air quality monitoring, solutions, reporting & warning · air quality monitoring, solutions,...

18
Air Quality Monitoring, Solutions, Reporting & Warning EPA AQS & AIRNow Data Feeds Air Quality Index Ozone, NOx, SO 2 , CO, Particulates Automatic Gas Chromotographs WRF Model Forecasting & much more... Portable Ozone Photometer Portable Zero Air Source Multi Gas Calibrator Portable Multi Gas Calibrator

Upload: dinhdung

Post on 11-Jun-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Air Quality

Monitoring, Solutions,

Reporting & Warning

EPA AQS & AIRNow Data Feeds

Air Quality Index

Ozone, NOx, SO2, CO, Particulates

Automatic Gas Chromotographs

WRF Model Forecasting & much more...

Portable Ozone Photometer

Portable Zero Air Source

Multi Gas Calibrator

Portable Multi Gas Calibrator

Page 2: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems
Page 3: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 11, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 11, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/[email protected]

Air Quality Monitoring, Reporting & Warning Systems

Instrumentation, Software & Systems that Provide Environmental Solutions for the Planet

LEADS‰ Environmental Monitoring System (EMS) , the ideal commercial solution for collecting, integrating, and processing meteorological, air and water quality data, is built on Numerical Weather Prediction Models. Our solutions deliver accurate data in real-time, providing agencies with the most proficient environmental monitoring systems available.

What does LEADS® EMS bring to the table?

Automatic Data Flagging Air Quality Pages

Graphical Validation Water Quality Pages

Automatic Alerting Specialized Reports

Web based Reports Hourly Air Quality Index

Comprehensive Environmental Tools for Decision Makers in One Integrated System

Portable Ozone Photometer

Sulfur Dioxide

Sulfur Dioxide Pollution Rose Web Page

Ozone Potential Web Page

Ozone

Fire Weather Forecasting

Page 4: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation Air Quality Monitoring, Reporting, Warning 2

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 11, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq [email protected]

Customized Web-Based Data Products

In Texas, Indiana, Virginia, Nevada and across the USA over 500 continuous-monitoring air and water quality sites are powered by Sutron’s Air Quality Division systems producing data that is continuously available on-line to view or plot within seconds. The Texas Commission on Environmental Quality, which operates one of the largest monitoring networks in the world, has collected air and water quality data for over 15 years using Sutron’s* systems. Among the data display and reporting products provided are the following:

Air Quality Monitoring Station Dynamic Monitoring Site Metadata

Animated Pollution Displays & Pollution Rose Maps showing Maximum Concentration, Percentiles

Web Based Graphical Plots

Calibration/Span by Parameters & Web-Based Calibration/Span Reports

Graphical Editing & Validation Tools

Data Validation Cal/Span History

Web-Based Operator Logs with Sorts & History

Web-Based Data Query Tools

Web-Based Plots

Ozone Monitoring Forecasts

Public Web Pages with Graphical AIQ Report

EPA Air Quality Index Report

Data Reports by Site, Daily, Monthly, Annually

Air Quality Monitoring InstrumentationSutron’s Air Quality Division also manufactures measuring equipment.

Multi-Gas Dilution Calibrator & Portable Multi-Gas Dilution Calibrator

Heated Sample Manifold

Methane Catalytic Oxidizer

Ozone Transfer Standard Photometer

Permeation Ovens (2 models) & Portable Permeation Oven

Zero Air System & Portable Zero Air System

Capabilities System Integration

Air Quality Monitoring W/Automatic Calibrations

Trace Gas Monitoring

Water Quality Monitoring (Surface and Well)

EMRS Air & Water Quality Alert Systems

Ozone Alerting

Automatic Gas Chromatographs(with Data Flagging)

FTIR Systems

Single and Multi-Can Canisters

NEXTRad, Satellite Imagery, Weather User Interface

Additional Services Site Selection and Site Prep

Site Installation and Retrofits

Instrument Selection and Installation

Communication Systems

Shelters

Station Plumbing

Calibration Systems

Meteorological Instruments & Towers

Solar Powered Monitoring Stations

Clark County, Nevada Air Quality Monitoring Web Page

Texas Commission on Environmental QualityWeb Page

Chlorophyll

Ammonia

Methane

Carbon Monoxide

Tribal Environmental Exchange Network

Page 5: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Portable Gas Dilution Calibrator

Users can easily perform gas dilution, ozone and GPT calibration with this Intelligent Portable Gas Dilution Calibrator from Sutron’s Sabio Division.

The Model 2010 is small in size and weight for users who require true portability. Enhanced with the latest technology, the Model 2010 calibration system is an uncomplicated and effective way to precisely control the dilution of gas standards for calibration of ambient air and source monitoring analyzers.

