ametek flue gas analyser

16
Thermox Flue Gas Analyzers PROCESS & ANALYTICAL INSTRUMENTS DIVISION

Upload: akamalapuri388

Post on 19-Oct-2015

135 views

Category:

Documents


7 download

DESCRIPTION

Catalouge Flue Gas Analyzer, Exhaust gas compliance NEPSHA

TRANSCRIPT

  • 1ThermoxfiFlue Gas Analyzers

    PROCESS & ANALYTICAL INSTRUMENTS DIVISION

  • 2FOR MORE DETAILED INFORMATION ONCOMBUSTION EFFICIENCY AND FLUE GASANALYSIS, ASK FOR OUR FREE BOOKLETCOMBUSTION EFFICIENCY HANDBOOK.

    AMETEK/Thermox product line offers the greatest selection of flue gasanalyzers in the industry for measuring oxygen, combustibles, methane,and fuel-rich atmospheres. Our commitment is to provide the rightanalyzer for every application to ensure the highest levels of efficiency,cost savings, and safety for our customers.

    Flue Gas Measurement and Combustion EfficiencyTypically, imbalance in the fuel/air mix results in incomplete combus-tion. Heat is wasted due to the presence of excess air and harmful NOxemissions are vented into the atmosphere.

    Thermox Site Survey AnalysisThermox can demonstrate the fuel savings and emissions reductiongained by measuring oxygen and combustibles by performing a sitesurvey analysis using a portable analyzer. Our service personnel alongwith your operational assistance will perform air-to-fuel ratio steptesting to determine the safest, lowest excess O2 set point. Combustionefficiency gain and NOx emissions reduction will be measured andrecorded. A report will detail opportunities to increase your combus-tion efficiency and maximize fuel savings. In most cases, a pay-backperiod of less than two months can be demonstrated for the combus-tibles option. At the same time we can also evaluate the condition ofyour existing Thermox analyzers and provide maintenance, retrofit and/or new analyzer purchase suggestions.

    Contact Thermox Service or Sales to arrange a site survey analysis ofyour facility today by calling (412) 828-9040.

    Thermox improves your combus-tion process by operating withminimum excess air. Thermox-combined oxygen and combus-tibles analyzers allow measure-ments of oxygen and combustiblesand/or methane gases from asingle sample point. Once levelsare determined, the supply ofexcess air can be controlled on acontinuous basis minimizingNOx emissions, heat loss, andincompletely burned fuel loss.When combustion efficiency ismaximized, the process operationis more cost-effective.

    AMETEK, INC. IS A LEADING GLOBAL

    COMPANY ENGAGED IN THE DESIGN,

    MANUFACTURE AND SALE OF ELEC-

    TRONIC INSTRUMENTS AND ELECTRIC

    MOTORS THROUGH MORE THAN 41

    MANUFACTURING PLANTS IN THE

    UNITED STATES AND NINE OTHER

    COUNTRIES. AMETEK PRODUCTS ARE

    USED BY INDUSTRIES SUCH AS AERO-

    SPACE, AUTOMOTIVE, FLOOR CARE,

    BUSINESS MACHINES, MEDICAL, COM-

    PUTER MARKETS, AS WELL AS PRO-

    CESS INDUSTRIES WORLDWIDE.

    Combustion Efficiency

    ThermoxTHE RIGHT ANALYZER FOR YOUR APPLICATION

  • 3IQ and Series 2000 Host

    The IQ sensor incorporates the necessary intelligence for proper opera-tion. The electronics contained in the sensor include all Series 2000features: auto calibration, analog output, alarm, self-diagnostics, andnetwork communications. A system trouble alarm signal alerts the userif the self-diagnostics determines a problem requiring service attention.

    The IQ or smart sensor line provides local and field communications.The new design allows access to process variable and diagnostic datapreviously inaccessible using conventional analog outputs. IQ analyz-ers are available with HART or RS-485 AMETEK protocol communica-tions options. User interface with the IQ sensor is achieved through aHART- compatible device, IQ Link Terminal, Series 2000 Host or web-enabled RS247 gateway system.

