Page 1: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

ACSM Update

Phil Croteau

August 24, 2019 AMS Users Meeting – Lund 1

Page 2: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Aerosol Chemical Speciation Monitor (ACSM)

Thermal Vaporization &

Electron Impact

Ionization

Aerodynamic lens

Particle Beam Generation

Particle Composition

Data AcquisitionComputer

TOF

or

Q-M

S

Turbo Pumps

AerosolSamplingIn from

cyclone

Out to sampling pump

Filter

3-way Valve

Ng, et al., Aerosol Sci. Technol., 45:7, 770-784, 2011

AMS Users Meeting – Lund August 24, 2019 2

No Size

Page 3: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Quadrupole, Q-ACSM Time-of-Flight, ToF-ACSM

AMS Users Meeting – Lund August 24, 2019 3

Page 4: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Von der Weiden et al., Atmos. Meas. Tech., 2, 479-494, 2009.

3/24/2019 CIAAS 2019 4

Aerosol

Sampling

Particle Beam

Generation

Particle

Composition

PM2.5 Capable ACSM Development and Lab EvaluationXu et al., Aerosol Sci. Technol., 51:1, 69-83, 2017 DOI:10.1080/02786826.2016.1241859

Page 5: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

PM2.5 ACSM Inlet Hardware

August 24, 2019 AMS Users Meeting – Lund 5

Page 6: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

ACSM Calibration Project Overall goal – Simplify and decrease cost of ACSM calibration

August 24, 2019 AMS Users Meeting – Lund 6

ACSMDMA

CPC

Atomizer

Aerosol

Diluter

By-pass

filter

Drier

Mixing Tube

Very Important!!

Page 7: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Results – Calibration

• Evaluated several approaches to calibration – landed on ultrasonic nebulization of NH4NO3 and integrating OPC measured size distribution to calculate mass for further development

August 24, 2019 AMS Users Meeting – Lund 7

Page 8: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Results – CalibrationStandard ACSM CalibrationUN/OPC ACSM Calibration

August 24, 2019 AMS Users Meeting – Lund 8

Page 9: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Since last time – Maybe we can do what we’ve always done but with lower cost components

August 24, 2019 AMS Users Meeting – Lund 9

Low resolution radial

ACSMDMA

CPC

Atomizer

Aerosol

Diluter

By-pass

filter

Drier

Mixing Tube

Small, water-based

3D Printed

Page 10: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

3D Printed atomizer – collaboration with Daniel Cziczoand Michael Rösch (MIT)

August 24, 2019 AMS Users Meeting – Lund 10

Page 11: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Particle size selection – Spider DMAStavros Amanatidis (CalTech) and Susanne Hering (ADI)

August 24, 2019 AMS Users Meeting – Lund 11

Page 12: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Particle Counting – MAGIC CPC

August 24, 2019 AMS Users Meeting – Lund 12

Page 13: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Testing MAGIC as aerosol source and Spider as counter

AtomizerBipolarIons (Kr)

TSI Long DMA +3776 CPC

R=10, 3.9/0.39

TSI Long DMA +3776 CPC

R=4.4, 3.9/0.89

Magic SpiderR=4.1, 1.1/.26

AMS

300 nm at 5 concentrations 300 nm, no ions189 nm at 1 concentration (Spider only)122 nm “

81 nm “55 nm “August 24, 2019 AMS Users Meeting – Lund 13

Page 14: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

AMS measured PTOF size distributions of Dmob = 300nm particles from TSI and Spider DMA. Roughly normalized.

August 24, 2019 AMS Users Meeting – Lund 14

Page 15: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

ACSM Calibration comparing Spider DMA and TSI DMA

August 24, 2019 AMS Users Meeting – Lund 15

Page 16: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Vertically mounted ACSM – Wade Robinson

• 19” Rack mounted design• Standardize PM2.5 inlet sampling

plumbing• Better for integration into

monitoring stations

August 24, 2019 AMS Users Meeting – Lund 16

Page 17: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

Q-ACSM – Pfeiffer will stop producing PrismaPlus and move to new PrismaPro

• Expect similar performance

• More analog and digital I/O

• Communication standard different• Requires significant rewrite

of ACSM DAQ

August 24, 2019 AMS Users Meeting – Lund 17

Page 18: ACSM Update - Cooperative Institute for Research in ...cires.colorado.edu/.../CroteauACSMUpdate2019.pdfTOF-ACSM X –almost exists •TOF-ACSM with a higher resolution mass spectrometer

TOF-ACSM X – almost exists

• TOF-ACSM with a higher resolution mass spectrometer (res ~>2000)• Same body as HTOF but with same dynode detector used in TOF-ACSM

August 24, 2019 AMS Users Meeting – Lund 18