chapter - 2 simultaneous determination of...
TRANSCRIPT
34
CHAPTER - 2
SIMULTANEOUS DETERMINATION OF ORLISTAT AND SIBUTRAMINE
HYDROCHLORIDE RELATED SUBSTANCES IN APSULES
FORMULATION AND IDENTIFICATION AND CHARACTERIZATION OF
IMPURITIES
2.1 OBJECTIVE:
To develop simultaneous determination method of orlistat and
sibutramine hydrochloride related substances in capsules Formulation,
Identification of impurities and Characterization of sibutramine
hydrochloride impurities.
2.2 INTRODUCTION:
Orlistat [135] is known as tetrahydrolipstatin, a drug prepared to
treat obesity [136]. It prevents absorption of fats from the human diet
and it is primary function, thereby reducing caloric intake. It is preferred
to use with a physician-supervised reduced calorie diet. Orlistat is a
potent natural inhibitor of pancreatic lipases isolated from the bacterium
streptomyces toxytricini and the saturated by-product of lipstatin [137].
However, due to stabilisation, orlistat rather than lipstatin was developed
as an anti-obesity drug [138]. The primary route of removal is through
the feces. It obstructs approximately 30% of dietary fat from being
absorbed [139] by over-the-counter dose of 60 mg [140, 141]. Higher
doses do not produce more strong effects [142]. Weight loss may increase
the risk of orthopedic problems and mineral loss. Therefore, inhibition of
35
dietary fat during absorption and result to inhibit intestinal lipase
activity during orlistat treatment could have a severe effect on dietary
mineral absorption and balance due to formation of insoluble mineral
soaps within the intestine [143].
Sibutramine hydrochloride monohydrate administered for the
treatment of obesity, as a hunger suppressant. It is a serotonin-
norepinephrine centrally-acting reuptake inhibitor chemical orally
structure related to amphetamine [144], although its mechanism of
action is clear [145- 148].
2.3 LITERATURE REVIEW:
Bennett P. et al. were reported a rapid, sensitive and specific
analytical method for orlistat determination in human plasma. In this
method reconstituted extracts of orlistat were analyzed by HPLC carried
out by using a 50 x 2mm i.d., 3 μm size of the particle, phenyl column.
The mobile phase used was acetonitrile and 2 mM ammonium acetate in
the ratio of 90:10, ion spray tandem mass spectrometry detector was
used for detection. The calibration graphs were found linear in the range
0.20 to 10 ng/mL [149].
Wieboldt R. et al. were reported the use of a quadrupolar ion trap for
quantitation in a bioanalytical laboratory. The orlistst evaluation was
completeded with the cross-validation of an LC-MS-MS quantitative
method first validated on a triple quadrupole mass spectrometer [150].
36
Mohammadi A. et al. were stated a stability-indicating high performance
liquid chromatographic method for the determination of orlistat in
capsules. In this method an isocratic separation was got by using a
Perfectsil ODS-3, 250 mm × 4.6 mm i.d., 5 μm particle size columns with
a flow rate of 0.7 ml/min by using a UV detector to monitor at 210 nm.
The mobile phase contains mixture of
methanol:acetonitrile:trifluoroacetic acid in the ratio of 82.5:17.5:0.01,
v/v/v respectively. Separation was achieved for all degradation products
and parent compound in an overall analytical run time of approximately
15 min [151].
Hassan K. A. et al were reported a HPLC method with UV detection for
determination of orlistat, a Nova- Pack C18 column and an isocratic
mobile phase of 0.1% phosphoric acid solution and acetonitrile in the
ratio of 10:90, v/v used respectively. Detection was carried out at 205
nm. Orlistat was eluted at 6 min without interfering peaks from placebo.
The method was found linear over the range of 10-160-μg/ml orlistat
[152].
Hakala K.S. et al were reported liquid chromatography with ion trap
mass spectrometry method used for the identifition and structural
characterization of by-products of the anti-obesity drug sibutramine.
Metabolites and their isomers formed via dehydrogenation,
hydroxylation, and demethylation, addition of CO2, acetylation and
glucuronidation were observed in MS experiments. However, evident
37
enantioselective formation was identified for two hydroxyl by-products
and two glucuronide conjugates, showing that the hydroxyl/glucuronic
acid moiety in those structures [153].
Radhakrishna T. et al were reported two isocratic liquid
chromatography methods for the purity standardisation and quantitative
estimation of sibutramine HCl, using 4-chloro aniline and lovastatin as
internal standards, respectively. The ratio of variances was close to unity,
correlation coefficient was more than 0.999. The accuracy estimated as
relative mean error for the intra-day assay is ±1.7%. The enantiomeric
separation of sibutramine by chiral chromatography method was
elaborated and this method was shown capable of partetioning the two
enantiomers with a selection of 1.4 and a resolution of 4.0. These two
methods are most useful in the quality control and found to be stability
indicaters [154].
Segall A.I. et al were reported the estimation of sibutramine
hydrochloride in the presence of its oxidatively-induced degradation
products by reversed-phase HPLC. The above method was validated with
respect to stability-indicating method parameters by forced degradation.
The chromatographic conditions includes C18 column and mobile phase
used was mixture of methanol, water, triethylamine (80:20:0.3) and the
pH was adjusted to 4.5 with 85% phosphoric acid, and operated at the
flow rate of 1.1 mL per minute. The eluent was monitored at 225 nm
[155].
38
Singh A.K. et al were reported UV-derivative spectrophotometry and
HPLC methods and compared for the estimation of sibutramine in drug
products. The UV-derivative spectrophotometry and HPLC methods were
found to be statistically exact and precise, there was no variation
observed between the proposed UV-derivative spectrophotometry and
HPLC methods. An alpha-1-acid glycoprotein column used for the
enantiomeric separation of sibutramine. The R- and S-sibutramine were
eluted in less than 5 min with baseline separation of the
chromatographic peaks (alpha = 1.9 and resolution = 1.9) [156].
