Best Practices: Transonic® Hemodialysis Surveillance

Hemodialysis

• Vascular Access Flow

• Dialysis Adequacy

• Cardiac Function

Page 2: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

AMERICASTransonic Systems Inc.34 Dutch Mill RdIthaca, NY 14850U.S.A.Tel: +1 607-257-5300Fax: +1 [email protected]

EUROPETransonic Europe B.V.Business Park Stein 2056181 MB ElslooThe NetherlandsTel: +31 43-407-7200Fax: +31 [email protected]

ASIA/PACIFICTransonic Asia Inc.6F-3 No 5 Hangsiang RdDayuan, Taoyuan County33747 Taiwan, R.O.C.Tel: +886 3399-5806Fax: +886 [email protected]

JAPANTransonic Japan Inc.KS Bldg 201, 735-4 Kita-AkitsuTokorozawa Saitama359-0038 JapanTel: +81 04-2946-8541Fax: +81 [email protected]

www.transonic.com

Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion, interventional radiology and research applications. In addition, Transonic provides pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems.

HemodialysisCover (DL-100-fly) Rev A 2014

Hemodialysis

“The Transonic® Flow-QC® Hemodialysis Monitor has benefited numerous dialysis-dependent patients by reducing and, in many cases, eliminating the agony of a clotted AV graft or fistula, thereby facilitating correction of access stenoses on an elective basis that prevents missed dialyses and the need for placing temporary catheters.”

Depner, T, MD, UC Davis

“The comprehensive cost is markedly reduced due to the decreased number of hospitalizations, catheters placed, missed treatments, and surgical interventions. Vascular access blood flow monitoring along with preventive interventions should be the standard of care in chronic hemodialysis patients.”

McCarley, P et al, Kid Int 2001; 60(3).

“Transonic flow monitoring is the cornerstone of my Vascular Access Management Program (VAMP).”

Spergel, LM, MD, FACS, Dialysis Management Medical Group, San Francisco

Flow-based Vascular Access Management

ULTRASOUND DILUTION TECHNOLOGY

Vascular Access SurveillanceFlow-based surveillance per KDOQI Guidelines alerts a patient care team to patients at risk for underdialysis, thrombotic events and cardiac failure.

“The [HD03] allows you to proactively manage your patients as part of a multidisciplinary vascular access care program to reduce complications and costs of end-stage renal disease.” Duda, CR et al, Nephrology News & Issues, 2000; 14(5).

Dialysis AdequacyEnsure adequate dose delivery by direct, accurate measurement of pump blood flow and access recirculation in AV accesses and catheters.

“Recirculation and delivered blood flow measurements provide important information about the most appropriate and effective therapies for your patients.”Sands, JJ et al, ASAIO, 1996; 42(5). Depner, TA et al, ASAIO 1995; 41(3).

Cardiac Function Cardiac Output profiling with a Transonic® HD Monitor identifies patients with dangerously low Cardiac Index due to inadequate dry weight estimation.

“Integration of cardiac output measurements into an ESRD treatment program forestalls the devastating progression of cardiovascular disease.” MacRae, JM et al, Am J Kid Dis 2004; 43(5).

Krivitski Method® Access Flow Measurement: Dialysis lines are reversed at their needle connections to induce recirculation through the access. Vascular access flow can then be calculated from the change in blood concentration detected by the paired flow/dilution sensors. To measure recirculation, saline is introduced into the venous line with the dialysis lines in normal position. Recirculation is calculated from the change in blood concentration between the venous sensor and arterial sensor.

Page 3: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

HD03Surveillance(DL-101-ds) Rev B 2014

Hemodialysis

Transonic® HD03Hemodialysis Monitor“Gold Standard”Surveillance

The battery-operated, portable HD03 Monitor & Flow/dilution Sensors Measure:

• Delivered Blood Flow • Vascular Access Recirculation • Vascular Access Flow• Cardiac Output (optional)

Measurements take less than 10 minutes per patient.

Results are displayed immediately on the Monitor.

Page 4: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

www.transonic.com

HD03 MonitorMeasurementsPerformed on AV fistulas, grafts and catheters during routine dialysis.

Easy to UseMeasurements are operator independent. Software guides user step-by-step through procedure.

Easy Set-upSimply clip the arterial and venous flow/dilution sensors onto the respective blood lines.

PortableRechargeable battery permits easy mobility between patients.

Safety/Infection ControlTouch-screen input prevents cross contamination. The screen can be cleaned with a dilute solution of bleach or soap.

H4FX Flow/Dilution SensorsPaired sensors pass ultrasound waves through dialysis tubing to measure blood flow and other parameters

• Sensors clip onto tubing connected to the patient’s blood lines.

• Saline can be released directly from saline bag or infused into the dialysis circuit.

Administrative Software Data ManagementA removable Data Transfer Module can be uploaded to a computer with HD03 Administration software, and information is synchronized between the Monitor and computer.

