basic principles and application in medicine basic principles and application in medicine october,...
TRANSCRIPT
Basic principles and application in MedicineBasic principles and application in Medicine
October, 2008 J.Brnjas-Kraljević
6 October 20036 October 2003The Nobel Assembly at Karolinska Institutet has today decided to award The Nobel Prize in Physiology or has today decided to award The Nobel Prize in Physiology or Medicine for 2003 jointly toMedicine for 2003 jointly to Paul C Lauterbur Paul C Lauterbur and and Peter MansfieldPeter Mansfieldfor their discoveries concerningfor their discoveries concerning "magnetic resonance imaging" "magnetic resonance imaging"
Paul Lauterbur (born 1929), Urbana, Illinois, USA, discovered the possibility to create a two-dimensional picture by introducing gradients in the magnetic field. By analysis of the characteristics of the emitted radio waves, he could determine their origin. This made it possible to build up two-dimensional pictures of structures that could not be visualized with other methods.Peter Mansfield (born 1933), Nottingham, England, further developed the utilization of gradients in the magnetic field. He showed how the signals could be mathematically analyzed, which made it possible to develop a useful imaging technique. Mansfield also showed how extremely fast imaging could be achievable. This became technically possible within medicine a decade later.
GlossaryGlossary
magnetic field – field intensity – tesla (T) Earths magnetic field <70 T – in medicine 0,5 – 3,0 T homogeneoushomogeneous – the same intensity in each space point
constantconstant – unchangeable intensity upon time radiofrequentradiofrequent – frequency of regular change of
magnetic field intensity, in medicine 100 kHz – 10 GHz field gradientfield gradient – regularity in the field intensity changes
in linear dimensions of the space - (T/m) – in medicine is better if more steep – 30 mT/m (0,3 mT/cm)
pulsepulse – is the measure of energy transfer to the system – time interval of RF-magnetic field that transferees the energy on spin system and induces excitation
nuclear spinnuclear spin – intrinsic property of the material particle – describes the magnetic property of nuclei with odd number of nucleons, in medicine nuclei with spin number ½ ; determines number of possible energy states in magnetic field: if ½ than two energy states
magnetic momentmagnetic moment – physical parameter - the measure of magnetic properties of nuclei with spin; the base of NMR
resonanceresonance – process of maximal energy transfer between
two systems – described by characteristic frequencycharacteristic frequency relaxationrelaxation – processes by which the excited system is
after ending of perturbation returned to the ground energy state – described by characteristic timecharacteristic time
Magnetic ResonanceMagnetic Resonance measured are magnetic properties of atomic nuclei in
sample placed in the strong external magnetic field
- the changes in the state of the system are controlled
- resonantresonant absorptionabsorption - the processes of returning to the equilibrium are followed – relaxationrelaxation emissionemission
if the structure of molecules is determined - it is spectroscopy methodspectroscopy method
in medical diagnostic - as spectroscopyspectroscopy (MRS) or as imagingimaging (MRI)
HistoryHistory 1944. F.Bloch i E.Purcell – nuclear magnetic resonance 1971. R. Damadian – differentiates T1 i T2 in tumors 1973. P.Lauterbur – the first MRI 1975. R.Ernst – distinguishing the signals by phase and
frequency - presentation by Fourier transform – the base of all modern MRI
1977. P. Lauterbur – independent by R.Damadian – MRI of the whole body
P.Mansfield – echo method (EPI) – 5 min/image – today 5 s/image
1986. NMR microscopy – resolution 10 m in volume of 1 cm3
1987. EPI method – cardiac cycles in real time C. Dumoulin – angiography - MRA – without contrast agents 1993. functional MRI 1995. spectroscopy in vivo 1998. combination with other imaging methods 2003. N.P. to P. Lauterbur and P. Mansfield
What is NMR ?What is NMR ? What is MRI?What is MRI? What is fMRI?What is fMRI? What is looked at, what is seen, what is What is looked at, what is seen, what is
measured?measured? How is it measured?How is it measured?
We are interestedWe are interested inin
wwe measuree measure
cell macromoleculesand water
watermolecules
watermolecule Hydrogen
atomHydroge
n nucleus
volume of
heterogenic tissue
Interaction of the Interaction of the nuclear magnetic moment of nuclear magnetic moment of
hydrogenhydrogenand magnetic fieldand magnetic field
hydrogen nucleus has spin – its magnetic properties are described by
magnetic moment,magnetic moment, , , and intrinsic magnetic field
in external magnetic field magnetic moment experience two possible states: parallelparallel or antiparallelantiparallel to the field direction – we talk about two possible states of energy
the volume of hydrogen placed outside the magnetic field – magnets are randomly oriented in space – volume is not magnetized
the same volume in the external magnetic field – energy states occupancy is determined by Boltzmann,s distribution
– there is more nuclei parallel to the field – volume is magnetized
the top of single magnetic moment precesses in magnetic field with Larmor frequency, because of giromagnetic constant characteristic for the nucleus
kT
E
eN
N
21
21
0B
B0M0
nono magnetic field magnetic field
- randomly oriented magnetic moments - no macroscopic macroscopic magnetizationmagnetization
homogeneous, constant homogeneous, constant magnetic field Bmagnetic field B00
more magnetic moments are in the magnetic field direction - macroscopic magnetizationmacroscopic magnetization in the direction of B0 field is measured
Very important: nuclei, atoms or molecules are not oriented, but magnetic moments!