Chen J. et al were reported a sensitive and selective method for the
estimation of the active primary amine metabolite of sibutramine, N- di-
desmethylsibutramine in human plasma, based on HPLC–MS–MS. ODS
column was used for chromatographic separation using a mobile phase
contains acetonitrile–0.1% trifluoroacetic acid (55:45, v/v) at a flow- rate
of 0.3 ml/min. Multiple reaction monitoring using precursor to product
ion combinations at m/z 252 [157].
Chandorkar J.G. et al were reported a rapid reproducible and specific
reverse phase HPLC method for the determination of Sibutramine and it‘s
impurities in Bulk as well as formulation. The analyte was separated by
using an ODS, C8, RP Column, 4.6 x 250 mm, 5 μm, mobile phase used
was sodium dihydrogen phosphate and acetonitrile at the flow rate of 1.0
mL/min, at wavelength of 230 nm . HPLC system used was Jasco
39
make, UV visible detector of model UV 1575 & Jasco make HPLC pump
of model PU 1580 [158].
Katarzyna M.S. et al were reported quality studies of
sibutramine using HPLC-ED, HPLC-MS/MS and X- ray powder
diffraction (XRPD) methods. Especially the XRPD assisted with an optical
microscopy seems to be useful as a fast screening method of general
sample composition of such preparations. First of all it can discriminate
between capsules containing pure herbal materials and those with some
chemical additives. In the case of mixtures of different chemical additives
further studies such as HPLC-ED and HPLC-MS/MS are helpful and
experience proved that the most often used additive in different herbal
preparations is sibutramine [159].
Young U.S. et al were reported a simple HPLC method for estimation
of active by-products of sibutramine, N- mono-desmethyl metabolite and
N-di-desmethyl metabolite in the rat administred serum sibutramine
HCl. Rat serum was injected directly onto the precolumn. Mobile phase
employed was mixture of methanol, acetonitrile, 20 mM ammonium
phosphate buffer pH 6.0 with phosphoric acid in the ratio 8.3:4.5:87.2
v/v/v and detected at 223 nm [160].
Ding L. et al were reported a sensitive LC-MS method for
simultaneous estimation of sibutramine and its by-products by HPLC on
a reversed phase C18 column with a mobile phase comprising of 10 mM
40
ammonium acetate buffer adjusted to pH 3.5 with acetic acid and
methanol in the ratio 25:75 v/v. These analytes were determined by
using electrospray ionization in a single quadrupole mass spectrometer.
Selected-ion monitoring mode to target ions at m/z: 280 for sibutramine
[161].
2.4 THEORETICAL ANALYSIS
2.4.1 Sample Information [162]
Orlistat is a molecule having the chemical name as [(1S)-1-[(2S,3S)-3-
hexyl-4-oxo-oxetan-2-yl]methyl] dodecyl] (2S)-2-formamido-4-methyl-
pentanoate, molecular formulae of C29H53NO5 and reported molecular
weight 495.735 g/mol. Orlistat is practically insoluble in water, freely
soluble in ethanol (96%) and acetonitrile organic polar solvents, organic
non-polar solvents, and in water. Orlistat is non-ionic basic compound
contains chiral centers and with a negative optical rotation in ethanol at
529 nm and a single diastereomeric molecule. Orlistat pKa and log P are
6.5 and 7.61 respectively.
Fig. 2.01
Orlistat Structural Formula
41
Sibutramine [163] is a molecule having the chemical name as 1-[1-(4-
chlorophenyl) cyclobutyl]- N,N,3- trimethylbutan- 1-amine and having a
molecular formulae of C17H26ClN and reported molecular weight 279.85
g/mol. Sibutramine hydrochloride is soluble in acetone, acetonitrile and
sparingly soluble in methanol. Sibutramine hydrochloride is ionic basic
compound. The pH, pKa and log p values for Sibutramine hydrochloride
are 8.5, 5.2 and 5.05 respectively.
42
Fig. 2.02
Sibutramine Hydrochloride Monohydrate Structural Formula
Sample selected was in solid state and complex mixture of
sibutramine hydrochloride, orlistat and other excipients. Both active
ingredients are in capsules formulation and which are soluble in
acetonitrile and other excipients are not soluble in acetonitrile, so
acetonitrile and water combination as a solvent can be selected for
extraction of these drugs from capsule formulation. Sibutramine HCl and
orlistat are having UV absorption.
In summary, reverse phase chromatographic separation is suitable for
method development. Columns suitable for sibutramine hydrochloride
and orlistst compounds in reversed phase chromatography are C18, C8,
phenyl, and cyano.
43
Table 2.01
Initial HPLC Method Development conditions for Orlistat and
Sibutramine HCl
Separation Variable Preferred Initial choice
Column
Dimensions(length,ID) 15x0.46 cm
Particle size 5 μma
Stationary phase C8 or C18
Mobile Phase
Solvents A and B Buffer-acetonitrile
% B 80-100%b
Buffer(compound, pH,
concentration)
25 mM potassium
phosphate, 2.0<pH<3.0c
Additives(e.g.,amine modifiers, ion-pair reagents)
Not recommended initially
Flow rate 1.0-2.0 ml/min
Temperature 25 °C
Sample size 20 μL
2.5 EXPERIMENTAL INVESTIGATIONS
44
2.5.1 Experiment No. 1
Potassium dihydrogen phosphate GR grade, Orthophosphoric acid GR
grade, Acetonitrile HPLC grade, Water milli Q, Sibutramine hydrochloride
WS and Orlistat WS chemicals and reagents were used for experiments.