Powerful• Documents and trends interventions and access history• Generates High Risk Thrombosis “Alert” List• Permits schedule planning• Calculates individual patient and clinic statistics• Displays comprehensive Patient Status Reports

HD03 Features & Benefits

The portable HD03 Monitor on a rolling stand.

Page 5: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

Hemodialysis

Best Practices inHemodialysis CareAccess Patency • Dialysis Adequacy • Cardiac Function

Hemodialysis

HemodialysisWhite Paper DL-200-wp, Rev A 2015 USltr

The singular purpose of an AV vascular access is to serve as a conduit for sufficient blood flow to sustain hemodialysis delivery. Inadequate flow causes underdialysis; too much flow can lead to cardiac problems. Each has associated morbidities and can lead to serious complications that even include death.

Transonic’s ultrasound dilution technology is the universally recognized gold standard measurement technology for hemodialysis surveillance.

The

method uses Transonic Flow-QC® Hemodialysis Monitors and Flow/dilution Sensors to directly: • Measure Dialysis Adequacy:

(delivered blood flow and recirculation) for on-the-spot identification and correction of dose delivery problems in AV access and central venous catheters;

• Trend Vascular Access Flow: to detect flow limiting problems wherever they occur in a vascular access. Flow-based surveillance per KDOQI Guidelines alerts the patient care team to patients at risk thrombotic events.

• Track Cardiac Function: to identify patients with dangerously low Cardiac Indices at risk for cardiac failure from inadequate dry weight estimation.

“Adequate blood flow in peripheral hemodialysis fistulae and grafts is vital to the success of hemodialysis and to the survival of the patient. Reduction in flow . . . presages failure of the access device itself. Access flow can therefore be considered a fundamental property of the access that should be monitored.” Depner, TA et al

Page 6: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

B e s t P r a c t i c e s i n H e m o d i a l y s i sFlow-QC Protocol: AV Graft and Fistula Surveillance

Dialysis Adequacy: Delivered Blood Flow; RecirculationThe Transonic Hemodialysis Monitor is used to optimize efficient dialysis delivery through measurement of delivered pump blood flow and recirculation in an AV access and central venous catheters. These measurements are used to:

• Verify true delivered blood flow; • Test the calibration of the blood pump; • Avoid underdialysis through inadvertent reversal of the dialysis lines; • Detect and quantify access recirculation in an AV access, or central venous catheter; • Help determine proper needle placement; • Maximize catheter function; • Identify sources of large negative arterial blood line pressure that causes underdialysis;• Determine the most appropriate blood pump setting for a low flow access when access flow

can’t be increased.

A Flow-based Access Management Protocol includes an initial dialysis adequacy study, access patency surveillance and a cardiac function assessment.

Dialysis Adequacy/Vascular Access Patency Cardiac Function

Initial Dialysis Adequacy & Vascular Access Patency

Measure initial Qb, AR, AF,

Nephrologist Analysis• Establish Access Flow levels • Establish testing schedule (monthly per KDOQI)

After flow restoration procedure

StartNew patient or revised

vascular access

YES

Vascular Access Patency Study

(monthly, or other interval)

Nephrologist AnalysisSchedule appropriate flow restoration procedure or refer patient for duplex scan and/or fistulogram.

Is AF above the critical threshold?

NO

Nephrologist Review

Acceptable

Initial Cardiac Function Study Hourly CO tests (CHP Study) performed

when cardiac complications are suspected during hemodialysis.

Prescription Concern Cardiovascular

Concern

Baseline Cardiac Studies A second CHP study and third, one

month later, establishes reliable aver-age cardiac function

parameters for the patient.

Nephrologist Analysis Set cardiac baseline values,

warning levels, testing schedule.

Follow-up Cardiac Function Study

CHP study performed after a weekend break.

Nephrologist Review

Acceptable

Acceptable

Further Studies, and/or Treat-ments

Cardiovascular Concern

Nurses’ AnalysisOptimize dialysis delivery immediately from Qb, AR results. Repeat at established intervals.

Page 7: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

B e s t P r a c t i c e s i n H e m o d i a l y s i s

Fig. 2: Access Flow Trend: This flow history of a patient’s AV access over one year shows that the onset of stenoses were identified by decreases in flow below 600 mL/min. Interventions, indicated by the inverted arrows, resulted in immediate increases in access flow.

Hemodialysis Surveillance: Trending Vascular Access Flow

“A hemodynamically significant stenosis is the substrate for thrombosis by reducing flow, increasing turbulence, and increasing platelet activation and residence time

against the vessel wall.” KDOQI Guidelines 20061

As a corner stone for a comprehensive Vascular Access Management Program, Transonic Flow-QC hemodialysis surveillance measures:

• Actual function in AV grafts and fistulas in order to identify failing accesses and avert underdialysis and/or thrombosis;

• Excludes access dysfunction quickly as cause of underdialysis;• Identifies a mid-access obstruction and high-flow versus low flow accesses to select ideal

treatment plan for flow-restricting versus re-vascularization correction procedures;• Indicates effectiveness of interventions through post-intervention surveillance;• Permits access surveillance to be performed by the clinic’s staff who then can alert the

nephrologist to possible onset of access dysfunction.Transonic Flow-QC surveillance implements KDOQI Guidelines.