It can be visualized like:
ordered state of equilibrium system in the magnetic field is described by - macroscopic magnetizationmacroscopic magnetization in the direction of the magnetic field
process of resonance resonance will be realized by energy equal to the difference of the two states and it will promote more nuclei in the higher energy state – resulting in change of amount and direction of macroscopic magnetization
this process is realized with the energy of radiofrequent magnetic field - frequency being characteristic for the observed nucleus
when the RF-field frequency is equal to Larmor-frequency of the nucleus the interaction of magnetic moment and the field changes the Boltzmann,s distribution
higher the difference of occupancy in the equilibrium more precise are the measurements of the resonance
direction of the vector or – the visualization of two possible
energy states of magnetic moment in B0 field difference in occupancy is bigger for the field of higher intensity macroscopic magnetizationmacroscopic magnetization is bigger for bigger difference
quant energy h will be absorbed if E =
h
that is the value of the field where the signal is measured
Theory – quantum physicsTheory – quantum physics
nucleus number of
protons neutrons
spin / MHz/T
11HH 1 0 1/2 99,98 42,58 2H 1 1 1 0,015 6,54 31P 0 1 17,25 23Na 2 1 3/2 11,27 14N 1 1 1 3,08 13C 0 1 10,71 19F 0 1 40,08
1/2
1/2
100
1,11
100 0,015
0,0004
0,0024
0,094
abundance biologicalabundance
1/2 100
0,63
Properties of the nucleusProperties of the nucleus
natural abundance - fraction of isotope in the elementbiological abundance – fraction of the element in the tissue
Resonance conditionResonance condition
states separation: E = E E = E+1/2+1/2 - E - E-1/2 -1/2 depends on external magnetic field
by absorption of energy quant higher energy state
basic relation of magnetic resonance
= = B B00
– Larmor frequency
02
hE B
02
hEh B
Radiofrequent magnetic field Radiofrequent magnetic field BB11 resonance absorption of time dependent magnetic field energy
B1(t)=B1maxis sin t
B1 is perpendicular to B0, and magnetic induction is 10-4 B0
B1 frequency = Larmor frequency of atomic nucleus
BB11
B0
M0
= = cconst.onst. the same nuclei have
different Larmor frequency if in different magnetic fields
if the inhomogeneity of the field is controlled – the base of NMR as imaging method
BB0 0 = const.= const. different nuclei have
different Larmor frequency, because differ
in spectra their lines are separated
BB00
by end of excitation the system returns to the equilibrium state defined by Boltzmann distribution – process of relaxationprocess of relaxation
two mechanisms of relaxation – both are the source of information on dynamic properties of the system
in magnetic resonance - 4 basic parameters: - macroscopic magnetization, - chemical shift, - relaxation time T1, - relaxation time T2
RelaxationRelaxation
nuclear magnetic moment – bar magnet
B0= 0 – because of Brownian motion randomly oriented
B0 0 - magnetic moments precess with Larmor frequency around field direction: more are in + Z, less in – Z direction
phase of precession are different: macroscopic magnetization is in magnetic field B0 direction; no component in perpendicular plane
absorption of RF- field energy, forces the macroscopic magnetization to simultaneous precession about the axes of both fields
the motion is represented by spiral path from Z axis to XY plane and towards –Z axis
QuasiclassicalQuasiclassical modelmodel
Macroscopic Macroscopic magnetizationmagnetization
sample in B0 is magnetized
in the direction of magnetic field (axis +Z) macroscopic magnetization M0 is measured - determined by:
and has only longitudinal component and expressed by measurable parameters:
)(1 2121
11
N
izi
N
iziV
02
222
0 BkT12
IhNM
2121 NNN
hencehence, , macroscopic magnetizationmacroscopic magnetization
increases with increasing magnetic field strength: good instruments work on higher fields is inversely proportional with temperature: the best is to measure on law temperatures, unsuitable
in medical applications depends on density of nuclear spins of interest: in medicine hydrogen from water molecules (free or
bound) and there is plenty of them in tissues; or hydrogen in fat
Appearance of transversal magnetization
in equilibrium no transversal magnetization,transversal magnetization, MMxyxy, because of different precession phases of magnetic moments
the action of magnetic field B1 forces the equalization of the phases and the appearance of transversal magnetization
because of resonance energy absorption the longitudinal component is decreasing, Mz < M0