The Model 2010 calibrator makes use of embedded microprocessor technology to enhance the accuracy and control of the featured calibration system. The unit can be operated manually, automatically, or semi-automatically by remote control. It also has the capability to produce and store in memory 20 calibration sequences. with up to 20 levels of source/dilution in each sequence.

Users can use either the front panel membrane keypad, external keyboard, or serial port to input calibration parameters.

The instrument’s pneumatic system uses Mass Flow Controllers to precisely meter gas streams and implements multiple types of curve correction algorithms to linearize calibrations.

Calibrator design exceeds US EPA calibration method requirements. Dilution components are calibrated with standards and test equipment traceable to the National Institute of Standards and Technology (NIST).

Standard Features

Pnuematics

Diluent Mass Flow Controller, 0-10 SLM

Source Mass Flow Controller, 0-100 SCCM

2 Calibration Gas Input Ports and 1 Purge Port

4 Output Ports and 1 Vent Port

1 Diluent Gas Input Port

Electronics

Bright Active Matrix Color Display

42-Button Membrane Keypad

8 Bits Digital Input/Output

2 Serial Ports for Data Communications

External PC keyboard Input Port

CentronicsParallel Printer Port

Air Source Control Port

Electrical

Standard 98-264 VAC, 150-300 VA, 50/60 Hz Operation

Optional Features

Alternate Diluent Mass Flow Controller, Ranges 0-20 SLM

Alternate Source Mass Flow Controller, Ranges 0-1 SLM

Second Source Mass Flow Controller, Ranges 0-1 SLM

Additional Calibration Gas Input Ports

Multi-source Blending

Permeation Port

Additional Status I/O Bits up to 24 Total

Internal Instrument Solenoid Driver, 6 Outputs, 24V @ 3W ea

Internal Ozone Generator

Ozone Generator UV Optical Servo Control Loop

Side mounted carrying straps

High-impact transport case

Rack Mount Version Available in 5.25”H U-Chassis (2010D) with internal UV photometer

Page 6: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Portable Gas Dilution Calibrator

SPECIFICATIONSSpecifications subject to change without notice

Dilutions System

Input Dilution Gases 1 Standard

Input Source Gases 1 Standard, 1 Purge

Additional Inputs 4 Additional

Dilution Mass Flow Controller 0-10 SLPM

Optional Ranges 0-20 SLPM

Source Mass Flow Controller 0-100 SCCM

Optional Ranges 0-1 SLPM

2nd Source Mass Flow Controller

0 thru 2 SLPM

Flow Accuracy < ±1% full scale

Flow Repeatability < ±0.15% full scale

Linearity < ±0.5% full scale

Input Pressure 12-35 PSIG

Output Manifold 4 Outputs Standard

Optional Outputs 4 Additional

Optional Internal Ozone Generator

Output 0.05-1.5 ppm @ 5 SLPM

Accuracy ± 2% of Set Point or

± 3 ppb @ 5 SLPM

Nominal Flow 100 SCCM, ± 1 SCCM

UV Lamp Temperature 50°C, ± 0.1°C

Calibrator interface

Operation Manually through 42-button membrane keypad, keyboard, serial ports

Calibration Definitions

20 factory-defined calibration sequences with up to 20 points per sequence

Calibration Types

Gas Diluiton, Ozone, GPT

Gas Definitions 1 Dilutent Gas20 Source Gases

Auto Calibrations

20 timer driven cal routines performing user-defined calibration sequences on a 7-day calendar

Digital Inputs/Outputs

8 status I/0 bits control calibrator functions by contact closures or TTL logic & output bits monitor calibrator functions

Optional I/O Additional 16 status I/O bits for a total of 24 I/O bits

Communications RS232

Calibrator System

Operating Temperature

5°C to 40°C

Dimensions 6.2 (15.6 cm) H 14.3 (36.4 cm) W12.4 (31.6 cm) D

Average Weight 19 lbs. (8.6 kg)

Input Voltage 98-264 VAC, 100 VA, 50/60 Hz Operation

Page 7: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 11, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/[email protected]

Portable Zero Air SourceModel 2020

Overview

The Model 2020 Portable Zero Air Source meets a wide range of requirements for the environmental industry. The unit achieves maximum proficiency when it operates with the Sabio Model 2010 Gas Dilution Calibrator (for testing of pollution monitoring analyzers).

The Model 2020 uses an internal drying system for removal of moisture and is coupled with filter canisters for scrubbing pollutants. Optional features such as the internal Carbon Monoxide Oxidizer or external Methane Oxidizer further the Model 2020’s clean air capabilities.

The small package size and weight make the Model 2020 suitable for portable applications. The unit is ideal for field auditors who must travel with their testing equipment and for limited bench space environments.