Vascular Access PatencyThe Kidney/ Dialysis Outcomes Quality Initiative (KDOQI) Clinical Practice Guidelines for Vascular Access call for access flow monitoring by stating “prospective surveillance of AV grafts and fistulas for hemodynamically significant stenosis, when combined with correction, improves patency and decreases the incidence of thrombosis.”

5 Canadian, Australian and European Guidelines also call for surveillance during hemodialysis to forestall stenosis formation and prolong the life of the access. Intra-access measurements (ultrasound dilution technology) are cited as the preferred method for surveillance.1

Transonic hemodialysis surveillance tracks a patient’s vascular access flow over time. If access flow decreases below a critical threshold, fistulograms or interventions can be scheduled proactively to delay access failure. Such early intervention with minimally invasive restorative flow procedures reduces morbidity and costs.7 The clinic can continue to administer dialysis, collect and analyze data, and reduce its dependence on outside services for lab tests.

Page 8: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

AMERICASTransonic Systems Inc.34 Dutch Mill RdIthaca, NY 14850, U.S.A.Tel: +1 607-257-5300Tel: +1 800-353-3569 (US, CAN)Fax: +1 [email protected]

EUROPETransonic Europe B.V.Business Park Stein 2051185 MB ElslooThe NetherlandsTel: +31 43-407-7200Fax: +31 [email protected]

ASIA/PACIFICTransonic Asia Inc.6F-3 No 5 Hangsiang RdDayuan, Taoyuan County33747 Taiwan, R.O.C.Tel: +886 3399-5806Fax: +886 [email protected]

JAPANTransonic Japan Inc.KS Bldg 201, 735-4 Kita-AkitsuTokorozawa Saitama359-0038 JapanTel: +81 4-2946-8541Fax: +81 [email protected]

www.transonic.com

Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion, interventional radiology and research applications. In addition, Transonic provides pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems.

Hemodialysis

References1 Guidelines for Vascular Access 2006 Updates,

http://www.kidney.org/professionals-/kdoqi/guideline_upHD_PDVA/va_guide4.htm.

2 European Renal Association-European Dialysis and Transplant Association (ERA-EDTA), European Best Practice Guidelines on Haemodialysis: Guideline 5.2. Nephrol Dial Transplant 2007; 22(Suppl 2): ii99-ii100.

3 Polkinghorne KR et al, “KHA-CARI Guideline: Vascular access – central venous catheters, arteriovenous fistulae and arteriovenous grafts.” Nephrology 2013; 18(11): 701-5.

4 Nesrallah GE et al, “Canadian Society of

Nephrology guidelines for the management of patients with ESRD treated with intensive

hemodialysis,” Am J Kid Dis. 2013; 62(1): 187-98.5 Krivitski NM, Depner TA, “Development of a

Method for Measuring Hemodialysis Access Flow: From Idea to Robust Technology,” Seminars in Dialysis 1998; 11(2): 124-130. (Transonic Reference # HD317A)

6 Spergel LM,“Transonic Flow Surveillance — The Cornerstone of My Vascular Access Management Program (VAMP©)” or Clinical Applications of the Transonic Flow Monitor in the Hemodialysis Facility. (Transonic Focus Note # DL-58-fn)

7 McCarley P et al, “Vascular Access Blood Flow Monitoring Reduces Access Morbidity and Costs,”

Kidney Int 2001; 60(3): 1164-1172. (Transonic Reference # HD317A)

8 Depner TA et al, “Pressure Effects on Roller Pump Blood Flow during Hemodialysis,” ASAIO Trans 1990; 31: M-456-M459. (Transonic Reference # HD10A

9 Krivitski NM et al, “Measurement of Cardiac Out-put and Central Blood Volume during Hemodialysis: Sources of Error,” 25th International Congress of Nephrology Abstract 1999; 333. (Transonic Refer-ence HD95A)

10 Tucker T et al, “Central Hemodynamic Profiling (CHP) during Outpatient Hemodialysis (HD),” J Am Soc of Nephrol Abstr 2002; 13: 209A. (Transonic Reference # HD268A)

ConclusionIn the current outcomes-driven climate of proactive end-stage renal disease (ESRD) care, Transonic flow measurements are integral to successful and comprehensive vascular access management during hemodialysis. The measurements inform and guide the nephrologist and hemodialysis care team as they seek to optimize dialysis delivery, maintain a healthy access, and oversee the cardiovascular health of their patients. Transonic flow-based surveillance during hemodialysis is the corner stone for Hemodialysis Best Practices.

Cardiac Function TestsCardiovascular complications in ESRD patients can be averted or forestalled through routine cardiac function tests with the Hemodialysis Monitor. Reported are Cardiac Output (CO), Cardiac Index (CI), Peripheral Resistance (PR), Central Blood Volume (CBV), Central Blood Volume Index (CBVI) and ratio of Access Flow to Cardiac Output (AF/CO).

