1.1 preface - shodhgangashodhganga.inflibnet.ac.in/bitstream/10603/31128/8/08_preface.pdf ·...
TRANSCRIPT
1
1.1 PREFACE
The number of new drugs being designed and introduced for therapy is
constantly increasing. Consequently, the dosage forms that include these drugs are
introduced into the market in huge numbers. So, there is always a necessity for
developing newer and efficient methods for determining these drugs in bulk samples
and formulations. The introduction of large number of newer drugs and their
formulations may also lead to widespread distribution of substandard or even
counterfeit drugs and their formulations in the market. Quality control and quality
assurance of pharmaceutical chemicals and their formulations are essential for
ensuring the availability of safe and effective drug formulations to the consumers and
safeguarding the general public against the hazards of substandard drugs.
Pharmaceutical analysis is indispensable in the process of quality control for statutory
certification of drugs and their formulations either by the industry or by the regulatory
authorities. Thus, constant development of new and improved analytical methods for
accurate determination of drugs in raw materials and in pharmaceutical dosage forms
is essential for quality control, pharmacokinetic, bioequivalence and toxicological
studies.
Pharmaceutical analysis deals with the analysis of not only the drugs but also
their formulations. It is also necessary to check the quality of the raw materials
including the bulk drugs that go into the formulation of the dosage forms. There are
several valid reasons for developing new analytical methods. The existing methods
may be erratic or unreliable i.e. having poor accuracy and precision. The existing
method may be time consuming or may be too expensive. The advent of new
techniques and improved instrumentation in the field of analysis may give way to
more sensitive, precise and accurate methods.
In order to develop a newer or improved analytical method, the analyst has to
set some goals. It is necessary to determine the analyte at trace levels accurately. The
method should be precise to the drug under study. The method should be simple
consuming minimum analysis time and using cheaper chemicals and materials. The
method should yield reproducible results, when carried out by different analysts and
in different laboratories. It should also be robust giving accurate results even there are
2
slight variations in the conditions of the method. It is not sufficient to develop and
optimize analytical methods using the pure standard drugs but, it is necessary that
these methods be validated appropriately and the methods should be applicable for
estimation of these drugs in their dosage forms.
High Performance Liquid Chromatography is the fastest growing analytical
technique for the analysis of drugs. Its simplicity and wide range of sensitivity and
short analysis time makes it ideal for analysis of many drugs in both dosage forms and
biological fluids. With the development of more sophisticated instrumentation,
efficient column materials1, 2 and moderate pricing, the HPLC technique has now
become more reliable and indispensable. In view of this the author has chosen to
develop HPLC methods for determination of some of the recent drugs.
The present study incorporated in the thesis was taken up by the author with
an aim to develop more efficient and validated new high performance liquid
chromatographic methods for estimation of fourteen important drugs namely
Eszopiclone, Lamotrigine, Dextromethorphan hydrobromide,Quinidine sulfate,
Fenofibrate, Pregabalin, Pramipexole dihydrochloride monohydrate,
Memantine, Donepazil hydrochloride, Levodopa, carbidopa, Entacapone in their
bulk samples as well as in dosage forms. The study design involves the development
of new reverse phase HPLC methods for estimation of the selected drugs either
individually or in combination with other drugs and validation of the methods thus
developed and testing their suitability for estimation of the drugs in their
pharmaceutical dosage forms. All the methods were carried out by adopting reverse
phase HPLC technique. The methods were validated as per ICH3 guidelines.
A literature survey on the analytical methods of Eszopiclone, Lamotrigine,
Dextromethorphan hydrobromide,Quinidine sulfate, Fenofibrate, Pregabalin,
Pramipexole dihydrochloride monohydrate, Memantine, Donepazil
hydrochloride, Levodopa, carbidopa, Entacapone revealed that a few HPLC
methods are available for their estimation in dosage forms in addition to other
techniques. Some of these methods have certain drawbacks like gradient elution
technique, long run time, less resolution and lack of sufficient sensitivity, precision
and accuracy. Furthermore, some methods lacked proper validation and
3
documentation. Hence, the author had attempted to develop simple, fast, accurate and
precise HPLC methods for determination of these drugs. The methods proposed by
the author are economical, quick and the solvents used in them are of moderate cost
and are thus easily affordable by the laboratories equipped with standard HPLC
systems. The proposed methods can be used as alternative methods to those reported
by the earlier workers and provide good choice for the routine determination of the
chosen drugs in their formulations and also in their clinical, pharmacokinetic and
biological studies.
The thesis incorporates the results of experimentation carried out by the author
for determination of the drugs listed above in pure form, validation of the method so
developed and applicability of the method for the estimation of the drugs in their
dosage forms by HPLC.
The thesis has been presented in two sections. Section 1 incorporates
introductory information about HPLC and its technique. This is followed by the
general guidelines and methodology to be followed for developing a new method for
estimation of drugs by HPLC. Later, the procedures adopted to determine various
parameters for validation of the developed method have been given.
Section 2 of the thesis deals with the details of the author’s experimentation
and results obtained in the HPLC method development for the assay of the following
fourteen selected drugs namely Eszopiclone, Lamotrigine, Dextromethorphan
hydrobromide,Quinidine sulfate, Fenofibrate, Pregabalin, Pramipexole
dihydrochloride monohydrate, Memantine, Donepazil hydrochloride, Levodopa,
carbidopa, Entacapone. The data in section 2 have been divided into nine chapters,
each chapter being devoted to one drug. The contents in each part have been
presented under the following heads.
1) Drug profile (s)
2) Review of the past work on the analytical methods
3) Experimental and results
4
a) Material and methods
b) Optimization of chromatographic conditions and method development
c) Validation of the proposed method
4) Summary of the results and Conclusion
5) References
The results obtained in these experiments have been thoroughly discussed.
The references cited in the text of the thesis have been given at the end of each part.
The following table shows the list of drugs taken up for the study and the source of
their monographs. The figures in the cells indicate the page number (*) / monograph
number (#).
S.No. Name of the
Drug *IP1 *USP2 *Martindale3
#Merck
Index4
#Drug
Bank5
1 Eszopiclone - - 1755 DB00402
2 Lamotrigine 1566 375 5368 DB00555
3 Dextrometharpha
n hydrobromide - - 1066 1392 DB00514
4 Quinidine sulfate 2016 - - DB00908
5 Fenofibrate 1337 3160 1307 679 DB1039
6 Pregabalin 1960 - - 1327 DB00230
7 - 4384 - 1755 DB00413
5
Name of the Drug *IP1 *USP2 *Martindale3 #Merck
Index4
#Drug
Bank5
Memantine - - 1165 1007 DB01043
Donepezil
hydrochloride 1249 2968 - 578 DB00843
Levodopa 1576 1636 1888 946 DB01235
Carbidopa - 1636 1888 291 DB00190
Entacapone - 3048 1159 613 DB00494
REFERENCES
1. Indian Pharmacopoiea 2010, Govt. of India, Ministry of Health and Family
Welfare. Delhi: Indian Pharmacopoeial Commission, Ghaziabad, 2010.
2. USP 35 NF 30, United States Phamacopoeia, The United States Pharmacopoeial
Convention, Rockville, MD, 2010.
3. Sweetman SC. Martindale, the Complete Drug Reference. 31st ed. London:
Pharmaceutical Press, 1996.
4. M.J. O’ Neil, The Merck Index: An encyclopedia of Chemicals, Drugs and
Biologicals, 13th Edn., The Merck and Co, NJ, 2001.
