Order by phone 01342 826836 Order on-line www.ecomcat.co.uk Order by fax 01342 826771 Website www.jencons.co.uk Introduction Biological Safety Cabinets 134 Take a look at our latest products on www.ecomcat.co.uk Biological Safety Cabinets All Biological Safety Cabinets purchased for use in the UK should comply with BS EN 12469:2000. The British Standards BS 5726-2 (1992) and BS 5726-4 (1992) give guidelines for siting, installation, commis- sioning and the use of cabinets. Our responsibilities as a supplier of safety cabinets are also detailed in BS 5726 (1992), parts 2 and 4. We, therefore, strongly recommend that, following consultation with your site safety officer and/or biological safety officer, you contact our Technical Office on 01342 826836 to arrange a FREE site survey prior to quotation. One of our Area Managers will be happy to attend to discuss the positioning and configuration (ducted or recirculating) of your Biological Safety Cabinet, paying full regard to BS 5726 (1992), parts 2 and 4 and the Advisory Committee for Dangerous Pathogens guidelines. Failure to do this may mean that the cabinet will not be able to comply with the British Standards, and the cost of alterations at the time of installation and commissioning may exceed the cost of the original cabinet. If you are planning a new building or laboratory refurbishment, our Technical Office will be happy to liaise with architects, builders or M&E contractors to ensure the correct fan ductwork and other site services are installed, and that sufficient make-up air has been allowed for in the room ventilation design. Types of Safety Cabinet Class I - operator protection ONLY Class II - operator and product protection Class III - operator and product protection for group 4 pathogens Note; vertical and horizontal laminar flow cabinets are for product protection only, and are, therefore, not classified as biological safety cabinets. Please see pages 558 to 562 for details. Configuration Cabinets can either be ducted to atmosphere, venting outside the building, or they can be recirculating, i.e. exhausting back into the laboratory through a double exhaust HEPA filter with no ductwork requirement. Factors to Consider 1. Siting Please ensure sufficient height from bench/floor to ceiling is available. Where height is restricted, a low- level option may be available. Please contact our Technical Office for further details. The positioning of a Safety Cabinet is extremely important as its performance can be affected by its proximity to the following: air conditioning units, ventilation ducts, doors, other safety cabinets, walls, other equipment or windows. Please contact our Technical Office for a FREE site survey. 2. Formalin Fumigation BS 5726-4 (1992) advises that cabinets are decontaminated by formalin fumigation every six months and/or prior to a service visit. Some customers prefer to fumigate on a more frequent basis. Great care has to be taken to ensure formalin is eliminated and disposed of in a safe manner. This is not a problem for ducted cabinets (assuming that your ductwork does not vent out onto a public walkway). However, for those requiring recirculating cabinets, there are several options available; Flexible ductwork Sometimes called ‘elephant trunking’, this length of ducting can be attached to the spigot of a recirculating cabinet and vented into a fume cupboard or out of a window. Mobile formalin evacuation unit This is a carbon filter on wheels that can be used when there is no provision to exhaust formalin on existing cabinets. The ‘formalin scrubber’ is therefore a useful item to have in a department, as it can be attached to any recirculating cabinet by the flexible ductwork supplied with the machine. Other methods Different manufacturers recommend and supply alternatives to the above. Scrubber units which fit on top of (or under) the Safety Cabinet and contain activated charcoal are alternatives to the mobile scrubber unit. Hydrogen Peroxide is a newer, viable alternative to the use of formalin. See pages 145 to 146. For advice on the best method for your laboratory please contact our Technical Office on 01342 826836.

Upload: lamduong

Post on 10-Apr-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukIntroduction

Biological Safety Cabinets

134 Take a look at our latest products on www.ecomcat.co.uk

Biological Safety CabinetsAll Biological Safety Cabinets purchased for use in the UK should comply with BS EN 12469:2000. TheBritish Standards BS 5726-2 (1992) and BS 5726-4 (1992) give guidelines for siting, installation, commis-sioning and the use of cabinets. Our responsibilities as a supplier of safety cabinets are also detailed in BS5726 (1992), parts 2 and 4.

We, therefore, strongly recommend that, following consultation with your site safety officer and/orbiological safety officer, you contact our Technical Office on 01342 826836 to arrange a FREE site surveyprior to quotation. One of our Area Managers will be happy to attend to discuss the positioning andconfiguration (ducted or recirculating) of your Biological Safety Cabinet, paying full regard to BS 5726(1992), parts 2 and 4 and the Advisory Committee for Dangerous Pathogens guidelines. Failure to do thismay mean that the cabinet will not be able to comply with the British Standards, and the cost ofalterations at the time of installation and commissioning may exceed the cost of the original cabinet.

If you are planning a new building or laboratory refurbishment, our Technical Office will be happy toliaise with architects, builders or M&E contractors to ensure the correct fan ductwork and other siteservices are installed, and that sufficient make-up air has been allowed for in the room ventilation design.

Types of Safety Cabinet Class I - operator protection ONLYClass II - operator and product protection Class III - operator and product protection for group 4 pathogens

Note; vertical and horizontal laminar flow cabinets are for product protection only, and are, therefore, not classified asbiological safety cabinets. Please see pages 558 to 562 for details.

ConfigurationCabinets can either be ducted to atmosphere, venting outside the building, or they can be recirculating, i.e.exhausting back into the laboratory through a double exhaust HEPA filter with no ductwork requirement.

Factors to Consider1. SitingPlease ensure sufficient height from bench/floor to ceiling is available. Where height is restricted, a low-level option may be available. Please contact our Technical Office for further details. The positioning ofa Safety Cabinet is extremely important as its performance can be affected by its proximity to thefollowing: air conditioning units, ventilation ducts, doors, other safety cabinets, walls, other equipmentor windows. Please contact our Technical Office for a FREE site survey.