“Hemodynamic stability is threatened and often severely compromised by hemodialysis largely because of the obligate fluid removal

during a short time span.” Depner, TA

Central Hemodynamic Profiling (CHP) during hemodialysis tracks the heart’s response to the stress of a dialysis treatment and identifies: 1) Prolonged high access flow that can lead to cardiomegaly and high output cardiac failure. 2) Low Cardiac Index (<2 L/min/m2) which offers the physician an opportunity to adjust dry weight

medication and length of dialysis to improve CI.10 3) Dramatic Cardiac Output drops during dialysis due to inaccurate dry weight estimation and/or

medication that places patients at high risk for cardiovascular complications and sudden death following the dialysis session.

Central Hemodynamic Profiling (CHP): CO measure-ments taken during a single hemodialysis session show Cardiac Index responses to the hemodialysis treatment.11

Page 9: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

Why Perform Surveillance?

Hemodialysis

7 Ways the HD03 Improves Outcomes for Your Patients

The Transonic® Gold Standard Hemodialysis Monitor Assures KDOQI Quality Compliance through: • Dialysis Adequacy Optimization • Vascular Access Surveillance • Cardiac Function Assessment

HD03Quality Assurance(DL-160-fly)RevA 2013

Page 10: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

1. Identifies Discrepancy Between Pump Setting & Delivered Blood Flow As A Result of:

HD03 Hemodialysis MonitorH e m o d i a l y s i s A d e q u a c y O p t i m i z a t i o n

4. Optimizes Dialysis in Dual-lumen Catheters

• Negative pressure effects of the roller pump• Condition of access• Needle size• Needle placement

• Kinked or occluded tubing• Calibration of dialysis machine• Change in type of dialysis tubing• Calibration of Transonic® Flow/dilution Sensors

Catheter recirculation is an early sign of catheter failure and usually depends on dialysis blood flow. The patient in Fig. 4 was dialysed at flows up to 300 mL/min without any recirculation. At flows higher than 300 mL/min, such as 450 mL/min shown in Fig. 5, 19% recirculation occurred.

Therefore, increasing delivered blood flow (Qb) did not proportionally increase the quality of dialysis.

Note: Discrepancies between pump flow and real deliv-ered flow can also be more dramatic with cath-eters than with vascular accesses.

2. Ensures Correct Needle PlacementWhen Transonic® Hemodialysis Monitoring first shows vascular access recirculation (Fig. 1), which disappears after the blood lines are reversed and the recirculation measurement is repeated (Fig. 2), the hemodialysis lines have been inadvertently reversed.

Transonic® Hemodialysis Monitor screenings show that dialysis occurs with the needles inadvertently reversed in more than 4% of cases.

blue — venous curvered — arterial curve

blue — venous curvered — arterial curvelines reversed

3. Confirms 0 % RecirculationIn contrast to measurement technologies that cannot sepa-rate vascular access recirculation from cardiopulmonary re-circulation and, therefore, show false positives, Transonic® Hemodialysis Monitoring can separate access recirculation from the cardiopulmonary (the red curve in Fig. 3) and can report zero % recirculation.

blue — venous curvered — arterial curve

blue — venous curvered — arterial curve

blue — venous curvered — arterial curve

Page 11: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

V a s c u l a r A c c e s s S u r v e i l l a n c e

7. Cardiac Output Check Indicates Potential Cardiac Overload

6. 0% Recirculation & Low Access Flow Pinpoints Stenoses Between NeedlesWhen a significant stenosis is located between the hemo-dialysis needles, hemodialysis pump flow simply bypasses the stenosis without producing any recirculation.

When low access flow (Fig. 9) is accompanied by 0% recir-culation (Fig. 8), a stenosis between the dialysis needles can be suspected. A stenosis between the needles can be confirmed by a color Doppler image.

In the case example on the right, vascular access flow measured more than 3 L/min (Fig. 10). Cardiac output exceeded 10 L/min (Fig. 11). When the vascular access was briefly occluded by the tip of the examiner’s finger, the patient’s pulse rate dropped from 112 to 88 per min. This patient had complained of chest pains and had been diagnosed with cardiomegaly.

The access was surgically revised by banding. Following the revision, access flow then measured 1700 mL/min. Cardiac output dropped to 7-8 L/min. The patient exhib-ited fewer post-dialysis hypotensive episodes, his dry weight decreased, his chest X-Ray cleared and he report-ed significant improvement in his previous symptoms.

blue — venous curvered — arterial curvelines reversed

red — arterial curve

blue — venous curvered — arterial curve

blue — venous curvered — arterial curvelines reversed

5. Recirculation with Low Access Flow Detects Significant Inflow/Outflow StenosesUnlike other technologies that can only identify outflow stenoses in AV accesses, HD03 Monitor surveillance can detect a stenosis wherever it occurs within the vascular access circuit: inflow, outflow or between the needles in both fistulas or grafts.