    Series 2000 Host

    The Series 2000 Host provides aninterface to the Thermox IQ and/or Series 2000-based analyzers.Up to 32 analyzers can be con-nected to a single Host using RS-485 with AMETEK protocol. TheHost can generate up to fouraveraged %O2 outputs or high/low-selects for a group of con-nected analyzers, or up to fourhigh-selects for combustiblesand/or CO.

    Wall Mount IQ Link

    ELIMINATES NEED FOR CONTROL UNIT

    IQ Links

    Handheld or dedicated wall mount IQ Link terminals are also available.The IQ Link provides a local interface using RS-232 communications toset up and display system parameters, initiate calibrations, system tests,and troubleshooting. With the handheld IQ Link you can disconnect andreconnect to other IQ analyzers. The handheld and dedicated wallmount unit also work in cooperation with the Series 2000 Host or RS247gateway system capable of linking up to 32 analyzers via an RS-485 bus.

    RS247

    The RS247 (Remote Service 24-hour-a-day, 7-day- a-week) gatewaysystem enables Thermox to assume a proactive service-provider role thatfosters a closer working relationship with customers. It also providescustomers located anywhere around the world with direct access toThermox resources.

    The RS247 gateway converts the RS-485 AMETEK protocol to an opensystem HSE (High-Speed Ethernet) protocol. The RS247 gateway can beset up on your Local Area Network and/or Wide Area Network allowingplant- or company-wide access to all of your Thermox IQ and Series 2000-based analyzers. Or, you can choose to access the RS247 gateway systemusing the World Wide Web through your IT system or dial-up network(local ISP). The system can send an E-mail to a primary, secondary andtertiary list when a system alarm or event occurs. It also provides trend-ing, service and calibration history.

    Per diem service contracts enable Thermox to take a full proactive servicerole for the analyzers in your facility, allowing greater flexibility withyour on-site resources.

    THERMOX IQ AND SERIES 2000-BASEDANALYZERS ARE EMC COMPLIANT TO89/336/EEC AND LOW VOLTAGE SAFETYCOMPLIANT TO 73/23/EEC. THEY AREALSO UL AND C-UL LISTED.

    Handheld IQ Link

    Series 2000 Host

  • 4ReliabilityHigh quality components are usedthroughout to ensure reliableperformance, including an inter-nationally approved universalinput power supply and twoindependent watchdog timers thatmonitor system operations. Inputand output electronic transientprotectors are standard.

    Automatic Timed CalibrationThe built-in real time clock allowsautomatic timed calibrations whencombined with the optionalremote calibration unit andcalibration module. This functionnot only saves time and labor butis also often a requirement foremissions compliance.

    Modular Design for Flexibility andServiceabilityThe IQ and Series 2000-basedanalyzers are built from modulesdesigned for serviceability. Thisapproach makes it easy to upgradeor add options such as automaticcalibration or combustibles and /ormethane measurements.

    Choice of MeasurementsYou can monitor oxygen only,oxygen and combustibles, oxygenand methane (IQ series only), oroxygen, combustibles, and meth-ane from a single analyzer. Withthe Excess Fuel Software option,the Series 2000 control unit canmeasure fuel-rich/reducingatmospheres (see the Excess FuelSoftware option).

    IQ and Series 2000 Features

    Ease of UseThe IQ and Series 2000 have aninformative full-text display andintuitive user interface. The four-line display has 20 characters perline with line 4 reserved forautomatically detected error anddiagnostic information.

    Software DiagnosticsA system test, accessible from theSetup menu, checks the A/D,RAM, EEPROM and keypad.

    POWERFUL FEATURES FOR YOUR APPLICATION

    track during calibration. Theoutput range is software scalable,and you can easily select between4-20 mA, 0-20 mA, 20-4 mA or 20-0mA. The degree of damping foreach output is programmable.

    AlarmsIndependent oxygen alarms areeach high or low selectable. Onealarm can be set to show when theunit is in calibrate or verify mode.Two additional alarms are pro-vided with the combustibles and/or methane option. Relays can beset to energize or de-energize onalarm. Independent service alarmsare also included. Dry contacts areprovided for each alarm.