User Benefits

The Model 2020 compressor can be activated remotely by all Sutron calibrators using a standard air source control cable. This allows the compressor to operate only when the calibrator requires zero air.

In addition, the Model 2020 is simple to operate. The unit has a front panel mounted pressure regulator with a locking mechanism and pressure gauge. Additionally, there is a front panel mounted condensate purge switch for manual purge. The internal air source control board can be adjusted for automatic purge frequency and duration. On power down or power failure the pressure tank is automatically drained of moisture and pressurized air.

Standard Features

Standard 12 LPM @ 15 PSIG

Condensate/purge valving for moisture removal. Operates automatically with internal circuitry or manually with the purge switch.

Front panel mounted pressure gauge.

Front panel mounted adjustable pressure regulator.

Front panel mounted purge switch.

Internal moisture removal system.

Media canisters for removal of NOX, S02, H2S, 03, low level organic compounds and CO.

Side mounted carrying straps (shown in picture)

Optional Features

Tilt-up front handle with tilt-up front feet

High-impact transport case

I nternal CO catalytic oxidizer

Model 2050 Methane Catalytic Oxidizer (separate portable chassis)

SPECIFICATIONSSpecifications subject to change without notice

AIR QUALITY

Standard unit NOX, NO, NO2, 03, SO2, H2S <0.5 ppb

Standard unit with CO catalytic Oxidizer:

NOX, NO, NO2, 03, SO2, H2S <0.5 ppb, CO <0.025 ppm

Standard unit with external methane catalytic oxidizer

NOX, NO, NO2, 03, SO2, H2S <0.5 ppb, Hydrocarbon ≤ 0.02 ppm

MECHANICAL

HeightWidthDepth

6.5 in. (16.5 cm) 13.9 in. (35.3 cm) 15.0 in. (38.1 cm)

Average Weight 23 lbs (10.4 kg)

ELECTRICAL

Standard 98-132 VAC

Optional 196-264 VAC, 50/60 Hz

Page 8: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Zero Air Source#1001

Overview

Sutron’s Zero Air Source is designed to meet the requirements of clean and dry air for dilution calibrators or other environmental applications. The units offer a wide range of standard features/options to accommodate a host of monitoring applications, including rack mount or portable environments

Equipped with dual coalescent filters and a permeation drier for removal of moisture from the source stream, further moisture removal is enabled by automatic drain removal using adjustable control circuitry, manual purging from the front panel, or upon completion of calibration, when controlled automatically by a Sutron dilution calibrator.

Air pollutants are removed using re-usable dual rear-mounted filter canisters. Other pollutants such as carbon monoxide or hydrocarbons can be removed with an internal oxidizer. Both types of internal oxidizers come equipped with front panel mounted temperature controllers that display the set-point as well as the temperature monitor.

SpecificationsSpecifications subject to change without notice

Air Quality Standard Unit NOX, NO, NO2, O3, SO2, H2S , < 0.5 ppb

Standard unit with CO catalytic Oxidizer

NOX, NO, NO2, O3, SO2, H2S < 0.5 ppb, CO < 0.025 ppm

Standard unit with hydrocarbon oxidizer

NOX, NO, NO2, O3, SO2, H2S < 0.5 ppb, Hydrocarbons < 0.02 ppm

Mechanical Rack Mount Unit Model 1001

Size 8.75 in. (22.2 cm) H X 19.0 in. (48.3 cm) X W 19.0 in. (48.3 cm) D,

Average Weight: 55-70 lbs. (25-32 kg)

High Output Portable Unit Model 1001P

Size 14.24 in. (36.2 cm) H X 21.0 in. (53.3 cm) X W 24.75in. (62.9 cm) D

Average Weight: 51-61 lbs. (23.1-27.7 kg) Portable Unit

Electrical Standard 120 VAC 60 Hz, 470 VA operation

Optional 240 VAC, 50 Hz, 340 VA operation

Standard Features

Standard 22 SLM at 25 psig air delivery (flow may decrease 2-4 SLM with addition of some options or changes in line frequency)

3.4 liter surge/condensate tank

Condensate/purge valve that operates both automatically and manually to drain condensate

Front panel mounted pressure gauge

Front panel mounted adjustable pressure regulator

Front panel mounted purge switch

Automatic Control via contact control or a switched low voltage DC power source

Internal moisture removal system

Media canisters for removal of NOX, SO2, H2S, O3, low level organic compounds and CO

Long-life brushless, oil-free compressor

Quiet operation

Optional

Internal or external CO catalytic oxidizer

Internal or external hydrocarbon oxidizer

Digital pressure meter

Output flow meter (Model 1001)

Rack-mount kit (Model 1001)Model 1001P - Portable

Smaller portable units available for lower output

applications (Model 2020).