The reference of sibutramine hydrochloride was kindly supplied by
Hetero Drugs Limited (Hyderabad, India) and capsules® was obtained in-
house sample. Capsules are claimed to contain 10 mg of sibutramine
hydrochloride, 120 mg of orlistat and following inactive ingredients. All
reagents were analytical or HPLC grade. Acetonitrile and methanol were
procured from Mecrk (Mumbai, India), the phosphoric acid from Merck
(Mumbai, India). Purified water was obtained by a Millipore® Dire-Q 3 UV
with pump (Molsheim, France)
The HPLC system (Agilent 1200 series, Santa Clara, USA) consisted of
a G1311A quaternary pump, G1322 vacuum degasser, G1316A
thermostat column compartment, G1329A standard auto sampler and
G1315B diode array detector set at 210 nm. The mobile phase was
acetonitrile and a solution of 50 mM potassium dihydrogen phosphate
buffer adjusted pH to 3.0 with 10% solution of phosphoric acid, (60:40;
v/v). Mobile phase was filtered through 0.45 μ membrane filter. The
analytical column, an inertsil ODS-3V, 250 mm x 4.6 mm, 5 µ, was
operated in ambient temperature (25° C). Washed the column with milli-
45
Q water and methanol for few minutes then saturated the column with
the mobile phase for 30-40 min.
Mobile phase flow rate was seted at 1.5 ml/min. Standard solution for
Sibutramine hydrochloride and orlistat solution was prepared having
concentration of 1200 μg/mL of Orlistat and 100 μg/mL of Sibutramine
respectively in mobile phase. Injected 20 μL standard solutions two times
and average detector response measured at 210 nm. Chromatograms
evaluated with respect to retention time, resolution and peak shape.
Only one peak was eluted at 5 min and second component peak was
not eluted upto 40 min. The eluted peak was confirmed as sibutramine
hydrochloride. Orlistat peak was not eluted in these chromatographic
conditions.
2.5.2 Experiment No.2
The mobile phase was acetonitrile and a solution of 50 mM
potassium dihydrogen phosphate buffer adjusted pH to 3.0 with 10%
solution of phosphoric acid, (90:10; v/v). Mobile phase was filtered
through 0.45 μ membrane filter. The analytical column, inertsil ODS-3V,
250 mm x 4.6 mm, 5 µ, maintained at 25 °C temperature. The mobile
phase flow rate was maintained at 1.5 ml/min. Standard Sibutramine
hydrochloride and orlistat solution was prepared at concentration, 1200
μg/mL of Orlistat and100 μg/mL of Sibutramine hydrochloride in mobile
phase. 20 μL standard solutions were injected two times and average
46
detector response taken at 210 nm. Chromatograms evaluated according
to retention time, resolution and peak shape.
Sibutramine eluted at 1.5 minutes and orlistat which were not eluted
up to 30 minutes and the next experiment carried out with methanol and
buffer as mobile phase.
2.5.3 Experiment No.3
The mobile phase was acetonitrile and a solution of 50 mM potassium
dihydrogen phosphate buffer set at pH to 3.0 with 10% solution of
phosphoric acid, (90:10; v/v). Mobile phase was filtered through 0.45 μ
membrane filter. The analytical column, inertsil ODS-3V, 250 mm x 4.6
mm, 5 µ, maintained at temperature 25 °C. The mobile phase flow rate
was maintained at 1.5 ml/min. Standard Sibutramine hydrochloride and
orlistat solution was prepared at concentration, 1200 μg/mL of orlistat
and 100 μg/mL of sibutramine hydrochloride in mobile phase. 20 μL
standard solutions were injected two times and average detector
response measured at 210 nm. Chromatograms evaluated with respect to
retention time, resolution and peak shape.
Sibutramine eluted at 1.5 minutes orlistat which was not eluted up
to 50 minutes. Hence it is concluded that the acetonitrile is favorable for
elution of orlistat rather than methanol, gradient elution required or
column polarity can be decreased.
47
2.5.4 Experiment No.4
The mobile phase was acetonitrile and a solution of 50 mM potassium
dihydrogen phosphate buffer adjusted pH to 3.0 with 10% solution of
phosphoric acid, (50:50; v/v). Mobile phase was filtered through 0.45 μ
membrane filter. The analytical column, Zorbax SB phenyl, 250 mm x
4.6 mm, 5 µ, maintained at temperature 25 °C. The mobile phase flow
rate was maintained at 1.5 mL/min. Standard Sibutramine
hydrochloride and orlistat solution was prepared at concentration, 1200
μg/mL of orlistat and 100 μg/mL of sibutramine hydrochloride in mobile
phase. 20 μL standard solutions were injected two times and average
detector response measured at 210 nm. Chromatograms evaluated with
respect to retention time, resolution and peak shape.
Sibutramine eluted at 15 minutes and orlistat not eluted up to 25
minutes peak shape obtained was not proper, hence buffer changed.
2.5.5 Experiment No.5
The mobile phase was acetonitrile and a solution of 50 mM sodium
perchlorate buffer adjusted pH to 2.1 with 10% solution of perchloric
acid. Buffer was filtered through 0.45 μ filter and gradient program
applied as given in below table. The analytical column, a Zorbax SB
phenyl, 250 mm x 4.6 mm, 5 µ, maintained at temperature 25 °C. The
mobile phase flow rate was maintained at 1.5 ml/min. Standard
Sibutramine hydrochloride and orlistat solution was prepared at
48
concentration, 1200 μg/mL of orlistat and 100 μg/mL of sibutramine
hydrochloride in mobile phase. 20 μL standard solutions were injected
two times and average detector response measured at 210 nm.