    Calibration and Verificationn Perform a full calibration or a

    verification that checks thecalibration without makingchanges

    n Select the process recovery timeand the calibration gas run time

    n View the previous calibrationand verification data.

    Additional Featuresn Password protectionn Programmable process pressure

    compensationn User Text allows a personalized

    identification text string to beplaced on lines 1, 2 or 3

    O2 Cell Lifetime ExtenderNatural aging, combined withprocess conditions, may cause theperformance of zirconium oxide todegrade over time. Some analyz-ers simply predict the approxi-mate lifetime remaining after eachcalibration. Our analyzers gofurther by actually extending thecell life through precise control ofcell operating parameters.

    3-in-1 DisplayWith the WDG-IVCM analyzer,you can display oxygen combus-tibles and methane together.

    Analog OutputIndependent current outputs foroxygen, combustibles, and/or themethane option are available. Youcan assign oxygen, cell tempera-ture, thermocouple mV, cell mV,combustibles or methane to eachcurrent output and choose hold or

  • 5IQ and Series 2000 Features

    The Series 2000 Control Unit isoffered in several packagingoptions. For many flue gas applica-tions, the control unit is locatedoutdoors or in an area that can beexposed to water. For these environ-

    FLEXIBILITY TO SUIT YOUR NEEDS

    CUSTOM PACKAGING IS ALWAYS AN OPTION.CALL US IF YOU HAVE SPECIAL NEEDS FORYOUR APPLICATION. OUR APPLICATIONSSPECIALISTS WILL BE PLEASED TO DISCUSSTHE OPTIMUM ANALYZER CONFIGURATION TOSUIT YOUR PROCESS.

    ments, the weatherproof NEMA 4(IP 56) or weatherproof stainlesssteel NEMA 4X (IP 56) option isavailable. Both are of a rugged,welded construction designed forlong life in heavy industrial service,and can be either wall- or panel-mounted. Hazardous area enclo-sures are also available, includingCENELEC and NEC approvedversions.

    Analyzer SelectionWHICH THERMOX ANALYZER IS RIGHT FOR YOUR APPLICATION?

    With the great variety of flue gas applications, such as boilers, furnaces, kilns, processheaters and incinerators, no single analyzer design can be suitable for all.

    Thermox has developed five types of analyzers to meet your application needs.

    Weatherproof NEMA 4 (IP56)

    Rack Mount

    Wall Mount

    Available options:1 Combustibles detector2 Methane detector3 Excess Fuel Software

    CLOSE-COUPLED EXTRACTIVEn WDG-IV 3

    n WDG-IVC 1

    n WDG-IVCM 1, 2

    n WDG IV/IQn WDG IVC/IQ 1

    n WDG IVM/IQ 2

    n WDG IVCM/IQ 1, 2

    n WDG IVC / 0-500 PPM 1

    n WDG IVC/IQ / 0-500 PPM 1

    n WDG IVCM/IQ / 0-500 PPM 1, 2

    CONVECTIVEn WDG-HPII 3

    n WDG-HPIIC 1

    n WDG HPII/IQn WDG HPIIC/IQ 1

    n WDG HPIIM/IQ 2

    n WDG HPIICM/IQ 1, 2

    INSITUn WDG-INSITUn WDG INSITU/IQn WDG HTIn WDG 210

    ACROSS-THE-STACK IRn EI 20/IQ

    EXTRACTIVEn CEM/O2

    3

    n CEM O2/IQn WDG HUMOX

  • 6WDG-IV and WDG-IVCCLOSE-COUPLED EXTRACTIVE FLUE GAS ANALYZERS

    Designed for fast response in a wide range of flue gas applications, thisanalyzer mounts directly on the combustion process to provide con-tinuous measurement of oxygen (WDG-IV) or oxygen and combustibles(WDG-IVC). It is suitable for gas streams up to 1300F (704C) withstandard 316 SS probes. Flue gas temperatures up to 3200F (1760C)can be measured using optional high temperature probes. For corrosivegas streams, sample wetted parts in Hastelloy or Inconel are available.The analyzer is provided in a range of mounting styles. For hazardousarea locations, purged or explosion-proof versions are available.