In the example on the right, access recirculation (Fig. 6), accompanied by low vascular access flow (Fig. 7), indicated the presence of a significant stenosis which was then confirmed by color Doppler and fistulogram.

blue — venous curvered — arterial curve

blue — venous curvered — arterial curvelines reversed

Hemodialysis Quality Assurance

Page 12: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

AMERICASTransonic Systems Inc.34 Dutch Mill RdIthaca, NY 14850U.S.A.Tel: +1 607-257-5300Fax: +1 [email protected]

EUROPETransonic Europe B.V.Business Park Stein 2056181 MB ElslooThe NetherlandsTel: +31 43-407-7200Fax: +31 [email protected]

ASIA/PACIFICTransonic Asia Inc.6F-3 No 5 Hangsiang RdDayuan, Taoyuan County33747 Taiwan, R.O.C.Tel: +886 3399-5806Fax: +886 [email protected]

JAPANTransonic Japan Inc.KS Bldg 201, 735-4 Kita-AkitsuTokorozawa Saitama359-0038 JapanTel: +81 04-2946-8541Fax: +81 [email protected]

www.transonic.com

Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion, interventional radiology and research applications. In addition, Transonic provides pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems.

Hemodialysis Adequacy

Summary: Flow-based Quality Assurance

Hemodialysis

• Tells actual function in AV grafts and fistulas in order to identify failing accesses and avert underdialysis and/or thrombosis;

• Indicates effectiveness of interventions (post-intervention surveillance) or limb ischemia; • Excludes access dysfunction quickly as cause of underdialysis;• Identifies a mid-access obstruction; • Identifies high-flow versus low flow accesses to select ideal treatment plan for correction (flow-

restricting versus re-vascularization procedure); • Permits access surveillance to be performed by the clinic’s staff who then can alert nephrologist to

possible onset of access dysfunction & referral for early intervention• Implements KDOQI Guidelines;

• Implements Best Practices for patient care with Gold Standard reference method.

• Tests calibration of the blood pump;

• Verifies true delivered blood flow and compares delivered blood flow to pump setting to identify flow disparity and avoid underdialysis. If disparity is significant, Flow-QC® assists in determining cause (blood pump calibration versus inflow restriction/excessive pre-pump negative arterial pressure);

• Detects and quantifies access recirculation in AV access and catheters;

• Identifies inadvertent reversal of dialysis lines to prevent recirculation and/or underdialysis;

• Determines proper needle placement;

• Identifies sources of large negative arterial blood line pressure (and its resulting underdialysis);

• Determines the most appropriate blood pump setting for a low flow access when it is not feasible to increase access flow;

• Provides delivered flow and recirculation measurements to maximize catheter function.

Vascular Access Measurements

Page 13: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

Sudden Cardiac Death & Chronic Kidney Disease

Hemodialysis

CKD(DL-200-wp)RevA2016USltr

“ESRD patients are prone to sudden death, stroke and myocardial infarction between dialysis sessions.”1

Page 14: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

Sudden Cardiac Death Sudden cardiac death (SCD) is defined as death from an unexpected circulatory arrest, occurring within an hour of the onset of symptoms, or an unwitnessed, unexpected death in patients known to be well within the previous 24 hours without an obvious non-cardiac cause.2

For the 20 million Americans that suffer from chronic kidney disease (CKD), SCD statistics are alarming. These CKD patients have a four to 20 times greater risk of sudden cardiac death (SCD) than do persons in the general population.3 As their glomerular filtra-tion rates decline indicating progressively lower kidney function, the risk of sudden cardiac death proportionally increases.2

Moreover, it is estimated that the number of patients with End-Stage Renal Disease (ESRD) on hemodialysis will reach one-half million by the year 2020.2,3 Sudden death among these patients accounts for 25% of all causes of mortality.4 In 1998, the American Journal of Kidney Disease underscored the gravity of this disease with the following title on the cover of their journal, “Cardiovascular Disease, An ESRD Epidemic.”5

Cardiovascular Collapse Not only is cardiovascular disease (CVD) the leading cause of morbidity and mortality in patients with end-stage renal disease (ESRD), but cardiovascular collapse is a major cause of complications during a hemodialysis treatment. Congestive heart failure (CHF) in ESRD patients results from cardiac overload, anemia, severe hypertension and cardiac dysfunction. CVD mortality rates are about 30 times that of the general population, and in adolescents, CVD mortal-ity rates are over 1,000 times that of their age-related peers.6

Multinational data from the Dialysis Outcomes and Practice Patterns (DOPPS) registry suggests that sud-den death accounts for the highest proportion of he-modialysis deaths in the United States (33%) whereas lower proportions were observed in Japan (23%), Australia/New Zealand (19%) and Canada (18%).7

Identifying SCD in CKD Patients However, determining the sudden nature of death is problematic particularly among End-Stage Renal Disease (ESRD) patients because:• Most sudden deaths among CKD patients are un-

witnessed. Therefore, clinical information collected around that time is often limited

• ESRD patients are often chronically ill with comor-bidities, and are frequently hospitalized.

CKD Sudden Death Pathophysiology In the general population, coronary heart disease that leads to ventricular fibrillation or sustained ven-tricular flutter is the major cause of sudden death.

