/- 2.00D

Is this good enough for refractive Is this good enough for refractive lens surgery?lens surgery?

Page 5: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Factors Needed to Factors Needed to Calculate IOL PowerCalculate IOL Power

Axial length of globe (distance Axial length of globe (distance from anterior corneal vertex to from anterior corneal vertex to fovea)fovea)

Corneal powerCorneal power Location of lens in eye (related to Location of lens in eye (related to

anterior chamber depth)anterior chamber depth)

Page 6: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Axial LengthAxial Length

Most important anatomical Most important anatomical variablevariable

Greater deviation away from 22.5 Greater deviation away from 22.5 the greater the IOL power the greater the IOL power calculated especially with short calculated especially with short eyeseyes

Page 7: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Axial Length MeasurementAxial Length Measurement

ContactContact• Very personal dependent Very personal dependent • Average error +/- .2 mm ( .50D)Average error +/- .2 mm ( .50D)

ImmersionImmersion• Technician unfriendlyTechnician unfriendly• Accurate +/- .1 mmAccurate +/- .1 mm

Page 8: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Contact ApplanationContact Applanation

Page 9: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Immersion ScanImmersion Scan

Page 10: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Measurement ContinuedMeasurement Continued

Buzard “Touch and Go”Buzard “Touch and Go”• Table mounted A-scanTable mounted A-scan• Flood eye with tearsFlood eye with tears• Advance probe toward eye until Advance probe toward eye until

retinal spike produced on oscilloscoperetinal spike produced on oscilloscope• Requires skilled and experienced Requires skilled and experienced

examinerexaminer

Page 11: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

IOL Master (Humphrey and IOL Master (Humphrey and Zeiss)Zeiss)

Uses optical interference (Partial Uses optical interference (Partial Coherence Interferometry) to Coherence Interferometry) to measure axial lengthmeasure axial length

Keratometry also performed by Keratometry also performed by machinemachine

Page 12: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

IOL MasterIOL Master

Page 13: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Corneal CurvatureCorneal Curvature

Error of 0.1 mm = 1 Diopter errorError of 0.1 mm = 1 Diopter error Sources of errorSources of error

• Contact lens wareContact lens ware• Refractive surgeryRefractive surgery

Page 14: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Anterior Chamber DepthAnterior Chamber Depth

Now refers to final position of IOL or Now refers to final position of IOL or the distance from the posterior the distance from the posterior vertex of the cornea to the anterior vertex of the cornea to the anterior surface of the IOLsurface of the IOL

ACD shallows 0.1 mm per decade ACD shallows 0.1 mm per decade because of lens growthbecause of lens growth

In myopia deepens 0.06 mm per 1 DIn myopia deepens 0.06 mm per 1 D Of less importance than pastOf less importance than past

Page 15: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Early Formulas (First Early Formulas (First Generation)Generation)

Anterior chamber depth (ACD) was Anterior chamber depth (ACD) was constant valueconstant value

Early lenses were iris supported which Early lenses were iris supported which produced small variations in Post Op produced small variations in Post Op ACDACD

Later with the introduction of PC IOL’s Later with the introduction of PC IOL’s formula was less accurateformula was less accurate• Difference of in the bag vs. sulcus was 1 Difference of in the bag vs. sulcus was 1

mm therefore 1 Dmm therefore 1 D

Page 16: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Next First Generation Next First Generation Regression Formula (SRK Regression Formula (SRK 1)1)

Used multiple regression analysisUsed multiple regression analysis Eliminated ACD variable and Eliminated ACD variable and

replaced it with A-constantreplaced it with A-constant• Given by manufacturer and is based Given by manufacturer and is based

on expected position in eye, haptic on expected position in eye, haptic and optic design, and refractive index and optic design, and refractive index of IOL materialof IOL material

Page 17: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Problems With SRK 1 Problems With SRK 1 FormulaFormula

Formula assumes 2.5 D refractive Formula assumes 2.5 D refractive change for each 1 mm of axial change for each 1 mm of axial length regardless the axial length length regardless the axial length of the globeof the globe

Tended to under estimate IOL Tended to under estimate IOL power in globes 25 to 29 mm longpower in globes 25 to 29 mm long

Page 18: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Second Generation Second Generation Regression FormulasRegression Formulas

SRK II recognized the non linear SRK II recognized the non linear relationship between axial length relationship between axial length and IOL powerand IOL power

Binkhorst II, Holladay, Donzis also Binkhorst II, Holladay, Donzis also addressed same problems addressed same problems

Page 19: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Third Generation FormulasThird Generation Formulas

Holladay 2, SRK/T, and HofferQHolladay 2, SRK/T, and HofferQ Normal range of 22.0 mm to 24.5 Normal range of 22.0 mm to 24.5

mm- All three do equally wellmm- All three do equally well Short eyes < 22.0 mm Hoffer Q Short eyes < 22.0 mm Hoffer Q

performed bestperformed best Long eyes (24.5 to26 mm) Holladay Long eyes (24.5 to26 mm) Holladay

formulaformula Very long eyes (>26 mm) SRK/TVery long eyes (>26 mm) SRK/T

Page 20: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

IOL Design and MaterialsIOL Design and Materials

Majority of lenses are convex-Majority of lenses are convex-plano, biconvex, or plano-convexplano, biconvex, or plano-convex

Vitreous pressure, haptic flexibility, Vitreous pressure, haptic flexibility, and final position of ccc by and final position of ccc by contraction of the lens capsule contraction of the lens capsule effect final refractive erroreffect final refractive error

Page 21: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Choice of Lens MaterialsChoice of Lens Materials

In normal, non allergic, disease free eye either In normal, non allergic, disease free eye either PMMA , silicone, or acrylic okPMMA , silicone, or acrylic ok

Eyes with silicone oil or anticipated vitro-Eyes with silicone oil or anticipated vitro-retinal surgery need heparin surface-modified retinal surgery need heparin surface-modified 100% PMMA -tend to retard adhesion of 100% PMMA -tend to retard adhesion of silicone oil to lenssilicone oil to lens

Uveitis- use heparin surface-modified lensesUveitis- use heparin surface-modified lenses Posterior capsule opacification - Prevent? with Posterior capsule opacification - Prevent? with

acrylic lenses (stick to pc and stop proliferation acrylic lenses (stick to pc and stop proliferation of epithelial cells)of epithelial cells)

Page 22: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Lens PositionLens Position

Plus lens- need more power as Plus lens- need more power as approach the retinaapproach the retina

Minus lens- need less power as Minus lens- need less power as approach the retinaapproach the retina

.Anterior iris plane, sulcus, capsule bag. .Anterior iris plane, sulcus, capsule bag. • For every 1 mm of displacement- 1 D of For every 1 mm of displacement- 1 D of

corrective changecorrective change• Example If a capsular bag lens is placed in Example If a capsular bag lens is placed in

the sulcus then the power is reduced by 1 Dthe sulcus then the power is reduced by 1 D

Page 23: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Good ScanGood Scan

Page 24: Choosing the Proper Power for the IOL Brannon Aden, MD Miles H. Friedlander, MD, FACS

Poor ScanPoor Scan