    APPLICATIONSn Ceramic kilnsn Coke ovensn Gas and light oil-fired boilersn Incineratorsn Ore roasting furnaces and convertersn Process heatersn Reformers

    Unique Close-Coupled Extractive Technique

    A sample is drawn from the process stream by means of an air-operatedaspirator and is immediately returned to the process. A portion of thisgas rises past the measuring cell and returns to the primary loop. Allsample wetted parts are heated to maintain the sample above thedewpoint of the gas.

    For more detailed information, ask for bulletin n P-260 (WDG-IV)n P-220 (WDG-IVC)n P-510 (WDG IV/IQ)n P-570 (WDG IVC / 0 to 500 PPM)n P-600 (WDG IVC/IQ) orn P-610 (WDG IVC/IQ / 0 to 500 PPM)

    Close-coupled extrative oxygen/combustibles analyzer

    Open WDG-IVC

    WDG-IV

  • 7WDG-IVCMCLOSE-COUPLED EXTRACTIVE FLUE GAS ANALYZER

    The 3-in-1 analyzer for measure-ment of oxygen, combustiblesand methane in natural gas-firedapplications.

    The WDG-IVCM analyzer is ideal for natural gas-fired power boilersor for those using gas during start-up and shutdown. With this ana-lyzer, you can monitor oxygen and combustibles for maximum fuelefficiency. In addition, the methane detector can be used for purge andlightoff cycles during start-up and shutdown.

    For more detailed information, ask for bulletin n P-230 (WDG IVCM)n P-600 (WDG IVCM/IQ or WDG IVM/IQ) orn P-610 (WDG IVCM/IQ / 0 to 500 PPM)

    NEMA 3R standard lift-off enclosure

    NEMA 4X optional hinged enclosure

    Process gas is drawn into the analyzers primary sample loop byan aspirator and returns to process. Gas enters the split flow blockthat is separated by a baffle and contacts the catalytic combustiblesand methane detector. It then passes to the zirconium oxide cell,where the oxygen concentration is measured, before returning tothe main sample loop and then back to the process.

    Convection Loop Operation

    Convection Loop

  • 8Designed for applications with high particulate levels in the gas stream,the analyzer mounts directly on the combustion process to providecontinuous measurement of oxygen (WDG-HPII) or oxygen and com-bustibles (WDG-HPIIC). This analyzer is suitable for gas streams up to1300F (704C) using the standard 316SS probe/filter assembly. Flue gastemperatures up to 2800F (1537C) can be measured using optionalhigh temperature probes. For corrosive gas streams, sample wettedparts in Hastelloy or Inconel are available. The analyzers can be pro-vided in a range of mounting styles. For hazardous area locations,purged or explosion proof versions are available.

    APPLICATIONS

    Convection Principle

    Since the sample inlet and return are inside the filter and at the sameabsolute pressure as the process, there is no flow through the filter. Theflue gas oxygen diffuses through the filter and the sample in the filterflows through the sensor plumbing because of a convective motiveforce. The sample in the vertical cell housing is heated to 1283F (695C)while the sample in the return vertical tube is at about 480F (249C).This temperature difference generates a density difference whichcirculates the gas through the system. Process gas enters and replen-ishes the gas inside the filter by diffusion only. Thus the filter neverblocks, making it ideal for high particulate applications.

    For more detailed information, ask forbulletin

    n P-210 (WDG-HPII)n P-270 (WDG-HPIIC)n P-530 (WDG HPII/IQ) orn P-590 (WDG HPIIC/IQ)

    A methane option is available usingthe split flow block. Ask forbulletin P-590 (WDG HPIIM/IQ orWDG HPIICM/IQ).

    WDG-HPII and WDG-HPIICCONVECTIVE FLUE GAS ANALYZERS

    WDG-HPII

    Open WDG-HPIIC

    Incineratorsn Garbagen Hazardousn Liquidsn Radioactiven Spent acidn Sludgen Tail Gas

    Furnacesn Annealingn Blast furnacen Carbonn Coke ovenn Galvanizingn Graphite anoden Meltingn Reformingn Reheatn Roastingn Sinteringn Wax burnout

    Boilersn Bagassen Coaln Heavy oiln Wood

    Kilnsn Limen Cement

  • 9WDG-InsituINSITU FLUE GAS OXYGEN ANALYZER

    The WDG-Insitu is a direct insertion-type oxygen probe where thezirconium oxide cell is directly in the stream of the products of combus-tion. The patented WDG-Insitu is designed for applications where theflue gas temperature does not exceed 1250F (677C) and where combus-tibles measurements are not required. The outer probe material isRA330, which is highly resistant to corrosion. For hazardous arealocations, explosion-proof and purged versions are available, includingCENELEC EEx d IIC. For high particulate applications, a filter can easilybe added.