2. Formalin FumigationBS 5726-4 (1992) advises that cabinets are decontaminated by formalin fumigation every six monthsand/or prior to a service visit. Some customers prefer to fumigate on a more frequent basis. Great carehas to be taken to ensure formalin is eliminated and disposed of in a safe manner. This is not a problemfor ducted cabinets (assuming that your ductwork does not vent out onto a public walkway). However,for those requiring recirculating cabinets, there are several options available;

Flexible ductwork Sometimes called ‘elephant trunking’, this length of ducting can be attached to the spigot of a recirculatingcabinet and vented into a fume cupboard or out of a window.

Mobile formalin evacuation unit This is a carbon filter on wheels that can be used when there is no provision to exhaust formalin onexisting cabinets. The ‘formalin scrubber’ is therefore a useful item to have in a department, as it canbe attached to any recirculating cabinet by the flexible ductwork supplied with the machine.

Other methodsDifferent manufacturers recommend and supply alternatives to the above. Scrubber units which fit ontop of (or under) the Safety Cabinet and contain activated charcoal are alternatives to the mobilescrubber unit. Hydrogen Peroxide is a newer, viable alternative to the use of formalin. See pages 145 to146. For advice on the best method for your laboratory please contact our Technical Office on 01342 826836.

Page 2: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsFaster

Take a look at our latest products on www.ecomcat.co.uk 135

566-800BH-EN 2000 Class II Biological Safety Cabinets withDouble FanThe BH-EN 2000 safety cabinets belong to the latest generation of laminar airflowsystems, combining rigid safety requirements with high quality construction.

The BH-EN 2000 cabinets have been designed as Class II Biohazard cabinets andcomply with EN-12469. They provide vertical laminar airflow with part of the air recir-culated via the main HEPA filter whilst the remainder is discharged through an exhaustHEPA filter. This cabinet is suitable for handling CDC 1-3, DPAG C-B1-B2 etiologicagents, providing operator, product and environmental protection.

The BH-EN 2000 is available in four widths 0.9m, 1.2m, 1.5m and 1.8m.

■ Sophisticated microprocessor based monitoring system■ Audible and visual alarms alert the operator if the cabinet is not operating in

complete safety■ Solid work surface in AISI-316 stainless steel ■ Double HEPA filtration on all models (optional additional exhaust filter) ■ Self-adjusting system to ensure a constant laminar airflow speed even in the

event of a progressive clogging of the filters ■ The negative pressure plenum “bio-dynamic sealing” ensures that all contami-

nated particles are kept inside the system and are automatically drawn to thepressure chamber to be captured by the main and exhaust filters

■ The fan system ensures that no part of the cabinet is under positive contaminat-ed pressure to the laboratory, thus protecting the environment and personnelfrom bio-contamination risks

■ Access from the front panel for ease of maintenance ■ Two motors for improved safety and performance - ensures constant downflow

in the event of filter clogging ■ Two vacuum or gas connections in the working area are provided, one

complete with solenoid valve, the other manual ■ Electrical socket with IP-44 protection ■ A seamless stainless steel spillage tray beneath the work surface collects

spilt liquids ■ An aluminium front closure panel ensures the sealing of the cabinet for

sterilisation/fumigation ■ An inlet allows the D.O.P. testing of the HEPA filters

New

BH-EN 2000 Specifications - all modelsControls and Microprocessor based monitoring system for controlling the cabinet. instrumentation The keyboard and the rear lit LCD display keep the user constantly informed of the cabinet conditions and operation, including: display of laminar

airflow velocity and frontal air barrier velocity; display of inside and outside temperature (which can affect internal airflow velocity); display ofresidual lifetime of HEPA filters, UV lamp and charcoal filter (if installed); display of total time of cabinet operation; display of power rating factorof the main and exhaust fans.

Alarms Out of range laminar airflow velocity and front air barrier velocity, front window opened, HEPA filters clogging, UV lamp (if installed) needs replacing,charcoal filter (if installed) saturated, electrical power-out, signal of other possible malfunctions and relevant remedies.

Lighting level Fluorescent tubes in built-in housing, placed outside the contaminated area.Construction Epoxy powder painted steel for improved strength and resistance to corrosion and atmospheric agents.

Work surface: AISI-316 stainless steel, solid work surface (one piece). Frontal slots in the access area provide air barrier protection preventing anyexchange of air between the inside of the cabinet and the environment. * Work chamber: in AISI-304 stainless steel Front window: Stratified safety glass, to give ease of access for large items. It is fitted with a gas strut to keep it open during maintenance orsanitisation operations.

Filters HEPA filters, tested with D.O.P. aerosol. Their efficiency - higher than 99.995% MPPS - 0.005% penetration - ensures a performance exceeding therequirements of EN 1822, U.S. Federal Standard 209 e and AS 1386-1976. The filtration system is designed to provide a barrier of air with mini-mum average speed over 0.4m/sec. according to EN-12469.

Fans Two centrifugal fans, direct driven motors, with protection factor IP-55, controlled by two automatic speed regulators and microprocessor systemshowing laminar flow and protection barrier work conditions.

Electrical supply 230V, 50Hz, single phase.Noise level <59dB(A) @ 1 metre

* On request the following work surfaces are available: AISI-316 stainless steel perforated work surface in 1 piece AISI-316 stainless steel solid work surface in 3, 4, 5 or 6 pieces (depending on the width of cabinet)

Page 3: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukFaster

Biological Safety Cabinets

136 Take a look at our latest products on www.ecomcat.co.uk

BH-EN 2000 Class II Biological Safety Cabinets with DoubleFan, continued…SpecificationsModel BH-EN 2003-D BH-EN 2004-D BH-EN 2005-D BH-EN 2006-DExt. dims. (wxdxh), mm 1090x780x1470 1395x780x1470 1700x780x1470 2005x1470x780Int. dims. (wxdxh), mm 885x580x660 1190x580x660 1495x580x660 1800x660x580Power, Kw 0.85 1.1 1.3 1.5Exhaust flow rate, 290 385 490 585m3/second

Cat. No 566-805 566-800 566-806 566-807

Price POA POA POA POA

The cabinets listed above are recirculating with double fan and double HEPA filter. Ducted cabinets are also available.Please contact our Technical Office on 01342 826836.