5. http://www.drugbank.ca/drugs
6
1.2. INTRODUCTION TO HPLC AND ITS TECHNIQUE
High-performance liquid chromatography1-2 (HPLC) is the fastest growing
analytical technique for analysis of drugs. Its simplicity, high sensitivity and
specificity make it ideal for the analysis of many drugs in pharmaceutical dosage
forms and biological fluids.
High Performance Liquid Chromatography is used to describe liquid
chromatography in which the liquid mobile phase is forced through the column at
high pressure and as a result the analysis time is reduced by 1-2 orders of the
magnitude relative to classical column chromatography. As it uses much smaller
particles of the adsorbent or stationary support it becomes possible for increasing the
column efficiency substantially. The importance of chromatography is increasing
rapidly in pharmaceutical analysis for exact differentiation, selective identification
and quantitative determination of structurally related compounds. Another important
field of application of chromatographic methods is the purity testing of final products
in a bulk drug industry. The reasons for the popularity of the method are its
sensitivity, its ready adaptability to accurate quantitative determinations, its suitability
for separating non-volatile and thermolabile species. Sensitive detectors have
transformed liquid column chromatography into high speed, efficient, accurate and
highly resolved method of separation.
The HPLC is the method of choice in the field of analytical chemistry, since
the methods developed using HPLC are specific, robust, linear, precise and accurate
and the limit of detection is low. HPLC technique offers the following advantages.
� Speed (many analyses can be accomplished within 20 min.)
� Greater sensitivity (various detectors can be employed)
� Improved resolution (wide variety of stationary phases can be used)
� Reusable columns (columns are expensive but can be used repeatedly)
� Easy sample handling and recovery.
� Automated instrumentation (less time and less labour)
7
� Precise and reproducible
� Computer monitored analysis and data interpretation.
PRINCIPLES OF SEPARATION
Adsorption chromatography employs high-surface area particles that
adsorb the solute molecules. Usually a polar solid such as silica gel, alumina or
porous glass beads and non-polar solvents such as heptane, octane or chloroform are
used as the stationary phase and mobile phase respectively in adsorption
chromatography. In adsorption chromatography, adsorption process is described by
competition model and solvent interaction model. Competition model assumes that
entire surface of the stationary phase is covered by mobile phase molecules as result
of competition for adsorption site. In solvent interaction model the retention results
from the interaction of solute molecule with the second layer of adsorbed mobile
phase molecules. The differences in affinity of solutes for the surface of the stationary
phase account for the separation achieved.
In partition chromatography, the solid support is coated with a liquid stationary
phase. The relative distribution of solutes between the stationary and the mobile phases
determines the separation. The stationary phase can either be polar or non polar. If the
stationary phase is polar and the mobile phase is non polar, it is called normal phase
partition chromatography. If the opposite case holds, it is called reversed-phase
partition chromatography. In normal phase mode, the polar molecule partition
preferentially in to the stationary phase and are retained longer than non-polar
compounds. In reverse phase partition chromatography, the opposite behavior is
observed.
8
INSTRUMENTATION
Schematic diagram of a typical HPLC unit
The individual components of HPLC and their functions are described below.
SOLVENT DELIVERY SYSTEM
The mobile phase is pumped under pressure from one or several reservoirs and
flows through the column at a constant rate. With micro particulate packing, there is a
high-pressure drop across a chromatography column. Eluting power of the mobile
phase is determined by its overall polarity, the polarity of the stationary phase and the
nature of the sample components. For normal phase separations, eluting power
increases with increasing polarity of the solvent but for reversed phase separations,
eluting power decreases with increasing solvent polarity. Optimum separating
conditions can be achieved by making use of mixture of two solvents. Some other
properties of the solvents, which need to be considered for a successful separation, are
the boiling point, viscosity, detector compatibility, flammability and toxicity.
The most important component of HPLC in solvent delivery system is the
pump, because its performance directly affects the retention time, reproducibility and
detector sensitivity. Among the several solvent delivery systems, (direct gas pressure,
pneumatic intensifier, reciprocating etc.) reciprocating pump with twin or triple
pistons is widely used, as this system gives less baseline noise, good flow rate and
reproducibility etc.
9
MOBILE PHASE
Typically, the mobile phases used for HPLC are mixtures of organic solvents
and water or aqueous buffers. Table given below lists the physical properties of
organic solvents commonly used for HPLC. Isocratic methods are preferable to
gradient methods. Gradient methods will sometimes be required when the molecules
being separated have vastly different partitioning properties. When a gradient elution
method is used, care must be taken to ensure that all solvents are miscible.
The following points should also be considered when choosing a mobile phase:
1. It is essential to establish that the drug is stable in the mobile phase for at least
in the duration of the analysis.
2. Excessive salt concentrations in the buffers should be avoided. High salt
concentrations can result in precipitation of the analytes, which can damage
the column.
3. The mobile phase should have a pH 2.5 and pH 7.0 to maximize the lifetime
of the column.
4. Cost and toxicity of the mobile phase can be reduced by using methanol
instead of acetonitrile where possible.
5. Absorbance of the buffer is to be minimized. Since trifluoroacetic acid, acetic
acid or formic acid absorb at shorter wavelengths, they may prevent detection
of products without chromophores above 220 nm. Carboxylic acid modifiers
can be frequently replaced by phosphoric acid, which does not absorb above
200 nm.
6. Volatile mobile phases should be used when possible, to facilitate collection of
products and LC-MS analysis. Volatile mobile phases include ammonium
acetate, ammonium phosphate, formic acid, acetic acid, and trifluoroacetic
acid. Some caution is needed as these buffers absorb below 220 nm.
10
PHYSICAL PROPERTIES OF COMMON HPLC SOLVENTS
Solvent MW BP RI
(25oC)
UV*
Cut-off
(nm)
Density
g/ml
(25oC)
Viscosity
cP
(25oC)
Dielectric
Constant
Acetonitrile 41.0 82 1.342 190 0.787 0.358 38.8
Dioxan 88.1 101 1.420 215 1.034 1.26 2.21
Ethanol 46.1 78 1.359 205 0.789 1.19 24.5
Ethyl acetate 88.1 77 1.372 256 0.901 0.450 6.02
Methanol 32.0 65 1.326 205 0.792 0.584 32.7
CH2Cl2 84.9 40 1.424 233 1.326 0.44 8.93
Isopropanol 60.1 82 1.375 205 0.785 2.39 19.9
n-propanol 60.1 97 1.383 205 0.804 2.20 20.3
THF 72.1 66 1.404 210 0.889 0.51 7.58
Water 18.0 100 1.333 170 0.998 1.00 78.5
* wavelength at which the absorbance of 1cm cell is 1.0
SOLVENT DEGASSING SYSTEM
The mobile phase should be degassed and filtered before use. Several methods
are employed to remove the dissolved gases in the mobile phase. They include heating
and stirring, vacuum degassing with an aspirator, filtration through 0.45 filter,
vacuum degassing with an air-soluble membrane, helium purging and ultra sonication
or combinations of these methods. HPLC systems are also provided an online
degassing system, which continuously removes the dissolved gases from the mobile
phase.
11
GRADIENT ELUTION DEVICES
HPLC column elution may be in the isocratic mode i.e., with a constant
composition of the mobile phase or in the gradient elution mode in which the mobile
phase composition varies during run. Gradient elution overcomes the problem of
analysis when dealing with a complex mixture of solutes.
COLUMNS
The heart of the system is the column. The choice of common packing
material and mobile phases depends on the physical properties of the drug. Many
different reverse phase columns will provide excellent specificity for any particular
separation. It is therefore preferable to routinely attempt separations with a standard
C8 or C18 column and determine if it provides good separations. If this column does
not provide good separation or the mobile phase is unsatisfactory, alternate methods
or columns should be explored. Reverse phase columns differ by the carbon chain
length, degree of end capping and percent carbon loading. Diol, cyano and amino
groups can also be used in the matrix for reverse phase chromatography.