Chromatograms evaluated according to retention time, resolution and
peak shape.
Table 2.02
Gradient Program for Experiment No.5
Sibutramine eluted in 11.4 minutes and orlistat not eluted up to 40
minutes, peak shape was proper and the forced degradation solutions
shown co-elution, hence tuning in gradient programme is required.
2.5.6 Experiment No.6
The mobile phase was acetonitrile and a solution of 50 mM sodium
perchlorate buffer adjusted pH to 2.1 with 10% solution of perchloric
Time (min) Buffer (%) Acetonitrile (%)
0 70 30
24 70 30
35 30 70
45 15 85
60 15 85
60 70 30
49
acid. Buffer was filtered through 0.45 μ membrane filter and applied
gradient program as given in below table. The analytical column, a
Zorbax SB phenyl, 250 mm x 4.6 mm, 5 µ, maintained at temperature 25
°C. The flow rate of mobile phase was maintained at 2.0 ml/min.
Standard Sibutramine hydrochloride and orlistat solution was prepared
at concentration, 1200 μg/mL of orlistat and 100 μg/mL of sibutramine
hydrochloride in mobile phase. 20 μL standard solutions were injected
two times and average detector response measured at 210 nm.
Chromatograms evaluated with respect to retention time, resolution and
peak shape.
Table 2.03
Gradient program for Experiment No.6
Time (min) Buffer (%) Acetonitrile (%)
0 60 40
20 40 60
25 20 80
35 20 80
40 60 40
45 60 40
Sibutramine hydrochloride and orlistat were eluted in 40 minutes
peak shape were proper and the forced degradation solutions shown a
well separation between the peaks.
50
The use of water instead of a solution of triethylamine and phosphoric
acid in the diluent mixture was verified and no modification in the
chromatographic profile or in the measured areas was observed.
2.5.7 Experiment No. 7
Applied Instrumental conditions given in experiment VI and injected
Orlistat known impurities. Impurities obtained at RT 3.793, 6.346,
24.417 and 33.287 were identified with known impurities spiking
method.
Impurities obtained at RT 6.644, 10.490 and 18.299 were isolated by
using preparative chromatographic technique and characterized these
isolated substance by using Mass spectroscopy, Neuclear magnetic
spectroscopy, IR, CHNS analyser and identified these impurities by
spiking method.
2.5.8 Experiment No.8 (Method Validation)
Specificity:
51
Diluent, standard solutions and sample solution of Sibutramine and
Orlistat capsules were prepared and injected into the HPLC as per
methology given in Experimental Results by using a photodiode detector.
A placebo solution of Sibutramine and Orlistat capsules was prepared
and injected into the HPLC along with selectivity solution as per
methology given Experimental results by using a photodiode array
detector.
The accelerated degradation conditions applied were: oxidant, acid
basic and light media. Sample were analysed against a freshly prepared
control sample. The peak purity was estimated using the tools of the
Waters software. Excipient solutions were used to the same degradation
conditions in order to demonstrate no interference. Specific details of the
experimental conditions for forced degradation are described below:
Effect of UV light:
1 ml of a solution of 1 mg/mL of sibutramine and 12 mg/mL of
orlistat in methanol was placed in a closed 1 cm quartz cell. The cells
were exposed to a UV chamber 100 x 18 x 17 cm with internal mirrors
and UV fluorescent lamp CRS F30W T8 emitting radiation at 254 nm for
15, 30, 60, 120 and 180 minutes. The same procedure was realized for
preparation for LC analysed samples, protected in aluminum foil (in
order to perotect from light) were submitted to identical conditions and
used as control. After the degradation treatment, the samples were
52
diluted to 100 μg/ml of sibutramine and 1200 μg/ml of orlistat with a
mixture of acetonitrile:methanol:water (6:3:1; v/v/v) and analysed
immediately.
Effect of Temperature (60 °C/24 h):
5 ml of a solution containing 1 mg/mL of sibutramine and 12 mg/mL
of orlistat in methanol was placed in a 10 ml volumetric flask at 60
°C/24 h for 24 h. After the degradation treatment, the samples were
diluted to 100 μg/ml of sibutramine and 1200 μg/ml of orlistat with a
mixture of acetonitrile:methanol:water (6:3:1; v/v/v) and analysed
immediately.
Effect of Humidity (25 °C/92% RH for 24 h):
5 ml of a solution containing 1 mg/mL of sibutramine and 12 mg/mL
of orlistat in methanol was placed in a 10 ml volumetric flask at 25
°C/92% RH for 24 h. After the degradation treatment, the samples were
diluted to 100 μg/ml of sibutramine and 1200 μg/ml of orlistat with a
mixture of acetonitrile: methanol: water (6:3:1; v/v/v) and analysed
immediately.
Effect of Oxidation:
Sibutramine and orlistat standards were dissolved in methanol (1
mg/ml of sibutramine amd12 mg/ml of orlistat) and 5 ml of this solution
53
was transferred to a volumetric flask, where hydrogen peroxide solution
(30%) was added until the final concentration of 10 %. The volume was
completed with methanol. After 20 hours the solution was diluted until
the final concentrations 100 μg/mL of sibutramine and 1200 μg/mL of
orlistst, filtered and analysed. Similar procedure was realized for the
capsules, when 25 ml of the initial solution 100 μg/ml of sibutramine
hydrochloride and 1200 μg/mL orlistat, obtained as described in sample
preparation for LC analysis, were transferred to a volumetric flask and
submitted to degradation. A control solution containing the excipients
was prepared under the same circumstances of the capsules.