Mz M0
Mxy
z
x
y
B0
B1
in NMR experiment always transversal magnetization is measured – as the induced electromotor force in detector coil
detector is placed in X-axis amount of Mz i Mxy depends on length of field B1 action. The
angle of decline from +Z is:
the amount of energy transferred on the system by the radiofrequent field is named pulse
tB max1
Characteristic pulsesCharacteristic pulses
/2 pulse
magnetization is rotated in Y-axis
pulse
magnetization is rotated in Z -axis
max12
2
Bt
z
x
y
z
x
y
max1Bt
Chemical shiftChemical shift observed nucleus is in B0 field not naked but in atom, so it
feels local magnetic fields of surrounding electrons - mainly from own atom
BBeff eff = B= B00 - B - Blocloc = B = B00 (1- (1- )) - shielding - depends on chemical composition of molecules
of observed nuclei effective field is always smaller than B0, because of
diamagnetic effect of electron
effeff = = (B(B0 0 - B- Blocloc)) hence, there is the shift in resonant frequency for the same
nuclei in static magnetic field, but in different molecules that is chemical shift, - defined by standard sample (ppm)
dtans
measdtans
610
chemical shift in water and fatchemical shift in water and fat
difference in resonant frequency is only 1 kHz for 42 MHz, but enough to differ that two molecules molecules are in the same static magnetic fieldsignal area is proportional to the number of resonating nuclei
intrinsic – defined by chemical surrounding of the nucleus induced – defined by the surrounding of the molecule -
solvent, pH, temperature, paramagnetic centers, secondary and tertiary structure in proteins, denaturation of proteins, different pathological processes
diagnostic value in spectroscopy spectroscopy in vivoin vivo
BB0 0 - B- Blolocc))
CHCH2
CH3
frekvencija/ Hz
CH
CH2
CH3
Relaxation processes –Relaxation processes – relaxation times relaxation times
relaxation processes relies the energy in surrounding
decrease of system energy interchange of energy among the observed nuclei
increase of entropy both processes are determined by dynamic properties of the
system in biological systems tissue differ in relaxation parameters processes are effective - signal of resonance is constantly
measurable, despite the small difference in energy state abundance
processes of relaxation are random therefore described by exponential function with characteristic times
relaxation parameters –
relaxation times T1 i T2
Spin-lattice relaxation - TSpin-lattice relaxation - T11
energy absorbed in the spin-system is released into the local magnetic field – induced by rotation of surrounding molecules
rotation is defined by correlation time:
c ~ 10-11 s for small molecule rot big
c ~ 10-8 s for big molecule rot small (Larmor frequency)
in surrounding of big molecules the relaxation of the spin-system is faster T1 shorter
in plain water relaxation is slow T1 longer
TT1 1 depends on temperature and viscosity of surrounding – it is the depends on temperature and viscosity of surrounding – it is the measure of molecular motionmeasure of molecular motion
tissues have different Ttissues have different T11
kT
a
3
4 3
c
Determination of TDetermination of T11 - - ---
/2/2 applying pulse longitudinal magnetization is changed from - M0 to + M0 :
T1 is determined from
1210
Tt
z eMtM
T1
M0(1-2e-1)
e
MMTM z
001 2)(
applying /2 pulse longitudinal magnetization increases from 0 to +M0
T1 is determined from
110
Tt
z eMtM
e
MMTM z
001 )(
Inversion recovery - IRInversion recovery - IR
longitudinal magnetization is by 180° pulse turned into Z- direction and than returns to equilibrium value
by 90° pulse applied before completed relaxation the transversal magnetization is proportional to the amount of relaxed spins
in detector coil FID is induced intensity of Fourier transform after one measurement is
S = S = kk r ( 1 - 2e r ( 1 - 2e-TI/T1-TI/T1 ) ) and after repetition
S = S = kk r ( 1 - 2e r ( 1 - 2e-TI/T1-TI/T1 + e + e-TR/T1-TR/T1) ) TR – time of repetitionTI – time between pulses
TI
signal
//22 pulsa pulsa magnetization is by 90o
pulse rotated into XY plane returns into equilibrium in detector coil FID is
measured intensity of FT signal
depends on time between pulses – TR
11 T
TR
ekS
T1 and T2
tissue T1 /s T2 /ms hydrogen density
CSF 0,8 - 20 110 - 2000 70 - 230 white matter 0,76 – 1,08 61 - 100 70 - 90 gray matter 1,09 – 2,15 61 - 109 85 - 125 membrane 0,5 – 2,2 50 - 165 5 - 44 muscle 0,95 – 1,82 20 - 67 45 - 90 fat 0,2 – 0,75 53 - 94 50 - 100
Spin-spin relaxation - TSpin-spin relaxation - T22
by termination of radiofrequency magnetic field action the magnetic moments interchange the energy