SAMPLE INTRODUCTION SYSTEM
The two methods of analyte introduction on the column are by injection into a
flowing stream and by stop flow injection. These techniques can be used with a
syringe or an injection valve. Automatic injector is a microprocessor-controlled
version of the manual universal injector. Usually, up to 100 samples can be loaded in
to the auto injector tray. The system parameters such as flow rates, gradient, run time,
volume to be injected, etc. are chosen, stored in memory and sequentially executed on
consecutive injections.
LIQUID CHROMATOGRAPHIC DETECTORS
The function of the detector in HPLC is to monitor the mobile phase as it
emerges from the column. Generally, there are two types of HPLC detectors, bulk
property detectors and solute property detectors.
12
a. Bulk property detectors: These detectors are based on differential measurement of
a property, which is common to both the sample and the mobile phase. Examples
of such detectors are refractive index, conductivity and dielectric constant
detectors.
b. Solute property detectors: Solute property detectors respond to a physical property
of the solute, which is not exhibited by the pure mobile phase. These detectors
measure a property, which is specific to the sample, either with or without the
removal of the mobile phase before the detection. Solute property detectors which
do not require the removal of the mobile phase before detection include
spectrophotometric (UV or UV-Visible) detector, fluorescence detectors,
polarographic, electro-chemical and radio activity detectors, where flame
ionization detector and electron capture detector both require removal of the
mobile phase before detection.
UV-Visible and fluorescent detectors are suitable for gradient elution, because
many solvents used in HPLC do not absorb to any significant extent.
Detectors
Optical detectors are most frequently used.
These detectors pass a beam of light through the flowing column effluent as it passes
through a low volume (~ 10 ml) flow cell.
The most commonly used detector in LC is the ultraviolet absorption detector.
A variable wavelength detector of this type, capable of monitoring from 190 to 460-
600 nm, will be found suitable for the detection of the majority samples.
Other types of detectors:
Refractive Index detector: Universal analyte detector. Solvent must remain the same
throughout separation. It is temperature sensitive.
Fluorescence detector: Excitation wavelength generates fluorescence emission at a
higher wavelength. Analytes must have fluorophore group. Very sensitive and
selective. Results vary depending upon separation conditions.
Mass Spectroscopic detectors: Mass to charge ratio (m/z). Allow identification of
specific compounds. Several types of ionization techniques include: electro-spray,
atmospheric pressure chemical ionization, electron impact. The detector usually
13
contains low volume cell through which the mobile phase passes carrying the sample
components.
INJECTORS
Sample introduction can be accomplished in various ways. The simplest
method is to use an injection valve. In more sophisticated LC systems, automatic
sampling devices are incorporated where sample introduction is done with the help of
auto samplers and microprocessors.
In liquid chromatography, liquid samples may be injected directly and solid
samples need only be dissolved in an appropriate solvent. The solvent need not be the
mobile phase, but frequently it is judiciously chosen to avoid detector interference,
column/component interference, and loss in efficiency or all of these. It is always best
to remove particles from the sample by filtering, or centrifuging since continuous
injections of particulate material will eventually cause blockage of injection devices
or columns.
DATA SYSTEMS
The main goal in using electronic data systems is to increase the accuracy and
precision of the analysis, while reducing operator attention. In routine analysis, where
no automation (in terms of data management or process control) is needed, a pre-
programmed computing integrator may be sufficient. For higher control levels, a more
intelligent device is necessary, such as a data station or minicomputer.
The advantages of intelligent processors in chromatographs:
� additional automation options become easier to implement;
� complex data analysis becomes more feasible;
� Software safeguards can be designed to reduce accidental misuse of the
system.
REFERENCES
1. Josefsson M, Zackrisson; A.L. Norlander B, J of Chromatography B, 1995, 672,
310-313.
2. Stopher D.A. Beresford A.P. Macrae, P.V. Humphrey M.J., of Cardiovasc.
Pharmacol, 1988, 12, Supp 7, S55-S59.
14
1.3. HPLC METHOD DEVELOPMENT AND OPTIMIZATION
The development of any new or improved method for the analysis of an analyte
usually depends on tailoring the existing analytical approaches and instrumentation.
Method development usually involves selecting the method requirements and on the
type of instrumentation. In the development stage of a HPLC method, decision
regarding the choice of column, mobile phase, detector and method of quantitation
must be addressed.
Once the instrumentation has been selected, it is important to determine the
chromatographic parameters for the analyte of interest. It is necessary to consider the
properties of the analyte(s) that may be useful to select the nature of the column to be
used, establish the approximate composition and pH of the mobile phase for
separation of the components wave length to be employed or mass/charge ratio to be
scanned at for detection of the compound, the concentration range to be followed and
choice of a suitable internal standard for quantification purpose etc. Such information
may be already available in the literature for the analyte or related compounds.
This is followed by optimization and preliminary evaluation of the method.
Optimization criteria must be determined with cognizance of the goals common to any
new method. Initial analytical parameters of merit like sensitivity (measured as
response per amount injected), limit of detection, limit of quantitation and linearity of
calibration plots are to be determined. As a precautionary measure, it is important that
method development be performed using only the analytical standards that are highly
pure and have been well identified and characterized and whose purity is known.
During the optimization stage, the initial sets of conditions that have evolved
from the first stages of development are improved or optimised in terms of resolution,
peak shape, plate counts, peak asymmetry, capacity, elution time, detection limits,
limit of quantitation, and overall ability to quantify the specific analyte of interest.
Results obtained during optimization must be evaluated against the goals of the
analysis set forth by the analytical figures of merit. This evaluation reveals if
additional improvement and optimization are needed to meet the initial method
requirements.
15
Optimization of the method should yield maximum sensitivity, good peak
symmetry, minimum detection and quantitation levels, a wide linearity range, and a
high degree of accuracy and precision. Other potential optimization goals include
baseline resolution of the analyte of interest from other sample components, unique
peak identification, on-line demonstration of purity and interfacing of computerized
data for routine sample analysis. Absolute quantitation should use simplified methods
that require minimal sample handling and analysis time.
Optimization of the method may follow either manual or computer driven
approaches. The manual approach involves varying one experimental condition at a
time, while holding all others constant and evaluating the changes in response. The
variables might include flow rate, mobile or stationary phase composition,
temperature, detection wavelength and pH. This univariate approach of system
optimization is usually time consuming and expensive. However, it may provide a
much better understanding of the principle involved and of the interactions of the
variables. In computer-driven automated method development efficiency is optimized
while experimental input is minimized. This approach can be applied to many types of
methods. It significantly reduces the time of analysis, energy, and cost of analysis.
SYSTEMATIC APPROACH TO THE REVERSE PHASE CHROMATOGRAPHIC
SEPARATION OF PHARMACEUTICAL COMPOUNDS
Classifying the sample
The first step in the method development is to characterize the drug whether it
is regular or special. The regular compounds are those that are neutral or ionic. The
inorganic ions, bio-molecules, carbohydrates, isomers, enantiomers and synthetic
polymers etc are called special compounds. The selection of initial conditions for
regular compounds depends on the sample type. The general approach for the reverse
phase chromatographic method development is based on the following considerations.