Effect of Acid Hydrolysis:
5 ml of 1 mg/mL of sibutramine and 12 mg/mL of orlistat reference
standard solution was transferred to a volumetricflask and HCl was
added until the final concentration of 1M HCl. After 5 hours and 1 and 6
days, one aliquot of the solution was neutralized with NaOH 1M and
diluted with acetonitrile and water (40:60, v/v) until the final
concentration of 100 μg/ml of sibutramine and 1200 μg/ml of orlistat for
LC analysis. Similar procedure was realized with the capsules, when 25
ml of the initial solution 1 mg/mL of sibutramine and 12 mg/mL of
orlistat (obtained as described in sample preparation for LC analysis)
were transferred to a volumetric flask and submitted to the degradation
and diluted with acetonitrile and water (40:60, v/v) until the final
54
concentration of 100 μg/ml of sibutramine and 1200 μg/ml of orlistat for
LC analysis. A control solution containing the excipients was prepared
under the same circumstances of the capsules.
Effect of Alkaline Hydrolysis:
5 ml of 1 mg/mL of sibutramine and 12 mg/mL of orlistat reference
standard solution was transferred to a volumetric flask and NaOH
(alkaline degradation) was added until the final concentration of 1M
NaOH. After 5 hours and 1 and 6 days, one aliquot of the solution was
neutralized with HCl 1M and diluted with acetonitrile and water (40:60,
v/v) until the final concentration of 100 μg/ml of sibutramine and 1200
μg/ml of orlistat for LC analysis. Similar procedure was realized with the
capsule, when 25 ml of the initial solution 1 mg/mL of sibutramine and
12 mg/mL of orlistat (obtained as described in sample preparation for LC
analysis) were transferred to a volumetric flask and submitted to the
degradation and diluted with acetonitrile and water (40:60, v/v) until the
final concentration of 100 μg/ml of sibutramine and 1200 μg/ml of
orlistat for LC analysis. A control solution containing the excipients was
prepared under the same circumstances of the capsules.
LOD and LOQ:
55
The qualification and detection limits were obtained based on signal-
to-noise approach. The background noise was obtained after injection of
the blank, observed over a distance equal to 20 times the width at half-
height of the peak in a chromatogram obtained by the injection 2 μg/mL
of each reference standards. The signal-to-noise ratio applied was 10:1
for the LOQ and 3:1 for LOD. The results were verified experimentally.
Based on the determination of signal to noise ratio and visual
observation for known impurities, six replicate injections were made for
LOD &LOQ.
Linearity and Range:
To test linearity, standard plots were construted with six
concentrations in the range 0.13-15.35 μg/mL of 4-chlorophenyl
acetonitrile, 0.12-14.84 μg/mL of Amine impurity, 0.11-14.67 μg/mL of
Nitrile impurity, 0.11-14.51 μg/mL of Ketone impurity, 0.81-88.80
μg/mL of ODI, 2.56-91.80 μg/mL of NNHUI, 0.95-91.80 μg/mL of ODLI,
5.07-90.84 μg/mL of CDZI in triplicates. The linearity was evaluated by
linear regression analysis that was calculated by the least square
regression.
Accuracy:
56
The accuracy was determined by the recovery of known amounts of 4-
chlorophenyl acetonitrile, Amine impurity, Nitrile impurity, and Ketone
impurity, ODI, NNJUI, ODLI and CDZI to the Sample in the beginning of
the preparative process. The added levels were LOQ, 50, 100 and 150%
of the specified limit in triplicate and then proceed with sample
preparation as described under experimental result.
Precision:
Six replicate injections of the standard preparation were madeinto the
HPLC used the methodology given in experimental result.
Six spiked sample preparations and one control sample preparation of
Sibutramine and Orlistat capsules were prepared and injected into the
HPLC using the method as described under experimental result.
Ruggedness:
Six spiked sample preparations and one control sample preparations
of Sibutramine and Orlistat capsules were analysed by a different
analyst, using different column, on different day and injected into a
different HPLC using the method as described in experimental result,
along with standard preparation.
Robustness:
57
Standard preparation, diluent, placebo preparation and sample
preparation in triplicate of the sample of Sibutramine and Orlistat
capsules were prepared as described in experimental result. The samples
along with standard and placebo were injected under different
chromatographic conditions.
Stability of analytical solution:
Standard solution, Sample solution were analysed initially and at
different time intervals at room temperature.
Systemsuitability:
The system suitability was verified through the evaluation of the
obtained parameters for the standard elution, such as theoretical plates,
peak asymmetry and retention factor, verified in different days of the
method validation.
2.6 EXPERIMENTAL RESULTS
On the basis of Sibutramine hydrochloride and Orlistat Analytical
method development experimental trials, RP-HPLC method described in
experiment No.7 was suitable for simultaneous determination of
Sibutramine hydrochloride and orlistat impurities in a single method.
RS method chromatographic conditions were applied for Identification as
well as quantification of related substance. Impurities were identified by
spiking known impurities in to sample preparation. Preparation of stock
58
solutions: Prepare solution having the concentration of sibutramine
hydrochloride 100 ppm and Orlistat 1200 ppm, in acetonitrile. Mix
standard solution: Prepared a solution having the concentrations of
sibutramine hydrochloride 100 ppm and orlistat 1200 ppm.
Sample Preparation:
Twenty capsules were weighed and crushed finely, powder equivalent
to 240 mg of Orlistat and 20 mg sibutramine hydrochloride (480mg) was
transferred to 200 ml volumetric flask and added 150 ml of diluent,
shake for 10 min and made up to volume with diluent, filtered the above
sample through 0.45 µ Teflon filter and injected. Concentration of
sibutramine hydrochloride was 100 ppm and orlistat was 1200 ppm.
Separately injected equal volumes of diluent, standard preparation in
six replicates and sample twice in to equilibrated HPLC system and
record chromatograms and measured the response in terms of peak area.