because of small inhomogeneities of magnetic field Larmor frequencies are different – phases of precession starts to differ
transversal magnetization decreases exponentially interchange of energy between spins is greater if the nuclei
are closer and less movable - T2 is considerably shorter in solid state
tissues have different Ttissues have different T22
for each nucleus in certain surrounding T2 T1
Determination of TDetermination of T22
applying /2 pulse the disappearance of transversal magnetization is
measured
T2 is determined from
20
Tt
xy eMtM
e
MTM xy
02 )(
Spin - EchoSpin - Echo
to determine T2 most used method is spin-echo: //2,2,
signal height depends on time between pulses (TE) and on repetition time (TR)
211 TTE
TTR
eekS
TE
How is spin-echo builtHow is spin-echo built
90° pulse induces transversal magnetization
it diminishes - moments are dephasing because of different - FID
after the time interval of the 180° pulse along Y –axis induces the phase coherence again after interval of 2 this is the echo signal
Bloch equationsBloch equations clasical presentation of macroscopic magnetic moment movement in
the magnetic fields
bases for T1 and T2 calculations
Relations ofRelations of T T11 and and TT22
in plain water T1 T2 ~ 3 s
tumor tissue has more water - T1 longer than for healthy tissues
in solid state T1 ~ min - h ; T2 ~ 10-6 s
differences in relaxation times adequate for contrast enhancement in MRI
different sequences of pulses necessary to repeat the sequence because the signals
are very small by good choice of field strength and sequence of pulses
the contrast of the tissues can be amazing although the density of the observed tissues is practically the same
Contrasts in MRIContrasts in MRI
biological parameters are the relaxation times
T1 i T2 are main parameters for production of contrasts
by adjustment of time interval between pulses i /2
for measurements the interval producing the biggest difference between measured signals from different tissues is chosen
further improvement of contrasts by changing the intervals between sequences
T1
A
T1
B
T2BT2A
sl.10.
TT1 1 - - source of contrastssource of contrasts
T1 difference source of contrast: sequence - - - - /2/2 - pulse flips the magnetization in Z- direction after time interval , Mz is greater for tissue with shorter T1
Mz,short> Mz,long /2 pulse flips that component into XY - plane:
Mxy,short> Mxy,long in detector coil the measured signal (S):
S(Mxy,short) > S(Mxy,long) is chosen so to satisfy Mxy,short-Mxy,long = max
TT2 2 - - source of contrastssource of contrasts
T2 difference - source of contrast: sequence /2/2 - - - -
- the appearance of spin echo (Hahn 1950) /2 pulse flips the magnetization into XY- plain after time interval , Mxy is bigger for tissue with longer T2
Mxy,long> Mxy,short pulse rephrases spins after - signal of spin echo – SSj j dependent on in detection coil measured signal:
Sj(Mxy,long) > Sj(Mxy,short)
must satisfy Mxyd-Mxyk= max
kj eS ~
ContrastContrast
T2T1
tissue T1 /s T2 /ms hydrogen density
gray matter 1,09 – 2,15 61 - 109 85 - 125
white matter 0,76 – 1,08 61 - 100 70 - 90
CSF 0,8 - 20 110 - 2000 70 - 230
fat 0,2 – 0,75 53 - 94 50 - 100
muscle 0,95 – 1,82 20 - 67 45 - 90
skin 0,5 – 2,2 50 - 165 5 - 44
MR spectroscopy (MRSMR spectroscopy (MRS)) in medicine we use nuclei with magnetic moment - in
characteristic molecules of tissues spectral lines belong to chosen nuclei in different molecules or
atomic groups spectra display chemical shift for the small volume excited in
the tissue changes in the place and/or intensity of lines or the
appearance of new lines point at metabolic and structural changes
Spectroscopy Spectroscopy in vivoin vivopoint resolved spectroscopy - - PRESSPRESS
with adequately chosen gradients of magnetic field B0 in X-, Y- and Z-direction we measure the signals from small volume spectrum is display of chemical shifts the concentration of single aminoacid can be determined the structure of small volume is determined in combination with imaging -fMRI
Magnetic resonance Magnetic resonance instrumentinstrument
constant and homogeneous magnetic field - electromagnet or superconductive magnet
in science - up to 14 T; in medicine – up to 2,3 T radiofrequent magnetic field -
frequency 600 MHz, or 64 MHz - induced in coil
intensity of B1 is 10-4 B0 detection coil + computer
registration
U
RF generator
signal
detector
B0
B1
Fourier transformFourier transform
mathematical procedure - enables differentiation of frequenciesshortens the time of signal detection enables huge number of repetition measurements