The regular samples like pharmaceuticals (either ionic or neutral) respond in
predicable fashion to changes in solvent strength (%B) and type (e.g. acetonitrile or
methanol) or temperature. A 10% decrease in %B increases retention by about three
fold and selectivity usually changes as either %B or solvent type is varied. An
increase in temperature causes a decrease in retention as well as changes in
16
selectivity. It is possible to separate many regular samples just by varying solvent
strength and type. Alternatively, varying solvent strength and temperature can
separate many ionic samples and some non-ionic samples.
The choice of the initial column, mobile phase and temperature is quite
important. The initial conditions for RP HPLC method developed are given in Table.
Separation variable Preferred initial choice
Column Packing C8 or C18 column, less acidic silica columns; if
temperatures > 50 C are planned, more stable,
sterically protected packings are preferred
Column configuration 15 X 0.46 cm column; 5µm particles
Flow rate 2.0 mL/min
Mobile phase Acetonitrile-water (neutral samples) or
acetonitrile-buffer (ionic samples); buffer is 25-
50 mM potassium phosphate at pH 2-3 (lower
pH preferrable if column is stable).
Temperatue 35 or 400 C
Sample size < 50 µL; 50-100 g
The Column and Flow rate
To avoid problems from irreproducible sample retention during method
development, it is important that columns be stable and reproducible. A C8 or C18
column made from specially purified less acidic silica and designed specifically for
the separation of basic compounds is generally suitable for all samples and is strongly
recommended. If temperatures >50 0 C are used at low pH, stearically protected
bonded-phase column packing is preferred. The column should provide reasonable
resolution in initial experiments, short run times and an acceptable pressure drop for
different mobile phases. A 5µ, 150 X 4.6 mm column with a flow rate of 2 mL/min is
good for different mobile phases as initial choice. These conditions provide
reasonable plate number (N=8000), a run time of < 15 min for a capacity factor k < 20
and a maximum pressure drop < 2500 psi for any mobile phase made from mixtures
of water, acetonitrile and/or methanol.
17
The mobile phase
The preferred organic solvent (B) for the mobile phase mixture is acetonitrile
(ACN) because of its favorable UV transmittance and low viscosity. However,
Methanol (MeOH) is a reasonable alternative. Amine modifiers like tetra hydro furan
(THF) are less desirable because they may require longer column equilibration times,
which can be a problem in method development and routine use of the method. They
may occasionally introduce additional problems like erratic base line and poor peak
shape. However, some samples may require the use of amine modifiers when poor
peak shapes or low plate number are encountered.
The pH of the mobile phase should be selected with two important
considerations. A low pH that protonates column silanols and reduces their
chromatographic activity is generally preferred. A low pH (<3) is usually quite
different from the pKa values of common acidic and basic functional groups.
Therefore, at low pH the retention of these compounds will not be affected by small
changes in pH and the reverse phase liquid chromatographic method will be more
rugged. For columns that are stable at low pH, a pH of 2 to 2.5 is recommended. For
less stable columns, a pH of 3.0 is a better choice.
Separation temperature
Mostly the temperature controllers operate best above ambient (>300 C).
Higher temperature operation also gives lower operating pressures and higher plate
numbers, because of decrease in mobile phase viscosity. A temperature of 35-400 is
usually a good starting point.
Sample size
Initially, a 25-50 µL injection (25-50 µg) can be used for maximum detection
sensitivity. Smaller injection volumes are required for column diameters of below 4.5
mm and /or particles smaller than 5 µm. The sample should be dissolved initially in
water (1mg/mL) or dilute solution of acetonitrile in water. For the final method
development stage, the best sample solvent is the mobile phase. The samples which
cannot be dissolved in water or the mobile phase should be dissolved initially in either
acetonitrile or methanol and then diluted with water or mobile phase before injection.
18
Equilibration of the column with the mobile phase
The analytical column is completely equilibrated with the mobile phase before
injecting the sample for analysis and retention data are collected for interpretation.
This is done for ensuring accurate retention data. Equilibration is required whenever
the column, mobiles phase or temperature is changed during method development;
usually by flow rate at least 10 column volumes of the new mobile phase before the
first injection. Some mobile phases may require a much longer column equilibration
time (e.g. mobile phases that contain THF amine modifiers such tri ethylamine and
tetra butyl amine and any ion pair reagent).
Column equilibration and reproducible data can be confirmed by first washing
the column with at least 10 columns volumes of the new mobile phase and injecting
the sample and then a second washing with at least 5 column volumes of the new
mobile phase and reinjection of the sample. If the column is equilibrated, the retention
times should not change by more than 0.02 min between the two runs.
Column Performance
The following values are used to assess overall system performance.
1. Relative retention
2. Theoretical plates
3. Capacity factor
4. Resolution
5. Peak tailing factor
6. Plates per meter
The chromatographic peak shape and plate number are calculated to assess
column performance. The asymmetry factor AS should fall between 0.9-10.5 and
number of theoretical plates should be >4000 for a 15 cm; 5 µm column at a flow rate
of 2 mL/min. The number of theoretical plates for well packed HPLC columns under
optimized test conditions is given in the Table.
19
Particle
Diameter (µm)
Column
length (cm) Plate Number N
10 15 6000-7000
10 25 8000-10000
5 10 7000-9000
5 15 10000-12000
5 25 17000-20000
3 5 6000-7000
3 7.5 9000-11000
3 10 12000-14000
3 15 17000-20000
Evaluating peak shape and plate number
The requirements for a given separation usually determine the type and
configuration of the column to be used. There are different suppliers for a given type
of column. These columns vary generally in performance. Therefore, certain
information concerning column specifications and performance is needed for use in
method development and their routine performance.
The column plate number (N) is an important characteristic of a column. N
signifies the ability of the column to produce sharp, narrow peaks for achieving good
resolution of band pairs with small α values. The table 2.2 shows the typical plate
numbers (small, neutral sample molecules) for well packed HPLC columns of various
lengths and particle sizes. A 15 or 25 cm column of 5 µ particles are preferred as a
starting point for method development. This configuration provides a large enough N
value for most separations and such columns are quite reliable. A column which gives
large N value can easily recognize closely over lapping peaks. Short columns of 3 µ
particles are useful for carrying out very fast separation (< 5 min). But these columns
are less used because they are more susceptible to sampling problems, more operator
dependent and more affected by band-broadening.
20
Peak asymmetry and Peak tailing
Columns and experimental conditions that provide symmetrical peaks always
preferred. Peaks with poor symmetry can result in inaccurate plate number and
resolution measurement, imprecise quantitaion, degraded resolution and poor
retention reproducibility.
Peak shape is measured in terms of peak asymmetry factor (AS) and peak
tailing factor (PTF). Peak asymmetry factor (AS) is measured at 10% of full peak
height. Good columns produce with AS values of 0.95 to 1.1. For accurate
measurement of symmetry bands should be measured with a magnified time scale.
Asymmetrical bands often appear symmetrical when observed in a compressed
chromatogram. Calculation of peak asymmetry factor and peak tailing factor are
represented is Fig.2.1. As per U.S.P., the peak asymmetry factor is calculated at 5% of
full height. Peak asymmetry and peak tailing factor are easily inter convertible as
shown below.
Retention
The time between the sample injection point and the analyte reaching the
detector is called the retention time tR.
21
Capacity factor k’
It measures how many times the analyte is retained to an unretained component.
Capacity factor, k’ = tR-t0/t0
Where tR = retention time of the peak
t0 = void time,
A k’ value zero means that the compound is not retained and elutes with the
solvent front. A k’ value of 1 means that the component is slightly retained by the
column while k’ value of 20 means that component is highly retained and spends
much time interacting with the stationary phase.
Selectivity α
Separation between tow components is only possible if they have different
migration rates through column. Selectivity or separation factor is a measure of
differential retention of two analytes It is defined as the ratio of the capacity factors
(k’) of two peaks.