System suitability parameters occurred during method validation were
Theoretical plates not less than 8000, tailing factor not more than 1.5,
relative standard deviation for six replicates of standard solution is not
more than 2.0%.
2.7 DISCUSSION OF RESULTS:
LOD and LOQ:
59
RSD is less than 33% at LOD level and less than 10% at LOQ level
for sibutramine, orlistat and known impurities.
Linearity and range:
The correlation coefficients are less than 0.9995 for Sibutramine,
Orlistat and known Impurities.
Precision:
system precision RSD is less than 5% and method precision RSD is
less than 10% for sibutramine, orlistat and known impurities.
Accuracy:
The mean recoveries for sibutramine, orlistat and known impurities
are within 90 -110 %.
Specificity:
Retention time of Sibutramine and Orlistat peaks and known peaks
in sample preparation is comparable with respect to retention time of
Sibutramine and orlistat and known impurities peaks in standard
preparation. Peak purity passes for Sibutramine and orlistat and known
impurities peaksin standard and sample preparations. No intereference
was observed at the retention time of Sibutramine and orlistat and
known impurities peaks. Peak purity passes for all degradation
conditions.
60
Ruggesness:
The RSD of twelve results obtained from two different analysts are
within 10 %.
Robustness:
Sibutramine, orlistat and all known impurities peaks were resolved
with each other and system suitability complies for all variable
conditions, the test method is robust for all variable conditions.
Stability in Analytical solution:
Standard and sample solutions are stable for 12 h at room
temperature
System Suitability:
Theoretical plates are less than 2000, tailing factor is less than 2.0
and relative standard deviation is less than 5.0 for six standard replicate
injections.
Table 2.04
Peak purity data for Orlistat and Sibutramine HCl
Sr. No. Name
Purity
Criteria
61
1 Sibutramine hydrochloride in
standard solution Pass
2 Sibutramine hydrochloride in
sample solution Pass
3 Orlistat in standard solution Pass
4 Orlistat in sample solution Pass
Table 2.05
RT, RRT, LOD and LOQ of Sibutramine Impurities
Name of Compound RT
(min) RRT RRF LOD
(μg/mL) LOQ
(μg/mL)
Sibutramine HCl 9.93 1.00 1.00 0.038 0.115
4 CPA 6.67 0.67 0.95 0.040 0.121
Amine impurity 7.92 0.80 1.00 0.038 0.115
Nitrile impurity 11.28 1.14 1.02 0.037 0.113
Ketone impurity 19.82 2.00 1.10 0.035 0.105
Table 2.06
RT, RRT, LOD and LOQ of Orlistat Impurities
Name of Compound RT
(min) RRT RRF LOD
(μg/mL) LOQ
(μg/mL)
62
Orlistat 30.82 1.00 1.00 0.32 0.99
Dimeric impurity 5.39 0.17 1.23 0.26 0.80
NN Dicyclohexyl urea 6.76 0.22 0.40 0.80 2.48
Delactone impurity 25.47 0.83 1.08 0.30 0.92
CBZ impurity 35.22 1.14 0.20 1.60 5.07
Table 2.07
Linearity Data of Orlistat and Sibutramine HCl at Limiting Conc.
Level (%) Orlistat Sibutramine HCl
63
Conc.
(µg/mL)
Area Conc.
(µg/mL)
Area.
LOQ 1.01 2023 0.12 2043
50 30.04 58998 5.33 86548
80 48.06 96501 8.52 140284
90 54.06 106842 9.59 155317
100 60.07 119550 10.65 173791
110 66.08 131768 11.72 191552
120 72.08 142312 12.78 206881
150 90.11 180257 15.98 262042
Slope 1993 16335
Intercept -211 -144
Correlation
coefficient
0.9999 0.9999
Table 2.08
Linearity of Sibutramine HCl Impurities
Level 4 CPA SAI SNI SKI
64
Conc.
(µg/mL) Area
Conc.
(µg/mL) Area
Conc.
(µg/mL) Area
Conc.
(µg/ml) Area
LOQ 0.13 2169 0.12 2125 0.11 1894 0.11 2112
50 5.12 81725 4.95 86014 4.89 85807 4.84 94725
80 8.18 132609 7.91 139424 7.82 139232 7.74 153548
90 9.21 147403 8.90 154046 8.80 153204 8.70 170677
100 10.23 165101 9.89 173241 9.78 173347 9.67 191170
110 11.25 183791 10.88 192661 10.76 192970 10.64 212600
120 12.28 197527 11.87 205602 11.74 205312 11.60 227798
150 15.35 251416 14.84 262980 14.67 263973 14.51 288476
Slope 16288 17607 17837 19836
Intercept -756 -547 -812 -413
C.C 0.9998 0.9998 0.9997 0.9999
65
Table 2.09
Linearity of Orlistat Impurities
Level ODI (%) NNHUI (%) ODLI (%) CBZI (%)
Conc.
(µg/mL) Area
Conc.
(µg/mL) Area
Conc.
(µg/mL) Area
Conc.