In the case of CKD patients, studies suggest funda-mental differences in the causes and pathology of sudden heart disease. In these patients, their coro-nary artery disease involves multi-vessel arterial stiffening and calcification rather than an ischemic myocardium that might trigger a terminal arrhythmia and death. Rather, CKD patients, especially those on hemodialysis, are subject to a wide array of potential arrhythmic triggers.

When Does Sudden Death OccurIt has been observed that SCD in CKD occurs most fre-quently on days when hemodialysis is being admin-istered, particularly on the first day after a three day weekend without dialysis.2 Moreover, as illustrated in the graph below, the probability of sudden death increases in the time period immediately following dialysis.8

Sudden Death Risks FactorsIn ESRD patients on hemodialysis, an increased risk of SCD has been linked to: • Rapid ultrafiltration rate • Low calcium during hemodialysis • Low potassium (hypokalemia) or high potassium (hyperkalemia) levels in the blood

“Given the heightened risk of SCD in the CKD population and the abysmal long-term survival

rate following a cardiac arrest, primary prevention of SCD should be a major priority.”2

Sudden Cardiac Death & Chronic Kidney Disease

Fig. 1: The graph illustrates the spike in the probability of sudden death immediately following dialysis. ““35% of deaths occurred in the 1st 12-hour interval… 27% of these deaths occurred during dialysis and 33% occurred in the hour after the dialysis treatment (8).”” Critically low CI levels (<2 L/min/m2) can occur in patients who do not feel well at the end of a dialysis session.

Page 15: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

Measure Cardiac Output — Save a Life!

Cardiac Output (CO)

Cardiac Index (CI)3.5

3.0

2.5

2.0

1.50 1 2 3 4 Dialysis Session

hours

Stable Group(N=27) Fluid Removed 2.63 ±1.32L

Unstable Group (N=24) Fluid Removed 3.58 ±1.13L

L/m

in

5

7

3

10 1 2 3 4 Dialysis Session

hours

------ =24% AVF CO

AVF CO ------ =33%

Sudd

en D

eath

Mon Wed Fri Mon

HD HD HD HD

Dialysis Sessions

Probability of Sudden Death

Access Flow (AVF)

Cardiac Output (CO)

Cardiac Index (CI)3.5

3.0

2.5

2.0

1.50 1 2 3 4 Dialysis Session

hours

Stable Group(N=27) Fluid Removed 2.63 ±1.32L

Unstable Group (N=24) Fluid Removed 3.58 ±1.13L

L/m

in

5

7

3

10 1 2 3 4 Dialysis Session

hours

------ =24% AVF CO

AVF CO ------ =33%

Sudd

en D

eath

Mon Wed Fri Mon

HD HD HD HD

Dialysis Sessions

Probability of Sudden Death

Access Flow (AVF)

Fig. 3: One third of CO is redirected from the systemic circulation to the AV fistula placing patients at cardiac risk.

Fig. 2: Inadequate dry weight estimation increases the risk of cardiac failure.1,9

“The ability to monitor cardiac output is one of the important cornerstones of hemodynamic assessment ...in

particular in patients with pre-existing cardiovascular comorbidities.”9

Proactive Patient Care Cardiovascular complications in ESRD patients can be averted or reduced through periodic exams by the patient’s nephrologist, and routine screening of cardiac function with the Transonic HD03 Monitor.

Cardiac Function Stressed The rapid removal of large volumes of fluid during hemodialysis severely tests the limits of a patient’s cardiac function. Cardiac output decreases an average of 20% during the hemodialysis treatment causing less and less blood flow to be available to sustain the body’s vital functions. HD03 cardiac output measurements monitor a heart’s response to fluid removal during dialysis.

Central Hemodynamic Profiling (CHP) Cardiovascular parameters can change dramatically during dialysis. Therefore, multiple cardiac measurements are advised during a dialysis session for patients at risk for SCD in order to create a CHP (Fig. 5) to track the heart’s response to the stress of a dialysis treatment. CHP Monitoring identifies:

• Dramatic 20-30% CO drop during dialysis due to inaccu-rate dry weight estimation and/or medication that places patients at high risk for cardiovascular complications and sudden death following the dialysis session (Fig. 2).

• Prolonged high levels of access flow (>1,600-2,000 ml/min) that can lead to cardiomegaly and high output cardiac failure. This can be identified by an access flow to cardiac output ratio (AVF/CO) exceeding 25-30% (Fig. 3).

• Cardiac Index of <2 L/min/m2.

Fig.3: HD03 Hemodialysis Monitor displaying a CO screen.

Deterioration of CO & CI during Hemodialysis

Case Report, courtesy of Dr. T.A. Depner

Routine Flow-QC® Cardiac Function screening commenced 40 minutes into the hemodialysis session for a patient with ischemic heart disease. The first Cardiac Output measurement was 4.3 L/min with a Cardiac Index of 2.5. When the cardiac function test was repeated two hours later, the patient’s Cardiac Output had dropped to 2.7 L/min and his Cardiac Index was 1.6. The nephrologist was alerted, the patient’s hemodialysis prescription was adjusted, and his cardiac condition was closely monitored.