α = k1’/k2’
Column efficiency (N) or number of theoretical plates
The term plate number N is a quantitative measure of the efficiency of the
column and is related to the ratio of the retention time and the standard deviation of
the peak width σ. Since ti is difficult to measure σ or Wb (width at base of the peak), a
relationship using width at half height or w1/2 is often used to calculate N as described
in the USP.
N = 5.54 (tR/ w1/2)2
Height equivalent of a Theoretical Plate (HETP) or Plate Height (H)
HETP = L/N
L= length of the column
N = number of theoretical plates
22
Resolution RS
It is the degree of separation of two adjacent peaks and is defined as the
difference in retention times of the two peaks divided by the average peak width. As
the peak width of adjacent peaks tends to be similar, the average peak width can be
equal to the width of one of the two peaks.
Resolution RS = tR1- TR2/ Wb1+wb2
Where,
Wb2 = Width of the base of component peak 2.
Wb1 = Width of the base of component peak 1.
Tailing factor T f
It is a measure of peak asymmetry. In this calculation peak width at 5% peak
height W0.05 is used.
T f = W0.05/2f
Tailing factors for most peaks should fall between 0.9 and 1.4 with a value of
1.0 indicating a perfectly symmetrical peak.
The following table furnishes the formulae for calculating the different system
performance parameters (Note: Where the terms W and t both appear in the same
equation they must be expressed in the same units).
Relative retention (Selectivity): α = ( t2 - ta ) / ( t1 - ta )
Capacity factor k' = ( t2 / ta ) - 1
Tailing factor T = W0.05 / 2f
Resolution R = 2 ( t2 - t1 ) / ( W2 + W1 )
Theoretical plates: n = 16 ( t / W )2
Plates per meter: N = n / L
HETP L/n
23
where
α = Relative retention.
t2 = Retention time of the second peak measured from point of
injection.
t1 = Retention time of the first peak measured from point of injection.
ta = Retention time of an inert peak not retained by the column
measured from point of injection.
n = Theoretical plates.
t = Retention time of the component.
W = Width of the base of the component peak using tangent method.
k' = Capacity factor.
R = Resolution between a peak of interest (peak 2) and the peak
preceding it (peak 1).
W2 = Width of the base of component peak 2.
W1 = Width of the base of component peak 1.
T = Tailing factor.
W0.05 = Distance from the leading edge to the tailing edge of the peak,
measured at a point 5 % of the peak height from the baseline.
f = Distance from the peak maximum to the leading edge of the peak.
N = Plates per meter.
L = Column length, in meters.
24
1.4. METHOD VALIDATION
Introduction
Validation is the process of ensuring that a test procedure performs within
acceptable standards of reliability, accuracy and precision for its intended purpose. It
is the act of confirming that a method does what it is intended to do.
It is difficult to completely separate method development and optimization
from validation; these areas often overlap. In the validation stage, an attempt should
be made to demonstrate that the method works with samples of the given analyte, at
the expected concentration in the matrix, with a high degree of accuracy and precision.
Complete method validation can occur only after the method is developed and
optimized. In validation studies, suitability of the final method for the given analyte
and a select sample matrix is demonstrated, using specified instrumentation, samples,
and data handling; ultimately, the method can be transferred from one laboratory to
another that is suitably equipped and staffed. A method that provides all or most of the
original method requirements is deemed optimized and becomes ready for validation.
There is no single validation approach that must always be employed for a
new method; the analyst’s primary concern should be to select an approach that will
prove to be a true validation. Acceptance of any new method by others in the field will
depend on the specific validation approaches used. It is the responsibility of the
individual analyst to select the correct validation method(s). Validation approaches
include the zero-, single-, and double-blind spiking methods; inter-laboratory
collaborative studies; and comparison with a currently accepted (compendium)
method.
The Zero-Blind Method
The zero-blind approach involves a single analyst using the method with
samples at known levels of analyte to demonstrate recovery, accuracy, and precision.
The method is subject to analyst bias, and though the method is, in general, fast,
simple, and useful, it leads to subjective results and doubt on the part of the unbiased
reviewer or end user. However, as a first approximation and a demonstration of
validation potential requiring minimal time, manpower, samples, and cost, a zero-
25
blind study is a good place to start the overall validation process. Clearly, if this
approach fails to validate a method, then there is no reason to proceed with further
validation of the method.
The Single-Blind Method
The single-blind approach involves one analyst preparing samples at varying
levels unknown to a second analyst, who also analyzes the samples. The results are
then complied and compared by the first analyst. Although this approach is unbiased
at the start, it loses its blindness at the most crucial stage − when both sets of data are
compared. While perhaps more valuable and believable than the zero-blind approach,
the single-blind approach still invites bias on the part of the first analyst to bring two
sets of data into better agreement. This approach is appropriate at the very start of the
method validation, after the single-blind approach has proven successful, but before
one decides to involve additional analysts or management.
The Double-Blind Method
The double-blind approach involves three analysts. The first analyst prepares
samples at known levels, the second does the actual analysis, and the third analyst (or
administrator) compares both sets of data received separately from the first two
analysts. Neither the first or second analyst has access to the set of data generated by
the other. This double-blind approach is the most objective approach, assuming no
bias on the part of the third analyst.
The Analysis of Standard Reference Materials
The analysis of a standard reference material (SRM) or an authenticated
sample is a generally accepted method of validation. The USP, NIST, and other,
private organizations specialize in preparing, guaranteeing, and marketing standard
reference materials of various analyte species in different sample matrices. It may be
necessary, however, to contract the preparation of a unique sample in a particular
matrix in order to utilize this procedure for method validation. When using SRMs, the
analyst must demonstrate that the method provides accurate and precise measurements
of the analyte in a particular sample matrix. Analyst bias can also be an issue,
especially when the analyst knows the amounts and levels of the SRM.
26
The Inter-laboratory Collaborative Study
The inter-laboratory collaborative study is perhaps the most widely accepted
procedure to validate any new analytical method, but it suffers from serious practical
drawbacks. The collaborative approach is costly and time consuming; it can take years
from start to finish. During that time, the analysts may have to expend considerable
effort coordinating the process, shipping samples and receiving results, statistically
analyzing and interpreting the results, and then finally interpreting and verifying the
data. Although the approach is operator dependent (generating laboratory-to-
laboratory variability), when all laboratories involved come up with over-lapping
quantitative values in comparison with known levels present, the method is generally
accepted as full validation. This approach is rarely employed when a method is being
described for the first time in the literature.
Comparison with a Currently Accepted Method
Comparison with a currently accepted analytical method is yet another
validation approach. This is usually done by a singly analyst, but it can be done by
two analysts using a split sample. This approach uses results from the currently
accepted method as verification of the new method’s results. Agreement between
results initially suggests validation. Disagreement is a serious cause for concern of
future acceptability of the new method. However, disagreement could also suggest
that the currently accepted method is invalid, creating additional problems. If the
analyst can prove that the currently accepted method is indeed invalid, the analyst
must then initiate an alternative approach to validate the new method.
The question will eventually arise as to how many samples should be analyzed
in any validation approach. In general, the more the better, and the greater the variety
of samples and variation in the concentration range the better. Ideally, the method
should be validated for the analyte using several different sample types, with several
of each type determined separately for statistical and validation purposes. A single,
zero-blind or a single, single-blind study is obviously less meaningful and less
acceptable than an inter-laboratory collaborative, true double-blind study of several
sample matrices at widely different concentration levels. Initial validation approaches
are generally less rigorous and demanding than ones performed for standard reference
material (SRM) development.
