amesa mercury sorbent trap system datasheet
TRANSCRIPT
AMESA-MTotal Mercury Analyser
Monitoring system for fully-automatic andcontinuous mercury analysis in smoke gas(without wet chemistry)
Already MCERTS and TÜV accredited for continuous dioxin and furan
monitoring, the AMESA - M has been adapted for mercury monitoring.
Consisting of a sample probe assembly and a control cabinet, the
AMESA - M extracts a sample from the flue gas stream under volume
proportional conditions.
Internal Data Logging Functionality The unit can store up to a four week period of sampling data which is
then available to download via a USB flash drive. Analysis can be achieved
with an onsite analyser or a certified laboratory.
Measuring Principle The sample gas is converted into mercury vapour by a combination of
thermal and dry chemical treatment. This is then continuously
measured in a photometer. Probe gas flow is measured after a gas cooler
at 2°C. The concentration is calculated and displayed as dry flue gas.
KEYBENEFITS
• Maintenance-free (6 months) dry
reactor
• High operational safety
• Easy maintenance
• Low cross sensitivities
• Thermal mass flow meter to control
mass flow independent of gas &
pressure
• Dry gas meter to determine sample
volume
Providing continuous Mercury monitoring, The AMESA - M is suited
for audit & control purposes “ “
11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com
Cert No. 996ISO 9001
TECHNICALSPECIFICATION
APPLICATIONS• Waste incinerations
• Sewage sludge incineration
• Contaminated soil burning plants
• Mercury mines and refineries
• Fluorescent light bulb recycling plants
MEASUREMENT PRINCIPLE • Measurements: total mercury
• Measuring ranges 0–45 to 0–500 μg/Nm³
• Measuring principle: UV-absorption
TEMPERATURE & PRESSURE • Flue gas temperature: 0–250 °C
• Flue gas pressure: -50 up to +50 hPa
• Ambient temperature: +5 bis +30°C
PHYSICAL • Dimensions: cabinet 1600 x 800 x 500 mm (H x W x D)
• Weight: 220 kg
• Duct diameter: >0.5 m
POWER • Measuring outputs: 2 x 0 / 4–20 mA/500 Ohm
• Digital outputs: 4 relay outputs, permissible load 250 V,
100 VA
• Digital inputs: 1 potential free outputs
• Supply voltage: 230 VAC, 50 Hz, 1200 VA
sample probe: 650 VA
sample line: 100 VA/m
SPAN & DRIFT • Accuracy: <1% of measuring range
• Detection limit: <1 μg/Nm³
• Reference point drift: <2% of measuring range/month
• Zero point drift: automatic zero correction
APPROVALS• MCERTS
• TÜV Certified
Complete Project Solutions
• Expert consultancy • Project management • In-house project design & build • Installation & commissioning
• After product service & support
Our complete project solutions are developed to assist you in IED compliance. We achieve this through;