Fig. 5: Central Hemodynamic Profiling (CHP): four measure-ments taken during a single hemodialysis session shows Cardiac Index responses to the hemodialysis treatment. Acceptable CI results range between 2.5 - 4.2 L/min/m2.

Page 16: Best Practices: Transonic Hemodialysis Surveillance Literature... · Best Practices: Transonic ® Hemodialysis Surveillance ... A Flow-based Access Management Protocol includes an

AMERICASTransonic Systems Inc.34 Dutch Mill RdIthaca, NY 14850U.S.A.Tel: +1 607-257-5300Fax: +1 [email protected]

EUROPETransonic Europe B.V.Business Park Stein 2056181 MB ElslooThe NetherlandsTel: +31 43-407-7200Fax: +31 [email protected]

ASIA/PACIFICTransonic Asia Inc.6F-3 No 5 Hangsiang RdDayuan, Taoyuan County33747 Taiwan, R.O.C.Tel: +886 3399-5806Fax: +886 [email protected]

JAPANTransonic Japan Inc.KS Bldg 201, 735-4 Kita-AkitsuTokorozawa Saitama359-0038 JapanTel: +81 4-2946-8541Fax: +81 [email protected]

www.transonic.com

Transonic Systems Inc. is a global manufacturer of innovative biomedical measurement equipment. Founded in 1983, Transonic sells “gold standard” transit-time ultrasound flowmeters and monitors for surgical, hemodialysis, pediatric critical care, perfusion, interventional radiology and research applications. In addition, Transonic provides pressure and pressure volume systems, laser Doppler flowmeters and telemetry systems.

Hemodialysis

1. Tucker, T et al, “Unrecognized Deterioration of Cardiac Function during Hemodialy-sis,” J Am Soc Nephrol 2002; 13: 213A

2. Pun PH, “The interplay between CKD, sudden cardiac death, and ventricular ar-rhythmias,” Adv Chronic Kidney Dis. 2014; 21(6): 480-8. (Transonic Reference # 10992R)

3. U.S. Renal Data System. USRDS 2007 Annual Data Report: Atlas of Chronic KidneyDisease and End-Stage Renal Disease in the United States. Bethesda, MD: NationalInstitutes of Health, National Institute of Diabetes and Digestive and Kidney Diseases; 2007.

4. U.S. Renal Data System. USRDS 2006 Annual Data Report: Atlas of Chronic Kidney Disease and End-Stage Renal Disease in the United States. Bethesda, MD: National Institutes of Health, National Institute of Diabetes and Digestive and Kidney Dis-eases; 2006.

5. Cardiovascular Disease — An ESRD Epidemic. Am J Kid Dis 1998; 32(5): Suppl 3.6 Locatelli F et al, “Cardiovascular Disease in Chronic Renal Failure; the Challenge

Continues,” Nephrol Dial Transplant 2000; 15(Suppl 5): 69-80. (Transonic Reference # HD9643R)

7 Jadoul M et al, “Modifiable practices associated with sudden death among hemodialysis patients in the Dialysis Outcomes and Practice Patterns Study,” Clin J Am Soc Nephrol. 2012; 7(5): 765-74 (Transonic Reference # HD11010R)

8 Bleyer AJ et al,“The Timing and Characteristics of Sudden Death in Hemodialysis Patients” J Am Soc Nephrol 2002; 13: SU-PO737.

9 Tucker T et al, “Central Hemodynamic Profiling (CHP) during Outpatient Hemodialysis (HD),” J Am Soc of Nephrol Abstr 2002; 13: 209A. (Transonic Reference # HD268A)

5 Krivitski NM, Depner, TA, “Cardiac Output and Central Blood Volume during, Hemodialysis: Methodology,” Adv Ren Replace Ther 1999; 6(3): 225-232. (Transonic Reference # HD8T)

7 Depner TA, Krivitski NM, “Central Blood Volume: A New Criterion for Predicting Morbid Events during Hemodialysis,” J Am Soc of Nephrol Abstr 1996; 7(9) 1511. (Transonic Reference # HD15A)

8 MacRae JM, “Vascular Access and Cardiac Disease: Is There a Relationship? Curr Opin Nephrol Hypertens 2006; 15(6): 577-82. (Transonic Reference # HD7382A)

9 MacRae JM et al, “Arteriovenous Fistula-associated High-output Cardiac Failure: A Review of Mechanisms, “Am J Kidney Dis 2004; 43(5): 17-22. (Transonic Reference # HD408A)

10 Nikiforov UV et al, “Validation of a New Method to Measure Cardiac Output during Extracorporeal Detoxification,” ASAIO J 1996; 42: M903-M905. (Transonic Reference # HD15V)

11 Kislouchine VV, Dean DA, “Validation of a Novel Ultrasound Dilution Method to Measure Cardiac Output during Hemodialysis,” ASAIO J 1996; 42: M906-M907. (Transonic Reference # HD16V)