Page 9: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Methane CatalyticOxidizerModel 1000M Methane Catalytic Oxidizer

Overview

The Model 1000M Methane Catalytic Oxidizer is a reliable method for the

removal of methane or carbon monoxide. Ideal for use with any of Sutron’s air

sources, this oxidizer is efficient in furthering the reduction of air pollutants

before the zero air.

The Model 1000M Methane Oxidizer meets industry requirements for a

thermal oxidizer with increased oven reliability and stability; faster rise times;

higher converter efficiencies; ease of use; and packaging tailored to meet the

customer’s needs.

The proprietary two-stage, multiple path tube design and highly efficientoven

design enable the Model 1000M Methane Oxidizer to efficiently oxidize at

relatively low temperatures. The oven temperature is precisely maintained by a

microprocessor-based temperature controller driving a high-capacity optically

isolated triac control module. The dual display temperature controller uses

pulse width modulation to regulate the heat cycle to prevent over shooting and

minimize thermal noise from on/off cycling. A thermal cutoff switch works with

the heating element to protect the electronic circuitry and oven from thermal

overload conditions.

Average Model 1000M Methane Oxidizer converter efficiency is 95 to 98

percent. Insertion of the Model 1000M into the analyzer sample loop results in a

decrease of analyzer sensitivity by about one percent of full scalesetting and an

increase in analyzer rise/fall time by five to 10 percent.

Optional Features / Spares

Catalyst Recharge Kit

Rack Mount Kit

Specifications

SPECIFICATIONSSpecifications subject to change without notice

Temperature Controller

Microprocessor-based

Features Proportional heater control

Digital display of set point

Digital display of over temperatures

Dimensions 5.25” (13.3 cm) H x 19” (48.3 cm) W x 15” (38.1cm) D

Weight 18 lbs (8.2 kg)

Communications RS232

Power Requirements

98-132 VAC(optional 196-264)360 VA,50/60 Hz

Note: The methane catalytic oxidizer is available

as an internal option for the Model 1001 and Model

1001P series Zero Air Sources.

Page 10: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Heated Sample ManifoldModel 1002/1004 Heated Sample Manifolds

OverviewCollecting and distributing homogenous ambient air samples for

gas analysis is easy with the Sutron Model 1002 and 1004 heated

Sample Manifolds. Able to work equally well in dry or humid and salty

environments, the manifold is well suited for collecting reactive air toxic

data.

For durability and reactive inertness, Model 1002 is constructed of 316

stainless steel. Model 1004 is constructed with highgrade glass. Particulate

build-up and sample degradation are minimized with the full-length,

heated, vertical design.

SPECIFICATIONSSpecifications subject to change without notice

Models 1002, 1004

Rain protection Sample cone

Sample heating Full length

Sample outlet ports Six ¼” (6.35 cm)

Temperature Over temperature cutout

Construction Fully wired & insulated

Blower High capacity 40 CFM with flexible connecting hose

Mounting Flange & gasket hardware for roof & flood mounting

Power Standard 98-132 VAC (optional 196-264 VAC) Operation, 250 VA, 50/60 HzAdd 50 VA for 24” (60.96 mm) sample manifoldextension

Options, Accessories Model 1003 Manifold ValveController

Custom Manifold lengths

Additional sample ports

Gas analyzer sample stream valves

24 volt DC blower

Bug Screen

Cleaning kit

Page 11: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Portable Permeation OvenModel 2505

Overview

Sutron’s Model 2505 Portable Permeation Oven generates precise gas concentrations for calibration of air quality analyzers, source monitors and other instruments. The Model 2505 may be combined with the Model 2010 and 4010 Gas Dilution Calibrators to form a very powerful intelligent calibration system.

Carrier air is provided by an internal air supply featuring an inlet filter, quiet internal pump and rear mounted scrubber canister. A precision regulator and critical orifice is used to regulate carrier flow through the permeation oven. A 0-15 psig pressure gauge is calibrated to provide carrier flow reference points.

The heart of the Model 2505 is a large capacity Teflon coated permeation chamber that can accommodate many types and sizes of permeation tubes concurrently. A front-panel-mounted knurled screw-in stopper is provided for easy access of the permeation chamber. The permeation chamber is housed in a precise temperature controlled insulated oven that features a carrier pre-heater, heaters and RTD sensor.

The oven temperature is precisel maintained by a microprocessor based temperature controller driving an optically isolated DC control module. Pulse width modulation is used to regulate the heat cycle to minimize thermal noise due to heater on/off cycling.

Dual displays on the temperature controller provide direct readout of set point temperature and actual oven temperature. Oven temperature is maintained within +/- 0.1 deg. C of setpoint.