27
Validation is an act of providing evidence that any procedures, process, equipments,
materials, activity or system perform as expected under given set of conditions and
also give the required accuracy, precision.
Assay procedures are intended to measure the analyte present in a given
sample. In the context of this document, the assay represents a quantitative
measurement of the major component(s) in the drug substance. For the drug product,
similar validation characteristics also apply when assaying for the active or other
selected component(s). The same validation characteristics may also apply to assays
associated with other analytical procedures (e.g., dissolution).
The objective of the analytical procedure should be clearly understood since
this will govern the validation characteristics which need to be evaluated. Typical
validation characteristics which should be considered are accuracy, precision,
(repeatability, intermediate precision), specificity, detection limit, quantitation limit,
linearity range, robustness and stability of analytical solution3.
28
METHOD VALIDATION PROCEDURES
1. SPECIFICITY
An investigation of specificity should be conducted during the validation of
identification tests, the determination of impurities and the assay. The procedures
used to demonstrate specificity will depend on the intended objective of the analytical
procedure.
It is not always possible to demonstrate that an analytical procedure is specific for a
particular analyte (complete discrimination). In this case a combination of two or
more analytical procedures is recommended to achieve the necessary level of
discrimination.
Identification
Suitable identification tests should be able to discriminate between compounds
of closely related structures which are likely to be present. The discrimination of a
procedure may be confirmed by obtaining positive results (perhaps by comparison
with a known reference material) from samples containing the analyte, coupled with
negative results from samples which do not contain the analyte. In addition, the
identification test may be applied to materials structurally similar to or closely related
to the analyte to confirm that a positive response is not obtained. The choice of such
potentially interfering materials should be based on sound scientific judgement with a
consideration of the interferences that could occur.
Assay and Impurity Test(s)
For chromatographic procedures, representative chromatograms should be
used to demonstrate specificity and individual components should be appropriately
labelled. Similar considerations should be given to other separation techniques.
Critical separations in chromatography should be investigated at an appropriate level.
For critical separations, specificity can be demonstrated by the resolution of the two
components which elute closest to each other.
In cases where a non-specific assay is used, other supporting analytical
procedures should be used to demonstrate overall specificity. For example, where a
29
titration is adopted to assay the drug substance for release, the combination of the
assay and a suitable test for impurities can be used.
The approach is similar for both assay and impurity tests:
Impurities are available
For the assay, this should involve demonstration of the discrimination of the
analyte in the presence of impurities and/or excipients; practically, this can be done by
spiking pure substances (drug substance or drug product) with appropriate levels of
impurities and/or excipients and demonstrating that the assay result is unaffected by
the presence of these materials (by comparison with the assay result obtained on
unspiked samples).
For the impurity test, the discrimination may be established by spiking drug
substance or drug product with appropriate levels of impurities and demonstrating the
separation of these impurities individually and/or from other components in the
sample matrix.
Impurities are not available
If impurity or degradation product standards are unavailable, specificity may
be demonstrated by comparing the test results of samples containing impurities or
degradation products to a second well-characterized procedure e.g.: pharmacopoeial
method or other validated analytical procedure (independent procedure). As
appropriate, this should include samples stored under relevant stress conditions: light,
heat, humidity, acid/base hydrolysis and oxidation.
For the assay, the two results should be compared;
For the impurity tests, the impurity profiles should be compared.
Peak purity tests may be useful to show that the analyte chromatographic peak is not
attributable to more than one component (e.g., diode array, mass spectrometry).
2. LINEARITY
The linear relationship should be evaluated across the range of the analytical
procedure. It may be demonstrated directly on the drug substance (by dilution of a
standard stock solution) and/or on synthetic mixtures of the drug product components,
30
using the proposed procedure. The latter aspect can be studied during investigation of
the range of the method.
Linearity should be evaluated by visual inspection of a plot of signals as a
function of analyte concentration or content. If there is a linear relationship, test
results should be evaluated by appropriate statistical methods, for example, by
calculation of a regression line by the method of least squares. In some cases, to
obtain linearity between assays and sample concentrations, the test data may need to
be subjected to a mathematical transformation prior to the regression analysis. Data
from the regression line itself may be helpful to provide mathematical estimates of the
degree of linearity.
The correlation coefficient, y-intercept, slope of the regression line and
residual sum of squares should be submitted. A plot of the data should be included. In
addition, an analysis of the deviation of the actual data points from the regression line
may also be helpful for evaluating linearity. For the establishment of linearity, a
minimum of 5 concentrations is recommended. Other approaches should be justified.
3. RANGE
The specified range is normally derived from linearity studies and depends on
the intended application of the procedure. It is established by confirming that the
analytical procedure provides an acceptable degree of linearity, accuracy and
precision when applied to samples containing amounts of analyte within or at the
extremes of the specified range of the analytical procedure.
The following minimum specified ranges should be considered:
For the assay of a drug substance or a finished (drug) product: normally from 80 to
120 percent of the test concentration
For content uniformity, covering a minimum of 70 to 130 percent of the test
concentration, unless a wider more appropriate range, based on the nature of the
dosage form (e.g., metered dose inhalers), is justified;
For dissolution testing: ± 20 % over the specified range;
31
e.g., if the specifications for a controlled released product cover a region from 20%,
after 1 hour, up to 90%, after 24 hours, the validated range would be 0-110% of the
label claim.
For the determination of an impurity: from the reporting level of an impurity1 to
120% of the specification;
For impurities known to be unusually potent or to produce toxic or unexpected
pharmacological effects, the detection/quantitation limit should be commensurate
with the level at which the impurities must be controlled;
Note: for validation of impurity test procedures carried out during development, it
may be necessary to consider the range around a suggested (probable) limit.
If assay and purity are performed together as one test and only a 100% standard is
used, linearity should cover the range from the reporting level of the impurities1 to
120% of the assay specification.
4. ACCURACY
Accuracy should be established across the specified range of the analytical
procedure.
Assay of Drug Substance: Several methods of determining accuracy are available:
a) Application of an analytical procedure to an analyte of known purity (e.g. reference
material);
b) Comparison of the results of the proposed analytical procedure with those of a
second well-characterized procedure, the accuracy of which is stated and/or
defined
c) Accuracy may be inferred once precision, linearity and specificity have been
established.
Assay of Drug Product: Several methods for determining accuracy are available
a) Application of the analytical procedure to synthetic mixtures of the drug product
components to which known quantities of the drug substance to be analysed have
been added;
32
b) In cases where it is impossible to obtain samples of all drug product components, it
may be acceptable either to add known quantities of the analyte to the drug
product or to compare the results obtained from a second, well characterized
procedure, the accuracy of which is stated and/or defined
c) Accuracy may be inferred once precision, linearity and specificity have been
established.
Assay of Impurities (Quantitation)
Accuracy should be assessed on samples (drug substance/drug product) spiked
with known amounts of impurities.
In cases where it is impossible to obtain samples of certain impurities and/or
degradation products, it is considered acceptable to compare results obtained by an
independent procedure. The response factor of the drug substance can be used.
It should be clear how the individual or total impurities are to be determined e.g.,
weight/weight or area percent, in all cases with respect to the major analyte.
Recommended Data
Accuracy should be assessed using a minimum of 9 determinations over a
minimum of 3 concentration levels covering the specified range (e.g., 3
concentrations /3 replicates each of the total analytical procedure).
Accuracy should be reported as percent recovery by the assay of known added
amount of analyte in the sample or as the difference between the mean and the
accepted true value together with the confidence intervals.
5. PRECISION
Validation of tests for assay and for quantitative determination of impurities
includes an investigation of precision.
