breath testing in the presence of chronic respiratory disease · breath test, a number of drivers...

6
Breath testing in the presence of chronic respiratory disease Dr Morris Odell*, Dr Christine McDonald**, John Farrar B.*** and Jeff Pretto** *Victorian Institute of Forensic Medicine **Austin and Repatriation Medical Centre ***Traffic Alcohol SectionVictoria Police ABSTRACT The presence of chronic respiratory disease is sometimes claimed as a defence by drivers who refuse or fail a breath test. Trials were conducted with two instruments used in Victoria in order to determine the effort required for a satisfactory test. Patients with restrictive and obstructive conditions participated in the trial which was conducted at the respiratory unit of a large hospital. The results are presented. They provide information useful in refuting defenses of inability to provide a breath sample due to respiratory incapacity. INTRODUCTION In Victoria, breath testing for alcohol is carried out under provisions of the Road Safety Act 1986 (Ref 7) . This act provides for compulsory breath testing of drivers under certain circumstances. While inability to provide a suitable sample results in a blood sample being required, refusal to provide a breath sample is an offence under the act and carries a heavy penalty. Notwithstanding the requirement for a blood sample to be provided in the event of a failed breath test, a number of drivers have attempted to use the presence of chronic respiratory disease as a defence against the charge of refusing a breath test. This trial was carried out in an attempt to quantify the minimum respiratory effort required for a satisfactory breath test in order that such defences could be evaluated in the light of formal respiratory function testing. - 329-

Upload: nguyenhanh

Post on 15-Dec-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

Breath testing in the presence of chronic respiratory

disease

Dr M orris Odell*, Dr Christine M cDonald**, John Farrar B.*** and Jeff Pretto**

* Victorian Institute o f Forensic Medicine

**Austin and Repatriation Medical Centre

***Traffic Alcohol SectionVictoria Police

ABSTRACT

The presence of chronic respiratory disease is sometimes claimed as a defence by drivers who

refuse or fail a breath test. Trials were conducted with two instruments used in Victoria in

order to determine the effort required for a satisfactory test. Patients with restrictive and

obstructive conditions participated in the trial which was conducted at the respiratory unit o f a

large hospital. The results are presented. They provide information useful in refuting defenses

o f inability to provide a breath sample due to respiratory incapacity.

INTRODUCTION

In Victoria, breath testing for alcohol is carried out under provisions of the Road Safety A ct

1986 (Ref 7) . This act provides for compulsory breath testing o f drivers under certain

circumstances. W hile inability to provide a suitable sample results in a blood sample being

required, refusal to provide a breath sample is an offence under the act and carries a heavy

penalty.

Notwithstanding the requirement for a blood sample to be provided in the event o f a failed

breath test, a num ber o f drivers have attempted to use the presence of chronic respiratory

disease as a defence against the charge of refusing a breath test. This trial was carried out in

an attempt to quantify the minimum respiratory effort required for a satisfactory breath test in

order that such defences could be evaluated in the light of formal respiratory function testing.

- 3 29-

Two instruments are used by Victoria Police for conducting breath tests. These are the Lion

Alcometer SD2, a portable hand-held device used for screening tests, and the Drager Alcotest

7110 which is a precision instrument used to provide evidentiary print-outs accepted by the

courts. The Drager Alcotest 7110 has been in use in Victoria since 1994 and is a substantially

different instrument to one with the same name used in other jurisdictions (Ref 2). Each of

the two instruments contains sensors and circuitry for the purpose of determining when an

adequate breath sample has been provided and will not produce a reading unless their criteria

have been satisfied.

The Lion Alcotest uses a simple timer to define the sampling interval. The timer is started

once the breath flow exceeds a trigger level and runs for 2.7 seconds. There is no check on

whether flow continues during the delay period. Flow characteristics are defined by the

disposable plastic tubular mouthpiece which are mass produced in large numbers with a high

degree of reproducibility. The Drager Alcometer is a complex microprocessor controlled

instrument which progressively displays a row of asterisks as the subject exhales, a

satisfactory effort being indicated by 16 asterisks appearing .