The oven temperature control system and carrier air supply are battery backed for 8 hours of operation.

Standard Features Permeation Oven

Teflon Coated Chamber: 3.09 cm inside chamber diameter, 13.4 usable length, 107 cubic cm volume. Nominal carrier gas flow is 100 sccm

Stainless Steel and Teflon gas handling materials

Oven operating range 5 deg. C above ambient to 60 deg. C,

Unit operating temperature 5-40 deg. C

SPECIFICATIONSSpecifications subject to change without notice

Mechanical 6.0” (15.2 cm) H 14.5” (36.8 cm) W 18.0” (45.7 cm) D

Weight 24 Lbs. (10.9 kg)

Electrical Standard: 98-264 VAC, 150 VA, 50/60 Hz

Battery Approximately 8 hours

Accessory High Impact Transport Case

Page 12: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Permeation OvenModel 2500C

OverviewSutron’s Model 2500C Permeation Oven Calibrator generates precise gas

concentrations for calibration of air quality analyzers, source monitors

and other instruments. The Model 2500C generates span and zero

reference points for calibration of analyzers. The span concentration

is adjusted using a precision front panel mounted potentiometer. The

calibration selection knob is used to determine whether a span point is

active, zero point is active or to place the unit in Standby mode for either

non-use or remote activation using digital I/O. An internal calibration

solenoid switches the instrument sample inlet from the Sample Manifold

Port to the Model 2500C permeation calibration facilities.

The Model 2500C contains an internal zero air supply for supplying

carrier and diluent air for diluting generated source gas. The zero

air source consists of an inlet particulate filter, air compressor, dual

coalescing filters with float drain valves, pollution scrubber canister and

pressure regulator with gauge. Dilution flow is maintained by a 10 SLM

mass flow controller with dual readouts and is front panel mounted for

visual confirmation of flow

Standard Features

Teflon Coated Chamber: 3.09 cm inside

chamber diameter, 13.4 usable length, 107

cubic cm volume.

Nominal carrier gas flow is 100 sccm.

Stainless Steel and Teflon gas handling

materials

Oven operating range 5 deg. C above ambient

to 60 deg. C.

Unit operating temperature 5-40 deg. C

Mass Flow Controller operating range: 1-10 SLM

Air Cleanup of NOX, SO2, H2S, O3 and

The heart of the Model 2500C consists of a large capacity permeation

chamber that can accommodate many types and sizes of permeation

tubes through a front panel accessed stopper. The well insulated chamber

is equipped with an RTD sensor connected to a microprocessor based

temperature controller with dual displays drives an optically isolated control

module to precisely maintain oven temperature within +/- o.1 deg. C of set

point.

SPECIFICATIONSSpecifications subject to change without notice

Mechanical 5.25” (13.3 cm) H 19” (48.3 cm) W 22” (55.9 cm) D

Weight 30 Lbs. (13.6 kg)

Electrical Standard: 98-132 VAC or 196-264 VAC, 50/60 Hz, 650 VA

Accessories Unit without internal zero air

Rack Mount Kit

Page 13: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 5, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Permeation OvenModel 2500

Overview

Sutron’s Model 2500 Permeation oven is a simple, effective tool for generating precise gas concentrations for calibrating air quality analyzers, source monitors and other instruments. When combined with Sutron’s Gas Dilution Calibrators, it creates a powerful and intelligent calibration system.

The Model 2500 Permeation Oven chamber is large enough for a variety of tasks and was designed to be used in a number of different settings.

Operators can use many types and sizes of permeation tubes concurrently in the Model 2500’s large-capacity gas permeation chamber. A front panel mounted precision pressure regulator, reference pressure gauge, temperature controller and oven inlet give the operator easy access to the unit controls at all times. Carrier air is provided by the Model 2500’s clean air system.

Model 2500 Optional Features

Rack Mount Kit

Dual Permeation Oven

Note: The permeation oven is available as an internal option for the Model 4010 Gas Dilution Calibrator.

Model 2505 Portable Permeation Oven

The Model 2505 enjoys the same basic features as the Model 2500, but it offers a battery backup for continuous control of the permeation oven while the unit is in transit. In addition, the portable unit is housed in a smaller chassis for easy transport and handling.