Repeatability
Repeatability should be assessed using:
a) a minimum of 9 determinations covering the specified range for the procedure (e.g.,
3 concentrations/3 replicates each) or
b) a minimum of 6 determinations at 100% of the test concentration.
33
Intermediate Precision
The extent to which intermediate precision should be established depends on the
circumstances under which the procedure is intended to be used. The applicant should
establish the effects of random events on the precision of the analytical procedure.
Typical variations to be studied include days, analysts, equipment, etc. It is not
considered necessary to study these effects individually. The use of an experimental
design (matrix) is encouraged.
Reproducibility
Reproducibility is assessed by means of an inter-laboratory trial. Reproducibility
should be considered in case of the standardization of an analytical procedure, for
instance, for inclusion of procedures in pharmacopoeias. These data are not part of the
marketing authorization dossier.
Recommended Data
The standard deviation, relative standard deviation (coefficient of variation) and
confidence interval should be reported for each type of precision investigated.
6. LIMIT OF DETECTION
Several approaches for determining the detection limit are possible, depending
on whether the procedure is a non-instrumental or instrumental. Approaches other
than those listed below may be acceptable.
Based on Visual Evaluation
Visual evaluation may be used for non-instrumental methods but may also be used
with instrumental methods.
The detection limit is determined by the analysis of samples with known
concentrations of analyte and by establishing the minimum level at which the analyte
can be reliably detected.
Based on Signal-to-Noise
This approach can only be applied to analytical procedures which exhibit baseline
noise.
Determination of the signal-to-noise ratio is performed by comparing measured
signals from samples with known low concentrations of analyte with those of blank
34
samples and establishing the minimum concentration at which the analyte can be
reliably detected. A signal-to-noise ratio between 3 or 2:1 is generally considered
acceptable for estimating the detection limit.
Based on the Standard Deviation of the Response and the Slope
The detection limit (DL) may be expressed as:
3.3 σ
DL = ---------
S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate
of σ may be carried out in a variety of ways, for example:
Based on the Standard Deviation of the Blank
Measurement of the magnitude of analytical background response is performed by
analyzing an appropriate number of blank samples and calculating the standard
deviation of these responses.
Based on the Calibration Curve
A specific calibration curve should be studied using samples containing an analyte in
the range of DL. The residual standard deviation of a regression line or the standard
deviation of y-intercepts of regression lines may be used as the standard deviation.
Recommended Data
The detection limit and the method used for determining the detection limit should be
presented. If DL is determined based on visual evaluation or based on signal to noise
ratio, the presentation of the relevant chromatograms is considered acceptable for
justification.
In cases where an estimated value for the detection limit is obtained by calculation or
extrapolation, this estimate may subsequently be validated by the independent
analysis of a suitable number of samples known to be near or prepared at the detection
limit.
35
7. LIMIT OF QUANTITATION
Several approaches for determining the quantitation limit are possible,
depending on whether the procedure is a non-instrumental or instrumental.
Approaches other than those listed below may be acceptable.
Based on Visual Evaluation
Visual evaluation may be used for non-instrumental methods but may also be used
with instrumental methods.
The quantitation limit is generally determined by the analysis of samples with known
concentrations of analyte and by establishing the minimum level at which the analyte
can be quantified with acceptable accuracy and precision.
Based on Signal-to-Noise Approach
This approach can only be applied to analytical procedures that exhibit baseline noise.
Determination of the signal-to-noise ratio is performed by comparing measured
signals from samples with known low concentrations of analyte with those of blank
samples and by establishing the minimum concentration at which the analyte can be
reliably quantified. A typical signal-to-noise ratio is 10:1.
Based on the Standard Deviation of the Response and the Slope
The quantitation limit (QL) may be expressed as:
10 σ
QL = ----------
S
where σ = the standard deviation of the response
S = the slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte. The estimate
of σ may be carried out in a variety of ways for example:
Based on Standard Deviation of the Blank
Measurement of the magnitude of analytical background response is performed by
analyzing an appropriate number of blank samples and calculating the standard
deviation of these responses.
36
Based on the Calibration Curve
A specific calibration curve should be studied using samples, containing an analyte in
the range of QL. The residual standard deviation of a regression line or the standard
deviation of y-intercepts of regression lines may be used as the standard deviation.
Recommended Data
The quantitation limit and the method used for determining the quantitation
limit should be presented. The limit should be subsequently validated by the analysis
of a suitable number of samples known to be near or prepared at the quantitation limit.
8. ROBUSTNESS
The evaluation of robustness should be considered during the development
phase and depends on the type of procedure under study. It should show the reliability
of an analysis with respect to deliberate variations in method parameters.
If measurements are susceptible to variations in analytical conditions, the analytical
conditions should be suitably controlled or a precautionary statement should be
included in the procedure. One consequence of the evaluation of robustness should be
that a series of system suitability parameters (e.g., resolution test) is established to
ensure that the validity of the analytical procedure is maintained whenever used.
Examples of typical variations are:
Stability of analytical solutions and extraction time.
In the case of liquid chromatography, examples of typical variations are:
Influence of variations of pH in a mobile phase, influence of variations in mobile
phase composition, different columns (different lots and/or suppliers), temperature
and flow rate.
9. SYSTEM SUITABILITY TESTING
System suitability testing is an integral part of many analytical procedures.
The tests are based on the concept that the equipment, electronics, analytical
operations and samples to be analyzed constitute an integral system that can be
evaluated as such. System suitability test parameters to be established for a particular
procedure depend on the type of procedure being validated.
37
References
1. Lloyd R. Snyder, Joseph J. Kirkland and Joseph L. Glajah
Practical HPLC method development, 2nd Edition, New York, 1997.
2. Satinder Ahuja and Michael W. Dong
Hand book of Pharmaceutical Analysis by HPLC, Vol. 6, 1st Edition, Elsevier
academic Press, 2005.
3. Validation of analytical procedures: Text and Methodology, ICH Hormonised
Tripartite Guideline Q2 (R1), Commission of the European Communities (2005).
38
1.5 STABIITY INDICATING ASSAY
THE ICH REQUIREMENTS OF STRESS TESTING FOR OF DRUGS
AND ESTABLISHMENTS OF STABILITY –INDICATING ASSAYS
Determining Inherent Stability
The ICH guidelines entitled “stability testing of new drug substances and
products” (QIAR) requires that stress testing be carried out to elucidate the inherent
stability characteristics of the active substances. It suggests that the degradation
products that are formed under a variety of conditions should be identified and
degradation pathways established. It is stated that testing should include the effect of
temperature, humidity, (where appropriate); oxidation, photolysis, and susceptibility
to hydrolysis across a wide range of pH values. The study of effect of temperature is
suggested to be done in 10ºC increments above the accelerated temperature test
condition (e.g.50ºC, 60ºC etc) and that of humidity at a level of 75%RH or greater.
The ICH guidelines QIAR also emphasizes that the testing of these features,
which are susceptible to change during storage and are likely to influence quality,
safety, and or efficacy, must be done by stability –indicating testing methods. The
ICH guidelines Q3B entitled impurities in new drug products emphasizes on
providing documented evidence that analytical procedure are validated and suitable
for the detection and quantification of degradation products. It is also required that
analytical method should validated to demonstrate that impurity unique to the new
drug substances do not interfere with or are separated from specified or unspecified
degradation products in the drug product .The ICH guidelines Q6A, which provides
notes for guidance on specification, also mention the requirement of stability –
indicating assays under universal test criteria for both the drug substances and
products.
The same is also a requirement in the guidelines Q5C on stability testing of
biotechnological /biological products. Since there are no single assays or parameter
that profiles the stability characteristics of such products, stability- indicating profile
39
that provides assurance on detection of change in identity, purity, and potency of the
product.