(µg/ml) Area
LOQ 0.81 2021 2.56 2004 0.95 2024 5.07 2017
50 29.60 72788 30.60 23671 30.60 64105 30.28 11847
80 47.36 118108 48.96 38409 48.96 103911 48.45 19205
90 53.28 131283 55.08 42263 55.08 114808 54.50 21347
100 59.20 147047 61.20 47820 61.20 129114 60.56 23910
110 65.12 163692 67.32 53233 67.32 143588 66.62 26590
120 71.04 175926 73.44 56638 73.44 153232 72.67 28491
150 88.80 223922 91.80 72820 91.80 195995 90.84 36080
Slope 2507 786 2121 396
Intercept -674 -229 -406 -49
C.C 0.9998 0.9997 0.9998 0.9999
66
Table 2.10 Recovery of Orlistat impurities at LOQ Level
ODI (%) NNHUI (%) ODLI (%) CBZI (%)
LOQ Spl. 1 103.9 98.3 105.5 95.5
LOQ Spl. 2 98.7 96.8 100.2 98.8
LOQ Spl. 3 101.6 103.5 101.9 102.4
Mean 101.4 99.5 102.5 98.9
S.D 2.6 3.5 2.7 3.5
%RSD 2.6 3.5 2.6 3.5
Table 2.11
Recovery of Orlistat Impurities at 50, 100 and 150 % Level
ODI (%) NNHUI (%) ODLI (%) CBZI (%)
50 Spl. 1 102.3 103.4 101.4 100
50 Spl. 2 99.5 102.7 101.1 100.8
50 Spl. 3 101.8 104.3 101.6 99
100 Spl. 1 98.3 99.4 98.6 100.4
100 Spl. 2 98 99.8 98.6 99.6
100 Spl. 3 98.9 98.3 96.9 99.2
150 Spl. 1 98.5 97.2 99.6 98.7
150 Spl. 2 101.2 99.2 97.3 99.7
150 Spl. 3 98.1 98.3 98.3 98.9
Mean 99.62 100.29 99.39 99.59
S.D 1.69 2.53 1.84 0.71
%RSD 1.70 2.52 1.85 0.72
67
Table 2.12 Recovery of Sibutramine HCl Impurities at LOQ Level
4-CPA (%) SAI (%) SNI (%) SKI (%)
LOQ Spl. 1 101.3 101.3 103.5 103.4
LOQ Spl. 2 99.7 99.8 102.2 99.5
LOQ Spl. 3 101.6 102.5 100.9 100.6
Mean 100.9 101.2 102.2 101.2
S.D 1.0 1.4 1.3 2.0
%RSD 1.0 1.3 1.3 2.0
Table 2.13 Recovery of Sibutramine HCl Impurities at 50,100 and 150%
Level
4-CPA (%) SAI (%) SNI (%) SKI (%)
50 Spl. 1 101.3 102.4 103.4 99.5
50 Spl. 2 99.5 104.7 100.1 101.46
50 Spl. 3 102.8 104.3 100.6 101.5
100 Spl. 1 99.3 95.4 99.6 100.4
100 Spl. 2 99 96.8 97.6 99.5
100 Spl. 3 98.9 99.3 99.9 98.8
150 Spl. 1 102.4 99.2 99.6 99.2
150 Spl. 2 101.2 99.7 97.3 101.3
150 Spl. 3 99.1 99.3 97.3 102.3
Mean 100.39 100.12 99.76 100.44
S.D 1.55 3.15 1.88 1.24
%RSD 1.54 3.14 1.89 1.24
68
Table 2.14
System Precision Data for Sibutramine Impurities
SH
(Area)
4-CPAI
(Area)
SAI
(Area)
SNI
(Area)
SKI
(Area)
Rep.1 173791 165101 173241 173347 191170
Rep. 2 172341 165098 174211 172234 191234
Rep. 3 173542 164981 174842 174121 188187
Rep. 4 173987 164892 173451 175236 188456
Rep. 5 174721 166234 173562 174675 189241
Rep. 6 172981 165990 173245 173425 189234
Mean (Area) 173561 165383 173759 173840 189587
S.D 825.24 575.47 639.28 1069.53 1319.24
% RSD 0.48 0.35 0.37 0.62 0.70
Table 2.15
System Precision Data for Orlistat Impurities
Orlistat (Area)
ODI (Area)
NNDHU (Area)
ODLI (Area)
CBZI (Area)
Rep.1 119550 47820 147047 129114 23910
Rep. 2 116341 47342 147745 128234 23923
Rep. 3 117825 47394 147253 129841 23876
Rep. 4 119145 47209 147846 129244 23985
Rep. 5 117871 47645 147980 128643 23908
Rep. 6 115629 47652 147823 128543 23890
Mean(Area) 117727 47510 147616 128937 23915
S.D 1528.86 230.98 374.28 579.43 37.90
%RSD 1.30 0.49 0.25 0.45 0.16
69
Table 2.16 Method Precision Data for sibutramine HCl impurities
SH (%) 4 CPA (%) SAI (%) SNI (%)
Spl. 1 1.01 1.06 1.15 1.15
Spl. 2 1.03 1.04 1.19 1.11
Spl. 3 1.01 1.02 1.14 1.12
Spl. 4 0.991 1.13 1.17 1.09
Spl. 5 0.987 1.19 1.15 1.08
Spl. 6 0.989 1.09 1.17 1.1
Mean (%) 1.003 1.088 1.162 1.108
S.D 0.017 0.063 0.018 0.025
%RSD 1.68 5.79 1.58 2.24
Table 2.17
Intermediate Precision Data for Sibutramine HCl Impurities
SH (%) 4 CPA (%) SAI (%) SNI (%)
Rep.1 1.05 1.09 1.05 1.11
Rep. 2 1.03 1.14 1.09 1.01
Rep. 3 1.05 1.12 1.17 1.02
Rep. 4 1.01 1.14 1.13 1.09
Rep. 5 1.07 1.19 1.14 1.08
Rep. 6 0.989 1.09 1.16 1.08
Mean (%) 1.033 1.128 1.123 1.065
S.D 0.030 0.038 0.045 0.040
%RSD 2.88 3.34 4.05 3.79
Table 2.18
70
Method Precision Data for Orlistat Impurities
ODI (%) NNDHU (%) ODLI (%) CBZI (%)
Spl. 1 1.04 1.05 1.12 1.13
Spl. 2 1.02 1.08 1.11 1.12
Spl. 3 1.03 1.07 1.1 1.14
Spl. 4 1.01 1.11 1.13 1.07
Spl. 5 1.07 1.12 1.16 1.18