    Simplified Construction For Easy ServiceWith the unique design of the WDG-Insitu oxygen probe, field replace-ment of all components is easily performed. While the outer protectiontube remains in the process, the inner tube structure, that houses theoxygen cell and furnace/thermocouple assembly, can be removed easily.With no need to return the probe to the factory, extensive down timeand expensive repairs on your oxygen insitu probe are no longernecessary.

    For more detailed information, ask for bulletin n P-300 (WDG-Insitu) orn P-490 (WDG Insitu/IQ)

    BD

    C E F G H I J

    K

    A

    Cross-section of Insitu probe

    Open WDG-Insitu

    WDG-Insitu

    A NEMA 4X junction box

    B 1/4 compression connectionfor calibration gas

    C 2 NPT (M) connection forflange adaption

    D Inner tube structure

    E Outer protection tube (no needto remove once installed)

    F Support spiders

    G Zirconium oxide cell lead(s)

    H Zirconium oxide cell heater/thermocouple

    I Zirconium oxide cell withinsulation sleeve

    J Ceramic sheath

    K Protective screen

  • 10

    The EI 20/IQ analyzer is an infrared (IR) across-the-stack analyzer using agas filter correlation measurement to determine the amount of CO in theprocess flue gas. It can be used in flue gas temperatures up to 1000F(537C) with maximum path length of 30 ft. The analyzer is designed for360,000 ppm-inches (10,000 ppm-meters). It can accept a type K thermo-couple input for process temperature compensation or a user-enteredtemperature value. An integral remote calibration option is available.

    FEATURES

    n NEMA 4X (IP 56) enclosuresfor the source and sensorsides

    n Window purge air connec-tions for instrument air;optional blowers or filterassemblies for high negativepressure processes

    n Source and sensor sideflanges and mirror opticalalignment adjustments

    n Down stack calibration/verification with 1.5 calibra-tion gas chamber/window

    For more detailed information,ask for bulletin P-540.

    EI 20/IQCARBON MONOXIDE ANALYZER

    EI 20/IQ Analyzerassembly

    n Resolution settings for 1 or10 ppm

    n Software selectable calibra-tion options

    n Extensive analog anddigital I/O capabilities

    n Bi-directional RS-485 serialcommunications port

    n No interconnecting cablenecessary between sourceand sensor enclosure

    n HART Communicationsoption available

  • 11

    For more detailed information,ask for bulletin n P-400 (CEM/O2) orn P-500 (CEM O2/IQ)

    APPLICATIONSn CEM applicationsn Emissions and compliancen Extractive sampling systemsn Instrumentation panels

    CEM/O2EXTRACTIVE FLUE GAS OXYGEN ANALYZER

    The cost-effective CEM/O2 analyzer provides advanced systemsfeatures that integrate into your overall emissions monitoring system.The analyzer complies with 40 CFR Part 60 and Part 75 Federal EPAperformance specification requirements along with regional require-ments for oxygen monitors.

    Custom designs, packaging, labeling, and software can be provided, aswell as OEM stocking and delivery agreements.

    The CEM/O2 analyzer is also available in a 19 rack mount version(integrated sensor and controller).

    CEM/O2

    Extrative oxygen analyzer

    WDG-HumoxPERCENT MOISTURE/OXYGEN IN FLUE GAS

    The WDG-Humox measures both percent moisture and percent oxygenin flue gases.

    APPLICATIONSThis 2-in-1 analyzer is designed for flue gas applications up to 3200F(1760C) containing up to 85% moisture. Ideal for direct-fired combus-tion application where oxygen and moisture content can continuouslychange. This ranges from utilities requiring the readings for emissionspurposes to processes using direct-fired combustion air that require aspecific moisture level in the process gas for manufacturing, dryingcontrol or other quality purposes.