Studies with instruments formerly used for breath analysis have concluded that persons

unable to provide a breath sample for analysis because of respiratory impairment would be so

severely affected that they would be incapable of driving, thus rendering the need for breath

testing superfluous. These have always required a respiratory effort which has been great

enough to exclude a significant number of drivers from being able to provide a breath

specimen. In fact, formal guidelines for assessment o f suitability for driving are quite tolerant

o f remarkably high levels of impairment.(Ref 8,9) The Victorian guidelines allow m otor car

drivers to exhibit a high level of impairment even to the extent of requiring intranasal oxygen.

The commercial vehicle guidelines are even less specific, only citing “severe respiratory

failure” as an exclusion. Thus it is possible for persons with various types o f advanced

respiratory disease to be driving cars and heavy commercial vehicles.

PROCEDURE

The study was conducted in two stages. The first stage involved a comparison of six Drager

units to determine whether there were significant differences in the volume required to

- 330-

achieve a satisfactory sample as indicated by 16 asterisks on the unit’s display panel. This

was done in order to check that the machines were substantially identical and that the

remainder o f the trial could be conducted using only one machine.

The minimum volume required to provide an adequate sample for the Drager units was

measured using a 3-litre graduated calibration syringe (Hans Rudloph model XXXX) with a

resolution of 25 millilitres. The procedure involved manually injecting air at room

temperature (23 degC) at a rate of approximately 100 - 150 ml/sec until an adequate sample

was provided (ie when the 16th asterisk was illuminated), and noting the injected volume.

This procedure was measured in triplicate in each of 6 Drager units. The mean minimum

sample volume thus measured was 840 ml with a range of 775 - 925 mis.

Staff from the Austin & Repatriation Medical Centre (ARMC ) attended a Police Random

Breath Testing Station to observe police breath testing procedures. This was done to ensure

that trial conditions mimicked field conditions as much as possible. Ethics committee

approval was obtained prior to patients attending the Respiratory Laboratory at the ARM C for

routine respiratory function testing being recruited for the study. Criteria for eligibility were:

1. No alcohol to have been consumed on the day o f the test

2. Known respiratory disability, as reflected by an FEV1 of less than or equal to 1.5 litres

(Refs 1,2,4) and one or more of the following ventilatory abnormalities:

a) an obstructive ventilatory defect secondary to asthma or COPD

b) a restrictive ventilatory defect secondary to neuromuscular disease affecting the

respiratory musculature

c) a restrictive ventilatory defect secondary to some form of diffuse interstitial lung

disease.

Informed consent was sought and the patient then asked to blow into each device using the

standard instructions used by the police in the field. In accordance with police practice,

subjects who failed an initial test were asked to repeat it up to 3 times. Information gathered

was:

• Age and sex of patient

• Type of respiratory problem

• Result o f respiratory function tests

-331 -

• Result o f attempt to blow into a Lion Alcometer (Pass/Fail)

• Result o f attempt to blow into a Drager Alcotest (Pass/Fail/no o f asterisks achieved)

Statistical analysis was performed using SPSS for W indows Version 6 .1

RESULTS

Sixty seven subjects participated in the trial. Details of the subjects are given in Table 1.

Table 1

Male sex 36 (53.7%)

Age range 33-82 years (mean 68.7, SD 9.7)

Diagnosis - COPD ("including asthma,

emphysema and chronic bronchitis)

55 (82.1%)

- interstitial lung disease 4 (5.9%)

- neuromuscular disease 3 (4.5%)

- other* 5 (7.5%)

Use of domiciliary oxygen therapy 3 (4.5%)

Drivers 25 (37.3%)

F E V 1 range (Litres) 0.54-2.16 (mean 1.06)

FVC range (Litres) 0.65-3.6 (mean 2.10)

(* includes Sarcoidosis, pleural disease and undiagnosed cases)