Model 2505 Specification Variances

Battery Backup: 10 hours operation when fully charged

Size Approximately: Height 6.2 in. (15.6 cm) x Width 14.3 in. (36.4 cm) x Depth 14.4 in. (36.6 cm)

· Weight 18 lbs (8.2 kg)

Model 2505 Optional Features

High Impact Transport Case

Dual Permeation Oven

Model 2500 SpecificationsSpecifications subject to change without notice

Temperature controller Microprocessor-based

Proportional heater control

Digital display of set point & oven temperatures

Temperature & set point Digital display

Temperature sensor RTD

Front panel mounted Pressure regulator, pressure guage, temperature controller and oven inlet

Glass Oven Size 3.20 cm inside chamber diameter, 13.4 cm usable length, 107 cubic cm volume

Teflon Coated Oven Size 3.09 cm inside chamber diam, 12.7 cm usable length, 123.6 cubic cm volume

Operating Range 2 deg. C above ambient to 60 deg. C, alternate ranges available upon request

Other Features Quiet, long life Diaphragm pump

Air inlet filter

Moisture removal system

Unit Operating Temperature:

5 deg. C to 40 deg. C

Warm Up: Less than 30 minutes

Electrical: 98-132 VAC (Optional 196-264 VAC), 150 VA, 50/60 Hz

Unit Size 5.25” (13.3 cm) x 19” (48.3 cm) W x 15” (38.1cm) D

Weight 18 lbs (8.2 kg)

Page 14: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 26, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Gas Dilution Calibrator4010L

Overview

Users can easily perform gas dilution, ozone, and GPT calibration with Sabio Instrument’s intelligent Model 4010 Gas Dilution Calibrator. The Model 4010 was designed to set a new standard of quality and performance for gas dilution calibrators. Enhanced with the latest technology, Sabio’s Model 4010 calibration system is an uncomplicated and effective way to precisely control the dilution of gas standards for calibration of ambient air and source monitoring analyzers.

The Model 4010 calibrator uses embedded microprocessor technology to enhance the accuracy and control of the featured calibration system. The unit can be operated manually, automatically, or semi-automatically by remote control to conduct calibrations. The calibrator has the capability to produce and store in memory 20 calibration sequences, with up to 20 levels of source/dilution in each sequence. Users can employ the front panel membrane keypad, an external keyboard, or use the serial port to input calibration parameters.

The pneumatic system for the Model 4010 utilizes Mass Flow Controllers to precisely meter all gas streams and implements multiple types of curve correction algorithms to linearize calibrations.

The Model 4010 design exceeds the United States EPA calibration method requirements. Dilution components are calibrated with standards and test equipment traceable to the National Institute of Standards and Technology (NIST).

Standard Features

Pnuematics

Diluent Mass Flow Controller, 0-10 SLM

Source Mass Flow Controller, 0-100 SCCM

4 Calibration Gas Input Ports and 1 Purge Port

5 Output Ports and 1 Vent Port

1 Diluent Gas Input Port

Electronics

Bright Active Matrix Color Display, 25 line x 80-character

55-Button Membrane Keypad

24 Bits Digital Input/Output

2 Serial Ports for Data Communications

External PC Keyboard Input Port

Centronics Parallel Printer Port

Air Source Control Port

Electrical

Standard 98-264 VAC, 150-300 VA, 50/60 Hz Operation

Optional Features

Alternate Diluent Mass Flow Controller, Ranges 0-20 SLM

Alternate Source Mass Flow Controller, Ranges 0-10,000 SCCM

Second Source Mass Flow Controller, Ranges 0-10,000 SCCM

Additional Calibration Gas Input Ports

Permeation Port

Additional Output Ports

Second Diluent Gas Input Port

Internal Ozone Generator

Ozone Generator UV Optical Servo Control Loop

Internal Ozone UV Absorption Photometer

Internal Permeation Oven

Master Output Solenoid

Internal Instrument Solenoid Driver, 6 Outputs, 24V @ 3W ea

Rack Mount Kit

Ethernet TCP/IP, RS-232

Page 15: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 26, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

September 26, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

SpecificationsSpecifications subject to change without notice

Dilutions System

Input Distribution Gases 1 Standard (2 Optional)

Input Source Gases 4 Standard, 1 Purge, Optional 2 Additional

Dilution Mass Flow Controller 0-10 SLPM, Optional 0-20 SLPM

Source Mass Flow Controller 0-100 SCCM, 0-10,000 SCCM

2nd Source Mass Flow Controller 0-10,000 SCCM

Flow Accuracy ≤ ± 1% Full Scale

Flow Repeatability ≤ ± 0.15% Full Scale

Linearity ≤ ± 0.5% Full Scale

Input Pressure 20-35 psig

Output Manifold 6 Outputs Standard, Optional 2 Additional

Optional Internal Ozone Generator

Output 0.05-1.5 ppm @ 5 SLPM

Accuracy ± 1% of Set Point or ± 2 ppb @ 5 SLPM

Nominal Flow 100 SCCM, ± 1 SCCM

UV Lamp Temp. 50° C, ± 0.1 ° C

Optional Internal UV Absorption Photometer

Standard Ranges 0-0.5, 1.0, 2.0 ppm

Linearity ± 1 ppb or ± 1% of full scale (greater of two)