    For more detailed information, ask for bulletin P-450.

    WDG-Humox

  • 12

    With the Excess Fuel Software option, you can monitor processescontaining fuel-rich atmospheres such as bright annealing or decarbur-izing applications, or for the fume stage of an incinerator.

    The Remote Calibration Unit and the Calibration Moduleallow automatic calibration of your analyzer.

    With calibration gases permanently connected to the RemoteCalibration Unit, you can program the Series 2000 or IQ-based analyzers to activate a calibration cycle at a timedinterval. A calibration can also be performed remotely ondemand by contact closure or via the RS-485 communicationsport. In addition, you can calibrate locally by manuallypressing keys on the Series 2000 or IQ Link.

    Excess Fuel Software OptionMONITOR COMBUSTION PROCESSES THAT OPERATE IN EXCESS FUEL ATMOSPHERES

    Automatic Calibration OptionSIMPLIFIES CALIBRATION OF ANALYZER

    Similarly, a verification can be performed in which the calibration gas valuesare recorded but no changes are made to the calibration settings in the ana-lyzer.

    Automatic calibration capabilities can be easily added to an analyzer at a laterdate by simply plugging a Calibration Module into the Series 2000 or IQanalyzer and installing a Remote Calibration Unit near the sensor.

    Remote Calibration Unit

    Theory of OperationIn the absence of molecular oxygen, the zirconium oxide cell responds to minute amounts of oxygen pro-duced by the dissociation of water or carbon dioxide at the high cell operating temperature. This dissociationis inhibited by the presence of combustibles in the sample gas. As the combustibles concentration increases,the oxygen concentration decreases, and the output signal from the cell becomes greater. This means that thecell keeps responding when there is no net oxygen in the flue gas. In fact, it becomes a sensor of net combus-tibles.

    FEATURESn User-selectable inputs include defining the hydro-

    gen-to-carbon ratio (H/C) of the fuel or selecting froma menu of common fuels

    n Alarms can be set for high or low readingsn Analog output and display scalable to cover com-

    bined excess oxygen to excess fuel/combustiblesranges

    n 2-point calibration in excess fuelrange

    APPLICATIONSn Bright annealingn Decarburizingn Furnace atmospheresn Heat treatingn Incineratorsn Water-to-hydrogen ratio

    The excess fuel software enables fuel-rich atmospheres to bemeasured. It displays and provides alarms and analog outputsfor the following:

    n Excess fuel n Oxides to fuel ration Combustibles n Fuel to oxides ratio

  • 13

    Hazardous Area Classification OptionsOPTIONS TO MEET VIRTUALLY ANY HAZARDOUS AREA CLASSIFICATION

    NEC Class I, Division 1Explosion-proof versions of the WDG sensor and Series 2000 ControlUnit are designed to meet NEC Class I, Division 1 requirements.

    NEC Class I, Division 2n The Series 2000 Control Unit and IQ Link are UL-approved for NEC

    Class I, Division 2, Groups A, B, C, and D classifications without theneed for a Z-purge apparatus.

    n The WDG-IV, WDG-IVC, WDG-IVM, HPII, HPIIC, and HPIIManalyzers with the Division 2 option are UL-approved for NEC ClassI, Division 2, Groups A, B, C, and D classifications.

    n The NFPA 496 Z-purge provides a fail-safe design by placing the flowmeasurement on the outlet of the enclosure.

    For more detailed information, ask for bulletin n P-250 (WDG-IV/IVC CENELEC) or

    n P-30C (WDG-Insitu CENELEC)

    CENELEC INSITUEEx d IIC T3-approved

    WDG-IV/IVC Class I,Division 2-approved

    CENELECn The CENELEC WDG-Insitu is certified to EEx d IIC T3 when used

    with a Series 2000 Control Unit, or to EEx d IIB T2 if used with a non-Thermox control unit.

    n The Series 2000 is certified to EEx d IIC T6.n The optional Remote Calibration Unit (RCU) is certified to EEx d IIC

    T5, and can be used with the CENELEC Insitu or CENELEC WDG-IV/IVC analyzers.