Spl. 6 1.09 1.11 1.11 1.11
Mean (%) 1.043 1.090 1.122 1.125
S.D 0.031 0.028 0.021 0.036
%RSD 2.95 2.53 1.91 3.22
Table 2.19 Intermediate precision Data for Orlistat Impurities
ODI (%) NNDHU (%) ODLI (%) CBZI (%)
Spl. 1 1.11 1.03 1.12 1.13
Spl. 2 1.12 1.09 1.09 1.05
Spl. 3 1.05 1.1 1.13 1.04
Spl. 4 1.08 1.08 1.13 1.09
Spl. 5 1.09 1.09 1.09 1.08
Spl. 6 1.11 1.11 1.13 1.09
Mean (%) 1.093 1.083 1.115 1.080
S.D 0.026 0.028 0.020 0.032
%RSD 2.36 2.59 1.77 2.99
71
Table 2.20 Solution Stability Data of Orlistat impurities
ODI (%) NNDHU (%) ODLI (%) CBZI (%)
Initial 1.03 1.03 1.01 1.05
After 8 h 1.05 1.04 1.05 1.08
After 12 h 1.08 1.09 1.05 1.09
After 24 h 1.03 1.02 1.03 1.03
After 48 h 1.18 1.07 1.12 1.08
Mean (%) 1.074 1.05 1.052 1.066
S.D 0.063 0.029 0.041 0.025
%RSD 5.84 2.78 3.94 2.35
Table 2.21
Solution stability Data of Sibutramine HCl Impurities
SDI (%) NNDHU (%) SDLI (%) CBZI (%)
Initial 0.98 1.08 1.06 1.08
After 8 h 1.01 1.1 1.07 1.02
After 12 h 1.03 1.09 1.02 1.09
After 24 h 1.08 1.06 1.07 1.02
After 48 h 1.1 1.02 1.13 1.08
Mean (%) 1.04 1.07 1.07 1.058
S.D 0.049 0.032 0.039 0.035
%RSD 4.76 2.96 3.68 3.30
72
Figure 2.03
Amine Impurity HMBC NMR Spectrum-1
73
Figure 2.04
Amine Impurity HMBC NMR Spectrum-2
74
Figure 2.05
Amine Impurity HMBC Nitrogen NMR Spectrum-3
75
Figure 2.06
4- Chloro phenyl acetonitrile HMBC NMR Spectrum-1
76
Figure 2.07
4- Chloro phenyl acetonitrile HMBC NMR Spectrum-2
77
Figure 2.08
Nitrile Impurity HMBC NMR Spectrum-1
78
Figure 2.09
Nitrile Impurity HMBC NMR Spectrum-2
79
Figure 2.10
4- Chloro Acetonitrile Impurity C-13 NMR Spectrum-1
80
Figure 2.11
Amine Impurity C-13 NMR Spectrum-1
81
Figure 2.12
4-Chloro acetonitrile C-13 NMR Spectrum-2
82
Figure 2.13
Ketone Impurity C-13 NMR Spectrum-1
Figure 2.14
83
Ketone Impurity C-13 NMR Spectrum-2
84
Figure 2.15
Nitrile Impurity C-13 NMR Spectrum-1
85
Figure 2.16
Nitrile Impurity C-13 NMR Spectrum-2
86
Fig. 2.17
4- Chloro phenyl acetonitrile Proton NMR Spectrum-1
87
Figure 2.18
Amine Impurity proton NMR spectrum-1
Figure 2.19
88
Amine Impurity Proton NMR Spectrum-2
Figure 2.20
89
Ketone Impurity Proton NMR Spectrum-1
Figure 2.21
90
Ketone Impurity Proton NMR Spectrum-2
Figure 2.22
91
Nitrile Impurity Proton MNR Spectrum-1
92
Figure 2.23
Nitrile Impurity Proton MNR Spectrum-2
93
Figure 2.24
Amine Impurity Mass Spectrum-1
94
Figure 2.25
4-ChloroPhenyl Acetonitrile Mass Spectrum-1
Figure 2.26
95
Amine Impurity Mass Spectrum-1
Figure 2.27
96
Ketone Impurity Mass Spectrum-1
97
Figure 2.28
Nitrile Impurity Mass Spectrum-1
98
Figure 2.29
Sibutramine + Orlistat Capsules Placebo Chromatograph
Figure 2.30
Sibutramine + Orlistat Capsules Sample Chromatograph
99
Figure 2.31
Mix Impurities Standard Chromatograph
Figure 2.32
Mix Impurities Standard Chromatograph
100
Figure 2.33
Impurities Spiked Sample Chromatograph
Figure 2.34
101
Ketone Impurity UV Spectrum
Figure 2.35
102
Nitrile Impurity UV Spectrum
Figure 2.36
103
4-Chlorophenyl Acetonitrile UV Spectrum
2.8 Summary, Conclusion and Recommendations.
104
The reversed phase LC method proposed was found to be simple, fast,
accurate, precise, linear, robust and specific and it is powerful tool to
investigate chemical stability of sibutramine and orlistat. The robustness
of the method was verified with small variation on pH, concentration of
triethylamine solution, concentration of organic phase, detector
wavelength, column manufacturer and analysis temperature. All the
parameters meet the criteria of the ICH guidelines for method validation
and its chromatographic retention time of 60 min allows the analysis of a
large number of samples in an adequate period of time. The method
could therefore be recommended for routine quality control analysis of
raw material and of capsules, as well as for routine quality control
analysis of raw material and of commercial cream, as well as for
protocols of sibutramine and orlistat stability study.