Precision ± 1 ppb

Zero Drift < ± 1 ppb for 24 hrs or 30 days

Flow Rate 1 Liter (nominal)

Analog Outputs 3 Ranges 5V, 1V, 100 mV

Optional Internal Permeation Oven

Chamber Materials Glass or Teflon

Chamber Size 1 Perm Tube (Any Size)

Adjustable Temperature 5° C Above Ambient Temperature to 65 °C

Temperature Accuracy ≤ ± 0.1 ° C

Calibrator Interface

Operation manually (55-button keypad, keyboard, serial ports

Calibration Definitions 20 factory-defined calibration sequences (< 20 points each)

Calibration Types Gas Dilution, Ozone, GPT

Gas Definitions 2 Diluent Gases, 20 Source Gas

Auto Calibrations 20 timer driven cal routines that perform user-defined calibration sequences on a 7 day calendar of events

Digital Inputs 24 status I/O bits for calibrator functionsDigital Input Types contact closure or TTL logic

Digital Outputs 24 status output bits for monitoring calibrator functions

Communications RS232, Ethernet TCP/IP

Calibrator System

Operating Temperature 5 deg. C to 40 deg. C

Dimensions 8.75” (22.2 cm) H x 17” (43.2 cm) W x 20” (50.8 cm) D

Average Weight 40 lbs. (18.1 kg)

Input Voltage 98-264 VAC, 150-300 VA, 50/60 Hz operation

Gas Dilution Calibrator4010L

Page 16: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

December 2, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Ozone Transfer Standard Model 2030 Ozone Photometer

Overview

The Model 2030 is a portable ozone photometer transfer standard for calibrating, certifying or auditing ozone analyzers and ozone generators.

Using Beer-Lambert’s Law, the Model 2030 measures the amount of ultraviolet (UV) light attenuated at 254 nm across the sample path for ozone.

The acquired sample is divided into two sample streams: a reference gas and a sample gas. Real-time comparison of the UV light intensity of the sample gas to the reference gas yields the precise concentration of ozone.

Advanced, easy to use, menu-driven software allows access to sample conditions, diagnostics and measured ozone concentrations. Users can operate the Model 2030 from the keypad or through the front panel buttons for quick access.

An optional internal ozone generator is available for generating ozone concentrations between 0.05- 1.5 PPM.

Standard Features

Microprocessor Controlled

33-Button Membrane Keypad

8 Line x 40 Character, 256 x 64 Pixel Graphics Vacuum Fluorescent Display

Optional Features

Internal Ozone Generator

Tilt-up Front Feet

Portable Side Handles

High-Impact Transport Case

Internal Zero Air Source

SpecificationsSpecifications subject to change without notice

Ranges 0-100, 0-200, 0-500 ppb or 0-1, 0-2, ppm

Zero Noise 0.6 ppb RMS 0.6 ppb RMS

Lower Detectable Limit 1 ppb

Zero Drift < 1.0 ppb/24 hr, < 2.0 ppb/7 days

Span Drift < 1% per month

Response Time 20 seconds, 10 seconds lag time

Precision 1 ppb

Linearity +/- 1% Full Scale

Sample Flow Rate 1 Liter/Minute

Operating Temperature 5 – 45 Deg. C

Power Requirements Universal Power Supply, 98-264 VAC, 100 VA, 50/60 Hz

Analog Outputs 0.1V, 1V, 2V, 5V, 10V, Standard DB-9 Connector

RS-232/RS-485 Port Standard DB-9 Connector

Physical Dimensions 6.2 in. (15.6 cm) H x 14.3 in. (36.4 cm) x 12.4 in. (31.6 cm) D

Average Weight 20 lbs. (9.1 kg)

Page 17: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

December 2, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Gas Dilution CalibratorModel 2010D

OverviewThe Model 2010D represents the latest generation of calibrator technology for ambient air monitoring, continuous emission monitoring (CEM), fugitive emissions, point source, odor and process monitoring.

Gas dilution, ozone, or gas phase titrations (GPT’s) are easily accomplished using this small, portable offering from Sabio.

The Model 2010D calibrator uses embedded microprocessor technology to precisely deliver and control gas concentrations along with multiple curve fits to linearize the desired output for the intended purpose. The unit can be operated manually or automatically to conduct calibrations, audits or certifications that exceed U.S. Environmental Protection Agency (EPA) and European Standard (EN) method requirements.

The Model 2010D can store 20 calibration sequences, with up to 20 concentration levels in each sequence. Menu driven software allows users access from the front panel keypad, external keyboard, or serial port for operation, feedback and real-time diagnostics.