For information and pricing on delivery, installation, commissioning and KI testing or to arrange a site survey, pleasecontact our Technical Office on 01342 826836.

Accessories for BH-EN cabinets

Cat. No. Description For Cabinets Price

566-803 Formalin vapouriser - see below for full details all models POA

566-808 UV light for sterilisation of the work chamber installed on all models POAthe front closure, controlled by two timers

566-809 Epoxy powder painted modular stand, 910mm working height BH-EN 2003 POA

566-801 Epoxy powder painted modular stand, 910mm working height BH-EN 2004 POA

566-810 Epoxy powder painted modular stand, 910mm working height BH-EN 2005 POA

566-811 Epoxy powder painted modular stand, 910mm working height BH-EN 2006 POA

566-812 Wheels for epoxy powder painted modular stand all stands POA

566-813 3-Drawers unit on pivotting wheels and removable epoxy all models POApowder painted vessels

566-814 Exhaust charcoal filter with relevant housing all models POA

566-815 Additional service connection for electric power all models POA

566-816 Additional service connection for gas/vacuum (manual tap) all models POA

566-824 Anti-blowback valve for connection to ductwork all models POA

566-802 Additional exhaust HEPA filter with relevant housing all models POA

Formalin Vapouriser/NeutraliserThis instrument has been designed to allow the fumigation/sterilisation in the Microbiological SafetyCabinets according to Annex J to the EN 12469 European standard.

■ Large capacity: ensures the correct fumigation for all models of Faster’s cabinets (a concentrationof 50mg/m3 formaldehyde is obtained by mixing 60ml formalin with 60ml water per each cubicmetre to be sterilised - full instructions and precise amounts are in the instruction manual).

■ Fully automatic procedure: the process is activated by pressing the key on the control panel andthe rest of the process takes place without any further intervention from the operator up to theend of the full cycle.

■ Safety: the chemical that neutralises the formaldehyde (ammonium bicarbonate) eliminates theformalin vapour with no ammonia residue. In the event of a power-out during the sterilisationcycle, the procedure will be completed correctly when the electric power is restored.

OperationThe formalin vapouriser consists of two containers. The first one contains a solution of 50% formalinand 50% distilled water and the second one contains a solution (e.g. ammonium bicarbonate) suitableto neutralise the formaldehyde vapours generated during the sterilisation process. Full instructions andcorrect volumes are given for each size of cabinet.

The device is fitted with a series of lights to highlight the various stages of sterilisation process.

Page 4: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsFaster

Take a look at our latest products on www.ecomcat.co.uk 137

566-818BHG Class II Biological Safety Cabinets withElectronically Operated Sash Window and Double FanThe BHG safety cabinets belong to the latest generation of laminar airflow systems,combining rigid safety requirements with high quality construction. The BHG cabinetshave been designed as Class II Biohazard cabinets and comply with EN-12469. Theyprovide vertical laminar airflow with part of the air recirculated via the main HEPA filterwhilst the remainder is discharged through an exhaust HEPA filter. This cabinet issuitable for handling CDC 1-3, DPAG C-B1-B2 etiologic agents, providing operator,product and environmental protection.

The BHG cabinets are provided with an electronically operated, vertically sliding safetyglass sash window (angled at 3 degrees to avoid vortices).

The BHG cabinet is available in three widths 0.9m, 1.2m and 1.8m.

■ Sophisticated microprocessor based monitoring system■ Audible and visual alarms alert the operator if the cabinet is not operating in

complete safety■ Solid work surface in AISI-316 stainless steel ■ Double HEPA filtration on all models (optional additional exhaust filter)■ Self-adjusting system to ensure a constant laminar airflow speed even in the

event of a progressive clogging of the filters ■ The negative pressure plenum “bio-dynamic sealing” ensures that all contami-

nated particles are kept inside the system and are automatically drawn to thepressure chamber to be captured by the main and exhaust filters

■ The fan system ensures that no part of the cabinet is under positive contami-nated pressure to the laboratory, thus protecting the environment and person-nel from bio-contamination risks

■ Access from the front panel for ease of maintenance ■ Two motors for improved safety and performance - ensures constant downflow

in the event of filter clogging ■ The electrically driven sash window (easily moved up and down by pressing the

keys placed on the front panel) can be completely closed or opened to makethe operator’s work easier

■ Two vacuum or gas connections in the working area are provided, one com-plete with solenoid valve, the other manual

■ Electrical socket with IP-44 protection ■ A seamless stainless steel spillage tray beneath the work surface collects spilt liquids ■ An inlet allows the D.O.P. testing of the HEPA filters

New

BH-EN 2000 Specifications - all modelsControls and Microprocessor based monitoring system for controlling the cabinet.instrumentation The keyboard and the rear lit LCD display keep the user constantly informed of the cabinet conditions and operation, including: display of laminar

airflow velocity and frontal air barrier velocity; display of inside and outside temperature (which can affect internal airflow velocity); display ofresidual lifetime of HEPA filters, UV lamp and charcoal filter (if installed); display of total time of cabinet operation; display of power rating factorof the main and exhaust fans.

Alarms Out of range laminar airflow velocity and front air barrier velocity, front window opened, HEPA filters clogging, UV lamp (if installed) needs replacing,charcoal filter (if installed) saturated, electrical power-out, signal of other possible malfunctions and relevant remedies.