Regulatory and Compendial Status of Stress Testing And Stability Indicating
Assays
The requirements of stress testing and establishment of stability indicating assays are
also covered in other international guidelines and compendia. The notes for guidelines
stability testing: stability testing of existing active substances and related finished
products (CPMP.QWP/122/02) issued by European committee for proprietary
medicinal products (CPMP) states that when a drug substance is described in an
official monograph (European pharmacopoeia or a European union member state), no
data required on the degradation products if they are named under the headings
“purity test “and or “impurities”; In this case no stress testing is required . That means
no forced decomposition required under the pharmacopoeial monographs. On other
hand, stress testing is to be done when no data available in the scientific literature or
the official pharmacopoeia. The route of stress testing for determining intrinsic
stability of drug is also mentioned as requirement in Canadian therapeutic products
directorate’s (TDP) draft guidelines entitled‘ stability testing of existing drug
substances and products ‘The requirements of establishment of stability indicating
assay is listed in guidelines on stability.
Testing of well established or existing drug substances and drug products issued by
CPMP, TDP, and WHO. Even the United States pharmacopoeia has a products should
be assayed for the potency of the use of a stability indicating assay. The requirement
is such explicit manner is, however, absent in other pharmacopoeias. Current ICH
guidelines on good manufacturing for active pharmaceutical ingredients (Q7A),
Which under adoption by WHO, also clearly mentions that the test procedure used in
stability testing should be validated and be stability indicating .The ICH guidelines do
not provide an exact definition of a stability indicating method.
Elaborate definitions of stability indicating methodology are, however provided in the
US-FDA stability guidelines of1998. Stability indicating methods, according to 1987
guideline, were defined as the quantitative analytical methods that are based on
characteristics structural, chemical, or biological properties of each ingredient of a
drug product that will distinguish each active ingredient from its degradation products
so that the active ingredient content can be accurately measured; This definition in
draft guidelines of 1998 read as:-
40
1. Validated quantitative analytical methods that can be detect change with time in
the chemical, physical, or microbiological properties of a drug substances and
drug products and that are specific so that the contents of active ingredients,
degradation products and other components of interest can be accurately measured
without interference The major change brought in the new guidelines are with
respect to a. introduction of the requirements of validation.
2. The requirements of analysis of degradation products and other components, apart
from the active ingredients.
Survey of The Literature For Reaction Conditions and Approaches
For Stress Testing
The practical aspect concerning the conduct of stress testing is addressed
neither by ICH nor by any other regulatory guidelines. Leaving the performance of
these studies to the discretion of the applicant. Therefore to determine the nature of
condition of stress testing, a survey was carried out of the monographs given in
various volumes of analytical profiles of drug substances .The monographs carry a
suitable reaction, which usually records the inherent stability behaviour of the drug
and the reaction condition used. Good amount of information on stress condition was
also found in the literature reports on establishment of stability indicating assays. The
information reveals that HCl at strength 0.1N is mostly used for stress decomposition
of drugs in acid condition. There are many studies where 1N has been exploited.
In some cases, HCl of higher normalities has been used. In few cases, mention is only
found of ‘acid’ without defining the type used. Also, large variation exists in the
reaction (temperature) condition, periods of study and the extent of decomposition.
While the temperature of study various between 25ºC and 116ºC the periods of
studies range from few minutes to as long as 2 months. The extent of drug
decompositions falls between two extremes, i.e. from nil to total degradation. This
clearly reflects that conditions for stress testing vary strongly, depending upon the
inherent stability of the drug. In alkaline degradation, sodium hydroxide either at
strength 0.1N or 1 N has been mostly employed for stress testing. Potassium
hydroxide or ammonium hydroxide is rarely used. Like the acidic degradation, a lot of
variation. Depending upon the inherent stability characteristics, some drugs did not
41
degrade even after refluxing in 0.1 N NaOH for one week while others underwent
almost complete degradation when kept at 25ºC in 0.1N alkali for just 210 min.
In another example, boiling with 1N NaOH for 5-20 min resulted in extensive
degradation or almost total loss of potency of Acyclovir, Cephalexin etc; while
boiling of Norfloxacin in alkali of same strength for 15 hr did not cause any change.
Not many reports could be found on stress studies in neutral pH. As may be seen
from the examples given in Table, the testing is generally done in water and no
significant degradation was observed; while temp. Ranged from 37-40ºC or refluxing
condition was employed. The slow rate of decomposition in neutral conditions in
most of the cases is understandable because reactions at neutral pH are non-catalytic
and hence very long periods under exaggerated temperature condition required to get
sufficient degradation products.
Approaches suggested in literature for performing stress testing
A few approaches have been purposed recently for performing forced
decomposition studies, as a sequel to introduction of requirements in ICH guidelines.
The first one by Hong and Shah suggests selection of stress testing condition, as
described in below Table. The concentration of the reagents and time of study are
given. It is recommended that stress tests should be carried out under a specific
condition for a time period enough to yield about 20-30% degradation would be too
strenuous and could possibly cause secondary degradation, yielding products of the
degradation product 42, which are not likely to be found under normal storage
condition.
Another approaches for forced decomposition studies have been put forth lately by
Reynolds etal. Following table provides the suggested protocol.
According to this approach, sufficient exposure of a drug substance degrade by ∼105
from its initial amount or after an exposure in excess of energy provided by
accelerated storage, whichever comes first .The duration of storage required at a given
temperature can be estimated by making conservative kinetic assumption For drug
products, non-drug substances related peaks should be distinguished from drug
substance related compounds which can be accomplished through comparative
analysis of stress samples drug substance plus excipients, and of excipients alone.
42
General Protocol For Stress Testing of Drug Substances and Products
Condition Drug Substances Drug Product
Solid Solution/
Suspension
Solid (Tablets,
capsules, blends)
Solution
i.v.,oral solution,
suspension)
Acid/Base √
Oxidative √ √ √
Photostability √ √ √
Thermal √ √ √
Thermal/
Humidity √ √
A further approach for automating multiple degradation experiments and performing
online HPLC analysis of the resultant impurities using combined diode array UV
detection and mass spectrometry has been exploited in another recent publication.
It involves the use of an automated workstation, wherein ten samples are refluxed
with stirring in a single heating block .The robot arm is equipped with a sampling
device capable of removing aliquots, during reflux experiment and them to HPLC
injector after suitable neutralization and dilution .The automation process does not
only benefit in terms of analyst time but also allow extra high quality data to be
collected with a minimal time delay between sample preparation and analysis .The
combination of HPLC and mass detection in the system also allows structural
information to be obtained on the degradants along with kinetic data on the
appearance and disappearance of degradants during the course of experiments.
43
REFERENCES
1) D.W. Reynolds, K.L. Fachieve, J.F. Mullaney, K.M. Alsante, T.D. Hatajik and
M.G. Motto, Available guidance and test practice for conducting forced
degradation studies, Pharm. Technol 20 (2002) pp48-56.
2) J.L.Sims, J.K. Roberts, A.G. Bateman, J.A.Carreiran and M.J. Hardy, an
automated workstation for forced degradation of active pharmaceutical
ingredients, J of Pharm. Science, 2002 (91) 884-892.
3) L. Chafetz, stability indicating assay method for drug and their dosage forms,
J of Pharm.Science, 1971 (60) 335-341.
4) ICH, Stability testing Photostability testing of new drug substances for
products, International conference on harmonization, IFPMA, Geneva 1996.
5) ICH, Stability testing of new drug substances & product, International
conference on harmonization, IFPMA, Geneva 1993.