applanation tonometry: applanation tonometry: measures iop by providing force which flattens the...

18
Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann, Perkins, Draeger, MacKay- Marg, and Tono-Pen and Pneumatonometer. based on Imbert-Fick law: P = f / A ( P-pressure ; f- Force ; A- area)

Upload: ethelbert-bates

Post on 18-Dec-2015

215 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Applanation tonometry:Applanation tonometry:

measures IOP by providing force which flattens the cornea.

Variable force applanation tonometers (Goldmann, Perkins, Draeger, MacKay-Marg, and Tono-Pen and Pneumatonometer.

based on Imbert-Fick law: P = f / A

( P-pressure ; f- Force ; A- area)

Page 2: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Applanation tonometers(1) Constant area (Goldmann, Mackay-Marg)(2) Constant force (Maklakoff)

Page 3: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,
Page 4: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Instructions to patientpress head firmly against chin and forehead

rest.look straight ahead and fixate on a target

(e.g. examiners opposite ear)breathe normally, do not hold your breathblink immediately prior to measurement to

moisten cornea.

Page 5: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Technique of measurementplastic biprism which contacts cornea

creates two semicirclesedge of corneal contact is visible after

placing fluorescein into tear film & viewing with cobalt blue light

manually rotate the dial calibrated in grams, force is adjusted by changing the length of a spring within the device.

inner margins of semicircles touch when 3.06 mm of cornea is applanated.

Page 6: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,
Page 7: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Measurement (cont.)position patient’s head with forehead

rest well above eyebrows, allowing raising of eyebrows.

anesthetic & fluorescein (0.25%) ,separately, or as mixture (preserved) placed in inferior cul-de-sac.

with maximal illumination of biprism the lamp is moved toward the eye until the tip of biprism contacts the apex of the cornea

stop moving forward when limbus shines with light, best observed with naked eye

Page 8: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Measurement (cont.)After contact, semicircles visible through

left (or right) ocular. Center in field of view.

Adjust vertically until semicircles equal in size.

Tension dial adjusted so that inner edge of upper and lower semicircles are aligned.

Multiply dial reading (grams of force) by 10 to obtain IOP (mmHg)

Read at median over which arcs glide to control for excursions due to ocular pulsations.

Page 9: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,
Page 10: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Measurement (cont.)If slit-lamp moved too far toward patient

the pressure arm will push against a spring which will press against the eye with a low inoffensive force.

Mires (flattened area) too large, moving dial doen’t alter appearance.

Solution: Draw back until regular pulsation noted and appearance of mires normalizes.

Page 11: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Measurement (cont.)Blue central area represents applanated

cornea, green semicircles are fluorescein-stained tears, inner border of ring is demarcation between flattened and non-flattened cornea.

Without staining of tears, bright reflection from air-cornea interface is seen; leads to underestimation of IOP.

Mires should be approximately 10% of circle width.

Page 12: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Errors in MeasurementThe fluorescein ring is too wide or too

narrow:Too wide: occurs if prism not dried after

cleaning or lids touch prism. Overestimates IOP. Solution: dry prism

Too narrow: inadequte fluorescein concentration may cause hypofluorescence. Underestimates IOP. Solution: patient blinks or additional fluorescein added.

Page 13: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Errors (cont.)thin corneas produces underestimatethick cornea d/t increased collagen gives

overestimate, if d/t edema gives underestimate.

inadequate vertical alignment of semicircles leads to overstimate of IOP.

distortion d/t irregular cornea influences accuracy, less useful with corneal scarring.

Page 14: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Errors (cont.)squeezing of eyelids, breath holding or

other Valsalva maneuvers, pressure on globe, excessive EOM force applied to restricted globe, vertical gaze, tight collars, retreating patient, inaccurately calibrated tonometer.

repeated tonometry may induce decline in estimated IOP.

Page 15: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Error d/t corneal curvatureincrease of 1 mmHg for every 3D

increase in corneal power.more fluid displaced under steep

cornea, increases contribution of ocular rigidity in overestimating IOP.

the steeper the cornea, the more cornea must be indented to produce standard area of contact.

>3D astigmatism produces elliptical rather than circular area

Page 16: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

Correction for astigmatismWith semicircles displaced horizontally,

IOP underestimated by 1 mmHg for every 4D of WTR astigmatism, vice versa for ATR astigmatism.

To minimize, prisms should be rotated so that axis of least corneal curvature is opposite red line on prism holder (i.e. align negative cylinder axis).

Can average reading with vertical and horizontal alignment of prism.

Page 17: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

SterilizationCDC recommendation (HIV, HSV, and

adenovirus): wipe tip clean and disinfect tip only with bleach (1:10 dilution x 5”, changed once daily).

Alternative is 3% H2O2, changed at least twice daily (affects tip less than bleach or ETOH).

Alternative #2: wiping tip with 70% ETOH

Page 18: Applanation tonometry: Applanation tonometry: measures IOP by providing force which flattens the cornea. Variable force applanation tonometers (Goldmann,

ReliabilityGoldmann applanation is standard against

which others measured.Good accuracy in gas-filled eyes.Inter- and intraobserver variability (>30%

varied by 2-3 mmHg), due to subjective nature of optical endpoint.

Assume error of 2 mmHg.