Lighting level Fluorescent tubes in built-in housing, placed outside the contaminated area.Construction Epoxy powder painted steel for improved strength and resistance to corrosion and atmospheric agents.

Work surface: AISI-316 stainless steel solid work surface (one-piece). Frontal slots in the access area provide air barrier protection preventing anyexchange of air between the inside of the cabinet and the environment.*Work chamber: in AISI-304 stainless steel sheet. Front window: Stratified safety glass. The 3 degree sloping fronted window optimises the visibility inside the work chamber. When the cabinet isoff, the window can be opened to 465mm (maximum) or can be completely closed.

Filters HEPA filters, tested with D.O.P. aerosol. Their efficiency - higher than 99.995% MPPS - 0.005% penetration - ensures a performance exceeding therequirements of EN 1822,U.S. Federal Standard 209 e and AS 1386-1976. The filtration system is designed to provide a barrier of air with mini-mum average speed over 0.4m/sec. according to EN-12469.

Fans Two centrifugal fans, direct driven motors, with protection factor IP-55, controlled by two automatic speed regulators and microprocessor systemshowing laminar flow and protection barrier work conditions.

Electrical supply 230V, 50Hz, single phase.Noise level <59dB(A) @ 1 metre

* On request the following work surfaces are available: AISI-316 stainless steel perforated work surface in 1 or 4 pieces. AISI-316 stainless steel solid work surface in 3, 4 or 6pieces depending on the width of cabinet.

Page 5: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukFaster

Biological Safety Cabinets

138 Take a look at our latest products on www.ecomcat.co.uk

BHG Class II Biological Safety Cabinets with ElectronicallyOperated Sash Window and Double Fan, continued…SpecificationsModel BHG 2003-D BHG 2004-D BHG 2006-DExternal dimensions (wxdxh), mm 1090x865x1470 1395x865x1470 2005x865x1470Internal dimensions (wxdxh), mm 885x606x655 1190x606x655 1800x606x655Power, Kw 1.0 1.3 1.7Exhaust flow rate, m3/second 290 385 585

Cat. No 566-817 566-818 566-819

Price POA POA POA

The cabinets listed above are recirculating with double fan, double HEPA filter and electrically operated sash. For information and pricing on delivery, installation, commissioning and KI testing or to arrange a site survey, pleasecontact our Technical Office on 01342 826836.

Accessories for BHG cabinets

Cat. No. Description For Cabinets Price

566-803 Formalin vapouriser – See page 136 for details all models POA

566-820 UV light for sterilisation of the work chamber installed on the all models POAfront closure, controlled by two timers

566-821 Epoxy powder painted modular stand, 910mm working height BHG 2003 POA

566-822 Epoxy powder painted modular stand, 910mm working height BHG 2004 POA

566-823 Epoxy powder painted modular stand, 910mm working height BHG 2006 POA

566-812 Wheels for epoxy powder painted modular stand all stands POA

566-814 Exhaust charcoal filter with housing all models POA

566-815 Additional service connection for electric power all models POA

566-816 Additional service connection for gas/vacuum (manual tap) all models POA

566-824 Anti-blowback valve for connection to ductwork all models POA

566-802 Additional exhaust HEPA filter with relevant housing all models POA

Cytosafe-N 2000, Cytotoxic Safety Cabinets The Cytosafe-N 2000 cabinet has been designed to protect the operator (and the environment) fromcytotoxic contamination and microbial hazards as well as the material being manipulated from contamination.

The Cytosafe-N 2000 has been manufactured according to EN-12469 and DIN-12980 standards, withapproximately 70% recirculated laminar air flow and 30% air exhausted through a HEPA filter. Designed forhandling cytotoxic drugs, Cytosafe-N 2000 is also suitable for the treatment of class CDC 1-3, DHSS-C/B1/B2 group 1,2,3 micro-organisms and wherever product personnel environment protection is required.

The Cytosafe-N 2000 cabinet is available in three widths 0.9m, 1.2m and 1.8m.

Applications:Cytosafe-N 2000 cabinet is especially suitable for applications such as preparation and handling ofcytotoxic drugs, preparation and manipulation of anti-neoplastic chemo- therapeutics, presence ofagents likely to cause genetic alterations or synergetic activities with other materials.

Since Cytosafe-N 2000 cabinets also meet BS EN-12469 as for Class II Biohazard cabinets, they can alsobe used for work including microbiology, virology, haematology, cell culture and recombinant DNA.

■ Sophisticated microprocessor based monitoring system■ Constant monitoring of work conditions to ensure the correct laminar airflow and a

protective air barrier■ Two direct motor driven centrifugal fans with protection factor IP-55■ An automatic regulation system to maintain the airflow speed at 0.4 m/sec in the work chamber

and counter the effects of gradual clogging of the three HEPA filters■ Solid work surface in AISI-316 stainless steel (perforated available as option*) ■ Three HEPA filters including one under the work surface■ D.O.P. test inlet, IP-44 electric power point, solenoid valve and gas tap, vacuum line ■ Safe and easy replacement of HEPA filters according to DIN-12980 standard ■ Cabinets can be easily serviced through the front panel

* A filter cover is required (must be ordered separately).

566-075New

Page 6: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsFaster

Take a look at our latest products on www.ecomcat.co.uk 139

Cytosafe-N 2000, Cytotoxic Safety Cabinets, continued…SpecificationsFilters Three HEPA filters (including one under the work surface) with efficiency higher than

99.995% MPPS -0.005% penetration - ensures performance exceeding requirements ofEN1822, U. S. Federal Standard 209 e, British Standard 5295, VDI-2083 and AS 1386-1979.Under the work surface HEPA filter (multidihedral type).