12 MacRae JM, “Vascular Access and Cardiac Disease: Is There a Relationship? Curr Opin Nephrol Hypertens 2006; 15(6): 577-82. (Transonic Reference # HD7382A)

13 van der Mark WAMA et al, “Decreased access resistance in haemodialysis patients with upper arm arteriovenous fistulae,” Nephrol Dial Transplant. 200; 23(6): 2105-6. (Transonic Reference # HD11002V)

14 Mavrakanas TA, Charytan DM, “Cardiovascular complications in chronic dialysis patients,” Curr Opin Nephrol Hypertens 2016; Sep 14, (Transonic Reference # HD10780R)

15. Kanbay M et al, “Sudden cardiac death in patients with chronic kidney disease: prevention is the sine qua non,” Kidney Blood Press Res. 2011; 34(4): 269-76. (Transonic Reference # HD10982R)

16 Ranpuria R et al,“Unruh M. Heart rate variability (HRV) in kidney failure: measure-ment and consequences of reduced HRV,” Nephrol Dial Transplant. 200; 23(2): 444-9.(Transonic Reference # HD10983R)

17 Green D et al, “Sudden cardiac death in hemodialysis patients: an in-depth review,” Am J Kidney Dis. 2011; 57(6): 921-9. (Transonic Reference # HD10985R)

18 Wasse H, Singapuri MS, “High-output Heart Failure: How to Define It, When to Treat It, and How to Treat It,” Semin Nephrol 2012; 32(6): 551-7. (Transonic Reference # HD9868AHR)

19. Makki N et al, “Do implantable cardioverter defibrillators improve survival in patients with chronic kidney disease at high risk of sudden cardiac death? A meta-analysis of observational studies.” Europace. 2014; 16(1): 55-62. (Transonic Reference # HD10986R)

20 Franczyk-Skóra B et al, “Prevention of sudden cardiac death in patients with chron-ic kidney disease,” MC Nephrol 2012; 13:162. (Transonic Reference # HD10987R)

21 Poulikakos D et al, “Risk of sudden cardiac death in chronic kidney disease,” J Car-diovasc Electrophysiol 2014; 25(2): 222-31. (Transonic Reference # HD10988R)

22 Alpert MA, “Sudden cardiac arrest and sudden cardiac death on dialysis: Epide-miology, evaluation, treatment, and prevention,” Hemodial Int. 2011; 15 Suppl 1:S22-9.(Transonic Reference # 10989R)

23 Shamseddin MK, Parfrey PS. “Sudden cardiac death in chronic kidney disease: epidemiology and prevention,” Nat Rev Nephrol, 2011; 7(3): 145-54. (Transonic Reference # 10990R)

24 Franczyk-Skóra et al, “Sudden cardiac death in CKD patients,” Int Urol Nephrol, 2015; 47(6): 971-82. (Transonic Reference # 10993R)

25 Bansal V, “Sudden Cardiac Death & Dialysis: Target Risk Factors,” Nephrology Times, 2011; 4(8): 3-5. (Transonic Reference # 10994R)

26 Assimon MM et al, “Ultrafiltration Rate and Mortality in Maintenance HD Patients,” Am J Kidney Dis, 2016 Aug 26. (Transonic Reference # 10995R)

27 Kramer H et al, “Ultrafiltration Rate Thresholds in Maintenance Hemodialysis: An NKF-KDOQI Controversies Report,” Am J Kidney Dis. 2016; 68(4): 522-32P (Transonic Reference # )

28. Banerjee D, “Sudden cardiac death in haemodialysis: clinical epidemiology and mechanisms,” J Electrocardiol 2016 Jul 28 (Transonic Reference # HD10996R)

29. Ramesh S et al, “Incidence of sudden cardiac death in adults with end-stage renal disease: a systematic review and meta-analysis,” BMC Nephrol, 2016; 17(1): 78. (Transonic Reference # HD10997R)

30 Di Lullo L et al, “Sudden cardiac death and chronic kidney disease: From pathophysiology to treatment strategies,”Int J Cardiol 2016; 217: 16-27. (Transonic Reference # HD10998R)

31 Basile C et al, “The impact of haemodialysis arteriovenous fistula on haemodynamic parameters of the cardiovascular system,” Clin Kidney J. 2016; 9(5): 729-34. (Transonic Reference # HD11001A)

32 Kiuchi MG, Mion D Jr., Chronic kidney disease and risk factors responsible for sudden cardiac death: a whiff of hope? Kidney Res Clin Pract. 2016; 35(1): 3-9. (Transonic Reference # HD11004R)

33 Potpara TS et al, “Cardiac Arrhythmias in Patients with Chronic Kidney Disease: Implications of Renal Failure for Antiarrhythmic Drug Therapy, “Curr Med Chem. 2016; 23(19): 2070-83. (Transonic Reference # HD11005R)

34. Pun PH et al, Primary prevention implantable cardioverter defibrillators in end-stage kidney disease patients on dialysis: a matched cohort study. Nephrol Dial Transplant. 2015; 30(5): 829-35. (Transonic Reference # HD11007R)

References: Sudden Death & Chronic Kidney Disease