There were no significant differences in respiratory capacities between the drivers and non­

drivers in the study (Table 2 ):

Table 2

Drivers Non-Drivers

M ean FVC (Litres) 2.18 1.96

M ean FEV 1 (Litres) 1.07 1.04

Mean FER 51.8 57.1

The subjects subjectively found the Drager unit much easier to blow into than the Lion unit

and this is reflected in the results. Sixty-two subjects (92.5%) of the experimental group were

able to provide a satisfactory evidentiary breath sample. All o f the subjects requiring

- 332 -

domiciliary intranasal oxygen were able to provide a satisfactory sample into the Drager but

only one was able to provide an adequate sample for the Lion. Only 5 subjects were not able

to provide a satisfactory evidentiary sample and their respiratory parameters were otherwise

undistinguishable from the rest of the group. Details of these subjects are given in Table 3.

Table 3

Age Sex FEV1

(L)

FVC

(L)

No of

asterisks

Diagnosis

67 F 0.63 0.85 1 2 Kyphoscoliosis/resp. failure

59 F 0.75 1.40 1 2 Bronchiectasis

64 F 0.95 1.75 1 2 Asthma

64 F 0.55 0.65 1 2 Post polio syndrome

82 M 1 . 0 1.15 1 1 Myasthenia gravis

The subjects unable to provide an adequate sample for the Lion Alcometer had the following

characteristics:

Table 4

Satisfactory sample Unsatisfactory

sample

P value

(t-test)

Number 18 (27%) 49 (73%)

Mean FE V 1 (Litres) 1.52 0.89 <0 . 0 0 1

Mean FVC (Litres) 2.45 1.97 0.02<P<0.01

FER (=FVC/FEV1)

(%)

65 50 0.001<P<0.01

Num ber o f drivers 10 (40% of drivers) 15 (60% o f drivers)

DISCUSSION

There was no significant difference in between respiratory impairment between the driving

and non-driving subjects indicating that respiratory disability was not the reason for the

decision not to drive.

- 333 -

These results are broadly similar to those obtained in other studies. The Lion Alcometer is

widely used and is accepted as evidentiary in some jurisdictions. It requires a greater

respiratory effort than more modern instruments and a significant number of drivers are not

able to provide a satisfactory breath sample for this device. Commonly measured respiratory

parameters may be helpful in assessing claimed inability to blow into this device.

The number o f subjects who could not provide a satisfactory sample for the Drager machine

was very low and not enough for formal statistical evaluation. It would appear however that

the Drager machine used in Victoria is able to be used at the limits o f respiratory function

likely to be encountered in drivers.

REFERENCES

1. Briggs J E, Patel H, Butterfield K, Honeybourne D "The effects of chronic obstructive airways disease on the ability to drive and use a roadside alcometer” Resp. Med. (1990) 84:43-46

2. Crockett AJ, Schembri D A, Smith D J, Laslett R, Alpers J H “M inimum respiratory function for breath testing in South Australia” J. Forens Sci Soc (1992) 32(4):349-56

3. Morris MJ “Alcohol breath testing in patients with respiratory disease” (editorial) Thorax (1990) 45:717-721

4. Gomm P J, Osselton M D, Broster C G, Johnson N M cl, Upton K “Study into the ability o f patients with impaired lung function to use breath alcohol testing devices” Med. Sci. Law (1991) 31:221-225

5. Gomm P J, Osselton M D, Broster C G, Johnson N M cl, Upton K “The Effect of Salbutamol on Breath Alcohol Testing in Asthmatics” Med. Sci. Law (1991) 31:226-228

6 . Marks P “Drink Driving Legislation: Medicine and the Law” Medico-Legal Journal 63/3:119-127

7. “Road Safety Act 1986” No 127/1986 Victorian Government Printer, Melbourne.

8 . “Interim General Driver Licensing Guidelines” Vicroads 1994

9. “Medical Examination o f Commercial vehicle Drivers” Federal Office of Road Safety 1994

- 334 -