Construction Corrosion resistant epoxy powder painted steel.Work surface: Solid AISI-316 stainless steel with radiussed angles and corners. Work chamber: AISI-304 stainless steel.

Fans Two centrifugal fans with direct drive motors. IP-55 protection factor. With automatic andindependent regulation to keep airflow speed constant (0.4m/sec) and counter theeffects of the gradual clogging of the three HEPA filters.

Lighting Fluorescent tubes in built-in housing, placed outside the contaminated area (>900 lux)Instrumentation The microprocessor monitoring system displays all relevant data with regard to the and controls operating functions, error messages and the different alarms.Alarms Out of range velocities, front window open, HEPA filters clogging, UV lamp worn out (if

installed), charcoal filter saturated (if installed), power outNoise level <60db(A) @ 1 metreElectrical supply 230V 50Hz, single phase** Other voltages on request.

SpecificationsModel Cytosafe-N 2003 Cytosafe-N 2004 Cytosafe-N 2006External dimensions (wxdxh), mm 1090x785x2245 1395x785x2245 2005x785x2245Internal dimensions (wxdxh), mm 885x580x660 1190x580x660 1800x580x660Power, kW 1.2 1.2 1.2Weight, kg (approx.) 242 329 463Exhausted air, m3/hr 290 390 580

Cat. No. 566-074 566-075 566-076

Price POA POA POA

The Cytosafe-N cabinets above are re-circulating, floor-standing models.

For information and pricing on delivery, installation and commissioning or to arrange a site survey, please contact ourTechnical Office on 01342 826836.

For special uses a version with an activated charcoal filter instead of the multidihedral HEPA filter, located under thework surface, is available. For details please contact our Technical Office on 01342 826836.

Accessories for Cytosafe-N cabinets

Cat. No. Description For Cabinets Price

566-077 UV lamp, factory fitted all models POA

566-815 Additional electric socket all models POA

566-816 Additional service connection for gas/vacuum (manual tap) all models POA

566-078 Filter cover to enable the installation of perforated work surface Cytosafe-N 2003 POA

566-079 Filter cover to enable the installation of perforated work surface Cytosafe-N 2004 POA

566-082 Filter cover to enable the installation of perforated work surface Cytosafe-N 2006 POA

For details on other options and accessories, please contact our Technical Office on 01342 826836.

Page 7: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

For more information on these and other products from Bioquellplease contact the Jencons-PLS Technical Office:

Tel: 01342 826836 Fax: 01342 826771

Email: [email protected]

Advanced HEPA and carbon filtration products

Complete range of sizes and options availablefrom the UK's leading cabinet manufacturer

After sales support, spares and expert technical advice

Largest network of service engineers in the UK

Unique Equipment Bio-Decontamination Service

Safe, affordable clean-air, biological control andcontainment solutions

from BIOQUELL

Page 8: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsMicroflow

Take a look at our latest products on www.ecomcat.co.uk 141

566-160ABS Class II Biological Safety Cabinets The Microflow range of Class II cabinets is designed to provide operator and product protection inaccordance with BS EN 12469.

The cabinets work on a recirculating principle with approximately 70% of the air drawn from theworking area being recirculated. Air then passes through the main HEPA filter mounted in the roof of thecabinet at a velocity in the region of 0.4m/sec. The unit is self balancing, drawing its make up air throughthe open front and then discharging through a high efficiency exhaust HEPA filter.

Cabinets can be ducted to the outside or operate in a recirculating mode. Recirculating models can havean optional carbon filter (option K) fitted for safe fumigation when using formalin. See page 142 fordetails on configuration options.

The Microflow ABS Class II safety cabinet has a unique combined sliding/hinged door mechanism. Thisoperates as a sliding front which can be fully closed to provide 100% protection during formalinfumigation without the need to tape the opening. The front door is also hinged to allow easy access tothe work area and services for loading and cleaning respectively.

■ Operator and product protection to BS EN 12469■ Continuous monitoring of cabinet status ■ Easy access, unique sliding/hinged door ■ Safety alarms for filter, airflow and window open ■ Microprocessor controlled ■ Automatic formalin sterilisation ■ Double exhaust HEPA filters for recirculating models ■ Solid stainless steel work surface ■ Services accessible from the front

ABS Class II Specifications - all modelsFilter, downflow 99.999% HEPA tested to EN 1822-H14Filter, exhaust 99.999% HEPA tested to EN 1822-H14Controls Microprocessor, mains, fan, lights, (UV and formalin vapouriser if fitted)Alarms Low air flow, filter change, front doorLighting 800 lux at work surfaceConstruction Powder coated mild steel outer with crevice-free 316 stainless steel inner liner

and work surfaceNoise level <65 dB(A) at 1mElectrical supply 230V 50Hz, single phase (other voltages available on request)

SpecificationsModel ABS1000 ABS1200 ABS1500 ABS1800External dimensions (wxdxh), mm* 1190x748x1380 1390x748x1380 1690x748x1380 1990x748x1380Internal dimensions (wxdxh), mm 1000x425x635 1200x425x635 1500x425x635 1800x425x635Weight, kg 170 190 230 250Power, kW 0.5 0.65 0.85 1.0Exhaust air volume, m3/hour 450 550 700 850

* Ducted models are 200mm higher. If you require a connection for Hydrogen Peroxide on the side of the cabinet,add 75mm to the width.

Cat. No Description Price

566-160 ABS1000 1.0m Recirculating/ducted cabinet 1, 2 POA

566-200 ABS1200 1.2m Recirculating/ducted cabinet 1, 2 POA

566-161 ABS1500 1.5m Recirculating/ducted cabinet 1, 2 POA

566-162 ABS1800 1.8m Recirculating/ducted cabinet 1, 2 POA

1. Cabinets can be ducted to the outside or operate in full recirculating mode. See page 142 for details onconfiguration options. One of these options must be specified in addition to the cabinet.