    CENELEC Series 2000 EnclosureEEx d IIC T6-approved

    CENELEC WDG-IV/IVCEEx d IIC T5-approved(shown with CENELEC RCU)

    WDG-Insitu Class I,Division 2-approved

  • 14

    All Thermox analyzers use the Series 2000 or IQ microprocessor-basedcontrol electronics. They are designed with an emphasis on reliability,serviceability, ease-of-use, and digital communications.

    In addition to our flue gas analyzers, Thermox offers analyzers for indus-trial gas and pre-mix gas applications. These analyzers use the Series 2000Control Unit.

    Pre-Mix Gas AnalyzersThe Thermox PreMix 2000 and CMFA-P 2000 (portable) analyzersaccurately and continuously measure the proportions of oxygen andfuel in pre-mix gases. They are used wherever efficient control isrequired for an open flame combustion process such as in glass andfiber glass manufacturing or in flame treating applications.

    Thermox Industrial Gas Oxygen Analyzers

    PreMix 2000 analyzer for measuring air/fuel proportions in pre-mix gases

    For nitrogen purity and otherindustrial applications,Thermox TM2000 and CG1000industrial gas oxygen analyzersprovide a quick response timeto process changes, and operatefrom 100% to 0.1 ppm O2.

    CMFA-P 2000 portable pre-mix/flue gas oxygen analyzer

    TM2000 field-mountedoxygen analyzer

    CG1000 portable/benchtopoxygen analyzer

    For more detailed information, ask forbulletin n P-410 (TM2000) or

    n P-420 (CG1000)

    For more detailed information, ask forbulletin n P-430 (PreMix) or

    n P-580 (CMFA-P 2000)

  • ANALYZER TYPE WDG-IV WDG-IVC WDG-IVCM WDG-HPII WDG-HPIIC WDG-INSITU CEM/O2(note 1) CCX CCX CCX CONV CONV INS EXT

    Gas MeasuredOxygen yes yes yes yes yes yes yesCombustibles no yes yes no yes no noMethane no no yes no no no noExcess Fuel (Option) yes no no yes no no yes

    Display RangeOxygen 0.1-100% 0.1-100% 0.1-100% 0.1-100% 0.1-100% 0.1-100% 0.1-100%Combustibles

    0-10,000ppm

    or 0-5%0-10,000ppm

    or 0-5% 0-10,000ppm

    or 0-5% Methane 0-5%

    Output RangeOxygen 0-1% to

    0-100%0-1% to 0-100%

    0-1% to 0-100%

    0-1% to 0-100%

    0-1% to 0-100%

    0-1% to 0-100%

    0-1% to 0-100%

    Combustibles 0-2,000ppm to 0-10,000ppm or0-1% to 0-5%

    0-2,000ppm to0-10,000ppm

    0-2,000ppm to 0-10,000ppm or0-1% to 0-5%

    Methane 0-5%

    Accuracy (note 2)Oxygen 0.75% m.v.

    or 0.05% O2

    0.75% m.v.or 0.05% O2

    0.75% m.v.or 0.05% O2

    0.75% m.v.or 0.05% O2

    0.75% m.v.or 0.05% O2

    1% m.v.or 0.05% O2

    0.75% m.v.or 0.05% O2

    Combustibles 2% f.s. 2% f.s. 2% f.s. Methane 5% f.s.

    Response Time

    Initial

  • 16

    n Boiler Feedwater Analyzer(Hydrazine, Reductant)

    n Photometric Analyzer Systems

    n Trace Moisture Analyzers

    n Diode Laser Analyzers

    n Quadrupole Mass Spectrometers

    n Trace Oxygen Analyzers

    OTHER PRODUCTS FROM

    AMETEKPROCESS & ANALYTICALINSTRUMENTS DIVISION

    150 Freeport RoadPittsburgh, PA 15238Tel: (412) 828-9040Fax: (412) 826-0399www.thermox.com

    PROCESS & ANALYTICAL INSTRUMENTS DIVISION

    Printed in U.S.A 1-01

    Bulletin SFC 2000