Pneumatics for the Model 2010D utilize Mass Flow Controllers (MFC’s) to accurately meter all gas streams for reliable and repeatable gas concentrations.

Standard Features

Pnuematics

Diluent Mass Flow Controller, 0-10 SLM

Source Mass Flow Controller, 0-100 SCCM

4 Calibration Gas Input Ports and 1 Purge Port

5 Output Ports and 1 Vent Port

1 Diluent Gas Input Port

Electronics

Bright Active Matrix Color Display

42-Button Membrane Keypad

24 Bits Digital Input/Output

2 Serial Ports for Data Communications

External PC Keyboard Input Port

Parallel Printer Port

Air Source Control Port

Electrical

Standard 98-264 VAC, 150-300 VA, 50/60 Hz Operation

Optional Features

Alternate Diluent Mass Flow Controller, Ranges 0-20 SLM

Alternate Source Mass Flow Controller, Ranges 0-10,000 SCCM

Second Source Mass Flow Controller, Ranges 0-10,000 SCCM

Additional Calibration Gas Input Ports

Permeation Port

Additional Output Ports

Second Diluent Gas Input Port

Internal Ozone Generator

Ozone Generator UV Optical Servo Control Loop

Internal Ozone UV Absorption Photometer

Permeation Oven

Master Output Solenoid

Instrument Solenoid Driver, 6 Outputs, 24V @ 3W ea

Rack Mount Kit

Ethernet TCP/IP, RS-232, RS-485 and USB

Page 18: Air Quality Monitoring, Solutions, Reporting & Warning · Air Quality Monitoring, Solutions, Reporting & Warning ... quality data, is built on ... EMRS Air & Water Quality Alert Systems

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

December 2, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

Sutron Corporation22400 Davis Drive Sterling, VA 20164 703.406.2800 www.sutron.com [email protected]

Air Quality Division2111 Sam Bass Road Round Rock, TX 78681 512.238.9359 www.sutron.com/[email protected]

December 2, 2014

Air Quality Division2548 Shell Road Georgetown, TX 28628 512.869.0544 www.sutron.com/aq.htm [email protected]

SpecificationsSpecifications subject to change without notice

Dilution System

Input Dilution Gases 1 Standard 1 Additional Option

Input Source Gases 4 Standard, 1 Purge2 Additional Option

Dilution Mass Flow Controller

0-10 SLPMOptional Range 0-20 SLPM

Source Mass Flow Controller:

0-100 SCCMOptional Ranges 0-10,000 SCCM

2nd Source Mass Flow Controller 0-10,000 SCCM

Flow Accuracy ≤ ± 1% Full Scale

Flow Repeatability ≤ ± 0.15% Full Scale

Linearity ≤ ± 0.5% Full Scale

Input Pressure 20-35 psig

Response Time at Output < 1 minute

Optional Internal Ozone Generator

Output 0.05-1.5 ppm @ 5 SLPM

Accuracy ± 1% of Set Point or ± 2 ppb @ 5 SLPM

Nominal Flow 100 SCCM, ± 1 SCCM

UV Lamp Temperature 50° C, ± 0.1 ° C

Optional Internal UV Absorption Photometer

Standard Ranges 0-0.5, 1.0, 2.0 ppm

Linearity ± 1 ppb or ± 1% of full scale (greater of two)

Precision ± 1 ppb

Zero Drift < ± 1 ppb for 24 hrs or 30 days

Flow Rate 1 Liter (nominal)

Analog Outputs 3 Ranges 5V, 1V, 100 mV

Optional Internal Permeation Oven

Chamber Materials Glass or Teflon

Chamber Size 1 Perm Tube (Any Size)

Adjustable Temperature 5° C Above Ambient Temperature to 65 °C

Temperature Accuracy ≤ ± 0.1 ° C

Calibrator Interface

Operation 42-button keypad, keyboard or serial ports

Calibration Definitions 20 factory-defined calibration sequences (< 20 points each)

Calibration Types Gas Dilution, Ozone, GPT (Gas Phase Titration)

Gas Definitions 2 Diluent Gases, 20 Source Gas

Auto Calibrations 20 timer driven cal routines that perform user-defined calibration sequences on a daily schedule

Digital Inputs 24 status I/O bits for calibrator functionsDigital Input Types Contact closure or TTL logicDigital Outputs 24 Status I/O bits

Communications Ethernet TCP/IP, RS-232, RS-485 & USB

Physical & Operational Specifications

Operating Temperature 5°C to 40°C

Dimensions 5.25” (13.3 cm) H x 16.2” (41.2 cm) W x 19.25” (48.9 cm) D

Average Weight 24 lbs. (10.9 kg)

Input Voltage 98-264 VAC, 150-300 VA, 50/60 Hz operation

Gas Dilution Calibrator2010D