2. Installation and commissioning are not included. For information and pricing on delivery, installation,commissioning and KI testing or to arrange a site survey, please contact our Technical Office on 01342 826836.

For suitable accessories please see page 144.

New

Page 9: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukMicroflow

Biological Safety Cabinets

142 Take a look at our latest products on www.ecomcat.co.uk

ABS Class II Biological Safety Cabinets, continued…

Class II Cabinet ConfigurationsRecirculating - Option F This cabinet exhausts into the laboratory. It is therefore necessary to have a double HEPA filter on theexhaust. If this is your chosen option, it is necessary to decide how the formalin vapour will be disposedof, following fumigation (unless you are decontaminating with Hydrogen Peroxide).

Recirculating with carbon filter - Option K This non-ducted, recirculating cabinet is designed for situations where there is no means of disposing offormalin following decontamination. This option is not required if you are decontaminating withHydrogen Peroxide.

Ducted <2m - Option D This cabinet exhausts via a length of ductwork which must be less than 2 metres in length. Normally,the ductwork runs to an exterior wall or window. It is therefore necessary to ensure that there is no riskof exposure to any passers-by, adjacent wards or laboratories.

Note; a ductwork run of <2 metres terminates in a high-velocity discharge spout as standard. If yoursituation demands a cowl (chinaman’s hat), there will be an added cost.

Please note that the cost indicated below does not include fitting, nor does it include any costs forbuilding work (e.g. making a hole in the wall or window). The cost for fitting will be included in aquotation following a detailed site survey. To arrange for a site survey and for advice on siting pleasecontact our Technical Office on 01342 826836.

Ducted >2m - Option C Where long distances over 2 metres are required, an option C cabinet should be selected. Please notethat a remote fan on top of the roof would also be required. If an outside contractor is to supply the fanand ductwork, it is imperative that they contact the Jencons-PLS Technical Office on 01342 826836, forthe appropriate specifications.

Please note that the cost indicated below does not include fitting, nor does it include any costs forbuilding work (e.g. making a hole in the wall or window). The cost for fitting will be included in aquotation following a detailed site survey. To arrange for a site survey and for advice on siting pleasecontact our Technical Office on 01342 826836.

Configuration Options for Class II Safety Cabinets

Cat. No Description For Cabinets Price

Option F Recirculation, double HEPA

566-028 Double HEPA filter with side exhaust hood and flap valve all models POA

Option K Recirculation, double HEPA with carbon filter

566-029 Double HEPA filter with flap valve and a carbon filter/fan all models POAsystem for formalin fumigation

Option D Ducted, <2m*566-027 Side exhaust hood with fan and flap valve all models POA

Option C Ducted, >2m**566-026 Side exhaust hood with flap valve for use with remote fan all models POA

* Includes exhaust fan module discharge spout, 2 off 90° bends, 2m duct hose, 2 off 45° bends, 2 sockets, softsleeve (160mm diameter), 500ml PVC solvent, 500ml MEK cleaner, 2 pipe bracket stand offs, 2 off pipe fixingbrackets and four woodscrews

** Remote fan (566-195) and ductwork not included. See page 144 for accessories.

Other configurations are available including: Low-level options where height is restricted; double HEPA ductedcabinets; special sized cabinets for robots. Please contact our Technical Office on 01342 826836 for details.

For suitable accessories please see page 144.

566-160New

Page 10: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsMicroflow

Take a look at our latest products on www.ecomcat.co.uk 143

566-023ABS Class I Biological Safety CabinetsThe Microflow ABS Class I biological safety cabinet offers operator protection in accordance with BS EN 12469.

■ Operator safety with alarms for filter and airflow■ Safe working environment with critical safety parameters monitored at all times■ Easy-to-clean, crevice-free, stainless steel construction■ Front screen has a dual sliding/hinged mechanism for easy access to services and the work area■ Unique sliding/hinged door gives 100% protection during fumigation process ■ Automatic sterilisation; the formalin vapouriser (if used) is controlled via the microprocessor panel

ABS Class I Specifications - all modelsFilter 99.999% HEPA tested to EN 1822-H14Controls Microprocessor, mains, fan, lights, (UV and formalin vapouriser if fitted)Alarms Low air flow, filter change, audible and visibleLighting 1000 lux at work surfaceConstruction Powder coated mild steel outer with crevice free 316 stainless steel

inner liner and work surface Noise level <65 dB(A) at 1mElectrical supply 230V 50Hz, single phase, other voltages available on request

Specifications

Model ABS1000 ABS1200 ABS1500

External dimensions (wxdxh), mm* 1190x748x1580 1390x748x1580 1690x748x1580Internal dimensions (wxdxh), mm 1000x640x635 1200x640x635 1500x640x635Weight, kg 189 209 234Power, kW 0.5 0.65 0.85Exhaust air volume, m3/hour 550 670 850

* If you require a connection for Hydrogen Peroxide on the side of the cabinet, add 75mm to the width.

Cat. No Description Price

566-023 ABS1000 1.0m Ducted cabinet 1, 2 POA

566-025 ABS1200 1.2m Ducted cabinet 1, 2 POA

566-024 ABS1500 1.5m Ducted cabinet 1, 2 POA

1. Includes anti blow-back valve. Class I Biological Safety Cabinets are supplied with an exhaust fan kit consisting ofdischarge fan assembly, 4 meters 160mm o.d. PVC ducting, 3 off 90º and 1 off 45º bends, 2 off sockets, 4 off pipefixing brackets, discharge span, 6 meters Hi-tuf cable and all necessary fan fixings.

2. A recirculating option is available; please contact our Technical Office on 01342 826836.

ABS Class III/I Biological Safety CabinetThe Microflow ABS Class III/I biological safety cabinet is a dual purpose cabinet designed for laboratorieswhere it is essential to have a Class III facility for work on Group 4 pathogens, but there may also be arequirement for Class I work.

■ Operator protection in Class I mode■ Operator and product protection in Class III mode ■ Transfer hatch and glove port panel■ Tested to BS5726 and BS EN 12469

Accessories■ Service connections (gas, electric and vacuum)■ Formalin vapouriser■ UV lights■ Adjustable stands

SpecificationsModel ABS1000 ABS1200External dimensions (wxdxh), mm 1821x748x1580 2021x748x1580Internal dimensions (wxdxh), mm 1000x640x635 1200x640x635Weight, kg 210 230Power, kW 0.65 0.75Exhaust air volume, m3/hour 630 790

For more details on the Microflow ABS Class III/I cabinet, please contact our Technical Office on 01342 826836.

Class III/I BiologicalSafety Cabinet

New

New

Page 11: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukMicroflow

Biological Safety Cabinets

144 Take a look at our latest products on www.ecomcat.co.uk

Accessories for ABS Biological Safety CabinetsAccessories for ABS Class II Biological Safety Cabinets

Cat. No. Description For Cabinet Price

566-180 Stand with feet ABS1000 POA

566-181 Stand with feet ABS1200 POA

566-186 Stand with feet ABS1500 POA

566-182 Stand with feet ABS1800 POA

566-183 UV tube and fittings (requires UV socket, see below) ABS1000 POA

566-184 UV tube and fittings (requires UV socket, see below) ABS1200 POA

566-187 UV tube and fittings (requires UV socket, see below) ABS1500 POA

566-185 UV tube and fittings (requires UV socket, see below) ABS1800 POA

566-188 UV socket, RHS* all models POA

566-213 ABS Double electrical socket, RHS* all models POA

566-216 Solenoid operated gas tap, RHS* all models POA

566-217 Vacuum tap, RHS* all models POA

566-203 Fixed formalin vapouriser all models POA

566-193 Flexible ducting for formalin evacuation all models POA

566-194 Formalin evacuation unit (can service several cabinets in succession). all models POAA mobile unit comprising a fan/filter box, complete with adaptor to clamp onto exhaust of safety cabinet. Filter included.

566-195 Remote exhaust Fan Pack (for >2m ductwork). Comprising; all models POAdischarge fan assembly, speed control board, 4mx160mm PVC ducting, 3 off 90° bends and 1 off 45° bends, 2 off sockets, 4 off pipe fixing brackets, 1 off discharge spout, 6m Hi-Tuf cables and all necessary fan fixings

566-198 Volt free contact all models POA

566-196 Installation and commissioning** all models POA

Accessories for ABS Class I Biological Safety Cabinets

Cat. No. Description For Cabinet Price

566-180 Stand with feet ABS1000 POA

566-181 Stand with feet ABS1200 POA

566-186 Stand with feet ABS1500 POA

566-183 UV tube and fittings (requires UV socket, see below) ABS1000 POA

566-184 UV tube and fittings (requires UV socket, see below) ABS1200 POA

566-187 UV tube and fittings (requires UV socket, see below) ABS1500 POA

566-188 UV socket, RHS* all models POA

566-213 ABS Double electrical socket, RHS* all models POA

566-216 Solenoid operated gas tap, RHS* all models POA

566-217 Vacuum tap, RHS* all models POA

566-203 Fixed formalin vapouriser all models POA

566-198 Volt free contact all models POA

566-196 Installation and commissioning** all models POA

* If Hydrogen Peroxide is used for fumigation, then all services must be fitted on the left hand side. For pricingdetails please contact our Technical Office on 01342 826836.

** Please note that installation costings do not include any alterations to the fabric of the building, e.g.; drillingholes through windows/ walls for ductwork. Nor do they include the cost of any scaffolding that may be required inorder to install ductwork above the first floor level.

For information and pricing on delivery, installation, commissioning and KI testing or to arrange a site survey, pleasecontact our Technical Office on 01342 826836.

Other accessories are available including stands with castors and high level UV lighting. Please contact our TechnicalOffice on 01342 826836 for details.

566-160New

566-023New

Page 12: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsBIOQUELL

Take a look at our latest products on www.ecomcat.co.uk 145

EBDS - Equipment Bio-Decontamination ServiceEBDS is a unique service from BIOQUELL usingthird generation hydrogen peroxide vapourtechnology to bio-decontaminate laboratoryequipment - including CO2 incubators and ClassII cabinets. One advantage of EBDS is that theuser does not have to decontaminate theequipment prior to an engineer’s visit. BIOQUELL’sservice engineers carry out the EBDS and thenservice the equipment.

EBDS uses BIOQUELL’s unique low temperature,residue-free hydrogen peroxide vapourtechnology to bio-decontaminate a broad rangeof laboratory equipment including:

■ Biological safety cabinets■ Other HEPA filtration enclosures■ CO2 incubators■ Incubators■ Centrifuges■ Robotic equipment■ Other laboratory equipment

Hydrogen Peroxide has proven efficacy against:

■ Bacterial spores (e.g. Geobacillus stearother-mophilus, Anthrax)

■ Bacteria (e.g. Staphylococcus, Pseudomonas,Mycoplasma)

■ Viruses (e.g. Adenovirus, Parvovirus, Poliovirus)■ Fungi (e.g. Aspergillus, Candida, Penicillium)

BIOQUELL also provides a Room Bio-Decontamination Service (RBDS) using itsunique hydrogen peroxide vapour technology(see page 146).

Key advantages of EBDSHealth & Safety■ Eradicate the use and handling of formaldehyde in the laboratory (by laboratory staff, engineers

or others)■ EBDS technology holds the equipment at negative pressure ensuring that the vapour cannot leak

into the laboratory■ Hydrogen peroxide is converted to water and oxygen at the end of the process – hence no

residues to affect research or employees

Efficiency■ No need to close the laboratory or decontaminate overnight – EBDS can take place whilst work

in the laboratory continues■ Significantly faster bio-decontamination cycles than using formaldehyde■ Excellent efficacy against a range of bacteria (including spores), viruses and fungi

Material compatibility■ First rate material compatibility characteristics – no damage to electronics or other sensitive

equipment■ Over 860 gassing cycles repeated back to back on a biological safety cabinet resulted in

zero defects

Cost savings■ The efficiency, benefits and Health & Safety advantages represent tangible cost savings

For more information and a quotation for your specific application, please contact our Technical Office on 01342 826836.

Page 13: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.ukBIOQUELL

Biological Safety Cabinets

146 Take a look at our latest products on www.ecomcat.co.uk

Room Bio-DecontaminationService (RBDS)Hydrogen peroxide bio-decontamination can beused in laboratories and other rooms in thehealthcare, pharmaceutical, biotechnology, foodand biomedical sectors.

Dealing with contaminated rooms has alwaysbeen difficult - safety and efficacy seemed topresent conflicting demands. Hydrogen peroxidevapour has been used extensively in the pharma-ceutical industry for many years, as the vapour iscompatible with most materials includingelectronics and has been proven to be highlyeffective - particularly against bacterialendospores.

Once the bio-decontamination process iscomplete, the vapour can be broken down tooxygen and water leaving no other residues.

BIOQUELL’s Room Bio-Decontamination Service(RBDS) offers these benefits to pharmaceutical,biotechnology, biomedical, healthcare, food andenvironmental applications. This unique offeringprovides a scheduled or emergency servicewherever bio-burden is suspected or needs to bede-activated.

High-level verification is assured as each cycle ischallenged with 6-log Geobacillus stearother-mophilus biological indicators, the same used tovalidate steam steriliser decontamination cycles.

Having proven efficacy against bacteria, virus,fungus and moulds, Hydrogen peroxide vapour isfast becoming the safer alternative method ofbio-decontamination, as opposed to traditionalmethods such as formaldehyde - which is aknown human carcinogen.

BIOQUELL’s Hydrogen peroxide technology offers:

■ Residue free, non carcinogenic and infinitely scalable

■ Fully validatable 6-log bio-reduction■ Excellent material compatability and

minimal down-time■ Consultancy and validation■ Room bio-decontamination service,

patented technology

For more information, and a quotation fordecontaminating your room, please contact ourTechnical Office on 01342 826836

New

Page 14: Biological Safety Cabinets - uk.vwr-cmd.comuk.vwr-cmd.com/ex/downloads/jencons/pdf/catalogue/bio_safe_cab.pdf · Our responsibilities as a supplier of safety cabinets are also detailed

Order by phone➤ 01342 826836 Order on-line➤ www.ecomcat.co.uk

Order by fax➤ 01342 826771 Website➤ www.jencons.co.uk

Biological Safety CabinetsK-SYSTEMS

Take a look at our latest products on www.ecomcat.co.uk 147

IVF Cabinets, Laminar Flow Cabinets and Anti-Vibration TablesThe K-SYSTEMS™ range of laminar flow products is divided into 3 different areas:

■ Workstations for IVF offering a range of built-in/integrated features and option■ Workstations with biohazard functionality (Class II)■ Low noise laminar flow hoods

All the above can be supplied with electrically height-adjustable stands toaccommodate users of different statures.

The IVF workstations are custom built to house items such as microscopes, hot tablesand even mini CO2 incubators. They are available in 1.0, 1.2, 1.5 and 1.8 metre widths.The work-surface can be temperature controlled and stereo microscopes and lightbases can be integrated into the base. The larger cabinets are designed for 2 workers– one side for an inverted microscope and the other for a stereo microscope or a stereomicroscope either side. An anti-vibration table can be integrated into the design toensure perfect microscopy techniques.

Standard Features■ Most makes and sizes of microscopes can be built-in■ Microscopes can be positioned to suit the user■ Choice of light base■ Vertical laminar flow utilising two filters – one pre-filter and a Class

H14 HEPA filter ■ Stainless steel work-surface with heated areas marked slightly darker■ The front of the work-surface is ergonomically shaped■ Noise level is just 51 dB(A)

Optional FeaturesSome optional features that can be built-in to the K-SYSTEMS™ workstations include:

■ Larger heated work-surface■ Electrically adjustable height stand (from 694 to 1094mm)■ Heated glass microscope stage■ AV/PC monitor can be integrated in rear panel■ Pass through hatch can be integrated in rear panel■ Mini CO2 incubator can be built-in

Laminar Flow HoodsK-SYSTEMS™ laminar flow hoods can be supplied as Class I (product protection only)or Class II (product and operator protection) models. Please note that these cabinetsare supplied without services and options are not available.

Anti-Vibration TablesK-SYSTEMS™ also manufacture anti-vibration tables that are ideal for use with, inparticular, research-style inverted microscopes such as the Nikon TE-2000. Vibrationsare eliminated by a complex anti-vibration system, achieved through an optimisedchoice of mass and support, ensuring a fast damping effect. The inverted microscopeis placed on an equilibrated and isolated platform that gives a smooth and roomyworkspace around the microscope for micromanipulation equipment, mini incubators,workbooks etc. The table is available in two widths (820 and 1200mm) and can bemade to suit all makes of inverted microscopes, including motorised models.

For further information of K-SYSTEMS™ equipment please contact our Technical Officeon 01342 826836

Laminar flow cabinet with built-in microscope New

Laminar flow cabinet with two built-in microscopes

New